CRYOSPRAY ABLATION SYSTEM, MODEL CU 8407020500

K070893 · CSA Medical, Inc. · GEH · May 1, 2007 · General, Plastic Surgery

Device Facts

Record IDK070893
Device NameCRYOSPRAY ABLATION SYSTEM, MODEL CU 8407020500
ApplicantCSA Medical, Inc.
Product CodeGEH · General, Plastic Surgery
Decision DateMay 1, 2007
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4350
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The CryoSpray Ablation™ System is intended to be used as a cryosurgical tool for destruction of unwanted tissue in the field of general surgery, specifically for endoscopic applications.

Device Story

CryoSpray Ablation System uses liquid nitrogen stored in a tank; propelled through a cryocatheter to perform cryo-ablation. Device destroys unwanted tissue via application of extreme cold. Catheter placed via visual observation. Modified catheter features closed distal tip with side hole for radial cryogen spray; otherwise identical to straight-tip predicate. Used in endoscopic applications; operated by clinicians. Output is controlled cryogen spray; affects clinical decision-making by enabling targeted tissue destruction. Benefits patient through minimally invasive tissue ablation.

Clinical Evidence

Bench testing only.

Technological Characteristics

Cryosurgical unit using liquid nitrogen. Catheter is a single-lumen design with a modified closed distal tip and side hole for radial spray. Materials are identical to predicate K060555. No electronic or software components described; mechanical delivery system.

Indications for Use

Indicated for destruction of unwanted tissue in general surgery, specifically for endoscopic applications.

Regulatory Classification

Identification

(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Special 510(k) Submission CSA TM System Directional Spray Catheter CU8407020500 Image /page/0/Picture/1 description: The image contains a sequence of characters that appear to be handwritten. The characters are K, O, 7, O, 8, 9, and 7. The characters are written in a bold, cursive style. There is a line underneath the first four characters. MAY - 1 2007 ## 510(k) Safety Summary March 2007 Submitted by: CSA Medical, Inc. Emerging Technology Center 1101 E. 33rd, Third Floor - #A305 Baltimore, MD 21218 0: 443.921.8053 Contact Person: Jennifer Cartledge, VP Development CSA Medical, Inc. (Direct Contact Number 864.506.0097) Name of Device - Trade Name: . CryoSpray Ablation System CU8407020500 Directional Spray Catheter - Common Name: Cryosurgical Unit, Cryogenic Surgical Device ● - Classification: Cryosurgical unit with Liquid Nitrogen, Class II . [21 CFR & 878.4350(a)]. - Establishment Registration Number: . 9062377 Predicate Devices ### Device #### Premarket Notification SprayGenix™ Cryo Ablation System KOROSSS Company History: CSA Medical, formerly CryMed Technologies received market clearance for the SprayGenix™ Cryo Ablation System (K060555). The name for the SprayGenix™ Cryo Ablation System was changed to the CryoSpray Ablation™ System as documented in the letter to file dated March 12, 2007. Device Description: The CryoSpray Ablation™ System is used to destroy unwanted tissue by application of extreme cold to a selected site. Liquid Nitrogen is stored in a tank and then propelled through a cryocatheter to perform the cryo-ablation procedure. The catheter, an accessory component of the CSA System, is placed in the appropriate position through the use of visual observation. The cryo-catheter applies the cryogen to a selected area and freezes the unwanted tissue. The catheter detailed in the K060555 submission is of a single lumen, straight tip design. CSAM 05-00002-00 {2}------------------------------------------------ Description of Modification: The only device modification presented in this Special 510k submission involves a tip configuration modification for the cryocatheter. It was desirable to allow the end user's the ability to direct the cryogen radially. While the identical materials are used as compared to the catheter listed in submission K060555, the end of cryo-catheter is closed and a side hole is located in the distal side the catheter, allowing a radial spray of the cryogen. All other dimensional specifications are identical to the catheter detailed in submission K060555. The surface area of the radial hole of the catheter is similar to the hole in the lumen opening of the straight spraying catheter. Refer to Attachments 5 and 6. ### Indications for Use: The CryoSpray Ablation™ System is intended to be used as a cryosurgical too! for destruction of unwanted tissue in the field of general surgery, specifically for endoscopic applications. The intended use of the modified catheter, as described in its labeling, has not changed as a result of the tip configuration modification. Refer to Attachment 3. Technical Characteristics: The modification to the catheter does not change the operating principals or mechanism of action for the CryoSpray Ablation System and it is substantially equivalent to the above listed predicate devices. #### Summary: Based on the principles of operation, design, materials and intended use, the modification to the Straight Catheter to produce a Directional Catheter results in a CryoSpray Ablation™ System that is substantially equivalent to devices currently marketed in the United States. Declaration of Conformity: Refer to Attachment 2. {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name in a circular arrangement around a stylized graphic. The graphic consists of three abstract human profiles facing to the right, stacked on top of each other, with flowing lines suggesting movement or connection. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 CSA Medical, Inc. % Ms. Jennifer Cartledge Vice President, Development 1101 East 33td Street, Third Floor - #A305 Baltimore, Maryland 21218 MAY - 1 2007 Re: K070893 Trade/Device Name: CryoSpray Ablation™ System Regulation Number: 21 CFR 878.4350(a) Regulation Name: Cryosurgical unit and accessories Regulatory Class: II Product Code: GEH Dated: March 26, 2007 Received: April 3, 2007 Dear Ms. Cartledge: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ Page 2 -- Ms. Jennifer Cartledge This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely fours, Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Special 510(k) Submission CSA TM System Directional Spray Catheter CU8407020500 # Indications for Use 510(k) Number (if known): _(x_07 6893 DEVICE NAME: CryoSpray Ablation™ System CSA Medical, Inc. INDICATIONS FOR USE: The CryoSpray Ablation™ System is intended to be used as a cryosurgical tool for destruction of unwanted tissue in the field of general surgery, specifically for endoscopic applications. Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use _X_ | or Over-The-Counter Use _____ | |----------------------|-------------------------------| | (per 21 CFR 801.109 | | (Division Sign-Off) Division of General, Restorative, and Neurological Devices | 510(k) Number | 1090893 | |---------------|---------| |---------------|---------| CSAM 05-00002-00
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...