K101825 · C2 Therapeutics · GEH · Sep 29, 2010 · General, Plastic Surgery
Device Facts
Record ID
K101825
Device Name
CRYOBALLOON ABLATION SYSTEM
Applicant
C2 Therapeutics
Product Code
GEH · General, Plastic Surgery
Decision Date
Sep 29, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The C2 Therapeutics CryoBalloon Ablation System is intended to be used as a cryosurgical tool for the destruction of unwanted tissue in the field of general surgery, specifically for endoscopic applications.
Device Story
CryoBalloon Ablation System is a cryosurgical tool for endoscopic tissue destruction. System components: catheter with balloon probe, handle with trigger-controlled refrigerant delivery valve, and liquid nitrous oxide cartridge. Physician-operated during endoscopic procedures (3.7mm accessory channel). Operation: balloon probe inflated and cooled with nitrous oxide; balloon contacts esophageal wall to ablate tissue via extreme cold. Nitrous oxide remains contained within balloon; gas exits through catheter proximal end. Device is single-patient, one-time continuous application use. Internal 9VDC battery powers cartridge heater. Output is thermal energy for tissue ablation; physician monitors procedure via endoscope to guide placement and assess tissue destruction. Benefits include targeted tissue ablation for general surgical applications.
Clinical Evidence
No clinical data provided. Safety and effectiveness supported by biocompatibility, performance, and animal test results.
Indicated for the destruction of unwanted tissue in general surgery, specifically for endoscopic applications.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
Predicate Devices
CryMed Technologies SprayGenix™ Ablation System (K060555)
Barrx Medical HALO360 Coagulation System (K051168)
Submission Summary (Full Text)
{0}------------------------------------------------
101825
CONFIDENTIAL
# Section 5-510(k) Summary or 510(k) Statement
t
#### I. General Information
Submitter: C2 Therapeutics, Inc. SEP 2 9 2010 303 Convention Way, Suite 1 Redwood City, CA 94063 Contact Person: Anne Worden Regulatory Consultant 925-895-1200 Summary Preparation Date: 6/30/2010
#### II. Names
Device Names: CryoBalloon Ablation System
Primary Classification Names: Cryosurgical unit with a nitrous oxide cooled balloon probe and accessories
#### III. Predicate Devices
- CryMed Technologies SprayGenix™ Ablation System K060555 .
- CSA Medical CryoSpray Ablation System K070893 .
- GI Supply Polar Wand Cryotherapy System K021387 .
- Barrx Medical HALO360 Coagulation System - K051168 .
## IV. Product Description
The CryoBalloon Ablation System ("System") is used to destroy unwanted tissue by application of extreme cold. Upon activation by a physician, the balloon probe at the end of the catheter is simultaneously inflated and cooled with nitrous oxide. The balloon probe comes in contact with the wall of the esophagus and ablates unwanted tissue. Nitrous oxide is fully contained within the balloon probe and does not contact the esophagus. The nitrous oxide gas exits the patient through the proximal end of the catheter. The CryoBalloon Ablation System is designed for one time, continuous application use (single patient) in conjunction with a therapeutic endoscope (3.7mm accessory channel ID). The System is comprised of the following main components:
- CryoBalloon Ablation Catheter (REF FG-1000) consists of a connector, catheter . shaft, balloon probe, and protective sheath. This is supplied sterile.
- CryoBalloon Ablation Handle (REF FG-1001) contains the cartridge heater and . refrigerant delivery valve which is controlled via the trigger. The unit is internally powered (9VDC non-replaceable lithium battery pack) and supplied non-sterile.
- CryoBalloon Ablation Cartridge (REF FG-1002) contains liquid nitrous oxide. The . Cartridge is supplied non-sterile and contains enough refrigerant for one to two ablation sites.
{1}------------------------------------------------
### V. Indications for Use
The C2 Therapeutics CryoBalloon Ablation System is intended to be used as a cryosurgical tool for the destruction of unwanted tissue in the field of general surgery, specifically for endoscopic applications.
#### VI. Rationale for Substantial Equivalence
The CryoBalloon Ablation System shares the same or similar indications for use, device operation, overall technical and functional capabilities, and therefore is substantially equivalent for use to the predicate devices as a cryosurgical unit with a nitrous oxide cooled balloon probe and accessories.
In addition, comparative performance test data demonstrated adequate device performance and safety.
| Comparison<br>Characteristic | K10 | K070893 & K060555 | K021387 | K051168 |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Brand Name | CryoBalloon<br>Ablation System | CryoSpray Ablation™<br>& SprayGenix™ Cryo | Polar Wand<br>Cryotherapy System | HALO 360<br>Coagulation System |
| Intended Use | | | | |
| Intended Use | Destruction of<br>unwanted tissue with<br>extreme cold | Destruction of<br>unwanted tissue with<br>extreme cold | Ablation of unwanted<br>tissue with extreme cold | RF for tissue<br>coagulation |
| Indications for<br>Use | Intended to be used as<br>a cryosurgical tool for<br>destruction of<br>unwanted tissue in<br>the field of general<br>surgery, specifically<br>for endoscopic<br>applications. | Intended to be used as a<br>cryosurgical tool for<br>destruction of unwanted<br>tissue in the field of<br>general surgery,<br>specifically for<br>endoscopic applications. | Used for ablation of<br>unwanted tissue in the<br>fields of dermatology,<br>gynecology, general<br>surgery, urology, and<br>gastroenterology. The<br>system may be used<br>with a variety of<br>cryogens, e.g. carbon<br>dioxide, nitrous oxide,<br>argon, krypton | Indicated for use in the<br>coagulation of bleeding<br>and non-bleeding sites<br>in the gastrointestinal<br>tract including but not<br>limited to the<br>esophagus. Indications<br>include Esophageal<br>Ulcers, Mallory-Weiss<br>tears, Arteriovenous<br>Malformations,<br>Angiomata, Barrett's<br>esophagus, Dieulafoy<br>Lesions, and<br>Angiodysplasia. |
| Key Technical Characteristics | | | | |
| Control | User | User | User | User |
| Method of<br>Action | Thermal | Thermal | Thermal | Thermal |
| Endoscopic<br>procedure | Yes | Yes | Yes | Yes |
See Table 9 - Substantial Equivalence Comparison of Intended Use and Technical Characteristics for comprehensive analysis of Technical Characteristics.
#### Safety and Effectiveness Information VII.
The review of the indications for use and technical characteristics provided demonstrates that the CryoBalloon Ablation System is substantially equivalent to the predicate devices.
{2}------------------------------------------------
Biocompatibility, performance and animal test results demonstrated the safety and effectiveness of the CryoBalloon Ablation System.
## VIII. Conclusion
The CryoBalloon Ablation System was found to be substantially equivalent to the predicate devices.
The CryoBalloon Ablation System shares identical indications for use, similar design features, and functional features with, and thus is substantially equivalent to, the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
SEP 2 9 2010
C2 Therapeutics, Inc. % Mr. Peter Garcia-Meza President and CEO 303 Convention Way, Suite 1 Redwood City, California 94063
Re: K101825
Trade/Device Name: CyroBalloon Ablation System Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical unit and accessories Regulatory Class: Class II Product Code: GEH Dated: June 30, 2010 Received: July 01, 2010
Dear Mr. Garcia-Meza:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
:
{4}------------------------------------------------
Page 2 - Mr. Peter Garcia-Meza
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersQffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance,
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
CONFIDENTIAL
Indications for Use Statement
510(k) Number (if known): K10 | 82 ・
Device Name: CryoBalloon Ablation System
Indications for Use:
The C2 Therapeutics CryoBalloon Ablation System is intended to be used as a cryosurgical tool for the destruction of unwanted tissue in the field of general surgery, specifically for endoscopic applications.
| Prescription Use | ✓ |
|-----------------------------|---|
| (Part 21 CFR 801 Subpart D) | |
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|-----------------------------------------------------------|---------|
| Division of Surgical, Orthopedic, and Restorative Devices | |
| 510(k) Number | K10/825 |
Page 1 of 1C2 Therapeutics Premarket Notification 510(k) Submission for: CryoBalloon Ablation System
Section 4: Indications for Use - Page 4-2
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.