ECHELON FLEX 45mm Powered Plus Articulating Endoscopic Linear Cutters, ECHELON ENDOPATH Endoscopic Linear Cutter Reloads, 45mm
K163454 · Ethicon Endo-Surgery, LLC · GDW · Jan 5, 2017 · General, Plastic Surgery
Device Facts
Record ID
K163454
Device Name
ECHELON FLEX 45mm Powered Plus Articulating Endoscopic Linear Cutters, ECHELON ENDOPATH Endoscopic Linear Cutter Reloads, 45mm
Applicant
Ethicon Endo-Surgery, LLC
Product Code
GDW · General, Plastic Surgery
Decision Date
Jan 5, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
ECHELON, ECHELON ENDOPATH™ and ECHELON FLEX families of endoscopic linear cutters and reloads are intended for transection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general, gynecologic, urologic, thoracic, and pediatric surgical procedures. They can be used with staple line or tissue buttressing materials. The instruments may also be used for transection and resection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney and spleen.
Device Story
Powered endoscopic linear cutter; performs simultaneous tissue cutting and stapling. Device consists of articulating shaft, handle with primary lithium battery, and interchangeable reload cartridges. Reloads contain six staggered rows of staples (three per side of cut line). Operator (surgeon) inserts reload, positions device at operative site, articulates shaft for lateral access, and fires instrument. Lock-out mechanism prevents firing without reload or with used reload. Output is stapled, transected tissue. Used in OR for general, gynecologic, urologic, thoracic, and pediatric procedures. Benefits include consistent staple line formation and tissue management during complex resections.
Clinical Evidence
Bench testing only. Verification and validation activities included testing for formed staple height, staple form quality, staple line integrity, staple line reinforcement compatibility, and system reliability. Risk analysis performed per EN ISO 14971:2012.
Technological Characteristics
Powered endoscopic linear cutter; 45mm staple line; 42mm cut line. Features articulating shaft, primary lithium battery power source, and six-row staple configuration. Reloads include gripping surface technology and color-coded staple heights for varying tissue thicknesses. Sterile, single-patient use. Design controls and risk management per EN ISO 14971:2012.
Indications for Use
Indicated for patients undergoing open or minimally invasive general, gynecologic, urologic, thoracic, or pediatric surgical procedures requiring tissue transection or anastomosis, including resection of liver parenchyma, pancreas, kidney, and spleen.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
ECHELON FLEX™ 60mm Powered Plus Articulating Endoscopic Linear Cutters (K140560, K160521)
ECHELON ENDOPATH™ Endoscopic Linear Cutter Reloads, 60mm (K140560)
Reference Devices
Echelon Endoscopic Linear Cutter Reload, Black (K131663, K112056)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, featuring a series of overlapping human profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 5, 2017
Ethicon Endo-surgery, LLC % Ms. Asifa Vonhof Ethicon Endo Surgery, Inc. 4545 Creek Road Cincinnati, Ohio 45242
Re: K163454
Trade/Device Name: Echelon Flex 45mm Powered Plus Articulating Endoscopic Linear Cutters, Echelon Endopath Endoscopic Linear Cutter Reloads, 45mm Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable Staple Regulatory Class: Class II Product Code: GDW Dated: December 7, 2016 Received: December 9, 2016
Dear Ms. Vonhof:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known)
K163454
#### Device Name
De nee Name
The ECHELON ENDOPATH™ and ECHELON FLEX families of endoscopic linear cutters and reloads
#### Indications for Use (Describe)
Indications of OSS (Desonib)
ECHELON, ECHELON ENDOPATH™ and ECHELON FLEX families of endoscopic linear cutters and reloads are intended for ECHEDON, ENDONY LINDON OF CRES and DEASE The instruments have application in multiple open or minimally invasive general, gynecologic, urologic, thoracies and procedures. They can be used with staple line or tissue buttressing gencials, The instruments may also be used for transection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney and spleen.
Type of Use (Select one or both, as applicable)
[x] Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
# A CONTRACT A CONTRACT FOR FOR FOR FOR FOR FOR FOR FOR
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
{3}------------------------------------------------
## 510(k) Summary
I. SUBMITTER
Ethicon Endo-Surgery, LLC 475 Calle C Guaynabo, PR 00969
Phone: 513-337-3118 Fax: 513-337-1122
Contact Person: Ms. Asifa Vonhof, MS, RAC Date Prepared: January 5, 2017
## II. DEVICE
- ECHELON FLEX™ 45mm Powered Plus Articulating Endoscopic Linear Cutters ●
- ECHELON ENDOPATH™ Endoscopic Linear Cutter Reloads, 45mm (+ Gripping ● Surface Technology)
| | Common or Usual Name: | Surgical Stapler with Implantable Staples |
|----------------------|---------------------------------------|-------------------------------------------|
| Classification Name: | Staple, implantable (21 CFR 878.4750) | |
| Regulatory Class: | II | |
| Product Code: | GDW | |
#### III. PREDICATE DEVICES
- ECHELON FLEX™ 60mm Powered Plus Articulating Endoscopic Linear Cutters . (K140560, K160521)
- ECHELON ENDOPATH™ Endoscopic Linear Cutter Reloads, 60mm (+ Gripping . Surface Technology) (K140560)
Reference device Echelon Endoscopic Linear Cutter Reload, Black (K131663, K112056)
{4}------------------------------------------------
#### IV. DEVICE DESCRIPTION
The ECHELON, ECHELON ENDOPATH and ECHELON FLEX families of endoscopic linear cutters and reloads are sterile, single patient use instruments that simultaneously cut and staple tissue. There are six staggered rows of staples, three on either side of the cut line. The ECHELON FLEX™ 45 mm Powered Plus instruments have a staple line that is approximately 45 mm long and a cut line that is approximately 42 mm long. The shaft can rotate freely in both directions and an articulation mechanism enables the distal portion of the shaft to pivot to facilitate lateral access to the operative site.
The instruments are packaged with a primary lithium battery pack that must be installed prior to use. There are specific requirements for disposing of the battery pack. Refer to the Battery Pack Disposal section.
The instruments are packaged without a reload and must be loaded prior to use. A staple retaining cap on the reload protects the staple leg points during shipping and transportation. The instruments' lock-out feature is designed to prevent a used or improperly installed reload from being refired or an instrument from being fired without a reload.
The ECHELON ENDOPATH™ Echelon Endoscopic Linear Cutter Reloads are sterile, single patient use devices. They come in five colors corresponding to different closed staple heights, with the intention of being used in different tissue thicknesses.
They are loaded into a Linear Cutter and deliver staples into the tissue when the instrument is fired. There are 70 staples in the 45mm length reload.
#### V. INDICATIONS FOR USE
ECHELON, ECHELON ENDOPATH™ and ECHELON FLEX families of endoscopic linear cutters and reloads are intended for transection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general, gynecologic, urologic, thoracic, and pediatric surgical procedures. They can be used with staple line or tissue buttressing materials. The instruments may also be used for transection and resection of liver parenchyma (hepatic vasculature and biliary structures), pancreas, kidney and spleen.
{5}------------------------------------------------
## VI. COMPARISON OF CHARACTERISTICS WITH THE PREDICATE DEVICES
Surgical stapling is the technological principle for both the subject and predicate device. It is based on the use of endoscopic instrumentation for transection, and/or creation of anastomoses.
The subject and predicate staplers have the following identical features:
- intended use
- indications
- contraindications
- technological characteristics
- design ●
- materials
- operational principles ●
The following differences exist between the subject and predicate staplers:
- Anvil length (45mm vs 60mm)
- Cut-line length (42mm vs 57mm) .
The subject and predicate reloads have the following identical features:
- intended use ●
- indications ●
- contraindications
- technological characteristics ●
- materials
- operational principles
- MR compatibility ●
The following differences exist between the subject and predicate reloads (marketed):
- Length of reload (45mm vs 60mm) ●
- . Number of staples (70 vs 88, to accommodate 45mm reload length)
- Sled component of white 45mm reload differs in design from the white 60mm reload ●
- . Wire diameter of the 45mm blue reload differs from the wire diameter of the 60mm blue reload
{6}------------------------------------------------
## VII. PERFORMANCE DATA
Signed Declarations of Conformity with Design Controls for verification and validation activities and the manufacturing facility are provided.
Risk analyses for each device modification are provided, according to EN ISO 14971:2012, Medical Devices - Application of Risk Management to Medical Devices.
During verification and validation testing, all acceptance criteria were met.
Non-Clinical Tests used to demonstrate substantial equivalence:
- Formed Staple Height ●
- . Staple Form Quality
- Staple Line Integrity
- Staple Line Reinforcement Compatibility ●
- System Reliability.
#### VIII. CONCLUSIONS
The risk profile of the device has not changed as a result of the described changes; furthermore, the performance of the modified device is consistent with the predicate device and does not raise any new questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.