INTUITIVE SURGICAL ONSITE FOR DA VINCI SURGICAL SYSTEMS
K121921 · Intuitive Surgical, Inc. · NAY · Oct 25, 2012 · Gastroenterology, Urology
Device Facts
Record ID
K121921
Device Name
INTUITIVE SURGICAL ONSITE FOR DA VINCI SURGICAL SYSTEMS
Applicant
Intuitive Surgical, Inc.
Product Code
NAY · Gastroenterology, Urology
Decision Date
Oct 25, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Software as a Medical Device
Intended Use
OnSite for da Vinci Surgical Systems is an accessory intended for use by trained Intuitive Surgical Field Service personnel for troubleshooting and servicing da Vinci Surgical Systems through either a wired or wireless Ethernet connection between the da Vinci Surgical System and the hospital's Internet Protocol (IP) infrastructure.
Device Story
OnSite is a software accessory for da Vinci Surgical Systems; enables remote access for Intuitive Surgical field service personnel. Connects via wired or wireless Ethernet to hospital IP infrastructure. Inputs include system diagnostic data; outputs include system configuration updates and fault diagnostic information. Used by trained service technicians for troubleshooting and maintenance. Facilitates remote system management; improves service efficiency; ensures system functionality. Does not perform surgery; strictly for service/maintenance context.
Clinical Evidence
Bench testing only. Software verification and in vitro validation testing conducted to confirm performance and ensure no adverse impact on da Vinci Surgical System functionality. All test cases passed.
Technological Characteristics
Software accessory; utilizes wired or wireless Ethernet connectivity to hospital IP infrastructure. Identical to predicate devices. No hardware materials or energy sources specified.
Indications for Use
Indicated for use by trained Intuitive Surgical Field Service personnel to obtain system information for fault diagnosis and to remotely enable/disable features, including configuration updates, via wired or wireless Ethernet connection to hospital IP infrastructure.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Predicate Devices
Intuitive Surgical® da Vinci® S™ Surgical System, Model IS2000 with da Vinci OnSite™ and da Vinci Connect™ (K081207)
Intuitive Surgical® da Vinci® S™ Surgical System, Model IS2000 with Connect™ OnSite™ and Wireless Connectivity Option (K101581)
Related Devices
K081207 — INTUITIVE SURGICAL DA VINCI S SURGICAL SYSTEM, MODEL IS2000, WITH DA VINCI CONNECT AND ONSITE · Intuitive Surgical, Inc. · Dec 19, 2008
K123840 — CONNECT FOR DA VINCI SURGICAL SYSTEM(S) · Intuitive Surgical, Inc. · Feb 14, 2013
K192876 — INTELLIO Tablet Application · Smith & Nephew · Feb 27, 2020
K251202 — da Vinci Surgical System (IS5000) · Intuitive Surgical, Inc. · Jul 16, 2025
K161271 — da Vinci Xi Surgical System · Intuitive Surgical, Inc. · Jul 11, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
/21/21
## 510(K) SUMMARY
OCT 25 2012
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the SMDA 1990 and 21 CFR 807.92.
## 510(k) Number:
| Date | June 29, 2012 |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter | Intuitive Surgical, Inc.<br>1266 Kifer Road<br>Sunnyvale, CA 94086 |
| Contact | Cherece L. Jones<br>Sr. Regulatory Affairs Specialist<br>Telephone: (408) 523 - 6925<br>Fax: (408) 523 - 1390<br>e-mail: cherece.jones@intusurg.com |
| Subject<br>Device | Trade Name: Intuitive Surgical® OnSite™ for da Vinci® Surgical<br>Systems |
| | Common Name: Endoscopic Instrument Control System,<br>Endoscopic Instruments, and Accessories |
| | Classification Name: General and Plastic Surgery |
| Predicate Devices | Intuitive Surgical® da Vinci® S™ Surgical System, Model IS2000<br>with da Vinci OnSite™ and da Vinci Connect™ (legally marketed<br>under K081207). |
| | Intuitive Surgical® da Vinci® S™ Surgical System, Model IS2000<br>with Connect™ OnSite™ and Wireless Connectivity Option<br>(legally marketed under K101581). |
| Device<br>Description | OnSite is a software accessory compatible with da Vinci Surgical<br>Systems and allows Intuitive Surgical, Inc. (ISI) personnel to<br>remotely, through either a wired or wireless Ethernet connection<br>between the da Vinci Surgical System and the hospital's Internet<br>Protocol (IP) infrastructure: |
p. 1/2
{1}------------------------------------------------
| | Protocol (IP) infrastructure: |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | • Obtain system information for the purpose of diagnosing faults |
| | • Enable/disable features including configuration updates |
| Intended Use | OnSite for da Vinci Surgical Systems is an accessory intended for use by trained Intuitive Surgical Field Service personnel for troubleshooting and servicing da Vinci Surgical Systems through either a wired or wireless Ethernet connection between the da Vinci Surgical System and the hospital's Internet Protocol (IP) infrastructure. |
| Technological<br>Characteristics | The technological characteristics of the subject device are identical to the predicate devices. |
| Performance<br>Data | Software Verification Testing and <i>In Vitro</i> Validation Testing were conducted to confirm that the new OnSite device performed as intended and that the proposed software changes had no adverse impact on the functionality of the <i>da Vinci</i> Surgical System. All test cases successfully passed demonstrating that the subject device is substantially equivalent to the predicate devices. |
| Conclusion | The modified OnSite software has the same Intended Use, device design and technological characteristics as the predicate devices (K081207 and K101581).<br><br>Performance data supports that the OnSite software, with expanded capabilities, is substantially equivalent to the predicate devices in terms of safety and effectiveness. |
P. 2/2
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a circular design with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is a stylized depiction of an eagle, which is a common symbol of the United States.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
OCT 25 2012
Intuituve Surgical, Incorporated % Ms. Cherece L. Jones Senior Regulatory Affairs Specialist 1266 Kifer Road Sunnyvale, California 94086 .
Re: K121921
Trade/Device Name: Intuitive Surgical® OnSite™ for da Vincio Surgical Systems Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: NAY Dated: October 11, 2012 Received: October 12, 2012
Dear Ms. Jones:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Ms. Cherece L. Jones
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson
Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## INDICATIONS FOR USE
长121921 510(k) Number if known:
Device Name: Intuitive Surgical® OnSite™ for da Vincia Surgical Systems
OnSite for da Vinci Surgical Systems is an accessory indicated for use by trained Intuitive Surgical Field Service personnel to: (1) obtain system information for the purpose of diagnosing faults, (2) remotely enable/disable features including configuration updates through either a wired or wireless Ethernet connection between the da Vinci Surgical System and the hospital's Internet Protocol (IP) infrastructure.
Prescription Use X (Per 21 CFR 801 Subpart D) Subpart C)
AND/OR
Over-the-Counter Use (Per 21 CFR 807
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nikke for mxm
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K121921
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.