K140314 · Yonmedical BV · GEH · Jun 12, 2014 · General, Plastic Surgery
Device Facts
Record ID
K140314
Device Name
WARTIE WART REMOVER
Applicant
Yonmedical BV
Product Code
GEH · General, Plastic Surgery
Decision Date
Jun 12, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
Wartie® Wart Remover is intended for over-the-counter treatment of common warts and plantar warts to be used in adults and children 4 years of age and older.
Device Story
Wartie® Wart Remover is a portable, over-the-counter cryosurgical device for home use. It consists of a pressurized canister containing 50ml of dimethyl ether (DME) and a custom applicator unit. The user activates the aerosol to deliver the cryogen to a metal nose piece, which acts as a cold retraction vehicle. The device is applied directly to the wart; the metal tip transfers cold to the skin, freezing the wart tissue. This process damages the wart cells, causing the wart to blister and fall off or be discarded by the body. The device features a secured locking ring for mechanical safety and thermal safety precautions during activation. It is intended for self-administration by patients or caregivers to treat common and plantar warts, providing pen-pointed accuracy for freezing.
Clinical Evidence
No clinical data provided. Safety and efficacy supported by bench testing, including comparative biological tissue skin test models against predicate devices, and literature regarding cryosurgical treatment of verrucae plantaris.
Technological Characteristics
Portable cryosurgical system. Materials: PP cap, POM cone/ring, nickel-coated bronze nose piece, aluminum canister. Energy: Thermal energy removal via cryogen (DME). Dimensions: Handheld aerosol canister. Connectivity: None. Sterilization: Not specified. Software: None.
Indications for Use
Indicated for OTC treatment of common and plantar warts in adults and children aged 4 years and older.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image contains the logo for YouMedical. The logo consists of a stylized graphic to the left of the text "YouMedical". The graphic is abstract and appears to be a stylized representation of interconnected shapes. The text "YouMedical" is in a sans-serif font and is positioned to the right of the graphic.
Sub. N.: Version : Date:
K140314 3.0 06-June-14
JUN 1 2 2014
# Premarket Notification 510(k) summary (As Required by 21 CFR 807.92)
YouMedical BV,
Submitter's Name: Address:
Contact Person: Date Prepared:
Device Trade Name: Proprietary Name:
Classification Name :
Common Name:
Rijnsburgstraat 9-11, 1059AT Amsterdam, The Netherlands. Kathleen Johnson 06th June, 2014 Wartie® Wartie® Wart Remover and other proprietary name OTC Wart Removal System Cryosurgical unit, accessories
Classification Panel : General & Plastic Surgery Classification Code : GEH
Regulation Number : 21 CFR 878.4350
Predicate Devices:
Substantial equivalence is claimed with the following devices:
| Name of Device | Manufacturer | Predicate Comparison | 510(k) Number |
|------------------------------------------------------|-----------------------------------------------------|---------------------------------------------------------|---------------|
| Wartner®<br>Wart<br>Removal<br>System | Wartner US<br>b.v | Intended<br>use,<br>technology,<br>materials,<br>label. | K032271 |
| Dr. Scholl's®<br>Freeze<br>AwayTM<br>Wart<br>Remover | Schering-<br>Plough<br>Healthcare<br>Products, Inc. | Intended<br>use,<br>technology,<br>materials,<br>label. | K031697 |
| Histofreezer®<br>Wart<br>Removal<br>System | OraSure<br>Technologies,<br>Inc. | Intended<br>use,<br>technology,<br>materials,<br>label. | K023487 |
Rijnsburgstraat 9-11 |1059 AT Amsterdam | The Netherlands Tel: +31 20-679-8529 l Fax: +31 20-524-8759 www.voumedical.com info@youmedical.com
{2}------------------------------------------------
Image /page/1/Picture/1 description: The image contains a logo with the text "YouMedical" next to a symbol. The symbol is a complex, abstract design featuring interconnected loops and curves, resembling a stylized representation of interconnectedness or a network. The text "YouMedical" is written in a bold, sans-serif font, with each letter clearly legible. The overall design is clean and professional, suggesting a modern and innovative approach.
Sub. N.: Version:
K140314 3.0 06-June-14
| Wart Freeze | Koninklijke | Intended | K130599 |
|-------------|-------------|-------------|---------|
| | Utermöhlen | use, | |
| | N.V. | technology, | |
| | | materials. | |
## Description of the Device:
The Wartie® Wart Remover is effective in the treatment of warts by providing an over the counter cryosurgery product (for the treatment of warts), to be used at home.
Wartie® Wart Remover consists of:
- . A pressurized canister filled with 50ml of the compressed liquid gas dimethy) ether.
- A custom application unit used to administer the cold delivered by the ◆ cryogen to the wart.
- Instructions for use in which you can read about the product how it works, . how to use the product to treat warts, warnings and limitations.
The Biological Safety Evaluation of the Wartie® Wart Remover assesses the chemical characterization of the product, reviews the literature data on safety aspects and assesses the need for biological safety testing according to ISO 10993. The biological safety evaluation has identified and discussed the materials coming into contact with the body during use. Special attention has been given to the nickel coating of the metal tip, which is in direct contact with the wart during application. The biological safety evaluation has shown that the levels of nickel dissociation of the product are within the applicable safety limits and safe for use as indicated. YouMedical BV has evaluated the total possible exposure of Wartie® Wart Remover considering the nature of the product and its evaluation and has established that the maximum possible exposure levels are well within the accepted limits. The biological safety evaluation has established the parameters of use of Wartie® Wart Remover and its equivalence to the predicate device.
Based on the chemical composition of contacting materials during treatment and their safety data, the short term contact during treatment, extremely low temperature of the tip during skin contact and the mode of action of Wartie® Wart Remover it can be concluded that the biological and performance safety risks of Wartie® Wart Remover for its intended use and application are low and therefore acceptable. Residual risks are identified and directly related to the mode of action (freezing) of the Wartie® Wart Remover and presence of nickel. The extent of residual risks are estimated lower or equal when compared to predicate products and therefore reqarded acceptable. Information on the residual risks is provided in the information to the user.
{3}------------------------------------------------
Image /page/2/Picture/2 description: The image contains the logo for YouMedical. The logo consists of a stylized symbol to the left of the word "YouMedical". The symbol appears to be an abstract design, possibly representing interconnectedness or a network. The text "YouMedical" is written in a bold, sans-serif font.
| Sub. N.: | K140314 |
|----------|------------|
| Version: | 3.0 |
| Date: | 06-June-14 |
Wartie® Wart Remover technology involves freezing a wart (common and/or platar) using a very cold releasing liquid gas (cryogen) into a patented-pending applicator which, in turn become very cold and is able to freeze the skin. The freezing of the skin causes damage to the cells housing and protecting the wart. The latter either develops a blister underneath and falls off, or the damaged skin cells are discarded by the body, thereby also effectively removing the wart and the virus.
Wartie® Wart Remover contains the following materials:
| Name | Material |
|------------|---------------------------|
| Cap | PP |
| Cone | POM |
| Nose Piece | Bronze coated with Nickel |
| Ring | POM |
| Can | Aluminum |
Wartie® Wart Remover is intended for topical treatment of common and-plantar warts, using the following application times, which vary on the basis of the wart diameter:
| | WART DIAMETER | TREATMENT<br>TIME |
|------------------------------------------------------------|----------------------------------|--------------------|
| Common warts and plantar<br>warts on toes and arch of foot | Smaller than 3/32 in<br>(2.5mm) | 10 seconds |
| Common warts and plantar<br>warts on toes and arch of foot | 3/32 - 3/16 in (2.5 - 5.0<br>mm) | 15 seconds |
| Common warts and plantar<br>warts on toes and arch of foot | Larger than 3/16 in<br>(5.0mm) | 20 seconds |
| Calloused Plantar warts on<br>heels and balls of feet | All sizes | 40 seconds or less |
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for YouMedical. The logo consists of a stylized graphic to the left of the company name. The graphic is an abstract design with interwoven lines, while the text "YouMedical" is written in a simple, sans-serif font.
Sub. N.: Version: Date:
K140314 3.0 06-June-14
## Intended Use and Indication for Use Statement:
Wartie® Wart Remover is intended for over-the-counter treatment of common warts and plantar warts to be used in adults and children 4 years of age and older.
## Application:
The product administers cryogens to the wart by way of an applicator.
## Technological Characteristics:
Wartie® Wart Remover is a portable cryosurgical systems comprised of a canister containing a cryogen and an applicator that applies the cold to the wart to be treated for the over-thecounter treatment of common and plantar warts. The Wartie® Wart Remover employs a metal applicator used as cold retraction vehicle. This allow for pen-pointed accuracy in freezing the skin. Wartie® Wart Remover makes use a secured locking ring in order to assure mechanical safety of the product, and thermal safety precautions are associated with activation of the aerosol. Finally, the product uses thermal energy removed from skin at the anatomical site of a common wart and/or plantar wart.
## Non-Clinical Testing:
The following tests were performed on the finished product:
- Stability test; .
- · Compatibility test.
## Bench testing:
The following test was performed on the finished product:
- Comparative bench testing between Wartie® Wart Remover and the predicate . devices using a biological tissue skin test model. Please note that the bench test has been conducted on Histofreezer® Wart Removal System, Wartner® Wart Removal System, Dr. Scholl's® Freeze Away TM Wart Remover, Wart Freeze (named in EU as Actifreeze), and Bozuuka Sub-Zero because aimed for both EU and US market. As indicated above, the device named "Actifreeze" in the bench test is available in US under the name "Wart Freeze", indivated as predicate device. Not being available in US, Bozuuka Sub-Zero cannot be selected as predicate device.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image contains the logo for YouMedical. The logo consists of a stylized graphic to the left of the text "YouMedical". The graphic is a complex, abstract design, possibly representing interconnectedness or a network. The text "YouMedical" is in a bold, sans-serif font, with the "You" slightly smaller than "Medical".
| Sub. N.: | K14031- |
|----------|-----------|
| Version: | 3. |
| Date: | 06-June-1 |
4 ﻡ (2 ) 4 )
14
The non-clinical and bench testing have supported the safety and efficacy of Wartie® Wart Remover.
Substantial Equivalence Information:
| | Wartie®<br>Wart | Wartner®<br>Wart | Dr. Scholl's®<br>Freeze Away | Histofreezer®<br>Wart Removal | Wart Freeze |
|-----------------------------------|--------------------------------------------------------------|-----------------------------------------------------------|-----------------------------------------------------------|-----------------------------------------------------------|-----------------------------------------------------------|
| | Remover | Removal<br>system | TM Wart<br>Remover | System | |
| FDA 510(k)<br>number | K140314 | K032271 | K031697 | K023487 | K130599 |
| intended use<br>comparison | ОТС<br>treatment of<br>common<br>warts and<br>plantar warts. | OTC treatment<br>of common<br>warts and<br>plantar warts. | OTC treatment<br>of common<br>warts and<br>plantar warts. | OTC treatment<br>of common<br>warts and<br>plantar warts. | OTC treatment<br>of common<br>warts and<br>plantar warts. |
| indications for<br>use comparison | OTC<br>treatment of<br>common and<br>plantar warts. | OTC treatment<br>of common<br>warts and<br>plantar warts. | OTC treatment<br>of common<br>and plantar<br>warts. | OTC treatment<br>of common and<br>plantar warts. | OTC treatment<br>of common<br>and plantar<br>warts. |
| target<br>population | Adults and<br>Children 4+. | Adults and<br>Children 4+. | Adults and<br>Children 4+. | Adults and<br>Children 4+. | Adults and<br>Children 4+. |
| anatomical<br>sites | Topical -<br>Wart. | Topical -<br>Wart. | Topical -<br>Wart. | Topical - Wart. | Topical - Wart. |
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for YouMedical. The logo consists of a stylized, abstract symbol to the left of the text "YouMedical". The symbol appears to be a geometric design, possibly representing interconnectedness or a network. The text "YouMedical" is in a bold, sans-serif font, with the "Y" capitalized and slightly larger than the rest of the letters.
Sub. N.: Version: Date:
K140314 3.0 06-June-14
| energy used<br>and/or<br>delivered | Thermal<br>energy<br>removed from<br>skin via a<br>metal<br>interface. | Thermal<br>energy<br>removed from<br>skin via a<br>sponge<br>containing<br>liquid gas. | Thermal<br>energy<br>removed from<br>skin via a<br>sponge<br>containing<br>liquid gas. | Thermal energy<br>removed from<br>skin via a<br>sponge<br>containing<br>liquid gas. | Thermal<br>energy<br>removed from<br>skin via a<br>polypropylene<br>applicator. |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| design | Device<br>requiring<br>activation and<br>application.<br>Metal tip<br>provides pen-<br>pointed<br>accuracy. | Device<br>requiring<br>assembly,<br>activation, and<br>application. | Device<br>requiring<br>assembly,<br>activation, and<br>application. | Device<br>requiring<br>assembly,<br>activation, and<br>application. | Device<br>requiring<br>activation and<br>application. |
| materials | DME from<br>aerosol can<br>applied to the<br>skin through<br>a metal core<br>via a Nickel<br>tip providing<br>pen-point<br>accuracy. | DME/Propane<br>from aerosol<br>can applied to<br>the skin<br>through a<br>sponge core<br>mounted on a<br>plastic<br>applicator. | DME/Propane<br>from aerosol<br>can applied to<br>the skin<br>through a<br>sponge core<br>mounted on a<br>plastic<br>applicator. | DME/Propane<br>from aerosol<br>can applied to<br>the skin<br>through a<br>sponge core<br>mounted on a<br>plastic<br>applicator. | DME from<br>aerosol can<br>applied to the<br>skin through a<br>polypropylene<br>applicator. |
| biocompatibility | Established<br>according to<br>ISO 10993. | Established<br>according to<br>ISO 10993. | Established<br>according to<br>ISO 10993. | Established<br>according to<br>ISO 10993. | Established<br>according to<br>ISO 10993. |
## Comparison of Technological Characteristics:
The Wartie® Wart Remover is substantially equivalent to the Wartner® Wart Removal System, Dr. Scholl's® Freeze Away TM Wart Remover, Histofreezer® Wart Removal System, Wart Freeze devices for the same indications. All indicated devices are portable cryosurqical systems comprised of a canister containing a cryogen and an applicator that applies the cold to the wart to be treated for the over-the-counter treatment of common and plantar warts. The Wartie® Wart Remover, however, employs a metal applicator while Wartner® Wart Removal System, Dr. Scholl's® Freeze Away TM Wart Remover and Histofreezer® Wart Removal System employ a Polyethylene sponge/foam applicator; and Wart Freeze employs a polypropylene applicator.
> Rijnsburgstraat 9-11 | 1059 AT Amsterdam | The Netherlands Tel: +31 20-679-8529 l Fax: +31 20-524-8759 www.youmedical.com info@youmedical.com
{7}------------------------------------------------
Image /page/6/Picture/2 description: The image shows the logo for YouMedical. The logo consists of a stylized graphic to the left of the text "YouMedical". The graphic is abstract and appears to be a stylized representation of a medical symbol or a stylized letter.
Sub. N.: Version . Date:
K140314 3.0 . 06-June-14
Wartie® Wart Remover differs from the predicate devices in that the material of the cold retraction vehicle differs. While Wartner® Wart Removal System, Dr. Scholl's® Freeze Away TM Wart Remover and Histofreezer® Wart Removal System apply a cryogen to a sponge, and , Wart Freeze applies a polypropylene applicator; Wartie® Wart Remover applies the cryogen to a metal nose piece which allows efficient heat transfer to the skin.
This difference in technology has been evaluated in bench testing to determine the equivalence of the products, and market experience and other available devices show that the heat transfer via a metal core is a used and approved technology.
The method of freezing skin via a metal applicator allows for pen-pointed accuracy in freezing the skin.
As mentioned earlier, bench testing has shown the performance of Wartie® Wart Remover to be equivalent with abovementioned OTC technologies in terms of safety and efficacy, thereby supporting its OTC status.
While the technology employed by Wartie® Wart Remover differs from its predicates in that the product uses pure DME gas applied via a metal applicator, bench testing has shown equivalent performance for safety and efficacy on biological tissue. DME gas is also used by Wart Freeze but with a polypropylene applicator.
Note: The limitation of 4 years and older is an industry-wide limitation based on clinical data available for comparable products. The predicate device Wartner® Wart Removal System -K032271, as well as other OTC products have the same restriction. The clinical study carried out with Histofreezer® has shown the technology to be appropriate for pediatric patients (above the age of 4) , A bench study, further discussed in section 18, has furthermore shown compared Wartie® Wart Remover with its predicate as well as other available products such as Histofreezer®, showing that its performance is within the parameters of Histofreezer® and, thereby, also applicable for patients above the age of 4. The bench study showing that Warties Wart Remover's metal probe does not adhere to the skin (also shown in section 18), thereby showing that there are no additional risks, further support the appropriate limitation to pediatric patients above the age of 4.
Rijnsburgstraat 9-11 |1059 AT Amsterdam | The Netherlands Tel: +31 20-679-8529 l Fax: +31 20-524-8759 www.youmedical.com info@youmedical.com
<sup>1</sup> Ricketti, J.C. et al. A Study of a New Method of Cryosurgical Treatment of Verrucae Plantaris. The Lower Extremity: Vol. 2 No. 3; 1995.
{8}------------------------------------------------
3.0
Image /page/7/Picture/1 description: The image shows the logo for YouMedical. The logo consists of a stylized, abstract symbol to the left of the word "YouMedical". The symbol appears to be a geometric design, possibly representing interconnectedness or a network. The word "YouMedical" is written in a bold, sans-serif font, with the "Y" and "M" capitalized.
# Premarket Notification 510(k) Summary Wartie
Sub. N. : K140314 Version: 06-June-14 Date:
## Summary_of Similarities:
The Wartie® Wart Remover has the following similarities with the Wartner® Wart Removal System, Dr. Scholl's® Freeze Away TM Wart Remover, Histofreezer® Wart Removal System and Wart Freeze:
- The same intended use and indication for use,
- The same warnings and contraindications,
- Available for over-the-counter use, and
- Destroy common warts and plantar warts by cryoablation.
In details:
- Wartie® Wart Remover and Wartner® Wart Removal System, Dr. Scholl's® Freeze . Away TM Wart Remover, Histofreezer® Wart Removal System and Wart Freeze are intended for over- the-counter treatment of common warts and plantar warts to be used in adults and children 4 years of age and older.
- Wartner® Wart Removal System, Dr. Scholl's® Freeze Away TM Wart Remover, Wart . Freeze and Histofreezer® Wart Removal System and Wartie® Wart Remover use a cryogen filled into an aerosol can in order to transform kinetic energy into thermal energy in order to remove a common wart or plantar wart by cryosurgery.
- . Wartie® Wart Remover makes use a secured locking ring in order to assure mechanical safety of the product. A similar feature is also employed by the Wartner® Wart Removal System.
- In the predicate devices identified above as well as Wart Remover, thermal . safety precautions are associated with activation of the aerosol.
- Wartie® Wart Remover, as the predicates mentioned, uses thermal energy removed . from skin at the anatomical site of a common wart and/or plantar wart.
- The label of the Wartie® Wart Remover has been developed to ensure consumer safety . and is equivalent to the predicate device identified for this purpose.
- The safety and warning statements for the OTC predicate devices and for all the other . predicate labeling device are similar.
{9}------------------------------------------------
Image /page/8/Picture/2 description: The image contains a logo for YouMedical. The logo consists of a stylized graphic to the left of the text "YouMedical". The graphic is abstract and appears to be a stylized representation of a medical symbol or a DNA strand.
Sub. N.: Version : Date:
K140314 3.0 06-June-14
## Conclusion:
-
Based on the information presented, it is concluded that the proposed product, Wartie® Wart Remover, is safe and effective for its intended use and is substantially equivalent to the predicate devices. It can be therefor considered substantially equivalent in intended use, indication for use, safety and affectivity profile, and labeling.
> Rijnsburgstraat 9-11 |1059 AT Amsterdam } The Netherlands Tel: +31 20-679-8529 l Fax: +31 20-524-8759 www.youmedical.com info@youmedical.com
{10}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized symbol resembling an abstract human figure with outstretched arms, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the perimeter of the circle.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 12, 2014
YouMedical BV % Ms. Kathleen Johnson Medical Device Approvals Incorporated P.O. Box 2042 Fairfield, lowa 52556
Re: K140314
Trade/Device Name: Wartie® Wart Remover Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical unit and accessories Regulatory Class: Class II Product Code: GEH Dated: May 14, 2014 Received: May 16, 2014
Dear Ms. Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register,
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{11}------------------------------------------------
Page 2 - Ms. Kathleen Johnson
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
.. . . . . . .
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{12}------------------------------------------------
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page,
## 510(k) Number (if known)
### K140314
#### Device Name
#### Indications for Use (Describe)
Wartie® Wart Remover is indicated for the over-the-counter treatment of common warts for patients aged 4 years and older.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
## PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.
Image /page/11/Figure/12 description: This image shows a form titled "FOR FDA USE ONLY". The form is for the "Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)". The form number is FORM FDA 3881 (8/13) and it is page 1 of 2.
Strictly Confidential
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.