K041514 · Sutures India Private Limited · GAP · Jun 3, 2005 · General, Plastic Surgery
Device Facts
Record ID
K041514
Device Name
TRUESILK NON-ABSORBABLE SILK SURGICAL SUTURE
Applicant
Sutures India Private Limited
Product Code
GAP · General, Plastic Surgery
Decision Date
Jun 3, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
TRUSILK" Non-absorbable Silk Surgical Suture is indicated for use in soft tissue approximation and/or ligation including use in cardiovascular, ophthalmic and neurological procedures.
Device Story
TRUSILK is a sterile, flexible, multifilament surgical suture composed of natural fibroin protein derived from Bombyx mori silkworms. The device is used by surgeons in clinical settings (e.g., OR) for soft tissue approximation and ligation. It is provided as a braided or twisted thread, either uncoated or with a biocompatible coating, and may be undyed or dyed with FDA-listed color additives (logwood extract). It is available with or without an attached needle. The suture provides mechanical support to hold tissue edges together or ligate vessels, facilitating healing. It is a single-use, non-absorbable device.
Clinical Evidence
Bench testing only. Performance verified against USP 26 requirements for diameter, length, knot pull tensile strength, needle attachment strength, extractable color, and sterility.
Technological Characteristics
Natural nonabsorbable multifilament silk (fibroin protein). Braided or twisted construction. Sterilized via Ethylene Oxide (EO). Meets USP 26 monographs for nonabsorbable surgical sutures. Biocompatible coatings used. Logwood extract colorant per 21 CFR 73.1410.
Indications for Use
Indicated for soft tissue approximation and/or ligation in cardiovascular, ophthalmic, and neurological procedures. No specific age or gender restrictions provided.
Regulatory Classification
Identification
Natural nonabsorbable silk surgical suture is a nonabsorbable, sterile, flexible multifilament thread composed of an organic protein called fibroin. This protein is derived from the domesticated species Bombyx mori (B. mori ) of the family Bombycidae. Natural nonabsorbable silk surgical suture is indicated for use in soft tissue approximation. Natural nonabsorbable silk surgical suture meets the United States Pharmacopeia (U.S.P.) monograph requirements for Nonabsorbable Surgical Suture (class I). Natural nonabsorbable silk surgical suture may be braided or twisted; it may be provided uncoated or coated; and it may be undyed or dyed with an FDA listed color additive.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{1}------------------------------------------------
K0415/4
page 1074
#### 3 2005 JUN
## SUTURES INDIA PVT.LTD SUBMISSION OF PREMARKET NOTIFICATION (510K) FOR NATURAL NONABSORBABLE SILK SURGICAL SUTURE
SECTION NO: 16 PAGE NO: -16-1
### 510K SUMMARY as required by: 21CFR 807.92
### A. APPLICANT INFORMATION
| Name | : SUTURES INDIA PVT. LTD |
|-----------------------|-------------------------------------------------------------------------------------------------|
| Address | : Sutures India Pvt. Ltd.<br>118, 3rd Phase, Peenya Industrial Area,<br>Bangalore-560058. India |
| PH.NO.<br>FAX NO | : 91-80-28395150 / 28370367 / 28377856<br>:91-80-28392280. |
| E mail<br>Web address | : sutures@vsnl.com<br>: www.suturesin.com |
| B. Contact Person | : L.G.Chandrasekhar<br>: MANAGING DIRECTOR |
| C. Date Prepared | : May 15,2004 |
### D. DEVICE TRADE NAME
- . Trade Name : TRUSILK Common name . : Nonabsorbable Surgical Suture, U.S.P.
- (Black Braided Silk)
- Classification Name :Natural Nonabsorbable Silk Surgical Suture. .
# E. PREDICATI: DEVICES
- Silkam Nonabsorbable surgical suture, 510(k) Number K990089, AESCULAP, . South San Fransisco, California.
- Silk Nonabsorbable surgical suture, 510(k) Number K960328. R.K.Medical . L.L.C. Danbury , CT06810
{2}------------------------------------------------
# SUTURES INDIA PVT.LTD SUBMISSION OF PREMARKET NOTIFICATION (510K) FOR NATURAL NONABSORBABLE SILK SURGICAL SUTURE
#### PAGE NO: -16-2 SECTION NO: 16
### F. DESCRIPTION OF THE DEVICE
TRUSH.K is a natural nonabsorbable silk surgical suture, sterile, flexible multifilament thread composed of an organic protein called fibroin. This protein is derived from the domesticated species Bombyx mori (B. mori) of the family Bombycidae. Natural Nonabsorbable silk surgical suture meets the United States Pharmacopeia (U.S.P.) monograph requirements for "Nonabsorbable Surgical Suture" . Natural Nonabsorbable silk surgical suture may be braided or twisted; it may be provided uncoated or coated; and it may be undyed or dyed with an FDA listed color additive. It will be available with and with out a standard needle attached.
### G. INTENDED USE OF THE DEVICE
Sutures India TRUSH.K Natural nonabsorbable silk suture, is indicated for use in soft tissue approximation and/ or ligation, including use in cardiovascular, ophthalmic and neurological procedures.
| Comparison items | Sutures<br>India<br>Pvt.Ltd | Aesculap | R.K.Medical |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|----------|-------------|
| Natural Nonabsorbable silk surgical suture is<br>a nonabsorbable, sterile,<br>flexible<br>multifilament thread composed of an organic<br>protein called fibroin. This protein is derived<br>from the domesticated species Bombyx mori<br>(B. mori) of the family Bombycidae. | Same | Same | Same |
| The sutures are inert, noncollageneous and<br>nonantigenic. | Same | Same | Same |
| Natural nonabsorbable silk surgical suture is<br>available undyed or dyed with logwood<br>extract | Same | Same | Same |
| Natural nonabsorbable silk surgical suture is<br>offered both uncoated, and treated with<br>biocompatible coatings to enhance its<br>handling properties. | Same | Same | Same |
| Natural Nonabsorbable silk suture is<br>indicated for use in soft tissue approximation<br>and/ or ligation, including use in<br>cardiovascular, ophthalmic and neurological<br>procedures. | Same | Same | Same |
| SECTION NO:16 | PAGE NO: -16-3 | | |
| Comparison items | Sutures<br>India<br>Pvt.ltd | Aesculap | R.K.Medical |
| Natural nonabsorbable sterile silk suture is<br>supplied for single use only. | Same | Same | Same |
| Natural nonabsorbable sterile silk suture is<br>sterilized by EO method | Same | Same | Same |
| Natural nonabsorbable silk suture is<br>packaged in the same or equivalent manner,<br>and has the same or equivalent labeling<br>claims as the predicate devices including<br>indications, warnings, cautions and<br>precautions | Same | Same | Same |
| Natural nonabsorbable silk suture meets the<br>Official Monograph of the United States<br>Pharmacopeia current edition USP26 for<br>extractable color | Same | Same | Same |
| Finished suture material meets the<br>performance requirements defined in the<br>United States Pharmacopeia current edition<br>USP26 for Diameter<861> | Same | Same | Same |
| Finished suture material meets the<br>performance requirements defined in the<br>United States Pharmacopeia current edition<br>USP26 for Tensile strength<881> | Same | Same | Same |
| Finished suture material meets the<br>performance requirements defined in the<br>United States Pharmacopeia current edition<br>USP26 for Needle attachment<871> | Same | Same | Same |
| Natural nonabsorbable silk suture meets the<br>Official Monograph of the United States<br>Pharmacopcia current edition USP26 for<br>Sterility | Same | Same | Same |
| Finished suture material packaged in a same<br>or equivalent manner with sterile single or<br>double packing having labeling conforming<br>to 21CFR and USP 26 | Same | Same | Same |
| Nonabsorbable Surgical Suture (Braided<br>Silk Black) is biologically compatible when | Same | Same | Same |
### COMPARISON TABLE SUTURES INDIA `S "TRUSILK" NATURAL SILK NONABSORBABLE SURGICAL SUTURE TO PREDICATE DEVICES
{3}------------------------------------------------
K04514
page 3 of 4
## SUTURES INDIA PVT.LTD SUBMISSION OF PREMARKET NOTIFICATION (510K) FOR NATURAL NONABSORBABLE SILK SURGICAL SUTURE
{4}------------------------------------------------
K041514 page 494
### SUTURES INDIA PVT.LTD SUBMISSION OF PREMARKET NOTIFICATION (510K) FOR NATURAL NONABSORBABLE SILK SURGICAL SUTURE
#### SECTION NO: 16 PAGE NO: - 16-4
### CONCLUSION
Sutures India's TRUSILK is Natural Nonabsorbable silk suture, composed of the same material, as are the predicated devices and has the same design, as do the predicate devices. The suture is manufactured in a manner typical of the industry and equivalent to that used to produce predicate devices. Further the subject device is offered with the same colorant Logwood extract at a concentration that conforms to the requirements of Title 21 CFR § 73.1410, as are of the predicate devices.
Testing of suture diameter, suture length, knot pull tensile strength and needle attachment strength, extractable color and sterility to methods outlined in USP 26 demonstrates Sutures India's TRUSILK, Natural Nonabsorbable silk suture meets or exceeds USP specifications and are equivalent in terms of the above mentioned predicate devices.
Robert F
L.G.Chandrasekhar Managing Director
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health and Human Services, USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a human figure in profile, with three overlapping figures suggesting community or support.
3 2005 JUN
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. L.G. Chandrasekhar Managing Director Sutures India Private Limited 472 D, 13th Cross, 4th Phase Peenya Industrial Area, Bangalore 560058, India
Re: K041514
Trade/Device Name: TRUSILK™ Non-absorbable Silk Surgical Suture Regulation Number: 21 CFR 878.5030 Regulation Name: Natural nonabsorbable silk surgical suture Regulatory Class: II Product Code: GAP Dated: May 3, 2005 Received: May 6, 2005
Dear Mr. Chandrasekhar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced we have teviewed your becalled by (4) personally equivalent (for the indications for use stated in above and nave determined the ad predicate devices marketed in interstate commerce prior to the chelosure) to regary mantowe of the Medical Device Amendments, or to devices that have been May 20, 1770, the chartinent and on the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the do not require upproval or a premails provisions of the Act. The general controls provisions of the device, dolor to the gallery annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found be subject to such additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that I case or advised that 12 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I DA ilas made a deceminations administered by other Federal agencies. You must comply with all the I caral states and regarding, but not limited to: registration and listing (21 CFR Part 807); labeling / (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
### Page 2 - Mr. L.G. Chandrasekhar
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120 . Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Signature
Miriam C. Provost, Ph.D. Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known): K041514
Device Name: TRUSILK™ Non-absorbable Silk Surgical Suture
Indications For Use:
TRUSILK" Non-absorbable Silk Surgical Suture is indicated for use in soft tissue approximation and/or ligation including use in cardiovascular, ophthalmic and neurological procedures.
Prescription Use
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
> Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.