FILASILK - Natural Non-absorbable Silk Surgical Suture, FILAMIDE - Non Absorbable polyamide surgical Sutures, MERICRON XL - Non Absorbable Polyester Surgical Suture, FILAPROP - Non Absorbable Polypropylene surgical Suture
K160623 · M/s. Meril Endo Surgery Private Limited. · GAP · May 31, 2016 · General, Plastic Surgery
Device Facts
Record ID
K160623
Device Name
FILASILK - Natural Non-absorbable Silk Surgical Suture, FILAMIDE - Non Absorbable polyamide surgical Sutures, MERICRON XL - Non Absorbable Polyester Surgical Suture, FILAPROP - Non Absorbable Polypropylene surgical Suture
Applicant
M/s. Meril Endo Surgery Private Limited.
Product Code
GAP · General, Plastic Surgery
Decision Date
May 31, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
FILASILK is intended for use as non-absorbable sutures in general soft tissue approximation and/or ligation including use in cardiac, vascular and ophthalmic procedures. FILAMIDE Non-absorbable Polyamide Surgical suture is indicated for use in soft tissue approximation and/ or ligation, including use in cardiovascular, ophthalmic and neurological procedures. MERICRON XL suture is intended for use in general soft tissue approximation and/or ligation including cardiovascular surgery, neurological and ophthalmic procedures. FILAPROP sutures are intended for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic and neurological procedures.
Device Story
Non-absorbable surgical sutures used for soft tissue approximation and ligation; available in various materials: natural silk (fibroin), polyamide (nylon), polyester (polyethylene terephthalate), and polypropylene. Sutures provided as monofilament or braided, undyed or dyed, and may be attached to stainless steel needles. Used by surgeons in OR settings for cardiovascular, ophthalmic, and neurological procedures. Devices function by providing mechanical tissue support during healing; performance verified against USP standards for diameter, tensile strength, and needle attachment. Benefits include secure tissue closure and reliable handling characteristics.
Clinical Evidence
Bench testing only. Performance evaluated per USP requirements including diameter (USP <861>), tensile strength (USP <881>), needle attachment (USP <871>), and suture length. Comprehensive biocompatibility testing performed, including in vitro cytotoxicity, skin sensitization, intracutaneous reactivity, acute/sub-chronic systemic toxicity, intramuscular implantation, bacterial reverse mutation, mammalian erythrocyte micronucleus, in vitro hemolysis, and pyrogen testing.
Technological Characteristics
Materials: Silk (fibroin), Polyamide 6/6,6, Poly(ethylene terephthalate), Polypropylene. Construction: Monofilament or braided. Coating: Bees wax. Colorants: Logwood extract (21 CFR 73.1410), D&C Green No. 6 (21 CFR 74.3206), Phthalocyanine blue (21 CFR 74.3045). Sterilization: Not specified. Connectivity: None. Form factor: Suture strands, with or without stainless steel needles.
Indications for Use
Indicated for general soft tissue approximation and/or ligation in patients requiring cardiac, vascular, ophthalmic, or neurological procedures.
Regulatory Classification
Identification
Natural nonabsorbable silk surgical suture is a nonabsorbable, sterile, flexible multifilament thread composed of an organic protein called fibroin. This protein is derived from the domesticated species Bombyx mori (B. mori ) of the family Bombycidae. Natural nonabsorbable silk surgical suture is indicated for use in soft tissue approximation. Natural nonabsorbable silk surgical suture meets the United States Pharmacopeia (U.S.P.) monograph requirements for Nonabsorbable Surgical Suture (class I). Natural nonabsorbable silk surgical suture may be braided or twisted; it may be provided uncoated or coated; and it may be undyed or dyed with an FDA listed color additive.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring a symbol that resembles an abstract eagle or bird with three curved lines representing its wings or feathers.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 31, 2016
M/S. Meril Endo Surgery Private Limited Mr. Umesh Sharma Deputy General Manager - Quality Assurance 3fl, E1-e3, Meril Park, Survey No. 135/2/b & 174/2, Muktanand Marg Chala, Vapi, Gujaret 396191 INDIA
Re: K160623
Trade/Device Name: Filasilk - Natural Non-absorbable Silk Surgical Suture, Filamide -Non Absorbable Polyamide Surgical Sutures, Mericron XI - Non Absorbable Polyester Surgical Suture, Filaprop - Non Absorbable Polypropylene Surgical Suture
Regulation Number: 21 CFR 878.5030 Regulation Name: Natural Nonabsorbable Silk Surgical Suture Regulatory Class: Class II Product Code: GAP, GAR, GAT, GAW Dated: March 1, 2016 Received: March 4, 2016
Dear Mr. Sharma:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K160623
Device Name FILAMIDE
Indications for Use (Describe)
FILAMIDE is intended for use as non-absorbable sutures in general soft tissue approximation including skin tissue closure and/or ligation in cardiac, vascular and ophthalmic procedures.
Type of Use (Select one or both, as applicable)
| <span> <b> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </b> </span> |
|----------------------------------------------------------------------------------------------------------------------|
| <span> <b> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </b> </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K160623
Device Name FILAPROP
Indications for Use (Describe)
FILAPROP sutures are intended for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic and neurological procedures.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-----------------------------------------------------------------------------------------|
| | <span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
| | <span style="font-family: Arial;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K160623
Device Name FILASILK
Indications for Use (Describe)
FILASILK is intended for use as non-absorbable sutures in general soft tissue approximation and/or ligation including use in cardiac, vascular and ophthalmic procedures.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions search existing data sources. gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K160623
Device Name MERICRON XL
Indications for Use (Describe)
MERICRON XL suture is intended for use in general soft tissue approximation and/or ligation including cardiovascular surgery, neurosurgery and ophthalmic procedures.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{6}------------------------------------------------
## 510(k) Summary
#### l. Submitter
M/s. Meril Endo Surgery Private Limited Third Floor, E1 – E3, Meril Park, Survey No. 135/2/B & 174/2, Muktanand Marg, Chala, Vapi – 396191 Gujarat, India. Tel. No: +91-260-3052100, Fax: +91-260-3052125 Web site: www.merillife.com E-mail: umesh.sharma@merillife.com
Contact Person: Umesh Sharma Date Prepared: Feb 26, 2016
### II. Device
| Sr. No. | Trade Proprietary Name | / Common Name | Classification | Regulatory Class | Product Code | Regulation Number | Review Panel |
|---------|------------------------------------------------------------|------------------------------------------------------|------------------------------------------------------------------|------------------|--------------|-------------------|---------------------------|
| 1. | FILASILKTM<br>Natural Non-absorbable Silk Surgical Suture | Natural Non-absorbable Silk Surgical Suture | Suture, Non absorbable, Silk | II | GAP | 878.5030 | General & Plastic Surgery |
| 2. | FILAMIDETM<br>Non Absorbable Polyamide Surgical Sutures | Non-absorbable Polyamide Surgical Suture | Suture, Non absorbable, Synthetic, Polyamide | II | GAR | 878.5020 | General & Plastic Surgery |
| 3. | MERICRON XLTM Non Absorbable Polyester Surgical Suture | Non absorbable Poly (Ethylene Terephthal ate) Suture | Suture, Non absorbable, Synthetic, Poly (Ethylene Terephthalate) | II | GAT | 878.5000 | General & Plastic Surgery |
| 4. | FILAPROPTM<br>Non Absorbable Polypropylene Surgical Suture | Non-absorbable Polypropyl ene Surgical Suture | Suture, Non absorbable, Synthetic, Polypropylene | II | GAW | 878.5010 | General & Plastic Surgery |
{7}------------------------------------------------
## III. Predicate Device
| Sr. No. | Meril's Proprietary Name | Suture Trade / Predicate Device Name | Predicate Device 510(k) No. |
|---------|---------------------------------------------------------|--------------------------------------|-----------------------------|
| 1. | FILASILKTM Natural Non-absorbable Silk Surgical Suture | TRUSILK | K041514 |
| 2. | FILAMIDETM Non Absorbable Polyamide Surgical Sutures | TRULON | K041510 |
| 3. | MERICRON XLTM Non Absorbable Polyester Surgical Suture | TRUBOND | K041512 |
| 4. | FILAPROPTM Non Absorbable Polypropylene Surgical Suture | PROLENE | K133356 |
## IV. Device Description
#### a. FILASILK™ Natural Non-absorbable Silk Surgical Suture
FILASILK™ silk suture is a non-absorbable, sterile, surgical suture composed of an organic protein called fibroin. This protein is derived from the domesticated species Bombyx mori (B. mori) of the family Bombycidae. FILASILK™ sutures are processed to remove the natural waxes and gums. FILASILK™ suture is available undyed or dyed black with Logwood extract and coated with bees wax. FILASILK™ suture meets requirements
established by the United States Pharmacopoeia (U.S.P.) for non-absorbable surgical suture.
#### b. FILAMIDE™ Non Absorbable Polyamide Surgical Sutures
FILAMIDE™ suture is a non-absorbable, sterile, synthetic, monofilament surgical suture composed of polyamide 6 [(NH-CO-(CH2)s]n and Polyamide 6,6 (NH(CH2)g-NH-CO-(CH2)4-CO)n. FILAMIDE™ sutures are dyed with Logwood extract. Available in a broad range of suture sizes and lengths, FILAMIDE'M is either non-needled or attached to standard stainless steel needles of varying types and sizes.
#### c. MERICRON XL™ Non Absorbable Polyester Surgical Suture
MERICRON XL™ suture is a coated, braided, non-absorbable sterile surgical suture composed of poly (ethylene terephthalate). The empirical molecular formula of the polymer is (C10HoO4),. The suture is coated with bees wax which acts as a lubricant to mechanically improve the ease of passage through tissue and the overall handling quality of the suture.
MERICRON XL™ is available undyed or dyed green with D & C Green No.6. MERICRON XL™ is available in a range of gauge sizes and lengths, or attached to standard stainless steel needles of various types and sizes
{8}------------------------------------------------
## d. FILAPROP™ Non Absorbable Polypropylene Surgical Suture
FILAPROP™ suture is a monofilament, synthetic non-absorbable, sterile surgical suture composed of an isotactic crystalline stereoisomer of polypropylene, a synthetic linear polyolefin. The molecular formula of polypropylene is (C3Hs)n, FILAPROP™ sutures are dyed with phthalocyanine blue.
FILAPROP™ is available in a range of gauge sizes and lengths attached to standard stainless steel needles of various types and sizes
## V. Intended Use
#### a. FILASILK™ Natural Non-absorbable Silk Surgical Suture
FILASILK'" is intended for use as non-absorbable sutures in general soft tissue approximation and/or ligation including use in cardiac, vascular and ophthalmic procedures.
#### b. FILAMIDE™ Non Absorbable Polyamide Surgical Sutures
FILAMIDE™ Non-absorbable Polyamide Surgical suture is indicated for use in soft tissue approximation and/ or ligation, including use in cardiovascular, ophthalmic and neurological procedures.
### c. MERICRON XL™ Non Absorbable Polyester Surgical Suture
MERICRON XL™ suture is intended for use in general soft tissue approximation and/or ligation including cardiovascular surgery, neurological and ophthalmic procedures.
## d. FILAPROP™ Non Absorbable Polypropylene Surgical Suture
FILAPROP™ sutures are intended for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic and neurological procedures.
#### VI. Substantial Equivalence
The device design, material of construction, performance, packaging and intended uses are similar to the predicate device. Substantial equivalence is conducted based on the following parameters:
- 1. Product description
- 2. Intended use
- 3. Suture Size
- 4. Colorant Used Dye / Un-dyed
- 5. Suture Coating
- 6. Single use
- 7. Sterilisation method
{9}------------------------------------------------
- 8. Packaging
- 9. Label Claim
- 10. Performance
- a. Diameter USP <861>
- b. Tensile strength USP <88l>
- c. Needle attachment USP <871>
- d. Suture Length
- 11. Labeling and Instructions for use (IFU)
## VII. Performance Data
The Surgical Suture was subjected to the performance testing as per USP requirements. The safety and effectiveness of the Surgical Suture has been evaluated for the following performance and safety requirements.
- 1. Diameter USP <861>
- 2. Tensile strength USP <88l>
- 3. Needle attachment USP <871>
- 4. Suture Length
- 5. Biocompatibility
- a. In Vitro Cytotoxicity Study
- b. Skin Sensitization Study
- c. Intracutaneous Reactivity Test
- d. Acute Systemic Toxicity Study
- e. Sub Chronic Toxicity Study
- f. Intramuscular Implantation Test
- Bacterial Reverse Mutation Test g.
- h. Mammalian Erythrocyte Micronucleus Test
- In Vitro Hemolysis Test i.
- Pvrogen Test ....
#### VIII. Conclusion
Meril's Surgical Sutures are composed of the same materials, as are the predicate devices and has the same design, as do the predicate.
The subject devices are offered with the same colorants as predicate devices i.e. Logwood extract black used in FILASILK™ & FILAMIDE™ is in accordance to 21 CFR 73.1410, D&C Green No. 6 used in MERICRON XL™ is in accordance with 21 CFR 74.3206 & Phthalocyanine blue used in FILAPROP™ is in accordance with 21 CFR 74. 3045).
Performance testing of the Meril sutures complies the USP requirements i.e. suture diameter, suture length, knot pull tensile strength, extractable color, suture-needle
{10}------------------------------------------------
attachment and sterility. However, FILAMIDE™ & FILAPROP™ may be slightly oversize in diameter to U.S.P. requirement for some suture sizes.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.