HISTOFREEZER WART REMOVAL SYSTEM OR OTHER PROPRIETARY NAME
K023487 · OraSure Technologies, Inc. · GEH · Mar 20, 2003 · General, Plastic Surgery
Device Facts
Record ID
K023487
Device Name
HISTOFREEZER WART REMOVAL SYSTEM OR OTHER PROPRIETARY NAME
Applicant
OraSure Technologies, Inc.
Product Code
GEH · General, Plastic Surgery
Decision Date
Mar 20, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Histofreezer® Wart Removal System is indicated for over-the-counter treatment of common warts and plantar warts.
Device Story
Histofreezer® Wart Removal System is a portable cryosurgical device for OTC treatment of common and plantar warts. System consists of a canister containing a liquid mixture of compressed gases (dimethyl ether, propane, and isobutane) and custom applicators. User saturates applicator with cryogen from canister and applies directly to wart. Device functions by freezing tissue to remove warts. Intended for patient self-use; labeling provides instructions for self-diagnosis and guidance on when to consult a physician. Benefits include convenient, non-prescription wart removal.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and labeling comparison to predicate devices.
Technological Characteristics
Portable cryosurgical system. Components: canister containing compressed gas mixture (dimethyl ether, propane, isobutane) and custom applicators. Manual application method. No electronic components or software.
Indications for Use
Indicated for OTC treatment of common and plantar warts in patients.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
{0}------------------------------------------------
## PREMARKET NOTIFICATION 510(k) SUMMARY (As Required By 21 CFR 807.92)
K023487
MAR 2 0 2003
807.92 (a):
| 1. | Submitter's Name: | OraSure Technologies, Inc. |
|----|-------------------|--------------------------------------|
| | Address: | 150 Webster St., Bethlehem, PA 18015 |
| | Telephone Number: | (610) 882-1820 |
| | Contact Person: | R. Sam Niedbala, Ph.D., BCFE |
| | Date Prepared: | October 15, 2002 |
- 2. Device Name: Histofreezer® Wart Removal System or other proprietary name Proprietary Name: Usual Name: OTC wart removal system Classification Name: Cryosurgical unit, accessories
3. Device to Which Substantial Equivalence Is Claimed: Histofreezer® Device, by OraSure Technologies, Inc. (Primary Predicate) - K911420, K924114, K971392 Wartner Wart Removal System, by Wartner Medical Products (OTC claim) - K011708 Compound W Gel, by Medtech (labeling only) Clear Away Liquid, by Schering-Plough HealthCare Products, Inc. (labeling only) Clear Away One Step for Kids, by Schering-Plough HealthCare Products, Inc. (labeling only) Rite Aid Wart Liquid, by Rite Aid Corporation (labeling only) Suave Hairspray, by Helene Curtis (flammability label only)
- Description of Device: 4.
The Histofreezer® Wart Removal System is a cryosurgical system for the treatment of warts. It consists of:
- A canister filled with a liguid mixture of the compressed gases dimethyl ether, propane 0 and isobutane
- o Custom applicators
- O An illustrated description of how to use the product
- న. Intended Use Statement:
The Histofreezer® Wart Removal System is indicated for over-the-counter treatment of common warts and plantar warts.
- Comparison of Technological Characteristics: 6.
The Histofreezer® Wart Removal System for over-the-counter treatment of common and plantar warts is substantially equivalent to the Histofreezer® device for the same indications (for prescription use). Both devices are portable cryosurgical systems comprised of a canister containing cryogen and an applicator that is saturated with cryogen and then applied to the wart to be treated. The Histofreezer® Wart Removal System, however, is intended for OTC treatment of common warts and plantar warts.
{1}------------------------------------------------
The labeling of the Histofreezer® Wart Removal System has been developed to ensure the consumer has adequate directions for use and for safety. Histofreezer® labeling has also been developed to provide adequate information for the consumer to make a self diagnosis and to ensure that they contact their doctor if in any doubt, if stinging or aching persists after treatment, or if the wart does not improve after four treatments.
The safety and warning statements for the OTC predicate device (Wartner Wart Removal System) and for all of the labeling predicate devices is essentially similar.
### 7. Conclusion:
Based on the information presented above, it is concluded that the proposed Histofreezer® Wart Removal System is safe and effective for its intended use and is substantially equivalent to the primary predicate device. It is also substantially equivalent in intended use, safety, and labeling to the labeling predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three wavy lines, which are meant to represent the human services aspect of the department.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# APR 2 5 2003
R. Sam Niedbala, PH.D., BCFE Orasure Technologies, Inc. 150 Webster Street Bethlehem, Pennsylvania 18015-1389
Re: K023487
Trade/Device Name: Histofreezer® Wart Removal System Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical unit and accessories Regulatory Class: II Product Code: GEH Dated: February 5, 2003 Received: February 19, 2003
Dear Dr. Neidbala:
This letter corrects our substantially equivalent letter of March 20, 2003, regarding the intended use of the device mentioned above. The intended use for the device was over-the-counter, however, the indications for use form indicated prescription use.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent [(for the indications for use stated in the enclosure)] to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
### Page 2 - R. Sam Niedbala, PH.D., BCFE
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at their toll free number (800) 638-2041 or at (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Muriam C. Provost
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### STATEMENT OF INDICATIONS FOR USE
510(k) Number (if known): _ K023487
Device Name: Histofreezer® Wart Removal System or other proprietary name
Indications For Use: The Histofreezer® Wart Removal System is indicated for over-the-counter treatment of common and plantar warts.
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use
c:\fda\indica.doc
Miriam C. Provost
(Division Sign-Off) Division of General, Restorative and Neurological Devices
510(k) Number Kc23487
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.