VANGUARD REPROCESSED SUTURE PASSER

K053585 · Ascent Healthcare Solutions · NLM · Oct 11, 2006 · Gastroenterology, Urology

Device Facts

Record IDK053585
Device NameVANGUARD REPROCESSED SUTURE PASSER
ApplicantAscent Healthcare Solutions
Product CodeNLM · Gastroenterology, Urology
Decision DateOct 11, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Reprocessed Suture Passers are intended to pass sutures through soft tissue during endoscopic/laparoscopic surgery.

Device Story

Reprocessed Suture Passer set consists of one Carter-Thomason needle-point suture passer and two suture passer guides; used during endoscopic/laparoscopic surgery. Device operation: suture passer guide inserted into wound site; suture passer pushes suture through guide and tissue; suture retrieved to create loop encompassing full wound depth; guide removed to allow wound closure. Vanguard receives previously used devices; cleans, inspects, tests, marks, packages, and sterilizes for subsequent clinical use. Benefits include enabling full-depth wound closure in laparoscopic procedures.

Clinical Evidence

Bench testing only. Validation included cleaning, sterilization, packaging, performance, and biocompatibility testing to demonstrate that reprocessed devices are equivalent to OEM devices.

Technological Characteristics

Materials: stainless steel shaft, polycarbonate ring-grip handle and guides. Dimensions: 3mm diameter shaft. Principle: mechanical suture passage via needle-point passer and guide system. Sterilization: validated reprocessing cycle. Connectivity: none.

Indications for Use

Indicated for passing sutures through soft tissue during endoscopic or laparoscopic surgical procedures.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Summary of Safety & Effectiveness Kos 3585 | Submitter | Vanguard Medical Concepts, Inc. 5307 Great Oak Drive Lakeland, FL 33815 | |---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact | Trish Stephens R&D Project Manager 863-683-8680 x252 [voice] 863-904-1604 [facsimile] tstephens@safe-reuse [email] | | Date | December 22, 2005 | | Device | <ul><li>Trade Name: Vanguard Reprocessed Suture Passer</li><li>Common Name: Suture Passer, Carter-Thomason Needle-Point Suture Passer </li><li>Classification Name: Laparoscope, General &amp; Plastic Surgery, Reprocessed </li><li>21 CFR Section: 876.1500</li><li>Reprocessed - Class II</li><li>Product Code: NLM</li></ul> | | Predicate Devices | <ul><li>Trade Names: <ul><li>Inlet Medical, Inc. CloseSure® Procedure Kit</li></ul></li><li>510(k) numbers: <ul><li>K980123: Louisville Laboratories, Inc. (Inlet Medical, Inc.) Carter-Thomason Needle-Point Suture Passer Instrument Kit </li></ul></li></ul> | | Indications for Use | Reprocessed Suture Passers are intended to pass sutures through soft tissue during endoscopic/laparoscopic surgery. | OCT 11 2006Continued on next page {1}------------------------------------------------ ## 510(k) Summary of Safety & Effectiveness, Continued The Reprocessed Suture Passer is a set of previously used devices that have Device been cleaned, tested, inspected, marked with an identifier, packaged, labeled, Description and sterilized by Vanguard. A "Suture Passer" is a set of three (3) devices, consisting of one (1) Carter-Thomason needle-point suture passer and two (2) suture passer guides. The Carter-Thomason needle-point suture passer is a 3mm diameter stainless steel shaft with a polycarbonate ring-grip handle that opens one side of the pointed tip of the steel shaft when activated. The polycarbonate suture passer guides have a cylindrical hub at the proximal end and a tapered shaft at the distal end. The suture passer guides taper from the junction with the cylindrical hub to the tip. Each suture passer guide has two holes in the hub; the two holes open on opposite sides of the pilot guide shaft. A Suture Passer Guide is inserted into the wound site. A Suture Passer is used to push suture through the suture passer guide and then through the tissue of the wound site. The suture is then retrieved and pulled through the opposite side of the guide. This creates a "loop" that encompasses the full depth of the wound and upon removal of the guide allows full closure of the wound. Vanguard receives previously used Suture Passers from healthcare facilities; cleans, inspects, tests, packages, labels, and sterilizes the devices; and returns them to a healthcare facility for subsequent use. The Vanguard Reprocessed Suture Passers are essentially identical to the Technological currently marketed Original Equipment Manufacturer (OEM) devices. Characteristics Device materials, specifications, and technological characteristics are equivalent. Cleaning, sterilization, and packaging validations, and performance and Test Data biocompatibility testing demonstrate that the reprocessed devices perform as intended and are safe and effective. Based upon the information provided herein and the 510(k) "Substantial Conclusion Equivalence" Decision Making Process Chart, we conclude that Vanguard Reprocessed Suture Passers are substantially equivalent to their predicate devices under the Federal Food, Drug and Cosmetic Act. {2}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles three human profiles facing to the right, connected by flowing lines. #### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ## OCT 1 1 2006 Ascent Healthcare Solutions % Ms. Moira Barton Regulatory Affairs Manager 10232 South 51st Street Phoenix, Arizona 85044 Re: K053585 - Supplemental Validation Submission Trade/Device Name: See Enclosed List Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: NLM Dated: January 12, 1998 Received: January 14, 1998 Dear Ms. Barton: The above-referenced premarket notification (510(k)) was cleared by the Office of Device Evaluation (ODE) on March 9, 1998. We have received your supplemental validation data as required for reprocessed single-use devices by the Medical Device User Fee and Modernization Act of 2002. After reviewing your supplemental validation data, we have determined the devices listed in the enclosure accompanying this letter are substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market these devices, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your devices are classified (see above) into either class II (Special Controls) or class III (PMA) they may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your devices in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your devices comply with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's applicable requirements, including, but not limited to: registration and {3}------------------------------------------------ Page 2 -- Ms. Moira Barton listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. The FDA finding of substantial equivalence of your devices to legally marketed predicate devices results in classification for your devices and thus, permits you to legally market the devices. This letter will allow you to continue marketing the devices listed in the enclosure accompanying this letter. If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at 240-276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Va. Mcllroy Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### Page 4 - Ms. Moira Barton ### Device listing found to be substantially equivalent: Carter-Thomason CloseSure System - (1) Carter-Thomason Need-Point Suture Passer . , . 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - - (1) 5mm Suture Passer . . : - (1) 10/12mm Suture Passer Guide {5}------------------------------------------------ ### Indications for Use 510(k) Number (if known): K053585 Device Name: Vanguard Reprocessed Suture Passers Indications for Use: Reprocessed Suture Passers are intended to pass sutures through soft tissue during endoscopic/laparoscopic surgery. | Prescription Use | <u>X</u> | |-----------------------------|----------| | (Part 21 CFR 801 Subpart D) | | AND/OR Over-The-Counter Use_ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of General, Restorative, and Neurological Devices Page 1 of 1 **510(k) Number** / K053585 012
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...