The Scope Antifogging System is intended use for this device is to be used prior to and during endoscopic and laparoscopic procedures to prevent fogging of the endoscope and laparoscope lens.
Device Story
Single-use, sterile device; prevents endoscope/laparoscope lens fogging during surgery. Device contains antifogging surfactant solution and battery-powered warming circuit. Operation: user activates device via on-switch; solution warms to body temperature; user inserts scope into device port (up to 12mm diameter) to apply solution. Includes microfiber cloth and sponge-tipped trocar wipes for debris removal. Used in hospitals/surgery centers by physicians, nurses, and healthcare professionals. Benefits: maintains clear operative view by preventing condensation caused by temperature differentials between scope and body. Warming circuit life expectancy: 5 hours.
Clinical Evidence
Bench and animal testing performed. Bench testing: validated fogging prevention effectiveness, 5-hour warming circuit duration, and trocar wiping efficacy. Animal testing: conducted in anesthetized pig model using Karl Storz laparoscopes; confirmed consistent fogging prevention and debris removal using microfiber cloth and foam pad. Biocompatibility testing per ISO 10993-1 (cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, pyrogenicity). Electrical safety per IEC 60601-1 and EMC per IEC 60601-1-2.
Technological Characteristics
Materials: Polycarbonate housing, water/isopropanol/ethanol/surfactant solution. Dimensions: Accepts scopes up to 12mm. Energy: Battery-powered warming circuit. Sterilization: Gamma radiation (ISO 11137-2). Standards: ISO 10993-1 (biocompatibility), IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC).
Indications for Use
Indicated for use by healthcare professionals during endoscopic and laparoscopic procedures to prevent fogging of the scope lens.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services seal on the left and the FDA acronym and name on the right. The FDA acronym and name are in blue, with the acronym in a square and the name in a sans-serif font. The logo is simple and professional, and it is easily recognizable.
September 19, 2018
Xodus Medical, Inc. Mr. Paul Lloyd Vice President, Global QA/RA & Technology 702 Prominence Drive New Kensington, Pennsylvania 15068
#### Re: K182080
Trade/Device Name: Scope Antifogging System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ, OCT Dated: July 27, 2018 Received: August 2, 2018
# Dear Mr. Lloyd:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# David Krause -S
- for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182080
Device Name Scope Antifogging System
Indications for Use (Describe)
The Scope Antifogging System is intended use for this device is to be used prior to and during endoscopic and laparoscopic procedures to prevent fogging of the endoscope and laparoscope lens.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> ☒ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> □ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image is a company logo for Xodus Medical. The logo has a symbol on the left that is made up of two overlapping shapes, one in black and one in blue. To the right of the symbol is the company name, "XODUS MEDICAL", in bold black letters. Below the company name is the tagline "Making Surgery Safer" in blue letters.
# SECTION 5 - 510(k) Summarv
- 510(k) Owner: Xodus Medical, Inc. Westmoreland Business & Research Park 702 Prominence Drive New Kensington, PA 15068 Phone: 724-337-5500 Fax: 724-337-1131 Contact: Paul Lloyd (Vice President, Global QA/RA & Technology)
Establishment Registration Number: 2530138
Date Prepared: July 27, 2018
#### 1. Device Information:
Trade/Device Name: Scope Antifogging System Common Name: Endoscope Antifogging Device Classification Name: Endoscope and/or Accessories Regulation Number: 21 CFR 876.1500 Product Code: GCJ; OCT Regulatory Class: II
#### 2. Legally Marketed Devices to which Xodus Medical claims Substantial Equivalence:
Device Name: Clearify Visualization System Common Name: Endoscope Antifogging Device 510(k) Number: K062779 510(k) Owner: Covidien LLC Classification Name: Endoscope and/or Accessories Regulation Number: 21 CFR 876.1500 Product Code: GCJ; OCT Regulatory Class: II
*The Covidien predicate device is formerly known as New Wave Surgical's DHELP, FDA 510k K062779, acquired by Medtronic/Covidien and marketed as Clearify Visualization System under a name change only.
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for Xodus Medical. The logo features a stylized "X" formed by two overlapping shapes, one in black and the other in blue. To the right of the symbol is the text "XODUS MEDICAL" in bold, black letters, with the tagline "Making Surgery Safer" in a smaller, lighter blue font underneath.
#### 3. Device Description
The Scope Antifogging System is a single use, sterile device that is designed to apply an antifogging solution (surfactant) and warm the scope at or above the human body temperature when used prior to and during procedures that utilize an endoscope or laparoscope.
The Scope Antifogging System provides antifogging of the lens of the endoscope and laparoscope when inserted into the Scope Antifogging System for use in a laparoscopic procedure. Fogging occurs when the ambient temperature of the lens of the endoscope/laparoscope is cooler than the temperature of the human body and condensation forms on the lens. This is especially important in a surgical procedure where a surgeon must have a clear view of the operative site to safely perform surgery.
The Scope Antifogging System is controlled by an on switch to activate the circuit containing battery power source to indicate the activation of the antifogging solution warming circuit. The activation of the warming circuit life expectancy is 5 hours.
The antifogging solution is retained in the device and warmed when the device is activated in the "on" position. Subsequently, the endoscope/laparoscope is inserted into the device port for application of the warmed antifogging solution. The device's single scope insertion port of the endoscope/laparoscope accepts scopes up to 12mm in diameter.
The antifogging solution is included in with the Scope Antifogging System as fluid contained in two separate bottles. The solution is dispensed into the single port opening of the housing of the Scope Antifogging System prior to use.
Included with the Scope Antifogging System in the same packaging is a pair of sponge-tipped, trocar wipes. The trocar wipes are a single-use, radiopaque accessory used to wipe debris from the inside of the trocar during the surgical procedure. The trocar wipe sponge tip diameters are 7mm and 12mm.
Also included with the Scope Antifogging System in the same packaging is a single-use, microfiber cloth. The microfiber cloth is an accessory used as a method to wipe the endoscope/laparoscope lens of any debris buildup.
This product is sold sterile to healthcare professionals only.
The Scope Antifogging System has not been previously submitted for 510(k) clearance. However, the Xodus Medical Scope Antifogging System component, Antifog solution. is also a separately sold product, and is currently marketed by Xodus Medical with FDA clearance under 510k number K063587.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for Xodus Medical. The logo consists of a stylized "X" made up of two overlapping parallelograms, one in black and one in blue. To the right of the "X" is the text "XODUS MEDICAL" in bold, black letters. Below that, in a smaller blue font, is the tagline "Making Surgery Safer™".
#### 4. Intended Use:
The Xodus Medical Scope Antifogging System intended use is identified to be used prior to and during endoscopic and laparoscopic procedures to prevent fogging of the scope lens.
#### 5. Indications for Use:
The Scope Antifogging System is a single use, sterile device to be used prior to and during endoscopic and laparoscopic procedures to prevent fogging of the scope lens.
# 6. Description of Safety and Substantial Equivalence
The Xodus Medical Scope Antifogging System has the same intended use as the predicate device, Covidien Clearify Visualization System, which is, "Use prior to and during endoscopic and laparoscopic procedures to prevent the fogging of the scope lens". The Xodus Medical Scope Antifogging System and Covidien Clearify Visualization System are based on the same technological elements shown in the table below.
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Xodus Medical. The logo consists of a blue and black abstract symbol on the left, followed by the text "XODUS MEDICAL" in black, bold letters. Below the company name, the tagline "Making Surgery Safer" is written in a smaller, blue font with a trademark symbol.
# Comparison Table to Predicate Device
| Name of Device | Xodus Medical Inc.'s Scope | Covidien's Clearify |
|--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Antifogging System | Visualization System |
| Intended Use | Scope Antifogging System<br>intended to be used prior to and<br>during endoscopic and<br>laparoscopic procedures to prevent<br>the fogging of the scope lens. | Clearify Visualization System<br>intended to be used prior to and<br>during endoscopic and<br>laparoscopic procedures to<br>prevent the fogging of the scope<br>lens. |
| Usage | Hospitals and/or surgery centers<br>on or by the order of a physician | Hospitals and/or surgery centers<br>on or by the order of a physician |
| Materials | Antifogging Solution: Water,<br>Isopropanol, Ethanol, Surfactant.<br>Material difference in housing<br>using polycarbonate plastic. | FDA approved ShurClens, a<br>wound cleaning surfactant;<br>Water, alcohol, and surfactant.<br>Material difference in housing<br>using polyurethane foam. |
| Physical<br>Characteristics | Antifogging Solution of Clear/<br>colorless, odorless, water soluble<br>liquid.<br>Similar physical footprint-Xodus<br>lower profile; Shape is triangular<br>design. | Antifogging Solution of Clear/<br>colorless, odorless, water soluble<br>liquid.<br>Physical footprint with rounded,<br>oval design, higher profile. |
| Sterilization | Gamma radiation | Gamma radiation |
| Design/Packaging | Sterile barrier of polyethylene<br>plastic and Tyvek | Sterile barrier of polyethylene<br>plastic and Tyvek |
| Mechanism of<br>Action | Heating and dipping of distal end<br>of endoscope/laparoscope into<br>solution | Heating and dipping of distal<br>end of endoscope/laparoscope<br>into solution |
| Biocompatibility<br>Standards<br>Compliant | Yes, per ISO 10993-1 | Yes |
| Target Population | Physicians, Nurses, Health Care<br>Professionals, General Surgery | Physicians, Nurses, Health Care<br>Professionals, General Surgery |
| Power Source | Battery | Battery |
| Single Use | Yes | Yes |
| Sterility | Sterile | Sterile |
| 510(k) Number | TBD | K062779 |
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the Xodus Medical logo. The logo consists of a stylized "X" formed by two intersecting blue and black shapes. To the right of the "X" is the text "XODUS MEDICAL" in bold, black letters. Below "XODUS MEDICAL" is the tagline "Making Surgery Safer" in a smaller, blue font, with a trademark symbol next to the word "Safer".
#### 7. Performance Data:
# 7.1 Performance Testing - Bench
The following bench tests were performed for this submission:
# 7.1.1 Scope Lens Fogging Validation:
Lens fogging validation testing was performed using the Xodus Medical Scope Antifogging System and the predicate device. The testing performed was to verify the effectiveness of the Scope Antifogging System at preventing fogging from occurring on simulated equipment. A laparoscope and a glass mirror plate were used in the testing and subjected to a high humidity source. Using a control, the antifogging solutions of both products were applied to the simulated equipment, the laparoscope and glass mirror plate. The testing conclusions were the Xodus Medical Scope Antifogging System demonstrated the effectiveness of preventing the fogging of the simulated equipment consistently, as with the predicate device.
# 7.1.2 Life Expectancy Validation
Testing of the duration of activation of the warming circuit was performed. The Scope Antifogging System was activated and observed for duration of the activation period. Thermal characteristics were observed during the activation period and determined the activation to be a minimum for five (5) hours.
#### 7.1.3 Trocar Wiping Validation
Testing of the wiping of the inside of the trocar to clear any debris following application of blood and tissue (general smudging) from beef liver and extracting the scope was conducted.
The blood and tissue were applied to the scope interior of the trocar and successfully wiped away on three test cases using the trocar wipes of the Scope Antifogging System and the Covidien Clearify Visualization system trocar wipes. There were no visible signs of any debris left behind following the wiping of the inside of the trocar.
#### 7.2 Performance Testing -Animal
The following animal tests were performed for this submission:
#### 7.2.1 Scope Lens Fogging Validation
Testing of the Xodus Medical Scope Antifogging System and the predicate device was conducted with a live, anesthetized pig. A Karl Storz endoscopy system was used with a 5mm and 10mm Storz laparoscope. A
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the Xodus Medical logo. The logo consists of a stylized "X" formed by two intersecting shapes, one in black and one in blue. To the right of the "X" is the text "XODUS MEDICAL" in bold, black letters. Below "XODUS MEDICAL" is the tagline "Making Surgery Safer" in a smaller, blue font, with a trademark symbol next to the word "Safer".
series of exploratory procedures with four different Storz laparoscope types were used. Cleaning and application of antifogging solution to the scopes for determination of fogging prevention was performed in several instances. In all cases, the Scope Antifogging System demonstrated effectiveness of preventing the fogging of the laparoscopes consistently, as with the predicate device.
#### 7.2.2 Scope Lens Wiping of Debris Validation
Testing of the wiping of the scope lens following application of bile, blood, and tissue was conducted. The matter was applied to the scope lens and subsequently successfully wiped away successfully on all occasions using the microfiber cloth and the foam pad on the Scope Antifogging System.
#### 7.3 Biocompatibility
The Scope Antifogging System was successfully conducted to the ISO 10993-1 standard, Biological Evaluation of Medical Devices - Part 1: Evaluation and testing within a Risk Management System". The biocompatibility tests performed are shown below. These tests were determined to be applicable to the product according to the device classification according to ISO 10993-1 Appendix A. Table A.1.
The product category was determined to be "External Communicating Device": Contact determined to be "Tissue/Bone/Dentin; with Contact Duration "A-limited, ≤24hr"; and Biological Effect determined to be "Cytotoxicity, Sensitization, and Intracutaneous Reactivity". Additional biocompatibility tests included were include in the biocompatibility testing, the Acute Systemic Toxicity, and Pyrogenicity.
#### 7.4 Sterilization
The Scope Antifogging System shall be sterilized by gamma radiation based on the ISO 11137-2 standard; "Sterilization of Health Care Products - Part 2: Establishing the Sterilization Dose". The process shall be validated to ensure 10-6 sterility assurance or better is achieved.
The sterilization shall follow the same processes as currently marketed Xodus Medical FDA 510(k) cleared sterile products to include Antifog Solution 510(k) K063587, ES Pencils and Electrodes 510(k) K081647; Cautery Tip Cleaner 510(k) K053433, and Light Guard Light Handle 510(k) K053321.
{10}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Xodus Medical. The logo consists of a stylized "X" formed by two overlapping blue shapes on the left. To the right of the "X" is the text "XODUS MEDICAL" in black, with the tagline "Making Surgery Safer" in blue below it. The logo is clean and modern, with a focus on the company's name and mission.
#### 7.5 Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and EMC compatibility testing was successfully conducted on the Scope Antifogging System. The system complies with the IEC 60601-1-2 standard for EMC.
The Scope Antifogging system shall be electrical safety tested to the standard IEC 60601-1.
#### 8. Conclusion:
The design, characteristics, and performance of the Scope Antifogging System substantiate that the device is working as intended (use prior to and during endoscopic and laparoscopic procedures to prevent the fogging of the scope lens). The Xodus Medical Scope Antifogging System is substantially equivalent to its predicate device, Covidien's Clearify Visualization System.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.