Pledgeted COR-SUTURE QUICK LOAD

K160529 · Lsi Soultions · GAW · Sep 15, 2016 · General, Plastic Surgery

Device Facts

Record IDK160529
Device NamePledgeted COR-SUTURE QUICK LOAD
ApplicantLsi Soultions
Product CodeGAW · General, Plastic Surgery
Decision DateSep 15, 2016
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 878.5010
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Pledgeted COR-SUTURE™ QUICK LOAD® is intended for use in the approximation of soft tissue and prosthetic material.

Device Story

Pledgeted COR-SUTURE™ QUICK LOAD® is a sterile, non-absorbable surgical suture system designed for use with LSI SOLUTIONS® suturing devices. The device consists of monofilament polypropylene or PTFE-coated braided polyester suture with a PTFE pledget pre-attached. Stainless steel ferrules are attached to each end of the suture to facilitate rapid, reliable loading into compatible suturing devices. The system includes a detachable clear tube to prevent tangling. Used in clinical settings by surgeons, the device simplifies suture placement during soft tissue or prosthetic material approximation. The pre-attached pledget reduces procedural steps, potentially improving efficiency and consistency in tissue repair. The device is MR safe and intended for single-patient use.

Clinical Evidence

No clinical studies were required. Substantial equivalence was demonstrated through bench testing, including USP monograph testing for diameter, tensile strength, and needle/ferrule attachment. Biocompatibility was established per ISO 10993-1:2009, including cytotoxicity, sensitization, irritation, toxicity, genotoxicity, implantation, and hemocompatibility testing. Data was leveraged from the suture material supplier and the reference device.

Technological Characteristics

Materials: monofilament polypropylene or PTFE-coated braided polyester; PTFE pledget; stainless steel ferrule. Energy: N/A (mechanical). Form factor: sterile tray with suture release feature and anti-tangle tube. Sterilization: Ethylene oxide. Connectivity: None. Software: None.

Indications for Use

Indicated for the approximation of soft tissue and prosthetic material in patients requiring surgical suture closure.

Regulatory Classification

Identification

Nonabsorbable polypropylene surgical suture is a monofilament, nonabsorbable, sterile, flexible thread prepared from long-chain polyolefin polymer known as polypropylene and is indicated for use in soft tissue approximation. The polypropylene surgical suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be undyed or dyed with an FDA approved color additive; and the suture may be provided with or without a standard needle attached.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a bird in flight. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 15, 2016 LSI Solutions Mr. Kevin Bentley Executive Director of Regulatory Affairs and Quality 7796 Victor-Mendon Rd Victor, New York 14564 Re: K160529 Trade/Device Name: Pledgeted Cor-Suture Quick Load Regulation Number: 21 CFR 878.5010 Regulation Name: Nonabsorbable Polypropylene Surgical Suture Regulatory Class: Class II Product Code: GAW, GAS Dated: August 9, 2016 Received: August 11, 2016 Dear Mr. Bentley: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # David Krause -S Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K160529 Device Name Pledgeted COR-SUTURE™ QUICK LOAD® Indications for Use (Describe) The Pledgeted COR-SUTURE™ QUICK LOAD is intended for use in the approximation of soft tissue and prosthetic material. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) SUMMARY | Submitted By: | LSI SOLUTIONS®, Inc.<br>7796 Victor-Mendon Road<br>Victor, NY 14564 | |------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Kevin Bentley<br>Executive Director of Regulatory Affairs and Quality<br>Phone: (585) 869-6669<br>Fax: (585) 742-3398<br>Email: kbentley@lsisolutions.com | | Date Prepared: | February 24, 2016 | | Trade Name: | Pledgeted COR-SUTURE™ QUICK LOAD® | | Common Name: | Nonabsorbable Surgical Suture | | Classification Name: | Nonabsorbable poly(ethylene terephthalate) surgical suture<br>(21 CFR 878.5000, Product Code GAS)<br>Nonabsorbable polypropylene surgical suture<br>(21 CFR 878.5010, Product Code GAW) | | Device Classification: | Class II | {4}------------------------------------------------ #### Predicate Device: The predicate device, LSI SOLUTIONS® Suture Placement Devices and Accessories, specifically the Non-Absorbable QUICK LOAD® Suture, was selected to demonstrate substantial equivalence to the Pledgeted COR-SUTURE™ QUICK LOAD®. LSI SOLUTIONS® Suture Placement Devices and Accessories was cleared in 510(k) Premarket Notification K100593 on September 24, 2010. The Bard® PTFE Felt Pledget, which was cleared under 510(k) Premarket Notification K770835 on October 5, 1977, is used as a reference device in this submission. # Device Description: Each LSI SOLUTIONS® Pledgeted COR-SUTURE™ QUICK LOAD® sterile surgical suture is held in a customized tray, with suture release feature designed to enable the rapid, easy and reliable loading of suture into compatible LSI SOLUTIONS® suturing devices. Pledgeted COR-SUTURE™ QUICK LOAD® suture is available as a non-absorbable monofilament polypropylene and PTFE coated, braided, polyester surgical suture with attached PTFE pledget. A short length of modified surgical stainless steel tubing, called a "ferrule", is attached to each end of the suture. The Pledgeted COR-SUTURE™ QUICK LOAD® suture also includes a detachable clear suture tube to keep the suture from tangling. Pledgeted COR-SUTURE™ QUICK LOAD® surgical suture is packaged for single patient use. The polypropylene suture is dyed blue with the FDA approved colorant [phthalocyaninato(2-)] copper. The polyester suture is offered undyed, or dyed green with the FDA approved colorant D&C Green No. 6. For both polypropylene and polyester suture, there is no known significant change in tensile strength retention to occur in vivo. The Pledgeted COR-SUTURE™ QUICK LOAD® surgical suture is MR safe. #### Intended Use: The Pledgeted COR-SUTURE™ QUICK LOAD® is intended for use in the approximation of soft tissue and prosthetic material. {5}------------------------------------------------ The indications for use statement is identical to that of the predicate device, LSI SOLUTIONS® Suture Placement Devices and Accessories, specifically the Non-Absorbable QUICK LOAD® Suture. # Technological Characteristics (comparison to Predicate Device): The Pledgeted COR-SUTURE™ QUICK LOAD® is substantially equivalent in intended use and fundamental technological characteristics to the legally marketed predicate device. Both the modified and predicate devices are non-absorbable surgical suture available as either monofilament polypropylene or PTFE coated, braided polyester. The Pledgeted COR-SUTURE™ QUICK LOAD® and the predicate device both have a short length of modified surgical stainless steel tubing, called a ferrule, attached to each end of the suture to facilitate use with LSI SOLUTIONS® suture placement devices. The only technological difference between the Pledgeted COR-SUTURE™ QUICK LOAD® and the predicate device is that the modified device will be packaged as surgical suture with a pledget preattached directly on the suture for ease of use. The pledget is composed of polytetrafluoroethylene (PTFE) and has identical material characteristics to the reference device, Bard® PTFE Felt Pledget. The modified device is packaged and sterilized in an equivalent manner and has equivalent labeling claims to the predicate device, including indications, contraindications, warnings, cautions and precautions. The modified device is considered substantially equivalent and has the same technological characteristics to its predicate device through comparison in design, intended use, material composition, function, and range of sizes. # Performance Testing Summary: As recommended by the FDA Guidance for Industry, Class II Special Control Guidance Document: Surgical Sutures; Guidance for Industry and FDA, the Pledgeted COR-SUTURE™ QUICK LOAD® was subject to the requirements of the United States Pharmacopeia (USP) monograph for Non-Absorbable Sutures. Performance testing includes: - . Sutures - Diameter <861> - Sutures Tensile Strength <881> . {6}------------------------------------------------ - Ferrule Attachment which complies with the requirements for standard needle . attachments as defined in USP section Sutures – Needle Attachment <871>. Biocompatibility evaluation for the Pledgeted COR-SUTURE™ QUICK LOAD® was conducted in accordance with ISO 10993-1:2009: Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process and FDA 510(k) Blue Book Memorandum - #G95-1 Table 1 Initial Evaluation Tests for Consideration, to demonstrate that the Pledgeted COR-SUTURE™ QUICK LOAD® is substantially equivalent to the predicate device. The Pledgeted COR-SUTURE™ QUICK LOAD® is a permanent contact duration device that will be implanted in circulating blood, which is the identical type and duration of patient contact as the predicate device. Biocompatibility data has been leveraged from the manufacturer that supplies the surgical suture material to LSI SOLUTIONS®. The testing included the following: - Cytotoxicity ● - Sensitization . - Irritation . - Sub-chronic Toxicity and Sub-acute Toxicity . - Genotoxicity ● - Implantation . - Hemocompatibility . All biocompatibility testing passed. The Bard® PTFE Felt Pledget, which has identical material characteristics to the pledget used in the Pledgeted COR-SUTURE™ QUICK LOAD®, was tested and found biocompatible based on the materials, manufacturing processes, body contact, sterilization properties, acceptable biocompatibility test results and long clinical history of this type of use. The Pledgeted COR-SUTURE™ QUICK LOAD® will be sterilized using a validated ethylene oxide cycle. {7}------------------------------------------------ #### Clinical Testing: The technological characteristics, material, manufacturing and sterilization processes are the same as the predicate device, therefore, no clinical studies were required to demonstrate the safety or effectiveness of the modified device. ## Substantial Equivalence: With the addition of the pledget there are no new risks introduced and no negative impact to the safety and effectiveness of the surgical suture. The indications for use remain exactly the same. Both the Non-Absorbable QUICK LOAD® Suture and the Pledgeted COR-SUTURE™ QUICK LOAD® are indicated for use in the approximation of soft tissue and prosthetic material. The Pledgeted COR-SUTURE™ QUICK LOAD® is substantially equivalent to the currently marketed LSI SOLUTIONS® Suture Placement Devices and Accessories, specifically the Non-Absorbable QUICK LOAD® Suture, cleared in 510(k) Premarket Notification K100593 on September 29, 2010.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...