OptiFix AT Absorbable Fixation System with Articulating Technology - 30 Fasteners, OptiFix AT Absorbable Fixation System with Articulating Technology - 15 Fasteners

K170278 · C.R. Bard, Inc. · GDW · May 31, 2017 · General, Plastic Surgery

Device Facts

Record IDK170278
Device NameOptiFix AT Absorbable Fixation System with Articulating Technology - 30 Fasteners, OptiFix AT Absorbable Fixation System with Articulating Technology - 15 Fasteners
ApplicantC.R. Bard, Inc.
Product CodeGDW · General, Plastic Surgery
Decision DateMay 31, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4750
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The OptiFix AT Absorbable Fixation System with Articulating Technology is indicated for the approximation of soft tissue and fixation of surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair.

Device Story

OptiFix AT is a sterile, single-use, mechanical fixation system for hernia repair; used in open or laparoscopic surgery by surgeons. Device consists of a pistol-grip deployment handle and a shaft containing preloaded absorbable fasteners (15 or 30 count). Key features include 360-degree handle rotation and an articulating shaft tip to improve access to surgical sites. Operation involves manual mechanical actuation to drive fasteners through mesh into tissue. Fasteners are composed of Poly(L-lactide-co-glycolide) and provide long-term fixation before absorbing. Device facilitates mesh placement, potentially reducing surgical time and improving fixation precision in confined anatomical spaces.

Clinical Evidence

No clinical studies performed. Safety and feasibility supported by non-clinical bench testing (actuation torque, deployment, fastener gap height, ball burst, mesh compatibility, in-vitro degradation) and an animal study comparing the device to the secondary predicate (AbsorbaTack). Animal study evaluated mesh contracture, fastener seating, tissue in-growth, and inflammatory response. Biocompatibility testing conducted per ISO 10993-1.

Technological Characteristics

Mechanical deployment system for absorbable fasteners. Fasteners made of Poly(L-lactide-co-glycolide) dyed with D & C Violet No. 2. Shaft length 37cm, compatible with 5mm trocars. Features 360-degree handle rotation and articulating shaft tip. Sterilized via gamma irradiation. No electrical or software components.

Indications for Use

Indicated for soft tissue approximation and surgical mesh fixation during open or laparoscopic procedures, including hernia repair, in patients requiring such surgical intervention.

Regulatory Classification

Identification

An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized human figure. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 31, 2017 C.R. Bard Incorporated Ms. Katherine Earle Regulatory Affairs Specialist 100 Crossings Boulevard Warwick, Rhode Island 02886 Re: K170278 Trade/Device Name: Optifix AT Absorbable Fixation System With Articulating Technology - 30 Fasteners, Optifix AT Absorbable Fixation System With Articulating Technology - 15 Fasteners Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable Staple Regulatory Class: Class II Product Code: GDW Dated: April 28, 2017 Received: May 1, 2017 Dear Ms. Earle: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply {1}------------------------------------------------ with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # Jennifer R. Stevenson -S For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K170278 #### Device Name OptiFix AT Absorbable Fixation System with Articulating Technology - 30 Fasteners, OptiFix AT Absorbable Fixation System with Articulating Technology - 15 Fasteners #### Indications for Use (Describe) The OptiFix AT Absorbable Fixation System with Articulating Technology is indicated for the approximation of soft tissue and fixation of surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair. | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------------------------------|-------------------------------------------------------------------------| | <span></span> <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### 510(k) Summary I. SUBMITTER Davol Inc. 100 Crossings Boulevard Warwick, RI 02886 | Contact Person: | Katherine Earle | |-----------------|-------------------------------| | | Regulatory Affairs Specialist | | Phone: | (401) 825-8450 | | Fax: | (401) 825-8765 | | E-mail: | | katherine.earle@crbard.com Date Prepared: May 30, 2017 #### II. DEVICE | 510(k) Number: | K170278 | |-----------------------|----------------------------------------------------------------------------------| | Name of Device: | OptiFix <sup>TM</sup> AT Absorbable Fixation System with Articulating Technology | | Common or Usual Name: | Implantable Staple | | Classification Name: | Implantable Staple (21 CFR §878.4750) | | Regulatory Class: | II | | Product Code: | GDW | #### III. PREDICATE DEVICE The primary predicate device for this submission is the OptiFix" Absorbable Fixation System (K142873), cleared March 12, 2015; marketed by Davol, Inc. The primary predicate, hereinafter referred to as "the predicate" throughout the remainder of the submission, has not been subject to a design-related recall. The secondary predicate device for this submission is the Medtronic Covidien AbsorbaTack™ Absorbable Fastener System (K123109), cleared January 27, 2012. {4}------------------------------------------------ #### IV. DEVICE DESCRIPTION The OptiFix™ AT Absorbable Fixation System with Articulating Technology, hereinafter referred to as "OptiFix ""AT", is a sterile (via gamma) single use device that is comprised of a deployment component and an absorbable fastener component. Two different product ordering codes are to be packaged for distribution; each contains the same ergonomically designed deployment device. The variation will be the number of preloaded fasteners; either 15 or 30. The deployment shaft of the OptiFix™ AT device is 37cm in length and designed for use with 5mm trocars. The tip of the shaft can be articulated and the handle of the device can be rotated 360 degrees to facilitate access for fixation during surgery. The fasteners are designed with retention features and are manufactured from Poly (L-lactide-co-glycolide) and are dyed with D & C Violet No. 2 (<0.15% by weight). # V. INDICATIONS FOR USE The OptiFix™ AT Absorbable Fixation System with Articulating Technology is indicated for the approximation of soft tissue and fixation of surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair. The Indications for Use statement for the subject device is identical to that of the primary predicate device (OptiFix "-K142873), and similar to that of the secondary predicate device (AbsorbaTack"-K123109). The proposed and predicate devices are all intended for the fixation of surgical mesh/prosthetic to tissue during open or laparoscopic surgical procedures, such as hernia repair. # VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES The fixation of surgical mesh/prosthetic to tissue during open or laparoscopic surgical procedures, such as hernia repair, using a mechanical actuation force to guide and launch the fastener implant is the operating principle for the proposed and predicate devices. The proposed and predicate devices are based on the following similar elements: - Deployment component - mechanical component used to position and launch the fastener (implant) into tissue - Fastener component -absorbable component composed of well-characterized absorbable polymer, used to fixate surgical mesh to tissue. - O The fastener shape/design and dimensions of the subject device are similar to that of the primary predicate device (OptiFix" -K142873). - The fastener material for the subject device (Poly(1-lactide-co-glycolide)) is o identical to that of the secondary predicate device (AbsorbaTack" -K123109). {5}------------------------------------------------ The following differences exist between the proposed and predicate devices: - The handle configuration of the proposed device can be rotated 360° and the deployment ● shaft articulated to facilitate fixation, unlike the predicate devices which have no articulation in the deployment shaft and a static handle. - The deployment shaft of the proposed device is shorter in length (37 cm) compared to the deployment shaft of the primary predicate device (39cm). - The proposed absorbable fastener has minor geometric differences and is composed of a different material (Poly(L-lactide-co-glycolide)) than the primary predicate absorbable fastener (Poly(D,L)-lactide). ## VII. PERFORMANCE DATA The following performance data is provided in support of the substantial equivalence determination. #### Performance standards No performance standards have been established for this device under Section 514 of the Federal Food, Drug, and Cosmetic Act. ## Biocompatibility testing The biocompatibility evaluation for the OptiFix™ AT device was conducted in accordance with FDA Guidance, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". June 16, 2016; and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The proposed device is supported by the biocompatibility testing listed below. - Cytotoxicity - Sensitization ● - Intracutaneous (Irritation) - Systemic Acute Toxicity - Sub-Chronic Toxicity (13 weeks) - Chronic Toxicity (26 weeks) - Genotoxicity - Mouse Peripheral Blood Micronucleus Study - Genotoxicity - Bacterial Reverse Mutation Study - Genotoxicity - Mouse Lymphoma Assay - Implantation - 4, 8, 12, 26, 39, 52, 77 weeks - Pyrogenicity All samples tested met the acceptance criteria and the proposed OptiFix™ AT device is biocompatible for its intended use. {6}------------------------------------------------ ## Electrical safety and electromagnetic compatibility (EMC) The proposed OptiFix™ AT device is not an electro-mechanical medical device nor is it a medical system, therefore this section does not apply. # Software Verification and Validation Testing The proposed OptiFix " AT device does not contain software; therefore this section does not apply. # Product Testing The following non-clinical tests were completed for the proposed and primary predicate device. The proposed OptiFix™ AT device passed all the test requirements and demonstrated substantial equivalence to the test results of the primary predicate device. - · Performance and Functional testing of the OptiFix™ AT device - Actuation (trigger) Torque ● - Fastener Deployment ● - Fastener Gap Height - Ball Burst Testing - · Mesh compatibility testing of the OptiFix " AT device - · Resorption Profile of the fastener - In-Vitro Degradation ● - In-Vitro to In-Vivo Correlation All samples tested met the established acceptance criteria. # Animal Study The safety and feasibility of the OptiFix™ AT device was evaluated by animal and histological evaluation of the tissue in the treatment fixated areas. The evaluation used the secondary predicate, AbsorbaTack™ (K123109), as a control since the fastener is composed of the same material, Poly (L-lactide-co-glycolide) in an 82%/18% molar ratio, and has the same intended use as the OptiFix" AT device. The study demonstrated that the OptiFix AT device can safely, approximate soft tissue and fixate surgical mesh to tissues during open or laparoscopic surgical procedures, such as hernia repair. The following endpoints were evaluated: - Mesh contracture - · Fastener seating properties - · Tissue in-growth (T-peel analysis) - · Host inflammatory/fibrotic response # Clinical Studies Clinical studies were not performed for the submission of this proposed device nor were clinical studies performed for the primary predicate device (OptiFix-K142873). {7}------------------------------------------------ ## VIII. CONCLUSIONS The proposed OptiFix™ AT device is substantially equivalent to the legally marketed primary predicate device for the following reasons: - A) The same intended use as the primary predicate device. - B) Similar technological characteristics as the primary predicate device such as: Fastener shape/design, trigger handle, penetration depth, and shaft length. The differences in technology do not adversely impact the safety and performance of the device. - C) Identical principle of operation to deliver fasteners perpendicular to the tissue plane. The comparative analysis, see Table 5-1 below, as well as the bench and preclinical testing results demonstrate that the proposed OptiFix™ AT device performs as intended and is substantially equivalent to the primary predicate device that is currently marketed for the same intended use. | Device Features | Proposed device<br>OptiFix™ AT | Predicate device<br>OptiFix™ (K142873) | |------------------------------------------|-------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | Identical to predicate | Soft tissue fixation | | Indication For<br>Use | Identical to predicate | Indicated for the approximation of<br>soft tissue and fixation of surgical<br>mesh to tissues during open or<br>laparoscopic surgical procedures,<br>such as hernia repair. | | Fastener Material | Resorbable Polymer<br>Poly(L-lactide-co-glycolide)<br>(PLG) | Resorbable Polymer<br>Poly (D,L) Lactide | | Fastener Violet<br>Dye | Identical to predicate | D & C Violet No. 2<br>(<0.15% by weight)<br>conforms to 21 CFR §74.1602 | | Fastener Body<br>Contact | Identical to predicate | Long term implant (>30 days)<br>contacting tissue and/or bone | | Fastener<br>Shape/Design | Push Tack with retention features<br>on end; nested tip<br>Image: [purple fastener] | Push Tack with retention features<br>on end<br>Image: [purple fastener] | | Device Features | Proposed device<br>OptiFix™ AT | Predicate device<br>OptiFix™ (K142873) | | Fastener<br>Dimensions | 7.1 mm overall fastener length<br>Fastener head: 3.5 mm diameter | 6.7 mm overall fastener length<br>Fastener head: 3.7 mm diameter | | Fastener<br>Manufacturing<br>Method | Identical to predicate | Injection-Molded | | Fastener<br>Absorption Time | Identical to predicate | 360 days | | Fastener Quantity<br>per Device | Identical to predicate | 15 & 30 fasteners | | Fastener Delivery<br>System | Identical to predicate | Push – Impact tube pushes<br>fasteners forward over a<br>guidewire | | Deployment<br>Component<br>Handle design | Pistol/Gun shape with 360 degree<br>rotation option | Pistol/Gun shape with no rotation | | Deployment<br>Component Shaft | 37 cm in length, 6 cm of<br>articulation | 39 cm in length, no articulation | | Deployment<br>Component<br>Configuration | Image: OptiFix AT Deployment Component Configuration | Image: OptiFix (K142873) Deployment Component Configuration | | Deployment<br>Component Body<br>Contact | Identical to predicate | Transient use - Less than 24 hour<br>duration | | Device<br>Sterilization | Identical to predicate | Gamma Irradiation<br>(25 - 40 kGy) | | Device<br>Packaging | Identical to predicate | Device is placed into an SBS<br>cardboard insert which is inserted<br>into a foil pouch, sealed and then<br>placed into an SBS paperboard<br>printed shelf carton. | Table 5-1: Device Substantial Equivalence - General Characteristics {8}------------------------------------------------
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...