K162752 · Lifecell Corporation · FTM · Feb 24, 2017 · General, Plastic Surgery
Device Facts
Record ID
K162752
Device Name
ARTIA Reconstructive Tissue Matrix Perforated
Applicant
Lifecell Corporation
Product Code
FTM · General, Plastic Surgery
Decision Date
Feb 24, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
ARTIA Tissue Matrix is intended for use as a soft tissue patch to reinforce soft tissue where weakness exists and for the surgical repair of damaged or ruptured soft tissue membranes which require or bridging material to obtain the desired surgical outcome. The implant is intended for reinforcement in plastic and reconstructive surgery. ARTIA Tissue Matrix is intended for single patient, one time use only.
Device Story
ARTIA Reconstructive Tissue Matrix Perforated is a porcine-derived surgical mesh used as a soft tissue patch for reinforcement or repair in plastic and reconstructive surgery. The device acts as a scaffold for cellular and microvascular ingrowth. It is provided as a terminally sterilized sheet in various geometric configurations, featuring a 3mm diameter circular perforation pattern. The device is used by surgeons in clinical settings. It is implanted to provide structural support to damaged or ruptured membranes, facilitating tissue repair and improving surgical outcomes. The perforations are added during the final die-cutting stage of manufacturing.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing including mesh thickness, tensile strength, device stiffness, suture pull-out strength, burst strength (ASTM D6797-07), and tear resistance (ASTM D5735-95).
Technological Characteristics
Porcine dermal matrix; terminally sterilized via electron beam irradiation; provided in various geometric configurations; features 3mm diameter circular perforations; packaged in plastic holder and pouch; phosphate buffered aqueous solution with matrix stabilizers.
Indications for Use
Indicated for reinforcement of soft tissue where weakness exists and surgical repair of damaged or ruptured soft tissue membranes requiring reinforcing or bridging material in plastic and reconstructive surgery. For single patient, one-time use.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are connected and appear to be emerging from a single base. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 24, 2017
Lifecell Corporation Ms. Anuja Yardi Regulatory Affairs Specialist One Millennium Way Branchburg, New Jersey 08876
Re: K162752
Trade/Device Name: ARTIA Reconstructive Tissue Matrix Perforated Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: FTM, OXH Dated: January 25, 2017 Received: January 26, 2017
Dear Ms. Yardi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# David Krause -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162752
Device Name
ARTIA Reconstructive Tissue Matrix-Perforated
#### Indications for Use (Describe)
ARTIA Tissue Matrix is intended for use as a soft tissue patch to reinforce soft tissue where weakness exists and for the surgical repair of damaged or ruptured soft tissue membranes which require or bridging material to obtain the desired surgical outcome. The implant is intended for reinforcement in plastic and reconstructive surgery.
ARTIA Tissue Matrix is intended for single patient, one time use only.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|----------------------------------------------------------------------------------|
| | <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
| | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 7. 510(k) Summary
## 7.1 SUBMITTER
#### Name and Address of Submitter:
LifeCell Corporation One Millennium Way Branchburg, NJ 08876 USA
## Contacts:
Anuja Yardi Regulatory Affairs Specialist Phone: (908) 947-1018 Fax: (908) 947-1095
Mira Leiwant Senior Director, Quality Engineering and Regulatory Affairs Phone: (908) 809-7747 Fax: (908) 947-1095
# Prepared by and Date:
Anuja Yardi September 29, 2016
## 7.2 DEVICE
Name of Device: ARTIA™ Reconstructive Tissue Matrix Perforated Common or Usual Name: Surgical Mesh Classification Name: Surgical Mesh- (21 C.F.R. §878.3300) Device Class: Class II Product Code: FTM and OXH
# 7.3 PREDICATE DEVICE
## Predicate Device:
ARTIA Reconstructive Tissue Matrix (cleared as HPTM via K142326) The predicate has not been subject to a design-related recall.
{5}------------------------------------------------
## Reference Device:
Strattice Tissue Matrix Perforated (K150712) – LifeCell Corporation The reference device has not been subject to a design-related recall.
## 7.4 DEVICE DESCRIPTION
ARTIA Reconstructive Tissue Matrix Perforated (ARTIA Tissue Matrix Perforated) is a surgical mesh that is derived from porcine dermis and is processed and preserved in a patented phosphate buffered aqueous solution containing matrix stabilizers. ARTIA Tissue Matrix Perforated is designed to perform as a surgical mesh for soft tissue repair while presenting a scaffold for cellular and microvascular ingrowth. ARTIA Tissue Matrix Perforated consists of a terminally sterilized sheet of processed porcine dermal matrix provided in various geometric configurations and packaged in a plastic holder enclosed within a pouch. The device is sterilized via electron beam irradiation. The subject device has the same underlying scientific technology, principles of operation, Intended Use and Indications for Use as the cleared predicate device, ARTIA Tissue Matrix (K142326). This 510(k) premarket notification describes a new design feature of the subject device, which introduces a 3mm diameter circular pattern of perforations (holes) throughout the tissue matrix.
## 7.5 INDICATIONS FOR USE
ARTIA Tissue Matrix is intended for use as a soft tissue patch to reinforce soft tissue where weakness exists and for the surgical repair of damaged or ruptured soft tissue membranes which require the use of reinforcing or bridging material to obtain the desired surgical outcome. The implant is intended for reinforcement in plastic and reconstructive surgery.
ARTIA Tissue Matrix is intended for single patient, one time use only.
# 7.6 COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The subject and predicate devices are both surgical meshes derived from porcine dermis which undergo standardized, controlled manufacturing processes. Both devices consist of a terminally sterilized sheet of the processed porcine dermal matrix provided in various geometric configurations. The subject and the predicate devices share the same underlying scientific design and have the same Intended Use. Indications for Use and principles of operation and packaging configuration. Both are sterilized via electron beam irradiation.
{6}------------------------------------------------
The subject device includes a new design feature, 3mm diameter circular pattern of perforations (holes) in the tissue matrix. The spacing and size of the perforations were evaluated against the same established performance specifications as the predicate device, ARTIA Tissue Matrix (K142326). The subject device utilizes the existing manufacturing processes of the predicate ARTIA Tissue Matrix. Perforations are added to the tissue during the final die cutting stage at the time the device is cut to its final perimeter dimensions.
## 7.7 PERFORMANCE
The predicate device, ARTIA Tissue Matrix (K142326) has undergone appropriate biocompatibility assessments and testing, sterilization, and viral inactivation testing. The data demonstrates that the device is biocompatible and that the manufacturing process is capable of viral inactivation. The subject device, ARTIA Tissue Matrix Perforated and the predicate device, ARTIA Tissue Matrix (K142326) utilize the same raw materials (porcine dermis), processing components/solutions, and manufacturing processes, and are identical except for the presence of perforations. The only difference in the manufacturing process of the subject device is that during the final die cutting stage, perforations will be added to the tissue at the time the device is cut to its final perimeter dimensions. The cutting dies are provided sterilized from the vendor and utilize the same base materials, and the die cutting process is the same as the predicate device. The addition of holes in the product via die cutting does not change the material properties of the tissue. Therefore, no further biocompatibility testing or viral inactivation testing was performed.
#### Bench Testing:
Bench testing was performed to support substantial equivalence to the predicate device. Testing and/or evaluations included relevant elements of the FDA "Guidance for the Preparation of a Premarket Notification Application for a Surgical Mesh" issued March 2, 1999. Table 1 includes the applicable product characterization criteria used to demonstrate substantial equivalence.
{7}------------------------------------------------
| Bench Testing /Evaluation | Applicable Standard |
|---------------------------|---------------------|
| Mesh Thickness | N/A |
| Tensile Strength | N/A |
| Device Stiffness | N/A |
| Suture Pull-Out Strength | N/A |
| Burst Strength | ASTM D6797-07(2011) |
| Tear Resistance | ASTM D5735-95 |
Performance data demonstrates that ARTIA Tissue Matrix Perforated has similar mechanical properties and meets the established specifications as the predicate device. In addition, the ARTIA Tissue Matrix Perforated Surgical Mesh shares the technological characteristics of perforations with LifeCell's other legally marketed reference device, Strattice Tissue Matrix Perforated (K150712). This new technological characteristic does not affect safety and efficacy of the device or raise new questions of safety or efficacy. ARTIA Tissue Matrix Perforated meets the requirements to perform its intended use as a soft tissue patch and is substantially equivalent to the predicate device, ARTIA Tissue Matrix (K142326).
## Clinical Testing:
No clinical testing was included in this submission.
# 7.8 CONCLUSIONS
The subject device, ARTIA Tissue Matrix Perforated, meets the requirements to perform its intended use as a soft tissue patch and is substantially equivalent to the legally marketed predicate device, ARTIA Tissue Matrix (K142326).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.