VENTRALIGHT (TM) ST MESH WITH ECHO PS (TM) POSITIONING SYSTEM
K110820 · C.R. Bard, Inc. · FTL · Apr 1, 2011 · General, Plastic Surgery
Device Facts
Record ID
K110820
Device Name
VENTRALIGHT (TM) ST MESH WITH ECHO PS (TM) POSITIONING SYSTEM
Applicant
C.R. Bard, Inc.
Product Code
FTL · General, Plastic Surgery
Decision Date
Apr 1, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Ventralight™ ST Mesh is indicated for use in the reconstruction of soft tissue deficiencies, such as for the repair of hernias. The Echo PSTM Positioning System is intended to be used to facilitate the delivery of soft tissue prosthetics during laparoscopic hernia repair.
Device Story
Ventralight™ ST Mesh with Echo PS™ Positioning System combines a surgical mesh with a removable deployment balloon; used in laparoscopic hernia repair. Mesh consists of co-knitted polypropylene and polyglycolic acid fibers; visceral side features bioresorbable hydrogel coating (sodium hyaluronate, carboxymethylcellulose, polyethylene glycol) to minimize tissue attachment. Positioning system is an RF-welded nylon balloon attached via polycarbonate connectors; facilitates unrolling and placement of mesh. Introducer tool (stainless steel rod, T-cap) aids delivery through trocar sites. Device operated by surgeons in clinical/OR settings. Positioning system inflated via manual inflation assembly. Mesh allows fibroblastic response for tissue ingrowth; coating resorbs within 30 days. System simplifies laparoscopic mesh deployment, potentially reducing operative time and improving placement accuracy.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing and in vivo simulated use experiments. Biocompatibility testing performed in accordance with ISO 10993-1 standards. Stainless steel components evaluated per ASTM F899-09e1.
Technological Characteristics
Mesh: Co-knitted polypropylene and polyglycolic acid fibers; bioresorbable hydrogel coating (sodium hyaluronate, carboxymethylcellulose, polyethylene glycol). Positioning system: RF-welded nylon balloon with polycarbonate connectors. Introducer tool: 304 Stainless Steel (ASTM F899-09e1) rod and T-cap. Sterile, single-use. Non-electronic, mechanical deployment system.
Indications for Use
Indicated for patients requiring reconstruction of soft tissue deficiencies, specifically hernia repair. Used during laparoscopic procedures to facilitate delivery of soft tissue prosthetics.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
Composix™ L/P Mesh with Echo PS™ Positioning System (K102766)
Submission Summary (Full Text)
{0}------------------------------------------------
K110820
Page 1/3
## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
APR - 1 2011
This 510(k) Summary is provided per the requirements of section 807.92(c).
#### Submitter Information:
| Submitter's Name: | Davol, Inc., Subsidiary of C. R. Bard, Inc. |
|----------------------|---------------------------------------------|
| Contact Person: | Keti Sino |
| | Regulatory Affairs Specialist |
| Address: | 100 Crossings Boulevard |
| | Warwick, RI 02886 |
| Telephone: | (401) 825-8575 |
| Fax: | (401) 825-8763 |
| Email: | Keti.sino@crbard.com |
| Device Name: | |
| Trade Name: | VentralightTM ST Mesh with Echo PSTTM |
| | Positioning System |
| Common/Usual Name: | - Surgical Mesh |
| | - Endoscope & Accessories |
| | - Mesh Deployment Balloon |
| Classification Name: | - Mesh, Surgical, Polymeric |
| | - Mesh Deployment Balloon |
| | - Laparoscope, General & Plastic Surgery |
| Classification Code: | - Class II, § 878.3300, Product Code FTL |
| Subsequent Codes: | - Class II, § 878.3300, Product Code OQL |
| | - Class II, § 876.1500, Product Code GCJ |
## Predicate Device Names:
- Ventralight™ ST Mesh, K101851 (Davol Inc.), FDA cleared on 07/15/2010 .
- Composix™ L/P Mesh with Echo PS™ Positioning System, K102766 (Davol Inc.), . FDA cleared on 12/16/2010
#### Device Description:
The proposed Ventralight™ ST Mesh with Echo PSTM Positioning System is comprised of a hernia repair mesh with a pre-attached mesh positioning system (mesh deployment balloon). All sizes of the proposed product with the exception of the largest size will also include an Introducer Tool designed to aid the user in laparoscopic introduction of the mesh/positioning system assembly.
# PREMARKET NOTIFICATION FOR THE VENTRALIGHT™ ST MESH WITH ECHO PSTM POSITIONING SYSTEM
### -CONFIDENTIAL-SECTION 8
{1}------------------------------------------------
K110820 page 2/3
The Ventralight™ ST Mesh is a low profile, sterile, single use device indicated for use in the reconstruction of soft tissue deficiencies, such as for the repair of hernias. This device is co-knitted using polypropylene (PP) and polyglycolic acid (PGA) fibers to result in a two-sided mesh with a PP surface and a PGA surface. The mesh is coated on the PGA surface with a bioresorbable, chemically modified sodium hyaluronate (HA), carboxymethylcellulose (CMC) and polyethylene glycol (PEG) based hydrogel. The fascial side of the mesh allows a prompt fibroblastic response through the interstices of the mesh, allowing for complete tissue ingrowth, similar to polypropylene mesh alone. The visceral side of the mesh is a bioresorbable coating, separating the mesh from underlying tissue and organ surfaces to minimize tissue attachment to the mesh. Shortly after placement, the biopolymer coating becomes a hydrated gel that is resorbed from the site in less than 30 days.
The Ventralight™ ST Mesh described above will be packaged preassembled with the Echo PSTM Positioning System, a removable mesh deployment balloon, previously cleared under K102766. The positioning system is an RF welded nylon balloon that is pre-attached to the mesh via the use of polycarbonate connectors, and is designed to help facilitate laparoscopic deployment, including unrolling, positioning, and placement of the Ventralight™ ST Mesh. The positioning system is inflated via use of the included inflation assembly, which is also identical to that cleared with the predicate positioning system under K102766.
For user convenience, all sizes of the proposed product with the exception of the largest size also include an Introducer Tool which is intended to facilitate delivery to the operative site. The largest configuration of the proposed product, the 12" x 14" size does not include an Introducer Tool as this size is recommended for deployment directly through a trocar incision site. The Introducer Tool included with the proposed product is identical to the Introducer Tool cleared under K102766 which consists of a handle with a thin stainless steel rod and a T-cap. The device is split into two pieces along the length of the entire rod and attached on one end to the handle.
#### Intended Use:
The Ventralight™ ST Mesh is indicated for use in the reconstruction of soft tissue deficiencies, such as for the repair of hernias. The Echo PSTM Positioning System is intended to be used to facilitate the delivery of soft tissue prosthetics during laparoscopic hernia repair.
Summary of Similarities and Differences in Technological Characteristics, Performance and Intended Use:
The Ventralight™ ST Mesh and the predicate mesh have the same indication: reconstruction of soft tissues deficiencies such as the repair of hernias. Further, the proposed product has the same intended use, physical attributes, and performance characteristics, as the predicate Ventralight™ ST Mesh.
## PREMARKET NOTIFICATION FOR THE VENTRALIGHT™ ST MESH WITH ECHO PSTM POSITIONING SYSTEM
#### -CONFIDENTIAL SECTION 8
{2}------------------------------------------------
K110820 Page 3/3
The Echo PSTM Positioning System maintains the same intended use, physical attributes, and performance characteristics, as the cleared Echo PSTM Positioning System (K102766). Further, the included inflation assembly intended to be used for the purposes of inflating the positioning system is also identical to the inflation assembly cleared under K102766.
The proposed product will combine into one the currently marketed Ventralight™ ST Mesh cleared under K101851, preassembled with the currently marketed Echo PSTM Positioning System cleared as part of K102766. The proposed product will also include an Introducer Tool and inflation assembly, both identical to those previously cleared as part of the positioning system predicate (K102766).
#### Performance Data:
Biocompatibility testing in accordance to ISO 10993-1 standards was conducted on the proposed device (mesh and positioning system) and the results indicate that the device is biocompatible per these standards.
No biocompatibility testing was conducted on the Introducer Tool to be packaged with the proposed device as the 304 Stainless Steel material used for the rolling tines is a recognized biocompatible material as per ASTM F899 - 09e1 Standard Specification for Wrought Stainless Steels for Surgical Instruments. In addition, the polymer handle attached to the rolling tines and the T-cap is non patient contacting, and therefore, not subject to biocompatibility testing requirements.
Bench testing results and in vivo simulated use experiments demonstrate that the proposed device design meets product specifications and intended uses.
All test results provided in this submission support the safety and effectiveness of the device for its intended use and demonstrate that the proposed device is substantially equivalent to its predicate devices.
# PREMARKET NOTIFICATION FOR THE VENTRALIGHT™ ST MESH WITH ECHO PSTM POSITIONING SYSTEM
### -CONFIDENTIAL-SECTION 8
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans. The eagle is stylized and abstract.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
C. R. Bard, Inc. % Davol, Inc. Keti Sino 100 Crossings Boulevard Warwick, Rhode Island 02886
APR - 1 2011
Re: K110820 .
Trade/Device Name: Ventralight™ ST Mesh with Echo PS™ Positioning System Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: FTL, OQL, GCJ Dated: March 22, 2011 Received: March 23, 2011
Dear Keti Sino:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{4}------------------------------------------------
# Page 2 – Keti Sino
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Aj B. n h
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### INDICATION FOR USE STATEMENT
K 110820 510(k) Number (if known):
Device Name:
# Ventralight™ ST Mesh with Echo PSTM Positioning System
Ventralight™ ST Mesh is indicated for use in the reconstruction of soft tissue deficiencies, such as for the repair of hernias. The Echo PSTM Positioning System is intended to be used to facilitate the delivery of soft tissue prosthetics during laparoscopic hemia repair.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel Kraefer MXM
Division Sign-Off Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K110820
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.