K133414 · Sterilmed, Inc. · NLM · Jan 22, 2014 · Gastroenterology, Urology
Device Facts
Record ID
K133414
Device Name
REPROCESSED CLOSURE SYSTEM
Applicant
Sterilmed, Inc.
Product Code
NLM · Gastroenterology, Urology
Decision Date
Jan 22, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The reprocessed closure systems are intended to pass sutures through soft tissue during endoscopic/laparoscopic surgery.
Device Story
Reprocessed closure system consists of three components: one suture passer and two guides (5mm and 10/12mm). Suture passer features stainless steel shaft with pointed tip and polycarbonate handle with ring actuator. Guides are polycarbonate with cylindrical proximal head and tapered distal shaft containing holes for suture passer access. Device used in endoscopic/laparoscopic surgery to facilitate suture placement through soft tissue. Reprocessing involves cleaning and sterilization of predicate devices. Healthcare providers use the system to assist in tissue closure; output is the physical placement of sutures. Benefits include reuse of surgical instrumentation through validated cleaning and sterilization processes.
Clinical Evidence
Bench testing only. Evidence includes cleaning validation, sterilization validation (ISO 11135, USP <71>), biocompatibility (ISO 10993-18), ethylene oxide residuals (ISO 10993-7), packaging validation (ASTM D 4169, F 88, F 2096), and shelf life (ASTM 1980). Functional performance validated via simulated use, visual inspection, and fatigue testing.
Technological Characteristics
Materials: Stainless steel (shaft), polycarbonate (handle/guides). Energy: None (manual). Connectivity: None. Sterilization: Ethylene oxide. Standards: ISO 11135, ISO 10993, ASTM D 4169, ASTM F 88, ASTM F 2096, ASTM 1980.
Indications for Use
Indicated for passing sutures through soft tissue during endoscopic or laparoscopic surgical procedures.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Predicate Devices
Cooper Surgical Carter-Thomason CloseSure System® (K980123)
K182828 — Medline ReNewal Reprocessed Carter-Thomason Needle Point Suture Passer · Surgical Instrument Service and Savings, Inc. · Apr 15, 2019
K172666 — CrossBow Fascial Closure System, CrossBow Fascial Closure System with Adaptor · Sutureease, Inc. · Dec 1, 2017
K963329 — ETHICON ENDOSUTURE SYSTEM · ETHICON, Inc. · Oct 25, 1996
K160117 — AbClose - Port Site Closure Device · Medeon Biodesign, Inc. · Sep 9, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the word "STERILMED" in a sans-serif font. To the left of the word is a circular graphic that is not clearly visible. The text is horizontally oriented and appears to be a logo or brand name.
#### 510(K) PREMARKET NOTIFICATION SUBMISSION 06 NOVEMBER 2013 For Reprocessed Closure Systems
### II. SUMMARY AND CERTIFICATION
- 510(k) Summary A.
Submitter:
K133414
014
| Contact Person: | Jason Skramsted |
|------------------------|-------------------------------------------------------|
| | 11400 73 <sup>rd</sup> Avenue North |
| | Maple Grove, MN 55369 |
| | Phone: 763-488-3483 |
| | Fax: 763-488-4491 |
| | JAN 22 201 |
| Date Prepared: | 06 November 2013 |
| Trade Name: | Reprocessed Closure Systems |
| Regulation Name: | Laparoscope, General and Plastic Surgery, Reprocessed |
| Device Classification: | Class II, 21 CFR 876.1500 |
Sterilmed, Inc.
Product Code: NLM
| Predicate<br>Devices: | The reprocessed closure systems are substantially equivalent to the Cooper Surgical Carter-Thomason<br>CloseSure System® (K980123). |
|---------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Description: | The closure system is a set of three components; consisting of one suture passer, one 5mm guide and one<br>10/12mm guide. The suture passer has a stainless steel shaft with a pointed tip that opens on one side to<br>grasp the suture and a polycarbonate handle with a ring that actuates the tip. The 5mm and 10/12mm guides<br>are made of polycarbonate with a cylindrical head on the proximal end connected to a shaft that tapers to a tip<br>on the distal end. Each guide has two holes in the cylindrical head that pass through to opposite sides of the<br>shaft. The holes allow the suture passer to be passed through the guide. |
| Intended Use: | The reprocessed closure systems are intended to pass sutures through soft tissue during<br>endoscopic/laparoscopic surgery. |
| Technological<br>Characteristics: | The reprocessed closure systems are identical to the predicate devices in design, materials of construction,<br>and intended use. There are no changes to the clinical applications, patient population, performance<br>specifications, or method of operation. |
| Functional and<br>Safety Testing: | Representative samples of reprocessed closure systems were tested to demonstrate appropriate functional<br>characteristics. Process validation testing was performed to validate the cleaning and sterilization procedures<br>as well as device packaging. In addition, the manufacturing process includes visual and validated functional<br>testing of all products produced. |
| Summary of<br>Non-clinical<br>Tests<br>Conducted: | Specific non-clinical tests performed included: cleaning validation, sterilization validation (ISO 11135, USP<br><71>), biocompatibility testing (ISO 10993 Part 18), ethylene oxide residual testing (ISO 10993-7),<br>packaging validation (ASTM D 4169, ASTM F 88, ASTM F 2096), and shelf life validation (ASTM 1980).<br>In addition, validation of functional performance (bench testing) was performed through simulated use, visual<br>inspection, fatigue testing, and function testing. Performance testing shows the reprocessed closure systems<br>to perform as originally intended. |
| Conclusion: | Sterilmed concludes that the reprocessed closure systems are safe, effective, and substantially equivalent to<br>the predicate devices, Cooper Surgical Carter-Thomason CloseSure System® (K980123), as described in this<br>premarket notification submission. |
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a snake winding around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The logo is black and white and has a simple, clean design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Sterilmed Incorporated Mr. Jason Skramsted,RAC Regulatory Affairs Specialist 11400 73rd Avenue North Maple Grove, Minnesota 55369
#### January 22, 2014
Re: K133414
Trade/Device Name: Reprocessed Closure Systems Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: NLM, GCJ, HCF, GEJ Dated: November 6, 2013 Received: November 7, 2013
Dear Mr. Skramsted:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{2}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Joshua CONpper -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
For
Enclosure
{3}------------------------------------------------
Class and the Market Children and
## Indications for Use
510(k) Number (if known): K133414
Device Name: Reprocessed Closure Systems
Indications For Use:
The reprocessed closure systems are intended to pass sutures through soft tissue during endoscopic/laparoscopic surgery.
Prescription Use AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
ﻣﯿﮟ
# David Krause -S
. ·
(Division Sign-Off) Division of Surgical Devices 510(k) Number: K133414
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.