The RAM® COR-SUTURE® QUICK LOAD® surgical suture is indicated for use in the approximation of soft tissue and prosthetic materials. The COR-SUTURE® OUICK LOAD® surgical suture is indicated for use in the approximation of soft tissue and prosthetic materials.
Device Story
Non-absorbable surgical sutures (PTFE-coated braided polyester or uncoated monofilament polypropylene) used for soft tissue and prosthetic material approximation; includes needle caps at suture ends to facilitate use with LSI SOLUTIONS suture placement devices; RAM version includes needle cap holder for optional PTFE pledget; pledget provides buttressing and mechanical support in fragile tissue; single-patient use; sterile; EO-sterilized; MR safe; needle caps discarded post-procedure; suture and pledget are the only implantable components.
Clinical Evidence
No clinical data. Bench testing performed per USP Monograph for Non-Absorbable Sutures and biocompatibility evaluation conducted per ANSI AAMI ISO 10993-1:2009/(R)2013.
Technological Characteristics
Non-absorbable surgical suture; materials: PTFE-coated braided polyester or uncoated monofilament polypropylene; USP sizes 2-0 and 3-0; lengths 38-53 inches; includes surgical stainless-steel needle caps; EO sterilization; MR safe; non-metallic/non-conducting implantable components.
Indications for Use
Indicated for approximation of soft tissue and prosthetic materials in patients requiring surgical suture placement.
Regulatory Classification
Identification
Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol is a stylized representation of a human figure, while the FDA name is written in blue and includes the words "U.S. Food & Drug Administration".
November 24, 2020
LSI Solutions Christopher Miller Executive Director of Regulatory Affairs and Quality 7796 Victor-Mendon Rd. Victor, New York 14564
Re: K203081
Trade/Device Name: RAM COR-SUTURE OUCIK LOAD Surgical Suture COR-SUTURE QUICK LOAD Surgical Suture Regulation Number: 21 CFR 878.5010, 21 CFR 878.5000 Regulation Name: Nonabsorbable Polypropylene Surgical Suture, Nonabsorbable Poly(ethylene terephthalate) Surgical Suture Regulatory Class: Class II Product Code: GAW, GAS Dated: October 10, 2020 Received: October 13, 2020
Dear Mr. Miller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Cindy Chowdhury, Ph.D., MBA Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K203081
#### Device Name
RAM® COR-SUTURE® QUICK LOAD® Surgical Suture; COR-SUTURE® QUICK LOAD® Surgical Suture
Indications for Use (Describe)
The RAM® COR-SUTURE® QUICK LOAD® surgical suture is indicated for use in the approximation of soft tissue and prosthetic materials.
The COR-SUTURE® OUICK LOAD® surgical suture is indicated for use in the approximation of soft tissue and prosthetic materials.
Type of Use (Select one or both, as applicable):
| <div> <span> <svg height="16" width="16"> <polygon points="1 0, 16 0, 16 15, 1 15" style="fill:none;stroke:black;stroke-width:1"></polygon> <polyline points="2 7, 6 12, 14 3" style="fill:none;stroke:black;stroke-width:2"></polyline> </svg> Prescription Use (Part 201 CFR Subpart D) </span> </div> |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg height="16" width="16"> <polygon points="1 0, 16 0, 16 15, 1 15" style="fill:none;stroke:black;stroke-width:1"></polygon> </svg> Over-The-Counter Use (21 CFR 201 Subpart C) </span> </div> |
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K203081 – Special 510(k) Summary
| Submitted By: | LSI SOLUTIONS, Inc.<br>7796 Victor-Mendon Road<br>Victor, NY 14564 |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Christopher B. Miller<br>Executive Director of Regulatory Affairs and Quality<br>Phone: (585) 869-6665<br>Fax: (585) 742-8086<br>Email: cmiller@lsisolutions.com |
| Date Prepared: | October 10, 2020 |
| Trade Name(s): | RAM® COR-SUTURE® QUICK LOAD®<br>COR-SUTURE® QUICK LOAD® |
| Common Name: | Nonabsorbable Surgical Suture |
| Classification Names: | Nonabsorbable poly(ethylene terephthalate) surgical suture<br>Nonabsorbable polypropylene surgical suture |
| Classification Regulations: | 21 CFR 878.5000 (Nonabsorbable poly(ethylene terephthalate) surgical suture)<br>21 CFR 878.5010 (Nonabsorbable polypropylene surgical suture) |
| Product Codes: | GAS<br>GAW |
| Device Classification: | Class II |
{4}------------------------------------------------
## Predicate Device
The predicate device for establishing substantial equivalence is listed in Table 1, below. This predicate has not been subject to any type of recall.
| Table 1: Predicate Device for Subject RAM® COR-SUTURE® QUICK LOAD® and COR-<br>SUTURE® QUICK LOAD | | | |
|---------------------------------------------------------------------------------------------------|---------------------|---------|-----------------------|
| Predicate Device | Manufacturer | 510(k) | Clearance Date |
| Pledgeted COR-SUTURE™<br>QUICK LOAD® | LSI SOLUTIONS, Inc. | K160529 | September 15,<br>2016 |
## Device Description
LSI SOLUTIONS® offers both the RAM® COR-SUTURE® QUICK LOAD® surgical suture, which is a non-absorbable polytetrafluoroethylene (PTFE)-coated, braided polyester surgical suture, and the COR-SUTURE® QUICK LOAD® surgical suture, which is available as either a PTFE-coated, braided polyester surgical suture or a monofilament, uncoated polypropylene surgical suture. The RAM® COR-SUTURE® QUICK LOAD® surgical suture is also supplied with an optional pre-attached PTFE pledget that is used to aid in suture buttressing and to mechanically secure and support the suture in fragile tissue, increasing the suture holding strength. A short length of modified surgical stainless-steel tubing, called a "needle cap", is attached to each end of the suture for both the RAM® COR-SUTURE® QUICK LOAD® surgical suture and the COR-SUTURE® QUICK LOAD® surgical suture. Relative to the COR-SUTURE® QUICK LOAD® device, the RAM® COR-SUTURE® QUICK LOAD® device additionally includes a needle cap holder to secure the optional pledget and needle caps in the suture tray. Both the RAM® COR-SUTURE® QUICK LOAD® surgical suture and the COR-SUTURE® QUICK LOAD® surgical suture also include a detachable clear suture tube to keep the suture from tangling. The needle caps are removed from the sutures and discarded once suturing is completed as they are not intended for implantation in the patient; only the suture and pledget are implantable of components.
Both devices are packaged for single patient use and are provided sterile by means of a validated ethylene oxide (EO) cycle. The polyester RAM® COR-SUTURE® QUICK LOAD® surgical suture is offered undyed (white), dyed green with the FDA approved colorant D&C Green No. 6. or striped green/white; all of which are offered with a surgical suture diameter of USP size 2-0 and a length of 38 inches. The polyester COR-SUTURE® QUICK LOAD® surgical suture is available
{5}------------------------------------------------
undyed (white) or dyed green with the FDA approved colorant D&C Green No. 6 and will be offered with a surgical suture diameter of USP size 2-0 and a length of 38 inches. The polypropylene COR-SUTURE® QUICK LOAD® surgical suture is dyed blue with the FDA approved colorant [phthalocyaninato(2-)] copper and will be offered with a surgical suture diameter of USP size 3-0 and a length of 53 inches. For both polypropylene and polyester suture, there is no known significant change in tensile strength retention to occur in vivo. The RAM® COR-SUTURE® QUICK LOAD® and COR-SUTURE® QUICK LOAD® surgical sutures, with needle caps removed, contains implantable components which are considered MR safe. The implantable polyester and polypropylene sutures and PTFE pledget material are non-metallic and non-conducting materials and therefore, considered MR safe.
## Intended Use
The intended use remains the same as the predicate device, and both the subject RAM® COR-SUTURE® QUICK LOAD® and COR-SUTURE® QUICK LOAD® surgical sutures are intended for the approximation of soft tissue and prosthetic materials.
The indications for use remains the same as that of the predicate device, and both the subject RAM® COR-SUTURE® QUICK LOAD® and COR-SUTURE® QUICK LOAD® surgical sutures are indicated for use in the approximation of soft tissue and prosthetic materials.
## Technological Characteristics (comparison to Predicate Device)
The subject RAM® COR-SUTURE® QUICK LOAD® and COR-SUTURE® QUICK LOAD® surgical sutures are substantially equivalent in intended use and fundamental technological characteristics to the predicate device. Both the subject and predicate devices are non-absorbable surgical suture available as either uncoated monofilament polypropylene (diameter of USP size 3-0) or PTFE-coated, braided polyester suture (diameter of USP size 2-0). Both the subject and predicate devices have a needle cap attached to each end of the suture component to facilitate use with LSI SOLUTIONS® suture placement devices. The only significant difference between the subject and predicate devices is that the subject devices utilize different raw material sources for the polyester fibers and PTFE coating of the OEM PTFE-coated, braided polyester surgical suture and a different polypropylene resin source for the OEM uncoated polypropylene suture supplied
{6}------------------------------------------------
directly from the manufacturer, Teleflex® Medical OEM. The chemical composition of these raw materials, however, remain the same.
The subject RAM® COR-SUTURE® QUICK LOAD® and COR-SUTURE® QUICK LOAD® surgical sutures are packaged, sterilized, and labeled in a manner substantially equivalent to the predicate device, including labeling claims such as indications, contraindications, warnings, and precautions. The subject devices are considered substantially equivalent and have the same technological characteristics to the predicate device through comparison in design, intended use, material chemical composition, function, and size range.
## Non-Clinical Performance Testing Summary
### Benchtop Testing
As recommended by the FDA Guidance for Industry, Class II Special Control Guidance Document: Surgical Sutures; Guidance for Industry and FDA, June 3, 2003, the subject RAM® COR-SUTURE® QUICK LOAD® and COR-SUTURE® QUICK LOAD® surgical sutures were subject to the requirements of the United States Pharmacopeia (USP), Monograph for Non-Absorbable Sutures. Performance testing focused on sections from the USP General Chapter on Physical Tests.
These were the same requirements and test methods used in the assessment of the predicate devices. The subject RAM® COR-SUTURE® QUICK LOAD® and COR-SUTURE® QUICK LOAD® surgical sutures continue to conform to these requirements.
## Biocompatibility Testing
Biocompatibility evaluation was conducted in accordance with ANSI AAMI ISO 10993-1:2009/(R)2013 Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process and the FDA Guidance for Industry, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" Attachment A: Evaluation Endpoints for Consideration, September 4,2020, to demonstrate that the RAM® COR-SUTURE® QUICK LOAD® and COR-SUTURE® QUICK LOAD® surgical sutures' patient contacting materials are substantially equivalent to the predicate
{7}------------------------------------------------
device. The RAM® COR-SUTURE® QUICK LOAD® and COR-SUTURE® QUICK LOAD® surgical sutures are classified as permanent contact duration devices that will be implanted and will have contact with blood, which is identical to the classification of the predicate device. Biocompatibility results from Teleflex® Medical OEM were leveraged and all appropriate biological endpoints for consideration per ISO 10993-1 were assessed with passing results.
## Substantial Equivalence
We believe that the predicate and the subject devices have been demonstrated to be substantially equivalent. Although the raw material sources used in the OEM PTFE-coated, braided polyester surgical suture and the OEM uncoated polypropylene surgical suture have changed, their material chemical composition and design remain the same. The indications for use and fundamental scientific technology remains the same. Both the subject and the predicate devices are indicated for use in the "approximation of soft tissue and prosthetic materials", and there are no changes to the devices' chemical composition of raw materials. Biocompatibility and benchtop performance testing were conducted to evaluate all risks associated with the raw material change, and a risk analysis demonstrates there are no new risks and no negative impacts to the safety and effectiveness of the surgical suture. Therefore, the subject RAM® COR-SUTURE® QUICK LOAD® and the COR-SUTURE® QUICK LOAD® surgical sutures are substantially equivalent to the predicate LSI SOLUTIONS® Pledgeted COR-SUTURE® QUICK LOAD® suture, which was cleared in 510(k) Premarket Notification K160529 on September 15, 2016.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.