K023237 · Xylos Corporation · FTM · Mar 18, 2003 · General, Plastic Surgery
Device Facts
Record ID
K023237
Device Name
XYLOS SURGICAL MESH
Applicant
Xylos Corporation
Product Code
FTM · General, Plastic Surgery
Decision Date
Mar 18, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
XYLOS™ Surgical Mesh is intended for implantation to reinforce soft tissue including, but not limited to: defects of the abdominal and thoracic wall, muscle flap reinforcement, prolapse repair, reconstruction of the pelvic floor, hernias, suture line reinforcement and reconstructive procedures. The device is intended for one-time use.
Device Story
XYLOS™ Surgical Mesh is a sterile, implantable device derived from microbial cellulose. It undergoes chemical processing, including solvent dehydration, to achieve desired mechanical properties. The mesh provides tensile strength and suture retention for soft tissue reinforcement. It is supplied in various sizes, double-pouched, and gamma-sterilized for single-use. Surgeons implant the mesh during reconstructive or hernia repair procedures to support or reinforce weakened tissue. The device functions as a permanent or semi-permanent scaffold to facilitate tissue repair.
Clinical Evidence
Bench testing only. Evidence includes safety, biocompatibility, toxicity, pyrogenicity, sterility, and mechanical strength testing.
Indicated for soft tissue reinforcement in patients requiring repair of abdominal/thoracic wall defects, muscle flaps, pelvic floor reconstruction, prolapse repair, hernia repair, or suture-line reinforcement.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
Predicate Devices
Bio-Vascular, Inc. Peri-Guard® Pericardium (K983162)
Ethicon, Inc. Prolene Polypropylene Mesh, Nonabsorbable (K962530)
{0}------------------------------------------------
K023237
# 510(k) Summary
#### DATE PREPARED:
09SEP2002
#### SUBMITTED BY:
XYLOS Corporation 41 University Drive, Suite 400 Newtown, PA 18940
## CONTACT PERSON:
Patsy J. Trisler, J.D., RAC Senior Director, Medical Device Consulting PharmaNet, Inc. 815 Connecticut Avenue, NW, Suite 800 Washington, D.C. 20006 202-835-1346 (direct dial) 609-520-8107 (fax)
### DEVICE:
Mesh, Surgical Classification Name: Common/Usual Name: Surgical Mesh XYLOS™ Surgical Mesh Proprietary Name:
## DEVICE CLASSIFICATION:
Product Code/Classification Number: FTM (878.3300) Requlatory Class:
#### PREDICATE DEVICES:
Bio-Vascular, Inc. Peri-Guard® Pericardium K983162 Ethicon, Inc. Prolene Polypropylene Mesh, Nonabsorbable K962530 XYLOS™ XCell® Wound Dressing K974251
#### STATEMENT OF SUBSTANTIAL EQUIVALENCE:
XYLOS™ Surgical Mesh is substantially equivalent to the predicate devices, having similar intended use, technological characteristics, and performance.
#### INTENDED USE:
XYLOS™ Surgical Mesh is intended for implantation to reinforce soft tissue including, but not limited to: defects of the abdominal and thoracic wall, muscle flap reinforcement, prolapse repair, reconstruction of the pelvic floor, hernias, suture line reinforcement and reconstructive procedures. The device is intended for one-time use.
## DEVICE DESCRIPTION:
XYLOS™ Surqical Mesh is prepared from microbial derived cellulose that undergoes chemical processing inclusive of solvent dehydration. Units are cut in various sizes to meet surgical needs. Each unit is double-pouched, labeled, and gamma sterilized surgical mesh exhibits excellent tensile strength, suture retention and consistent thickness.
## PERFORMANCE DATA:
XYLOS™ Surgical Mesh functions as intended when subjected to safety, biocompatibility, toxicity, pyrogenicity, sterility and mechanical strength testing.
MAR 1 8 2003
000017
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three human profiles incorporated into its design. The eagle is positioned diagonally, facing upwards and to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle.
MAR 1 8 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Xylos Corporation c/o Ms. Patsy J. Trisler, J.D., RAC Senior Director, Medical Device Consulting PharmaNet, Inc. 815 Connecticut Avenue NW, Suite 800 Washington, D.C. 20006
Re: K023237
Trade/Device Name: XYLOS™ Surgical Mesh Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: II Product Code: FTM Dated: February 5, 2003 Received: February 7, 2003
Dear Ms. Trisler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2 - Ms. Patsy J. Trisler, J.D., RAC
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Muriam C. Provost
Po/ Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use FORM
# Page 1 of 1
**510(k) Number (if known)** KC23237
Device Name:
XYLOS™ Surgical Mesh
This device is intended for implantation to Indications for Use: reinforce soft tissue including, but not limited to: defects of the abdominal and thoracic wall, muscle flap reinforcement, prolapse repair, reconstruction of the pelvic floor, hernias, suture-line reinforcement and reconstructive procedures. The device is intended for one-time use.
# (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21CFR 801.109)
OR
Over-the-Counter-Use Optional Format 1-2-96
Muriam C. Provost
(Division Sign-Off) Division of General, Restorative and Neurological Devices
000037
510(k) Number K023237
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.