5MM FLARED CANNULA MODEL 420262, 8MM FLARED CANNULA MODEL 420319
K101743 · Intuitive Surgical, Inc. · NAY · Feb 4, 2011 · Gastroenterology, Urology
Device Facts
Record ID
K101743
Device Name
5MM FLARED CANNULA MODEL 420262, 8MM FLARED CANNULA MODEL 420319
Applicant
Intuitive Surgical, Inc.
Product Code
NAY · Gastroenterology, Urology
Decision Date
Feb 4, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Flared Cannulae are Intended to be used with the da Vinci Surgical System (da Vinci S (Model IS2000) or da Vinci SI (Model IS3000)) to serve as a port of entry during da Vinci procedures that do not require maintenance of Insufflation.
Device Story
Intuitive Surgical Flared Cannula (5mm and 8mm models) serves as a port of entry for endoscopic instruments during robotic-assisted surgery. Device consists of a tapered hollow tubular shaft with a flared distal tip and a bowl for attachment to the da Vinci Surgical System (IS2000/IS3000). Includes an integral receptacle for connection to a dispersive electrode grounding pad via a 2-pin connector. Used in clinical settings by surgeons trained on the da Vinci system. Provides a stable access point for instruments; does not maintain insufflation. Benefits include facilitating instrument exchange and electrical grounding during procedures.
Clinical Evidence
No clinical data. Performance demonstrated via bench testing and animal lab tests to confirm design output meets design input requirements and safety/efficacy equivalence to the predicate.
Technological Characteristics
Tapered hollow tubular shaft with flared distal tip and bowl receptacle. Includes 2-pin connector for dispersive electrode grounding pad. Mechanical accessory for robotic surgical systems. No energy source or software.
Indications for Use
Indicated for use as a port of entry during da Vinci surgical procedures that do not require maintenance of insufflation, utilizing the da Vinci S (IS2000) or da Vinci Si (IS3000) systems.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Predicate Devices
da Vinci Endoscopic Instrument Control System (Model IS1000) and Endoscopic Instruments (K990144)
K143217 — 12 mm & Stapler Bladeless Obturators · Intuitive Surgical, Inc. · Dec 3, 2014
K181395 — Universal Cannula Seal (5-12 mm) · Intuitive Surgical, Inc. · Jul 27, 2018
K153126 — da Vinci Xi Hasson Cone · Intuitive Surgical, Inc. · Dec 9, 2015
K101671 — SINGLE SITE FLEXIBLE BLADELESS OBTURATOR · Teleflexmedical, Inc. · Jun 25, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
K101743, Intuitive Surgical 5 mm and 8 mm Flared Cannula
## FEB - 4 2011
## 510(k) Summary
(As Required by 21 CFR 807.92(c))
| Date: | 09/10/2010 | |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|
| Submitter: | Intuitive Surgical, Inc.<br>1266 Kifer Road<br>Sunnyvale, CA 94086 | |
| Official Contact: | Meghna Sridharan<br>Sr. Regulatory Engineer<br>Ph: (408) 523- 2487<br>Fax: (408) 523-1390<br>Meghna.sridharan@intusurg.com | |
| Trade Name: | Intuitive Surgical Flared Cannula<br>• Model # 420262: 5mm Flared Cannula<br>• Model # 420319: 8mm Flared Cannula | |
| Common Name: | Endoscopic Instrument and Accessories | |
| Classification: | Endoscope and Accessories,<br>21 CFR 876.1500, NAY(LFL) | |
| Predicate Device: | Intuitive Surgical, Inc.<br>K990144, da Vinci Endoscopic Instrument<br>Control System (Model IS1000) and Endoscopic<br>Instruments | |
| Device Description: | The Intuitive Surgical Flared Cannula consists of<br>a tapered hollow tubular shaft with a flared<br>distal tip and a bowl with an Integral<br>receptacle. The hollow shaft serves as the port<br>of entry, the bowl is used to attach the<br>cannula to the da Vinci Surgical System and<br>the receptacle allows for attachment of<br>commercially available dispersive electrode<br>grounding pad via a 2-Pin grounding<br>connector | |
| Intended Use: | The Flared Cannulae are Intended to be used<br>with the da Vinci Surgical System (da Vinci S<br>(Model IS2000) or da Vinci SI (Model IS3000)) to<br>serve as a port of entry during da Vinci | |
| Intuitive Surgical, Inc. | Confidential | Page 3-2 |
P. 1 of 2
{1}------------------------------------------------
K101743, Intuitive Surgical 5 mm and 8 mm Flared Cannula
procedures that do not require maintenance of Insufflation.
Technological Characteristics: The subject device is equivalent in Intended use, design and technology as compared to the predicate device.
Performance Data: Performance tests (bench and animal lab tests) were conducted to demonstrate that the device Is substantially equivalent to the predicate device and that the design output meets the design input requirements. The results of the testing did not raise any new Issues of safety or efficacy.
Summary: on the technical characteristics, Based Intended use and performance test data, the Intuitive Surgical 5mm and 8mm Flared Cannula have been determined to be equivalent in in safety, efficacy, and performance to the predicate device.
Intuitive Surgical, inc.
{2}------------------------------------------------
Image /page/2/Picture/12 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The logo is black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609. Silver Spring, MD 20993-0002
Intuitive Surgical, Inc. % Ms. Meghna Sridharan Senior Regulatory Engineer 1266 Kifer Road Sunnyvale, California 94086
FEB - 4 2011
Re: K101743
Trade/Device Name: Intuitive Surgical® Flared Cannula Regulation Number: 21 CFR 876-1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: NA Y Dated: January 27, 2011 Received: January 28, 2011
Dear Ms. Sridharan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially_equivalent.(for the_indications_ for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
Page 2 - Ms. Meghna Sridharan
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
AZ
VS
for
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K101743, Intuitive Surgical 5 mm and 8 mm Flared Cannula
## Indications for Use
510(k) Number if known: K101743
Device Name: Intuitive Surgical Flared Cannula
INDICATION FOR USE:
The Flared Cannuiae are intended to be used with the da Vinci Surgical System (da Vinci S (Model IS2000) or da Vinci Si (Model IS3000)) to serve as a port of entry during da Vinci procedures that do not require maintenance of insufflation.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801 Subpart D)
AND/OR
Over-the-Counter Use _ (Per 21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R.P. Osdoba for mkm
Division Sign-O (Division Sign of Surgical, Orthopedic, and Restorative Devices
510(k) Number K101743
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.