K990912 · Mtp Medical Technical Promotion GmbH · GCJ · Jun 9, 1999 · Gastroenterology, Urology
Device Facts
Record ID
K990912
Device Name
MTP EXTRACTION BAG
Applicant
Mtp Medical Technical Promotion GmbH
Product Code
GCJ · Gastroenterology, Urology
Decision Date
Jun 9, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Intended Use
The Extraction Bag for use in Laparoscopy is intended for use by qualified surgeons during laparoscopic procedures, including cholecystectomy, appendectomy, nephrectomy, lymphadenectomy, intestinal resection, oophorocystectomy and myomectomy.
Device Story
MTP Extraction Bag is a single-use surgical device designed for tissue retrieval during laparoscopic procedures. Operated by qualified surgeons in an operating room setting, the device facilitates the containment and removal of tissue specimens (e.g., during cholecystectomy, appendectomy, nephrectomy, lymphadenectomy, intestinal resection, oophorocystectomy, or myomectomy) through small laparoscopic incisions. The device acts as a physical barrier to prevent tissue contact with the incision site during extraction. By enabling safe removal of resected tissue, the device assists surgeons in completing minimally invasive procedures, potentially reducing patient trauma and recovery time associated with larger incisions.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Materials: high density polyethylene and polyurethane film. Form factor: bag for tissue containment. Manual operation; no energy source or software components.
Indications for Use
Indicated for use by qualified surgeons during laparoscopic procedures, including cholecystectomy, appendectomy, nephrectomy, lymphadenectomy, intestinal resection, oophorocystectomy, and myomectomy.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Related Devices
K093194 — MGB DISPOSABLE RETRIEVAL BAG LAPBAG · Mgb Endoskopische Gerate GmbH (Berlin) · Mar 3, 2010
K210248 — Lab Bag with Fabric Pouch · Advanced Medical Design Co., Ltd. · Jul 30, 2021
K172789 — Specimen Retrieval Bag · Beijing Hangtian Kadi Technology R&D Institute · Feb 1, 2018
K132375 — GENISTRONG SINGLE-USE SPECIMEN RETRIEVAL BAG · Genicon · Jun 12, 2014
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a logo for "mtp technical medical promotion". The letters "m", "t", and "p" are stacked on top of each other, with "t" in white and the other letters in black. Below the letters, the words "technical medical promotion" are written in a smaller font. The logo is set against a dark background.
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS 129909122
JUN
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All g 1999 data included in this document is accurate and complete to the best of MTP's knowledge.
| Applicant: | mtp medical technical promotion gmbh<br>p.o. box 4529 • d-78510 tuttlingen<br>tel.: +49/7461/96630 - 0<br>fax: +49/7461/96630 - 25 |
|------------|------------------------------------------------------------------------------------------------------------------------------------|
|------------|------------------------------------------------------------------------------------------------------------------------------------|
Contact: Kevin Kennan (310) 410-2769
Device Identification: Common Name: Extraction Bag
> Trade Name: (optional) MTP Extraction Bag
Indication: The MTP Extraction Bag is intended for use by qualified surgeons in tissue extraction procedures during laparoscopic surgery
Device Description: The MTP Extraction Bag is intended for use by qualified surgeons in tissue extraction procedures during laparoscopic surgery. The body contact materials are high density polyethylene and polyurethane film.
Substantial Equivalence: The MTP Extraction Bag for tissue extraction procedures during laparoscopic surgery is substantially equivalent to the predicate device since the basic features, design and intended uses are the same.
Signed:
Kevin Kennan
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, topped by a single wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the caduceus.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUN 9 1999
MTP Medical Technical Promotion GmbH c/o Mr. Kevin A. Kennan P.O. Box 1954 Culver City, California 90232
K990912 Re: Trade Name: MTP Extraction Bag Regulatory Class: II Product Code: GCJ Dated: February 16, 1999 Received: March 18, 1999
Dear Mr. Kennan:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{2}------------------------------------------------
Page 2 - Mr. Kevin A. Kennan
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
- Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo for MTP, which stands for Technical Medical Promotion. The logo is in black and white and features the letters "mtp" in a stylized font. The word "technical" is written below the letters "mtp", and the word "medical promotion" is written below the word "technical". The logo is set against a black background.
510(k) Number (if known):
Device Name: Extraction Bag for use in Laparoscopy
Indications for Use: The Extraction Bag for use in Laparoscopy is intended for use by qualified surgeons during laparoscopic procedures, including cholecystectomy, appendectomy, nephrectomy, lymphadenectomy, intestinal resection, oophorocystectomy and myomectomy.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | <div></div> |
|-----------------------------------------|-------------|
| Division of General Restorative Devices | |
| 510(k) Number | K990912 |
| Prescription Use: | <div>X</div> | OR Over-The-Counter Use: | |
|----------------------|--------------|--------------------------|--|
| (Per 21 CFR 801.109) | | | |
(Optional Format 1-2-96)
000003
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.