The Plasma Shield is to be used prior to and during endoscopic and laparoscopic procedures to prevent fogging of the scope lens
Device Story
Plasma Shield is a cold plasma-based device used in operating rooms by physicians to prevent laparoscope lens fogging. The system comprises a reusable main unit and a sterile, single-use disposable part that interfaces with the scope. The device generates cold plasma to treat the scope lens, creating a super-hydrophilic surface that prevents fog formation. The disposable part includes an anti-fog solution and foam applicator. The device is battery-powered (Li-Ion) and compatible with 5mm to 10mm laparoscopes. By maintaining clear intra-abdominal visibility, the device assists the surgeon during procedures, potentially reducing the need to remove the scope for cleaning, thereby improving procedural efficiency and patient outcomes.
Clinical Evidence
No human clinical data. Bench testing included lens fogging validation, life expectancy verification, and usability simulated use testing. Animal testing (porcine model) compared Plasma Shield to the Clearify predicate, demonstrating that both were effective in preventing fogging compared to untreated scopes. Biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1), and EMC (IEC 60601-1-2) testing were performed.
Technological Characteristics
Cold plasma-based lens activation system. Components: Reusable main unit, sterile disposable part with anti-fog solution and foam. Materials: Biocompatible plastics. Energy: RF-based cold plasma, Li-Ion battery (11.1V/12.6V). Dimensions: 15x10x7 cm. Compatibility: 5mm-10mm laparoscopes. Sterilization: ETO for disposable parts. Software: Minor level of concern.
Indications for Use
Indicated for use prior to and during endoscopic and laparoscopic procedures to prevent fogging of the scope lens in patients undergoing such procedures.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
September 12, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Plasmatica LTD % Vaibhav Rajal Official Correspondent for Plasmatica LTD MDI Consultants Inc 55 Northern Blvd Great Neck, New York 11021
Re: K221533
Trade/Device Name: Plasma Shield Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: GCJ, OCT Dated: May 2, 2022 Received: May 27, 2022
# Dear Vaibhav Rajal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221533
Device Name Plasma Shield
Indications for Use (Describe)
The Plasma Shield is to be used prior to and during endoscopic procedures to prevent fogging of the scope lens
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Sponsor: PLASMATICA Subject Device: Plasma Shield, Model: .1.0 Document Name: FDA 510(k) Company Cover Letter
# 510(k) SUMMARY
The assigned 510(k) number is K221533
### 1. Submitter's Identification:
Plasmatica LTD. HaTa'asiya ST 25 Ra'anana, Israel. 4365413.
| Contact Person: | Mr. Adam Sagiv<br>Chief Executive Officer and Founder |
|-----------------|-------------------------------------------------------|
| Telephone: | +972-52-2232562 |
| Email: | Adam@plasmatica.com |
Date Summary Prepared: September 8, 2022
### 2. Name of the Device:
| Trade Name: | Plasma Shield |
|--------------------|-----------------|
| Regulation Number: | 21 CFR 876.1500 |
| Regulatory Class: | Class II |
| Product Code: | OCT, GCJ |
# 3. Information for the 510(k) Cleared Device (Predicate Device):
| FDA Product Code | Common Name | Regulation Number |
|------------------|----------------------------------------------|-------------------|
| OCT | Anti-Fog Solution And Accessories, Endoscopy | 876.1500 |
| GCJ | Laparoscope, General & Plastic Surgery | 876.1500 |
#### 4. Device Description:
The Plasma Shield is intended to be used prior to and during endoscopic and laparoscopic procedures to prevent fogging of the scope lens. The Plasma Shield device is designed for reducing laparoscopic lens fogging by performing cold plasma treatment of a laparoscope lens. The Plasma Shield device is a cold plasma-based device to be used in an operating room, prior to and during laparoscopic procedures. The cold plasma creates a super hydrophilic scope lens surface after treatment (aka activation). The device does not interact directly with the patient's body.
The Plasma Shield consists of these components and accessories:
- The main system is a Reusable Part (RP) that generates the cold plasma . required to treat the scope for defogging.
- A Disposable Part (DP) a single-use, sterilized component interfacing with the ● scope itself and used for each procedure (can be used multiple times within the
{4}------------------------------------------------
same procedure for the same patient). The DP assembly includes a folded surgical drape and packaged in a pouch.
A single-use, sterile anti-fog solution and foam is supplied together with the Disposable Part.
### 5. Indications for Use:
The Plasma Shield is to be used prior to and during endoscopic and laparoscopic procedures to prevent fogging of the scope lens
# 6. Comparison to the 510(k) Cleared Devices (Predicate Devices):
Table of Comparison to Legally Marketed Device and Discussion of Similarities and Differences:
The proposed subject device is being compared to the predicate devices identified under Section 1(i): Identification of Legally Marketed Device Which We Claim Substantial Equivalence;
| Parameters | Subject Device | Primary Device | Predicate | Reference Device | Predicate | Similarities or Differences |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|---------------------------------------------------------------------------------------------------------------------------------------------------|-----------|-----------------------------|
| Device Name | Plasma Shield | Scope System/See Sharp | Antifogging | DHELP/Clearify | | |
| Manufacturer | Plasmatica Ltd. | Xodus Medical, Inc. | | New Wave/Medtronic | | |
| Regulatory Pathway | 510(k) | 510(k) | | 510(k) | | Same |
| 510(k)/De Novo number | TBD | K182080 | | K062779 | | |
| Classification | II | II | | II | | Same |
| Product Code | GCJ, OCT | GCJ, OCT | | GCJ, OCT | | Same |
| Regulation Number | 876.1500 | 876.1500 | | 876.1500 | | Same |
| Regulation Name | Endoscope and accessories | Endoscope and accessories | | Endoscope and accessories | | Same |
| Regulation Medical Specialty Panel | Gastroenterology/Urology | Gastroenterology/Urology | | Gastroenterology/Urology | | Same |
| | General & Plastic Surgery | General & Plastic Surgery | | General & Plastic Surgery | | Same |
| Intended Use | The Plasma Shield is to be used prior to and during endoscopic and laparoscopic procedures to prevent fogging of the scope lens | The Scope Antifogging System is intended use for this device is to be used prior to and during endoscopic and laparoscopic procedures to prevent fogging of the endoscope and laparoscope lens. | | DHELP is intended to be used prior to and during endoscopic and laparoscopic procedures to prevent fogging of the endoscope and laparoscope lens. | | Similar |
| Technological Characteristics | 1) Mechanism of Action involves Cold Plasma | 1) Mechanism of Action involves heating and | | 1) Mechanism of Action involves heating and | | See NOTE 1 |
{5}------------------------------------------------
| | treatment<br>of<br>a<br>laparoscope lens.<br>2) Uses antifog solution.<br>3) Compact device that<br>enables intra-abdominal<br>visibility in laparoscopic<br>procedures.<br>4) Consists of a reusable<br>main system, disposable<br>part (sterile) with Anti-Fog<br>solution and sponge<br>applicator (both sterile).<br>5) Compatible with<br>laparoscopes of 5mm to<br>10mm. | dipping of distal end of<br>endoscope/laparoscope<br>into solution<br>2) Uses antifog solution<br>3) Compact device that<br>enables intra-abdominal<br>visibility in laparoscopic<br>procedures<br>4) Device is small, sterile,<br>and disposable with Anti-<br>Fog solution, sponge, and<br>trocar wipes (sterile).<br>5) Compatible with<br>laparoscopes of 5mm to<br>12mm. | dipping distal end<br>of<br>endoscope into solution.<br>2) Uses antifog solution.<br>3) Compact device that<br>enables intra-abdominal<br>visibility in laparoscopic<br>procedures.<br>4) Device is small, sterile,<br>and disposable with Anti-<br>Fog solution, sponge<br>applicator and trocar<br>wipes (sterile).<br>5) Compatible with<br>laparoscopes of 5mm to<br>10mm. | | | Li-Ion Battery pack of the<br>device) 12.6V and 0.5A<br>and DC power jack.<br>When the device is being<br>charged, it is turned off by<br>design. | | | |
|---------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|-----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|---------------|
| Performance<br>testing | Testing conducted with<br>live anesthetized pig. A<br>Stryker endoscopy<br>system was used with a<br>10mm and 5mm Stryker<br>scope, and Plasma Shield<br>was evaluated in a series<br>of experiments. Risk<br>analysis, Biocompatibility<br>(ISO 10993-1),<br>Sterilization,<br>Electrical and EMC based<br>on IEC 60601-1-2 and<br>60601-1, SW V&V, bench<br>testing, Life Expectancy<br>Verification,<br>Lens Fogging Validation,<br>usability simulated use<br>testing to demonstrate<br>effectiveness in a surgical<br>environment simulation<br>chamber. | Testing conducted with<br>live anesthetized pig. A<br>Karl Storz endoscopy<br>system was used with a<br>5mm and 10mm Storz<br>laparoscope. A series of<br>exploratory procedures<br>with 4 different Storz<br>laparoscope types were<br>used. Cleaning and<br>application of antifogging<br>solution to the scopes for<br>determination of fogging<br>prevention was performed<br>in several instances. Lens<br>Fogging Validation, Lens<br>Wiping of Debris<br>Validation, Life<br>Expectancy Validation,<br>Trocar Wiping Validation,<br>Biocompatibility, Electrical<br>Safety and EMC testing<br>were also conducted. | Testing conducted with<br>live anesthetized pig. A<br>Stryker 888 endoscopy<br>system was used with a<br>10mm and 5mm Stryker<br>laparoscope, and DHELP<br>was evaluated in series of<br>experiments with varying<br>temperatures and<br>surfactant concentrations<br>and by comparison to the<br>current defogging<br>methods. DHELP was<br>also used with 4 different<br>laparoscope types and on<br>Olympus, Stryker and<br>Storz camera systems to<br>make sure that it was<br>functionally compatible<br>with the most common<br>scopes and systems.<br>Temperature testing was<br>conducted on the DHELP<br>device. Risk Analysis and<br>biocompatibility testing<br>were also conducted. | See NOTE 2 | Electrical safety | Electrical safety and EMC<br>compatibility testing was<br>successfully conducted<br>on the Scope Antifogging<br>System. The system<br>complies with the IEC<br>60601-1-2 standard for<br>EMC.<br>The Scope Antifogging<br>system was electrical<br>safety tested under the<br>Standard IEC 60601-1. | Electrical safety and EMC<br>compatibility testing was<br>successfully conducted<br>on the Scope Antifogging<br>System. The system<br>complies with the IEC<br>60601-1-2 standard for<br>EMC.<br>The Scope Antifogging<br>system shall be electrical<br>safety tested under the<br>Standard IEC 60601-1. | No information available | |
| Clinical Studies | Not conducted | Not conducted | Not conducted | Similar | Materials,<br>Biocompatibility,<br>and Chemical<br>Safety | Biocompatible plastics<br>housing FDA approved<br>Anti-fog solution kit | Biocompatible plastics<br>housing<br>Anti-fog solution: Water,<br>Isopropanol, Ethanol,<br>Surfactant | Biocompatible plastics<br>housing FDA approved<br>ShurClens wound<br>cleaning surfactant | Similar |
| Compatibility<br>with the<br>environment and<br>other devices | Compatible with<br>endoscope and<br>laparoscope | Compatible with<br>endoscope and<br>laparoscope | Compatible with<br>endoscope and<br>laparoscope | Similar | Similarities to<br>Subject Device | NA | Intended Use and<br>compatibility with scopes<br>of 5mm to 12mm | Intended Use and<br>compatibility with scopes<br>of 5mm to 10mm | Similar |
| Energy used<br>and/or delivered | RF | Heat radiation | Heat (Infrared) radiation | See NOTE 3 | Weight | 1 kg max | No information available | 155 g | |
| Where used<br>(hospital, home,<br>ambulance, etc.) | Hospitals and/or surgery<br>centres on or by the order<br>of a physician | Hospitals and/or surgery<br>centres on or by the order<br>of a physician | Hospitals and/or surgery<br>centres on or by the order<br>of a physician | Same | Dimensions | 15*10*7 cm | No information available | 9*7*6 cm | |
| Design<br>performance | Cell batteries at 11.1V<br>and a BMS (battery<br>management system) and<br>a Li-Ion battery charger | Cell Battery<br>No charger | Cell Battery<br>No charger | See NOTE 4 | Sterilization | ETO | Gamma Radiation | Gamma Radiation | See NOTE<br>2 |
| Operating<br>environmental<br>conditions | Temp: 10-40°C<br>Rel Humid: 20-80%<br>Atm Pres: 90-106 kPa | Temp: 5-40°C<br>Rel Humid: 15-95%<br>Atm Pres: 70-106 kPa | No information available | Similar | | | | | |
{6}------------------------------------------------
#### Discussion of Similarities and Differences between the proposed subject device and both the primary and reference predicate devices.
#### Comparison of intended use
Plasma Shield and predicate have identical indications for use. Plasma Shield and predicate have an identical primary intended use.
#### Technological Characteristic Comparison to Predicate
#### Comparison in Detail(s):
NOTE 1: Plasma Shield does not raise any additional risk to the patient when applied on commercially available laparoscopes comprised of stainless steel and glass. The subject device mechanism of action utilizes Cold Plasma while the predicate and reference device mechanism of action uses heat for introducing the fogging effect. Cold plasma treatment has a long (over 25 years) history of safe use in medical applications, i.e., sterilization of medical devices made of stainless steel and glass.
{7}------------------------------------------------
In addition to a disposable part, the subject device has a reusable part, which is not the case of the predicate and reference devices. This fact raises no additional safety concern since the reusable part was fully validated for cleaning and is fully in compliance with the TIR30 reusable standard requirements. The reusable device is covered by a sterile drape in-between use.
### Cleaning Validation:
Manual cleaning process validation for the Plasma Shield reusable device, based on our Study, was performed and passed testing in accordance with TIR30.
#### Ingress tests:
A ingress test was performed for determining the viability of organisms after being exposed to the Single Use part of the Plasma Shield multiple times, at worst case system operation conditions, concluded that all test method acceptance criteria were met. Testing was performed in compliance with US FDA good manufacturing practice (GMP) regulations 21 CFR Parts 210, 211, and 820.
NOTE 2: The following tests for Electrical and EMC based on IEC 60601-1-2 and 60601-1. SW V&V, and usability simulated use testing were conducted in addition to the special controls conducted on the predicate device:
#### Software Verification and Validation Testing
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Minor" level of concern.
Usability Study Report
All the subjects succeeded to perform the device related tasks, with minimal number of requests for assistance and with no errors or inefficiencies. Following the summarized results, it can be concluded that the study success criteria was completely met, and that the Plasma Shield device can be easily operated by the potential end used safely according to the instructions manual.
Animal testing to demonstrate effectiveness in porcine model report: "Comparison of Plasma Shield vs. Clearify. Anti-fogging Devices, and Using Laparoscopy in Swine" was conducted and included in the submission.
The results of the risk analysis (according to ISO 14971) performed on Plasma Shield is detailed in the Plasma Shield Risk Analysis Report.
The subject device is categorized per Annex A of ISO 10993-1 as a medical device with no body contact. No further testing is deemed necessary. This difference has no impact on the device safety, since all device materials are well categorized, and biologically evaluated.
An ETO Sterilization validation (according to ISO 11135) was performed in the study titled EO Sterilization Validation for Clarity One of Plasmatica Ltd.
{8}------------------------------------------------
### NOTE 3:
With reqards to the use of radio frequency. Plasma Shield does not raise any additional risk to the patient. Plasma Shield conforms to the requirements of IEC 60601-1-2 and 6061-2-2 sections 202.7 and 202.8.
### NOTE 4:
The predicate, reference, and subject devices use reqular disposable batteries. The subject device, in addition, uses a Li-Ion battery charger, 12.6V and 0.5A and DC power jack. This fact raises no additional safety impact on the user. The Li-Ion battery is widely used in medical devices in a similar environment. It was tested and is fully compliant with the IEC 623166 standard under test report IEC 62133-2 Secondary cells and batteries containing alkaline or other non-acid electrolytes – Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems.
### Conclusions:
Comparison with the predicate and reference devices regarding functional characteristics shows that the subject device is intended for equivalent procedures as compared to the predicate and reference devices and makes use of similar functionality. The technological differences do not raise additional safety issues.
Substantial Equivalence Conclusion: The Sponsor has demonstrated through performance testing, design and features, and non-clinical testing, that the proposed subject device is substantially equivalent to the predicate and reference predicate devices.
The subject device Plasma Shield is Substantially Equivalent to the predicate and reference predicate devices.
# 7. Electrical and EMC Testing
The following testing supports testing information demonstrating safety and effectiveness of the Plasma Shield Device in the intended environment of use:
- FR Number 19-4 07/09/2014 ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, ● C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)
- FR Number 19-8 09/17/2018 IEC 60601-1-2- Edition 4.0 2014-02 Medical ● electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests
- FR Number 6-389 06/07/2021 IEC 60601-2-2 Edition 6.0 2017-03 Medical ● electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories. Sections 202.7, 202.8 ONLY
{9}------------------------------------------------
# 8. Performance Testing:
Bench Performance testing was conducted on the proposed subject device to determine the performance and efficacy of the device. In addition, Performance testing – animal was conducted on the proposed subject device to compare the anti-fogging effect of Plasma Shield vs. Clearify™ devices during laparoscopy in swine. The study demonstrated the anti-fogging effect for both the proposed subject device and the predicate device.
Validation of product performance in compliance with the following product specification requirements was conducted on the proposed subject device:
- PRSJ.02: Scope's lens surface characteristics immediately post treatment (aka . activation): Plasma-induced surface activation should create a hydrophilic scope lens surface to reduce scope lens fogging as effectively as the predicate device, thereby demonstrating substantial equivalence.
- . PRS 4.3: The Plasma Shield interfaces with laparoscopes of 5mm and 10mm diameter, with 0-45 anqles.
- . PRS H.07: The system shall perform lens activation for scopes of 0, 30, and 45 angles, and different materials
The Plasma activated lenses of all tested scope are Super Hydrophilic and comparison to untreated lenses show highly improved contact angles. Product performance comply with the specification requirements. Therefore, plasma surface activation induced hydrophilicity reduces scope lens foqqing, conferring substantial equivalence with the predicate device working mechanism. Comparison with the predicate and reference devices regarding functional characteristics shows that the subject device is intended for equivalent procedures as compared to the predicate and reference devices and makes use of similar functionality.
Comparison of Proposed Subject Device Plasma Shield Vs. Clearify Predicate Device, Anti-Fogging Devices was conducted during Laparoscopy in Swine to compare the antifogging hydrophilicity effect of Plasma Shield vs. Clearify™ devices during laparoscopy in swine in order to demonstrate substantial equivalence.
Overall, both Plasma Shield and Clearify scope pretreatment received a higher score compared to none treated scope following video evaluation in all experimental conditions tested. These suggest that pretreatment of the scope with Plasma Shield or Clearify was more effective in preventing fog formation on the scope during laparoscopy compared to no treatment in these experimental settings. It can be concluded that Plasma Shield scope pretreatment is comparable to Clearify in its effectiveness to prevent fogging on the scope during laparoscopy and both are better in grading than no treatment, therefore demonstrating substantial equivalence between the devices.
# 9. Usability Studies
Usability Testing was conducted on the proposed subject device as per IEC 62366-1:2015, per FDA Guidance "Applying Human Factors and Usability Engineering to Medical Devices" February 2016 and ANSI/AAMI NS4 2013 (R)2017. In addition, the proposed subject device is in compliance with the IEC 60601-1-6 - Collateral Standard -Usability test standard.
{10}------------------------------------------------
### 10. Sterilization and Product Shelf Life
The proposed subject device is composed of a reusable part (RP) and disposable part (DP). The DP is sealed and the sterile part contains electrodes intended to deliver the plasma to the scope. The following validations have been conducted on the proposed subject device:
| Test Name | Main<br>standard/guideline | Essence | DP<br>level | RP<br>level | System<br>level |
|-------------------------------------------|--------------------------------|-----------------------------------------------------------------------------------------|-------------|-------------|-----------------|
| Sterilization<br>validation | ISO 11135 | Conduct testing to ensure compliance to sterilization standard (ETO) | ✓ | | |
| Packaging<br>validation DP | ISO 11607-1 | Conduct packaging integrity, storage and transportation tests per appropriate standards | ✓ | | |
| Shelf-life<br>validation | ASTM F 1980:2016 | Conduct shelf-life tests per appropriate standard to verify required device shelf life. | ✓ | | |
| Packaging,<br>shipping and<br>environment | ASTM D4169-16<br>ASTM D4332-14 | Shipping performance and environmental testing | ✓ | ✓ | |
# 11. Product Cleaning/Disinfection Information
The proposed subject device is composed of a reusable part (RP) and disposable part (DP). The RP cleaning validation was performed to validate the cleaning process for the entire external surface of the device.
# 12. Software information:
Software verification and validation testing were conducted and supported by documentation according to the FDA's Guidance for Industry and FDA Staff "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. Based on FDA Software level of concern, the subject device is classified as "Minor" level of software concern.
The details about the software were clearly demonstrated in the software documentation and the risks related to the software products were mitigated
# 13. Risk Analysis
The proposed subject Plasma Shield Device complies with the ISO 14971: 2019 Medical devices - Application of risk management to medical devices" test standard. The testing ensures that the proposed subject device "Plasma Shield" meets the appropriate safety and effectiveness requirements.
{11}------------------------------------------------
# 14. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
Testing information demonstrating safety and effectiveness of the proposed subject Plasma Shield Device in the intended environment of use is supported by testing that was conducted in accordance with the FDA November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", DCRND, which outlines Electrical, Mechanical and Environmental Performance requirements.
- IEC 62133-2 Edition1.0 2017-02 Secondary cells and batteries containing alkaline or . other non-acid electrolytes – Safety requirements for portable sealed secondary cells. and for batteries made from them, for use in portable applications – Part 2: Lithium systems
- ISO 11607-1: 2019 Packaging for terminally sterilized medical devices Part 1: . Requirements for materials, sterile barrier systems and packaging systems
- ISO 11607-2:2019 Packaging for terminally sterilized medical devices Part 2: ● Validation requirements for forming, sealing and assembly processes
- . ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
- . ASTM F1886/F1886M - 2016 - Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
- . FR Number 14-482 06/27/2016 ASTM F88/F88M-15- Standard Test Method for Seal Strength of Flexible Barrier Materials
- . ASTM F1140/ F1140M-13 (2020) Standard Test Methods for Packages Internal Pressurization Failure Resistance of Unrestrained
- ASTM F1929-15 - Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
- ASTM F3039-15 Standard Methods in Nonporous Packaging or Flexible Barrier ● materials by Dye Penetration
- ASTM D4169-16 Standard Practice for Performance Testing of Shipping Containers ● and Systems. Sections 14, 15 NOT APPLICABLE
- . ASTM D4332-14 Standard Practice for Conditioning Containers, Packages, or Packaging Components for Testing
- ISO 10993-1:2018 Biological evaluation of medical devices Part 1: Evaluation and . testing within a risk management process **Additional compliance to US-FDA Guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" of 4th September 2020.
- . ISO 15223-1: 2016 Medical devices - Symbols to be used with medical device labels, labelling, and information to be supplied - Part 1: General requirements
- AAMI TIR12:2020 Designing, testing, and labeling reusable medical devices for . reprocessing in health care facilities: A guide for medical device manufacturers
- . AAMI TIR30:2011/(R)2016 A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable medical devices
None of the testing demonstrated any design characteristics that violated the requirements of the Reviewer Guidance or resulted in any safety hazards. It was our conclusion that the proposed subject Plasma Shield Device tested met all relevant requirements of the aforementioned tests.
{12}------------------------------------------------
### Conclusions:
The design, characteristics, and performance of the proposed subject Plasma Shield device substantiates that the device is working as intended (use prior to and during endoscopic and laparoscopic procedures to prevent the fogging of the scope lens).
The proposed subject Plasma Shield Device is substantially equivalent to both its primary predicate device and the reference predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.