K083737 · Barrx Medical, Incorporated · GEI · Feb 10, 2009 · General, Plastic Surgery
Device Facts
Record ID
K083737
Device Name
HALO ABLATION CATHETER
Applicant
Barrx Medical, Incorporated
Product Code
GEI · General, Plastic Surgery
Decision Date
Feb 10, 2009
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The HALO90 System (inclusive of the HALO90 Ablation Catheter) is indicated for use in the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to the esophagus. Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, and Angiodysplasia.
Device Story
HALO90 Ablation Catheter is a single-use electrosurgical device used with the HALO Energy Generator for tissue coagulation in the gastrointestinal tract. The system delivers energy to bleeding or non-bleeding sites (e.g., Barrett's Esophagus, ulcers, vascular lesions). Operated by clinicians in a clinical setting, the device functions via an electrosurgical coagulation principle. The current submission represents a name change for the previously cleared HALO Coagulation Catheter (K062723) to unify international and domestic branding; no changes were made to construction, materials, or principle of operation.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on the device being identical to the predicate, with no changes to construction, materials, or principle of operation.
Technological Characteristics
Electrosurgical coagulation system consisting of a single-use ablation catheter and an energy generator. Operates via electrosurgical coagulation. No changes in materials or construction compared to the predicate.
Indications for Use
Indicated for coagulation of bleeding and non-bleeding GI tract sites, including esophagus. Conditions include esophageal ulcers, Mallory-Weiss tears, arteriovenous malformations, angiomata, Barrett's esophagus, Dieulafoy lesions, and angiodysplasia.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
HALO® Coagulation Catheter model 90-9100 (K062723)
Submission Summary (Full Text)
{1}------------------------------------------------
## 510(k) SUMMARY
## BÂRRX Medical's HALO® Ablation Catheter
### FEB 1 0 2009
#### Submitter's Name, Address, Telephone Number, Contact السندر السياري والي الموسم الموسم الموسم الموسم الموسم الموسم الموسم الموسم الموسم الموسم الموسم الموسم الموسم الموسم الموسم الموسم الموسم الموسم الموسم الموسم الموسم الموسم Person, and Date Prepared:
BARRX Medical Inc. 540 Oakmead Parkway Sunnyvale, CA 94085
Phone: (408) 328-7302 (408) 328-7395 Facsimile:
Contact Person: Viorica Filimon
December 15, 2008 Date Prepared:
#### Name of device and Name/Address of Sponsor: 2.
HALO90 Ablation Catheter
BÂRRX Medical Inc. 540 Oakmead Parkway Sunnyvale, CA 94085
#### 3. Common or Usual Name(s):
Electrosurgical Coagulation System
#### 4. Classification Name:
Product code: GEI CFR Section: 878.4400 Electrosurgical, cutting & coagulation & accessories Device Class: II Classification panel: General & Plastic Surgery
#### Predicate Devices 5.
HALO® Coagulation Catheter model 90-9100 (K062723) manufactured by BÂRRX Medical Inc.
{2}------------------------------------------------
K083737 page 2 of 2
#### 6. Intended Use / Indications for Use
The HALO90 System (inclusive of the HALO90 Ablation Catheter) is indicated for use in the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to the esophagus. Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, and Angiodysplasia.
#### Technological Characteristics 7.
The HALO90 System consists of the following components:
- HALO" Energy Generator model 1100C-115C and 90-9100 and . accessories (output cable, and an optional footswitch).
- A single use HALO90 Ablation Catheter model 90-9100 .
The HALO® System performance and mode of operation did not change. This 510 (k) addresses the name change for the catheter.
### HALO® Ablation Catheter
There are no changes in construction, materials, principle of operation, intended use and indications for use, for the HALO® Ablation catheter model 90-9100 when compared with the predicate device HALO® Coagulation Catheter model 90-9100.
The following changes are subject to this 510k submission: HALO® Coagulation Catheter model 90-9100 was marketed in Europe and Canada as HALO90 Ablation Catheter, under the same indications for use for tissue coagulation. For unifying the international and domestic labeling we request FDA to allow the name change for HALO® Coagulation Catheter to HALO90 Ablation Catheter. There are no changes associated with the intended use, indication for use or the principle of operation.
### HALO90 Energy Generator
There are no changes associated to the HALO® Energy Generator software, hardware and accessories.
#### Substantial Equivalence 8.
The HALO90 Ablation Catheter model 90-9100 and the predicate devices HALO® Coagulation Catheter model 90-9100 are identical in construction except the product name changes:
o From HALO90 Coagulation Catheter to HALO90 Ablation Catheter
o From HALO90 Coagulation System to HALO90 System.
There are no differences were evaluated on bench and did not raise questions regarding safety and efficacy. Thus the devices are equivalent.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or a stylized human figure with outstretched arms.
### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Viorica Filimon Vice President, Quality and Regulatory Affairs BÂRRX Medical, Inc. 540 Oakmead Parkway SUNNYVALE CA 94085
## FEB 1 0 2009
Re: K083737
Trade/Device Name: HALO90 Ablation Catheter (model 90-9100) Regulation Number: 21 CFR §878.4400 Regulation Name: Electrosurgical, cutting &coagulation & accessories Regulatory Class: II Product Code: ·GEI Dated: December 15, 2008 Received: January 13, 2009
Dear Ms. Filimon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | (240) 276-0115 |
|----------------|----------------------------------|----------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | (240) 276-0115 |
| 21 CFR 892.xxx | (Radiology) | (240) 276-0120 |
| Other | | (240) 276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/odrb/industry.suppor/index.html.
Sincerely yours,
Laura B. Morris
anine M. M. orri Acting Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
### Indications for Use Statement
510(k) Number (if known): K083737
Device Name: HALO90 Abliation Catheter (model 90-9100)
Indications for Use:
The HALO® System (inclusive of HALO® Ablation Catheter) is indicated for use in the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to, the esophagus. Indications include Esophageal Ulcers, Mallory-Weiss tears, Arteriovenous Malformations, Angiomata, Barrett's Esophagus, Dieulafoy Lesions, and Angiodysplasia.
Prescription Use X (Part 21 C.F.R. 801 Subpart D)
AND/OR
Over-The-Counter Use (21 C.F.R. 807 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Heiber Perez
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 08 27 3 510(k) Number _
Page **_ of _**
HALO® Ablation Catheter
で
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.