K132025 · Kensey Nash Corporation Dba Dsm Biomedical · FTM · Oct 30, 2013 · General, Plastic Surgery
Device Facts
Record ID
K132025
Device Name
MESO BILAYER SURGICAL MESH
Applicant
Kensey Nash Corporation Dba Dsm Biomedical
Product Code
FTM · General, Plastic Surgery
Decision Date
Oct 30, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Meso Bilayer Surgical Mesh is intended for implantation to reinforce soft tissues where weakness exists in patients requiring soft tissue repair and reinforcement in plastic and reconstructive surgery including but not limited to the following procedures: reinforcement of primary closure such as suture line reinforcement and muscle flap reinforcement; hernia repair (e.g. hiatal, femoral, paracolostomy, umbilical.) Meso Bilayer Surgical Mesh is supplied sterile and for one time use.
Device Story
Meso Bilayer Surgical Mesh is a resorbable implant derived from porcine tissue and a synthetic absorbable polymer; designed to reinforce soft tissue where weakness exists. Supplied sterile and dry in double-layer packaging; requires hydration with saline or autologous body fluids (blood, bone marrow aspirate, platelet-rich plasma) prior to surgical implantation. Used by surgeons in plastic and reconstructive procedures. Device provides mechanical reinforcement to tissue; facilitates normal tissue healing and remodeling. Benefits include structural support for soft tissue defects.
Clinical Evidence
No human clinical data. Evidence includes biocompatibility testing per ISO 10993-1:2009 (cytotoxicity, sensitization, irritation, toxicity, genotoxicity, hemocompatibility), viral inactivation, residual chemical assessment, and biomechanical bench testing (tensile strength, burst, tear, suture retention). Animal implant studies confirmed normal tissue healing and remodeling capability.
Technological Characteristics
Resorbable bilayer surgical mesh composed of porcine tissue and synthetic absorbable polymer. Sterilized via Ethylene Oxide. Single-use device. Biocompatible per ISO 10993-1:2009. Mechanical properties evaluated via tensile, burst, tear, and suture retention testing.
Indications for Use
Indicated for patients requiring soft tissue repair and reinforcement in plastic and reconstructive surgery, including reinforcement of primary closure (suture line, muscle flap) and hernia repair (hiatal, femoral, paracolostomy, umbilical).
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the text "K132025 Page 1/4" at the top. Below this text is a logo for DSM, with the tagline "BRIGHT SCIENCE. BRIGHTER LIVING." The DSM logo is in bold, black letters. To the left of the DSM logo is a black and white graphic.
DSM Biomedical 735 Pennsylvania Drive Exton, PA 19341 USA
www.dsm.com/medical
## Submitted By:
Contact Person:
# DSM Biomedical 735 Pennsylvania Drive Exton, PA 19341 Susan Pileggi Regulatory Specialist
510(k) SUMMARY
E: Susan. Pileggi@dsm.com P: 484-713-2100 F: 484-713-2903 May 28, 2013
Meso Bilayer Surgical Mesh
OCT 3 0 2013
Date Prepared: Device:
| Trade Name: | |
|----------------------------|--|
| Common/Usual Name: | |
| Classification Name: | |
| Classification Regulation: | |
| Device Class: | |
| Device Code: | |
| Advisory Panel: | |
Surgical Mesh Mesh, Surgical 21 CFR 878.330 Class II FTM, OXH General and Plastic Surgery
Predicate:
K094061: Kensey Nash ECM Surgical Patch [Kensey Nash Corporation]
#### Device Description:
Meso Bilayer Surgical Mesh is a resorbable surgical mesh intended to reinforce soft tissue where weakness exists. The implant is derived from porcine tissue and a synthetic absorbable polymer. The material is supplied sterile in double-layer packages. The implant is packaged dry and prior to use is hydrated with saline or autologous body fluids such as blood, bone marrow aspirate, or blood concentrates such as platelet rich plasma.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the DSM logo with the text "BRIGHT SCIENCE. BRIGHTER LIVING." The image also contains the text "K132025 Page 2/4" at the top. The DSM logo is a stylized image of a molecule or chemical structure. The text is in a bold, sans-serif font.
#### Intended Use:
Meso Bilayer Surgical Mesh is intended for implantation to reinforce soft tissues where weakness exists in patients requiring soft tissue repair and reinforcement in plastic and reconstructive surgery including but not limited to the following procedures: reinforcement of primary closure such as suture line reinforcement and muscle flap reinforcement; hernia repair (e.g. hiatal, femoral, paracolostomy, umbilical.)
Meso Bilayer Surgical Mesh is supplied sterile and for one time use.
## Technological Characteristics:
The intended use, product design, material, and function of Meso Bilayer Surgical Mesh is substantially equivalent to the FDA cleared predicate device Kensey Nash ECM Surgical Patch (K094061).
| Characteristic | Meso Bilayer Surgical Mesh | Kensey Nash ECM Surgical Patch |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | Meso Bilayer Surgical Mesh is<br>intended for implantation to<br>reinforce soft tissues where<br>weakness exists in patients<br>requiring soft tissue repair<br>and reinforcement in plastic<br>and reconstructive surgery<br>including but not limited to<br>the following procedures:<br>reinforcement of primary<br>closure such as suture line<br>reinforcement and muscle<br>flap reinforcement; hernia<br>repair (e.g. hiatal, femoral,<br>paracolostomy, umbilical.)<br>Meso Bilayer Surgical Mesh is<br>supplied sterile and for one<br>time use. | The Kensey Nash ECM Surgical Patch is<br>intended for implantation to reinforce soft<br>tissues where weakness exists in patients<br>requiring soft tissue repair, reinforcement<br>in plastic and reconstructive surgery, and in<br>the urological, gynecological,<br>gastroenterological anatomy including but<br>not limited to the following procedures:<br>reinforcement of primary closure such as<br>suture line reinforcement and muscle flap<br>reinforcement; hernia repair (e.g., hiatal,<br>femoral, paracolostomy, umbilical),<br>urethral and vaginal prolapse repair, colon<br>and rectal prolapse repair, reconstruction<br>of the pelvic floor, bladder support, and<br>sacrocolposuspension.<br>The device is supplied sterile and intended<br>for one time use. |
-
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the text "K132025 Page 3/4" at the top. Below this text is a logo with a circular design on the left and the letters "DSM" in bold font on the right. Underneath the letters, the text "BRIGHT SCIENCE. BRIGHTER LIVING." is displayed. The image appears to be a page from a document or report.
| Characteristic | Meso Bilayer Surgical Mesh | Kensey Nash ECM Surgical Patch | |
|------------------------|----------------------------------|---------------------------------------|--|
| Origin | Porcine tissue | Porcine tissue | |
| Device | Resorbable double layer surgical | Resorbable single layer surgical mesh | |
| Characteristics | mesh | | |
| Biocompatibility Yes | | Yes | |
| Reusable | Single Use Device | Single Use Device | |
| Shelf Life | 6 months | 36 months | |
| Sterilization | Ethylene Oxide | Ethylene Oxide | |
| Method | | | |
| Packaging | Double peel packages | Double peel packages | |
## Biocompatibility and Performance Data:
Biocompatibility testing, biomechanical bench testing, characterization testing and in vivo performance testing have been conducted to evaluate the biological safety and biomechanical performance characteristics of Meso Bilayer Surgical Mesh.
Biocompatibility testing was completed on the finished sterile device in accordance with the requirements of ISO 10993-1: 2009, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. Testing included Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Genotoxicity, Hemocompatibility, Subacute Systemic Toxicity, and Chronic Systemic Toxicity. Other safety testing included a viral inactivation study and residual chemical assessment. Results indicate that the device's biocompatibility profile is acceptable.
Biomechanical testing included tensile strength burst testing, wet tear testing, and suture retention. Testing results indicate that the device is equivalent to the predicate device and meets the requirements for its intended use.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for DSM, which includes a stylized graphic to the left and the letters "DSM" in bold font. Below the letters, the text "BRIGHT SCIENCE. BRIGHTER LIVING." is printed. The text "K132025 Page 4/4" is at the top of the image.
Animal implant studies were performed to confirm the functionality and tissue response characteristics of proposed device. Results indicate a normal tissue healing response and confirm the device's remodeling capability.
## Substantial Equivalence:
Performance testing has confirmed that the Meso Bilayer Surgical Mesh is substantially equivalent to the predicate device with regard to materials, intended use, operation, and function, and technological characteristics, pursuant to section 510(k).
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized graphic of three horizontal lines that curve upwards and to the right, resembling a stylized human figure.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Kensey Nash Corporation dba DSM Biomedical Ms. Susan Pileggi Regulatory Specialist 735 Pennsylvania Drive Exton, Pennsylvania 19341
October 30, 2013
Re: K132025
Trade/Device Name: Meso Bilayer Surgical Mesh Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: FTM, OXH Dated: October 17, 2013 Received: October 18, 2013
Dear Ms. Pileggi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 - Ms. Susan Pileggi
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### Indications For Use Statement
510(k) Number: _K132025
Device Name: Meso Bilayer Surgical Mesh
Indications For Use:
Meso Bilayer Surgical Mesh is intended for implantation to reinforce soft tissues where weakness exists in patients requiring soft tissue repair and reinforcement in plastic and reconstructive surgery including but not limited to the following procedures: reinforcement of primary closure such as suture line reinforcement and muscle flap reinforcement; hernia repair (e.g. hiatal, femoral, paracolostomy, umbilical.)
Meso Bilayer Surgical Mesh is supplied sterile and for one time use.
| Prescription Use | X |
|-----------------------------|---|
| (Per 21 CFR 801 Subpart D) | |
| AND/OR Over-The-Counter Use | |
| (Per 21 CFR 801 Subpart C) | |
Concurrence of CDRH, Office of Device Evaluation (ODE)
# David Krause -S
(Division Sign-Off) Division of Surgical Devices 510(k) Number: K132025
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.