Matristem Surgery Matrix RS, PSM, PSMX, Matristem Pelvic Floor Matrix
K141084 · Acell, Inc. · FTM · Jun 17, 2015 · General, Plastic Surgery
Device Facts
Record ID
K141084
Device Name
Matristem Surgery Matrix RS, PSM, PSMX, Matristem Pelvic Floor Matrix
Applicant
Acell, Inc.
Product Code
FTM · General, Plastic Surgery
Decision Date
Jun 17, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
MatriStem Surgical Matrix RS (2-layer) are intended for implantation to reinforce soft tissue where weakness exists in patients requiring urological, or plastic & reconstructive surgery. Reinforcement of soft tissue within urological, gastroenterological, and plastic & reconstructive surgery includes, but is not limited to, the following procedures: hernia and body wall repair, colon and rectal prolapse repair, and esophageal repair. MatriStem Surgical Matrix PSMX (6-layer) is intended for implantation to reinforce soft tissue where weakness exists in patients requiring gastroenterological or plastic & reconstructive surgery. Reinforcement of soft tissue within gastroenterological and plastic & reconstructive surgery includes, but is not limited to, the following procedures: hernia and body wall repair, colon and rectal prolapse repair, and esophageal repair. MatriStem Pelvic Floor Matrix (6-layer) is intended for implantation to reinforce soft tissue where weakness exists in patients requiring urological or gynecological surgery. Reinforcement of soft tissue within urological and gynecological surgery includes, but is not limited to, the following procedures: pubourethral repair, transabdominal reconstruction of pelvic floor and vaginal prolapse repair, bladder support, and sacrocolposuspension. By providing pubourethral support, Matristem Pelvic Floor Matrix may be used for the treatment of urinary incontinence resulting from urethral hypermobility and intrinsic sphincter deficiency.
Device Story
MatriStem devices are porcine-derived extracellular matrix (ECM) scaffolds (urinary bladder matrix) used for soft tissue reinforcement. Supplied as multi-layer sheets in various configurations (2, 3, or 6 layers), they are implanted by surgeons during urological, gastroenterological, gynecological, or plastic/reconstructive procedures. The scaffold provides a structural matrix to support tissue repair and reinforcement. Devices are terminally sterilized via electron beam irradiation. Clinical benefit is derived from the mechanical reinforcement of weakened soft tissue, aiding in the treatment of conditions like hernia, prolapse, and urinary incontinence. No electronic components or software are involved.
Clinical Evidence
Bench testing only. No clinical data provided. Biocompatibility testing per ISO 10993-1 included cytotoxicity, sensitization, irritation, acute systemic toxicity, pyrogenicity, subacute/subchronic toxicity, implantation, genotoxicity, hemocompatibility, and LAL endotoxin. Mechanical testing included tensile strength, suture retention strength, delamination strength, tear strength, and stiffness.
Technological Characteristics
Porcine-derived urinary bladder matrix (ECM) scaffold; multi-layer sheet configuration; terminally sterilized via electron beam irradiation; biocompatibility per ISO 10993-1; mechanical properties include tensile, suture retention, delamination, tear, and stiffness.
Indications for Use
Indicated for patients requiring soft tissue reinforcement in urological, gastroenterological, gynecological, or plastic & reconstructive surgery, including hernia/body wall repair, prolapse repair, esophageal repair, and pubourethral support for urinary incontinence due to urethral hypermobility or intrinsic sphincter deficiency.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo, which features a stylized human figure. The logo on the right is the FDA U.S. Food & Drug Administration logo, with the FDA acronym in a blue square and the full name of the agency in blue text.
August 10, 2021
ACell, Inc. % Salman Elmi, VP/Deputy General Counsel, Chief Compliance Officer 6640 Eli Whitney Drive Columbia, MD 21046 Telephone: 410-953-8500 Email: salelmi4ace11.com
Re: K141084
Trade/Device Name: Matristem Surgery Matrix RS, PSM, PSMX, Matristem Pelvic Floor Matrix Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: FTM, OXH
Dear Mr. Salman Elmi:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated June 17, 2015. Specifically, FDA is updating this SE Letter to remove the indications for use for transvaginal repair of pelvic organ prolapse from the SE package and labeling in accordance with FDA's final rule, dated January 5, 2016 in which such devices were upclassified from Class II to Class III.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Jason Roberts, OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices, 240-402-6400, jason.roberts(@fda.hhs.gov.
Sincerely,
# Jason Roberts -S
Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Image /page/1/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of three faces in profile, one behind the other, with a stylized bird above them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 17, 2015
ACell Incorporated Mr. Miles Grody Senior Vice President, General Counsel 6640 Eli Whitney Drive, Suite 200 Columbia, Maryland 21046
Re: K141084
Trade/Device Name: MatriStem® Surgical Matrix RS, PSM, PSMX, and MatriStem® Pelvic Floor Matrix Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: FTM, OXH Dated: June 16, 2015 Received: June 17, 2015
Dear Mr. Grody:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{3}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
for
Sincerely yours,
# David Krause -S
- Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## 510(K) SUMMARY
| Submission Date: | July 6, 2018 |
|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer Name: | ACell, Inc. |
| Submitted by: | 6640 Eli Whitney Drive<br>Columbia, MD 21046 |
| Contact Person: | Andrea Pilon Artman<br>Sr. Regulatory Affairs Manager<br>ACell, Inc.<br>Phone: (410) 953-8549<br>Email: andreaartman@acell.com<br>Fax: 410-715-4511 |
### DEVICE NAME AND CLASSIFICATION
| 510 (k) Number: | K141084 |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------|
| Trade/Proprietary Name: | MatriStem® Surgical Matrix RS, PSM, PSMX,<br>MatriStem® Pelvic Floor Matrix |
| Common/Usual Name: | Surgical Mesh, ECM, Surgical Scaffold |
| Regulation Name: | Surgical Mesh |
| Device Class: | Class II, 21 CFR 878.3300 |
| Product Code: | FTM, OXH |
| Reviewing Panel: | General & Plastic Surgery |
| Predicate Devices: | ACell UBM Surgical Mesh (K040621)<br>ACell Surgical Mesh ML and ML Plus (K041140)<br>SIS Plastic Surgery Matrix (K034039) |
### DEVICE DESCRIPTION
MatriStem® Surgical Matrix RS, PSM, PSMX, and MatriStem® Pelvic Floor Matrix devices are composed of porcine-derived extracellular matrix scaffolds, specifically known as urinary bladder matrix. The devices are supplied in multi-layer sheet configurations in sizes up to 10 cm x 15 cm, and packaged in double peel-open pouches. The devices are terminally sterilized using electron beam irradiation.
### INDICATIONS FOR USE
MatriStem® Surgical Matrix RS (2-layer) and PSM (3-layer) are intended for implantation to reinforce soft tissue where weakness exists in patients requiring urological, gastroenterological, or plastic & reconstructive surgery. Reinforcement of soft tissue within urological, gastroenterological, and plastic & reconstructive surgery includes, but is not limited to, the following procedures: hemia and body wall repair, colon and rectal prolapse repair, and esophageal repair.
MatriStem® Surgical Matrix PSMX (6-layer) is intended for implantation to reinforce soft tissue where weakness exists in patients requiring gastroenterological or plastic & reconstructive surgery. Reinforcement of soft tissue within gastroenterological and plastic & reconstructive surgery
{5}------------------------------------------------
includes, but is not limited to, the following procedures: hernia and body wall repair, colon and rectal prolapse repair, tissue repair, and esophageal repair.
MatriStem® Pelvic Floor Matrix (6-layer) is intended for implantation to reinforce soft tissue where weakness exists in patients requiring urological or gynecological surgery. Reinforcement of soft tissue within urological and gynecological surgery includes, but is not limited to, the following procedures: pubourethral support, urethral repair, transabdominal reconstruction of pelvic floor and vaginal prolapse, bladder support, and sacrocolposuspension. By providing pubourethral support, MatriStem Pelvic Floor Matrix may be used for the treatment of urinary incontinence resulting from urethral hypermobility and intrinsic sphincter deficiency.
# EQUIVALENCE TO MARKETED DEVICES
MatriStem® Surgical Matrix RS, PSM, PSMX, and MatriStem® Pelvic Floor Matrix have the same intended use as predicate surgical meshes, which is to reinforce or repair soft tissue. The additional indications are identical to those included in the identified predicate devices. The technological characteristics of the MatriStem® Surgical Matrix RS, PSM, PSMX, and MatriStem® Pelvic Floor Matrix are substantially similar to the cleared predicates, as all are comprised of animal tissuederived, collagen extracellular matrix (ECM) scaffolds supplied in a sheet configuration that are packaged and terminally sterilized. The available sizes of the subject device (8 - 150 cm) are consistent with the range of sizes of the predicate devices (16 - 1350 cm -). The minor differences between the MatriStem® Surgical Matrix RS. PSM, PSMX, and MatriStem® Pelvic Floor Matrix and the identified predicates do not raise different questions of safety or efficacy and performance testing demonstrates that the devices have comparable performance to the predicates.
# Biocompatibility Testing
MatriStem® Surgical Matrix RS, PSM, PSMX, and MatriStem® Pelvic Floor Matrix underwent the following biocompatibility testing on sterilized devices per ISO-10993-1 standard: cytotoxicity. sensitization, irritation/intracutaneous reactivity, acute systemic toxicity, pyrogenicity, subacute and subchronic toxicity and implantation, genotoxicity, hemocompatibility, and LAL endotoxin. The results of these tests provided evidence that the MatriStem® Surgical Matrix RS, PSM, PSMX, and MatriStem Pelvic Floor Matrix meet biocompatibility requirements of the ISO standard.
# Mechanical Testing
The MatriStem Surgical Matrix RS, PSM, PSMX, and MatriStem® Pelvic Floor Matrix material were tested for the following: tensile strength, suture retention strength, delamination strength, tear strength, and stiffness test. The mechanical testing provided evidence that MatriStem® Surgical Matrix RS, PSM, PSMX, and MatriStem® Pelvic Floor Matrix provide adequate mechanical strength for their respective applications.
# CONCLUSION
Based on direct testing and comparison to predicate devices, MatriStem® Surgical Matrix RS, PSM, PSMX, and MatriStem® Pelvic Floor Matrix do not raise different questions of safety and effectiveness and the results support a determination of substantial equivalence through this 510(k) Premarket Notification.
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141084
#### Device Name
MatriStem Surgical Matrix RS, PSM, PSMX, and MatriStem Pelvic Floor Matrix
#### Indications for Use (Describe)
MatriStem Surgical Matrix RS (2-layer) are intended for implantation to reinforce soft tissue where weakness exists in patients requiring urological, or plastic & reconstructive surgery. Reinforcement of soft tissue within urological, gastroenterological, and plastic & reconstructive surgery includes, but is not limited to, the following procedures: hernia and body wall repair, colon and rectal prolapse repair, and esophageal repair.
MatriStem Surgical Matrix PSMX (6-layer) is intended for implantation to reinforce soft tissue where weakness exists in patients requiring gastroenterological or plastic & reconstructive surgery. Reinforcement of soft tissue within gastroenterological and plastic & reconstructive surgery includes, but is not limited to, the following procedures: hernia and body wall repair, colon and rectal prolapse repair, and esophageal repair.
MatriStem Pelvic Floor Matrix (6-layer) is intended for implantation to reinforce soft tissue where weakness exists in patients requiring urological or gynecological surgery. Reinforcement of soft tissue within urological and gynecological surgery includes, but is not limited to, the following procedures: pubourethral repair, transabdominal reconstruction of pelvic floor and vaginal prolapse repair, bladder support, and sacrocolposuspension. By providing pubourethral support, Matristem Pelvic Floor Matrix may be used for the treatment of urinary incontinence resulting from urethral hypermobility and intrinsic sphincter deficiency.
| Type of Use (Select one or both, as applicable) | | | | |
|--------------------------------------------------|---------------------------------------------|--|--|--|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | | | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | | |
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.