K223052 · Synovis Life Technologies, Inc. · FTM · Apr 7, 2023 · General, Plastic Surgery
Device Facts
Record ID
K223052
Device Name
Peri-Guard and Supple Peri-Guard
Applicant
Synovis Life Technologies, Inc.
Product Code
FTM · General, Plastic Surgery
Decision Date
Apr 7, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
PERI-GUARD and SUPPLE PERI-GUARD are intended to be used as a surgical mesh. PERI-GUARD is intended for use as a prosthesis for the surgical repair of soft tissue deficiencies which include: defects of the abdominal and thoracic wall, gastric banding, muscle flap reinforcement, and hernias (including diaphragmatic, femoral, incisional, inguinal, lumbar, scrotal, and umbilical hernias). SUPPLE PERI-GUARD is intended for use as a prosthesis for soft tissue deficiencies which include: defects of the abdominal and thoracic wall, gastric banding, muscle flap reinforcement, and hernias (including diaphragmatic, femoral, incisional, inguinal, lumbar, scrotal, and umbilical hernias).
Device Story
Bovine pericardium-derived surgical mesh; cross-linked with glutaraldehyde; treated with sodium hydroxide. Used as prosthesis for soft tissue repair; abdominal/thoracic wall defects; hernia repair. Implanted by surgeons in clinical settings. Provides structural reinforcement for soft tissue deficiencies. MR safe. Subject of 510(k) for modified sterilization (gamma irradiation) and packaging (double foil pouch).
Clinical Evidence
No human clinical trials. Evidence includes bench testing (suture retention, burst strength, collagenase digestion, denaturation temperature, water permeability, pliability) and two animal studies: 6-month porcine incisional hernia repair model (n=not specified) showing equivalent tensile/burst strength and safety; 3-week rat calcification study showing no significant mineralization differences. Biocompatibility testing per ISO 10993-1:2018 confirmed safety for permanent implant.
Technological Characteristics
Bovine pericardium cross-linked with glutaraldehyde and sodium hydroxide. Gamma irradiation sterilization (ISO 11137). Double foil pouch packaging (ISO 11607-1). MR safe. Permanent implant. No software or electronic components.
Indications for Use
Indicated for surgical repair of soft tissue deficiencies including abdominal/thoracic wall defects, gastric banding, muscle flap reinforcement, and various hernias (diaphragmatic, femoral, incisional, inguinal, lumbar, scrotal, umbilical).
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
Predicate Devices
PERI-GUARD Repair Patch and SUPPLE PERI-GUARD Repair Patch (K983162)
Reference Devices
Synovis PERI-GUARD Repair Patch and SUPPLE PERI-GUARD Pericardium Patch (K142447)
PERI-STRIPS DRY Staple Line Reinforcement with VERITAS Collagen Matrix with SECURE GRIP Technology (K192615)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Synovis Life Technologies, Inc. Megan Sajjad Sr. Manager Regulatory Affairs A Subsidiary of Baxter International Inc.) 2575 University Ave. W. St. Paul, Minnesota 55114
Re: K223052
Trade/Device Name: Peri-Guard and Supple Peri-Guard Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: FTM Dated: March 27, 2023 Received: March 28, 2023
Dear Megan Sajjad:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Deborah A. Fellhauer -S
Deborah Fellhauer, RN, BSN Assistant Director DHT4B:Division of Infection Control and Plastic Surgery Devices OHT4:Office of Surgical And Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223052
Device Name PERI-GUARD Repair Patch SUPPLE PERI-GUARD Repair Patch
### Indications for Use (Describe)
PERI-GUARD is intended for use as a prosthesis for the surgical repair of soft tissue deficiencies which include: defects of the abdominal and thoracic wall, gastric banding, muscle flap reinforcement, and hernias (including diaphragmatic, femoral, incisional, inguinal, lumbar, scrotal, and umbilical hernias).
SUPPLE PERI-GUARD is intended for use as a prosthesis for soft tissue deficiencies which include: defects of the abdominal and thoracic wall, gastric banding, muscle flap reinforcement, and hernias (including diaphragmatic, femoral. incisional, inguinal, lumbar, scrotal, and umbilical hernias).
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><input type="checkbox"/>Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><input type="checkbox"/>Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510k SUMMARY: PERI-GUARD Repair Patch and SUPPLE PERI-GUARD Repair Patch
#### I. SUBMITTER
Synovis Life Technologies, Inc. (Synovis) (A Subsidiary of Baxter International Inc.) 2575 University Avenue West St. Paul, MN 55114-1024 Phone: 651-796-7410 Fax: 224-270-4119
Contact Person: Megan Sajjad, Sr. Manager, Regulatory Affairs
Date prepared: March 27, 2023
## II. DEVICE
Device Trade Names: PERI-GUARD and SUPPLE PERI-GUARD
Common Name: Repair Patch
Regulation: 21 CFR 878.3300 - Surgical Mesh
Product Codes: FTM (Mesh, Surgical), OWV (mesh, surgical, collagen, diaphragmatic hernia), OXB (mesh, surgical, collagen, thoracic, chest wall reconstruction);
## III. PREDICATE DEVICE
PERI-GUARD Repair Patch and SUPPLE PERI-GUARD Repair Patch, K983162
Manufacturer: Synovis Life Technologies, Inc., (A Subsidiary of Baxter International Inc.)
Additionally, Synovis PERI-GUARD Repair Patch and SUPPLE PERI-GUARD Pericardium Patch (K142447) and PERI-STRIPS DRY Staple Line Reinforcement with VERITAS Collagen Matrix with SECURE GRIP Technology (K192615) serve as Reference devices for this 510k.
## IV. DEVICE DESCRIPTION
PERI-GUARD Repair Patch (PERI-GUARD) and SUPPLE PERI-GUARD Repair Patch (SUPPLE PERI-GUARD) are derived from bovine pericardium procured from cattle originating in the United States. The pericardium is cross
{4}------------------------------------------------
linked with glutaraldehyde and treated with 1 molar sodium hydroxide for a minimum of 60 minutes at 20-25℃ (68-77°F).
PERI-GUARD and SUPPLE PERI-GUARD are terminally sterilized using gamma irradiation and packaged within a sterile double-pouch system. The contents of the unopened, undamaged outer pouch are sterile.
PERI-GUARD and SUPPLE PERI-GUARD are MR Safe.
PERI-GUARD and SUPPLE PERI-GUARD utilize animal tissue; patient must be informed prior to any procedure.
See Table 1 for PERI-GUARD and SUPPLE PERI-GUARD product models and sizes.
| PERI-GUARD Model Number | Size (cm) |
|--------------------------------|-----------|
| PG0404 | 4 x 4 |
| PG0608 | 6 x 8 |
| PG0814 | 8 x 14 |
| PG1016 | 10 x 16 |
| PG1225 | 12 x 25 |
| SUPPLE PERI-GUARD Model Number | Size (cm) |
| SPG0404 | 4 x 4 |
| SPG0406 | 4 x 6 |
| SPG0608 | 6 x 8 |
| SPG0814 | 8 x 14 |
## Table 1 - PERI-GUARD and SUPPLE PERI-GUARD Product Models and Sizes
# V. INTENDED USE/INDICATIONS FOR USE
### Statement of Intended Use:
PERI-GUARD and SUPPLE PERI-GUARD are intended to be used as a surgical mesh.
## Indications for Use:
PERI-GUARD is intended for use as a prosthesis for the surgical repair of soft tissue deficiencies which include: defects of the abdominal and thoracic wall, gastric banding, muscle flap reinforcement, and hernias (including diaphragmatic, femoral, incisional, inguinal, lumbar, scrotal, and umbilical hernias).
SUPPLE PERI-GUARD is intended for use as a prosthesis for soft tissue deficiencies which include: defects of the abdominal and thoracic wall, gastric
{5}------------------------------------------------
banding, muscle flap reinforcement, and hernias (including diaphragmatic, femoral, incisional, inguinal, lumbar, scrotal, and umbilical hernias).
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
PERI-GUARD and SUPPLE PERI-GUARD are substantially equivalent to the predicate PERI-GUARD and SUPPLE PERI-GUARD devices based on having the same fundamental technology and intended use. The changes between the subject and predicate device include a change in sterilization method and redesigned packaging. The safety and performance of the subject devices has been evaluated through non-clinical testing.
The subject and predicate devices are identical in the following respects:
- Intended Use
- Patch material ●
- Viral inactivation processing steps ●
- Strength, chemical and physical specifications ●
The following technological differences exist between the subject and predicate devices:
- Modified sterilization method from chemical sterilization to gamma . irradiation
- Modified packaging from a jar to a double foil pouch ●
### PERFORMANCE DATA VII.
The following bench testing performance data were provided in support of the substantial equivalence determination:
- Visual
- Suture retention
- Dimensional
- Burst strength
- Collagenase digestion ●
- Denaturation Temperature ●
- Water Permeability
- . Pliability
- . Chemical and heavy metal residuals
- Bioburden ●
- Pyrogenicity/Endotoxins ●
- . Temperature Excursion Testing
- Biocompatibility Assessment ●
{6}------------------------------------------------
The results of performance testing demonstrate that the modified PERI-GUARD and SUPPLE PERI-GUARD devices are substantially equivalent to the predicate devices.
### Animal Studies
Two animal studies were performed to support the substantial equivalence of the chemically sterilized and gamma irradiated PERI-GUARD and SUPPLE PERI-GUARD devices.
| Test | Results | Conclusions |
|-----------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Implantation:<br>Porcine incisional<br>hernia repair model<br>for 6 months per<br>ISO 10993-6 | No test material-related<br>changes among clinical<br>observations, bodyweights,<br>veterinary physicals, clinical<br>pathology parameters,<br>mortality, significant<br>macroscopic or microscopic<br>differences, tensile or burst<br>strength, or clinical<br>performance between the<br>subject and predicate devices at<br>Day $113\pm3$ or Day $203\pm3$ . | The subject and predicate<br>devices performed<br>equivalently and had an<br>equivalent safety profile in the<br><i>in vivo</i> porcine model for<br>incisional hernia repair. |
| 3-week Rat<br>Calcification<br>Potential Study | No significant mineralization<br>was noted microscopically for<br>either the subject or predicate<br>device. | No significant difference in<br>calcification potential. |
Biocompatibility
The results of biocompatibility testing demonstrate the biocompatibility of the product in accordance with ISO 10993-1:2018 requirements for a permanent (long term) implant with tissue, bone or (non-direct) blood contact. As per FDA Guidance, testing included cytotoxicity, sensitization, acute systemic toxicity, pyrogenicity, genotoxicity, implantation, hemolysis, and subchronic toxicity endpoints.
### Shelf Life
Synovis has performed aging testing to support a 1 year shelf life claim.
### Sterilization and Packaging
Sterilization validation was conducted according to ISO 11137 parts 1 and 2.
The modified packaging was designed and evaluated in accordance with ISO 11607-1. The integrity of the sterile barrier is supported by testing conducted in accordance with ASTM F88-15 and ASTM F2096-11.
{7}------------------------------------------------
### Validation Studies
Human factors testing was conducted to confirm that the re-designed packaging allows for aseptic transfer of the product to the sterile field without compromising sterility, and that the modified packaging design does not impact the surface of the tissue patch.
# VII. CONCLUSION
The subject PERI-GUARD and SUPPLE PERI-GUARD devices share the same intended use and technological characteristics as the predicate devices. The physical, functional and performance specifications for the devices are substantially equivalent. Testing supports that the subject devices are as safe and effective as the predicate devices when used according to their labeling.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.