Visera Elite II Video System Center, Visera Elite II Endoeye 3D
K193026 · Olympus Medical Systems Corp. · GCJ · Apr 3, 2020 · Gastroenterology, Urology
Device Facts
Record ID
K193026
Device Name
Visera Elite II Video System Center, Visera Elite II Endoeye 3D
Applicant
Olympus Medical Systems Corp.
Product Code
GCJ · Gastroenterology, Urology
Decision Date
Apr 3, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
OLYMPUS OTV-S300: This video system center is intended to be used with OLYMPUS camera heads, endoscopes, monitors, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis, treatment, and video observation. VISERA ELITE II VIDEO TELESCOPE ENDOEYE 3D WA50080A, WA50082A: This instrument has been designed to be used with a video system center, light source, documentation equipment, monitor, hand instruments, electrosurgical unit, and other ancillary equipment for endoscopy and endoscopic surgery within the thoracic and abdominal cavities including the female reproductive organs. The device is also indicated for visualization during transanal and transvaginal natural orifice surgery.
Device Story
VISERA ELITE II system comprises OTV-S300 video center and WA50080A/WA50082A 3D video telescopes. System captures optical images via distal CCD sensors; transforms optical signals into electrical signals for processing; displays 3D/2D endoscopic images on compatible monitors. Used in OR/clinic by surgeons for endoscopic diagnosis/surgery. OTV-S300 integrates processor, light source, and 3D mixer; supports White Light (WLI) and Narrow-Band Imaging (NBI) modes. 3D observation provides spatial depth perception to optimize surgical speed, accuracy, and precision. Surgeons view output via 3D monitor with polarization glasses. System aids visualization, potentially shortening surgical learning curves.
Clinical Evidence
No clinical or animal studies performed. Substantial equivalence supported by bench testing, including image quality (brightness, intensity, color, SNR), optical properties (distortion, resolution, FOV), thermal safety, mechanical stress resistance, and reprocessing validation.
Technological Characteristics
System includes OTV-S300 processor/light source and WA50080A/WA50082A 3D telescopes. Features LED illumination, dual CCD sensors for 3D, and NBI optical enhancement. Complies with ANSI/AAMI ES 60601-1, IEC 60601-2-18, and IEC 60601-1-2. Biocompatibility per ISO 10993-1. Reprocessing validated per FDA guidance.
Indications for Use
Indicated for patients undergoing endoscopic diagnosis, treatment, or surgery within thoracic and abdominal cavities, including female reproductive organs, and for visualization during transanal and transvaginal natural orifice surgery.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
April 3, 2020
Olympus Medical Systems Corp. Lisa Boyle Regulatory Affairs Specialist II Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610
Re: K193026
> Trade/Device Name: VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300, VISERA ELITE II VIDEO TELESCOPE ENDOEYE 3D WA50080A, WA50082A Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: GCJ, HET, GCM, FET, NWB Dated: February 25, 2020 Received: February 28, 2020
Dear Lisa Boyle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Jason Roberts, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K193026
### Device Name VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300 VISERA ELITE II VIDEO TELESCOPE ENDOEYE 3D WA50080A, WA50082A
Indications for Use (Describe)
### OLYMPUS OTV-S300
This video system center is intended to be used with OLYMPUS camera heads, endoscopes, monitors, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis, treatment, and video observation.
## VISERA ELITE II VIDEO TELESCOPE ENDOEYE 3D WA50080A, WA50082A
This instrument has been designed to be used with a video system center, light source, documentation equipment, monitor, hand instruments, electrosurgical unit, and other ancillary equipment for endoscopy and endoscopic surgery within the thoracic and abdominal cavities including the the a female reproductive organs. The device is also indicated for visualization during transanal and transvaginal natural orifice surgery.
Type of Use (Select one or both, as applicable)
| <span> <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|---------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "OLYMPUS" in a bold, sans-serif font. The word is in a dark blue color. There is a thin, horizontal yellow line underneath the word. The registered trademark symbol is located to the right of the word.
April 3, 2020
K193026
# 510(k) Summary
# 5.1 GENERAL INFORMATION
- 510(k) Submitter: OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-cho, Hachioji-shi, Tokyo, Japan 192-8507
- Contact Person: Lisa M. Boyle Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610, USA Phone: 484-896-3676 Fax: 484-896-7128 Email: 1isa.boyle@olympus.com
# 5.2.1 DEVICE IDENTIFICATION
| ■ Device Name | VISERA ELITE II VIDEO SYSTEM CENTER |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------|
| ■ Model Name | OLYMPUS OTV-S300 |
| ■ Common Name | ENDOSCOPIC IMAGING SYSTEM |
| ■ Regulation Number | 876.1500 |
| ■ Regulation Name | Endoscope and accessories |
| ■ Regulatory Class | II |
| ■ Product Code | FET; Endoscopic Video Imaging System/Component,<br>Gastroenterology-Urology<br>NWB; Endoscope, Accessories, Narrow Band Spectrum |
| ■ Manufacturing site | Shirakawa Olympus Co., Ltd.<br>3-1 Okamiyama, Odakura, Nishigo-mura,<br>Nishishirakawa-gun, Fukushima 961-8061, Japan |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are slightly stylized, with a slight gradient effect. There is a thin yellow line underneath the word. The registered trademark symbol is located to the right of the word.
K193026
Page 2 of 12
# 5.2.2 DEVICE IDENTIFICATION
| ■ Device Name | VISERA ELITE II VIDEO TELESCOPE ENDOEYE 3D |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ■ Model Name | WA50080A, WA50082A |
| ■ Common Name | ENDOSCOPIC IMAGING SYSTEM |
| ■ Regulation Number | 876.1500 |
| ■ Regulation Name | Endoscope and accessories |
| ■ Regulatory Class | II |
| ■ Product Code | GCJ; Laparoscope, General & Plastic surgery<br>HET; Laparoscope, Gynecologic (And Accessories)<br>GCM; Endoscope, Rigid<br>NWB; Endoscope, Accessories, Narrow Band Spectrum |
| ■ Manufacturing site | Olympus Winter & Ibe GmbH<br>Kuehnstr.61,<br>22045 Hamburg, Germany |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are slightly rounded and have a three-dimensional effect. The word is positioned horizontally and takes up most of the frame. There is a thin yellow line below the word.
K193026
# 5.3 PREDICATE DEVICE
| • Predicate device of OTV-S300 | | |
|--------------------------------|---------------------------------|------------|
| Primary Predicate Device 1 | | |
| Device name | 510(k) Submitter | 510(k) No. |
| OLYMPUS OTV-S190 | OLYMPUS MEDICAL<br>SYSTEMS CORP | K111425 |
| Additional Predicate Device 2 | | |
| Device name | 510(k) Submitter | 510(k) No. |
| OLYMPUS CLV-S190 | OLYMPUS MEDICAL<br>SYSTEMS CORP | K111425 |
| Additional Predicate Device 3 | | |
| Device name | 510(k) Submitter | 510(k) No. |
| MAJ-Y0154 | OLYMPUS MEDICAL<br>SYSTEMS CORP | K123365 |
## ● Predicate device of WA50080A, WA50082A
| Primary Predicate Device 1 | | |
|----------------------------|--------------------------|------------|
| Device name | 510(k) Submitter | 510(k) No. |
| WA50050A/52A | Olympus Winter& Ibe GmbH | K190744 |
# 5.4 Recalled Predicate (K111788)
The Endoeye HD II Video Telescope cleared in predicate K111788 was the subject of a class II design related recall due to a damaged temperature sensor in the distal end of the endoscope. The issue has been successfully resolved and design changes have been cleared via K190744.
The installed heating element warms the distal tip to minimize or eliminate fogging of the lens during the procedure and aids in the prevention of endoscope removal to clean the lens due to fogging during a procedure. The ENDOEYE 3D does not include this feature.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are a sans-serif font and are evenly spaced. There is a yellow line underneath the word, which appears to be a design element. The registered trademark symbol is located to the right of the word.
# 5.5 DEVICE DESCRIPTION
# ● General Description of the subject device
The subject VISERA ELITE II consists of the following primary components:
- VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300
- VISERA ELITE II VIDEO TELESCOPE ENDOEYE 3D WA50080A/ WA50082A
# • Principle of Operation
The fundamental technology for the 3D function is identical to the predicate device.
# A) VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTY-S300
The subject device has both a processor function and light source function. By switching on the illumination lamp in the subject device provides light through the endoscopes directly for endoscopic observation.
This device also constantly regulates the endoscopic image brightness from the video system center. The observation mode can be switched by the optical filter and LED extracting the specific wavelengths. There are two modes: WLI (White light imaging) mode for normal observation and NBI (Narrow-band imaging) mode for an optical image enhancement technology.
By driving the CCD equipped in the endoscopes, the images are transduced into electrical signals from the optical signals, and the subject device displays the endoscopic images on the monitors.
# B) VISERA ELITE II VIDEO TELESCOPE ENDOEYE 3DWA50080A/ WA50082A
The subject devices are designed to transfer optical images from a body cavity via a lens system directly to an imager for further electrical signal transmission to a video processor. The video telescope is designed for examination, diagnosis, and visualization of treatment (treatment can only be performed by using endoscopic accessories in combination). For illumination of body cavities, the transfer of light from a supply unit to the body cavity is achieved by means of a light guide. The subject device, ENDOEYE 3D, is a design variant of the predicate ENDOEYE HD II video endoscope offering a 3D observation mode. The 3D effect enables the surgeon to perceive significant spatial information which is beneficial in terms of optimized speed, accuracy and precision of surgical procedures by users of all skill levels. It may also shorten the learning curve for surgical tasks.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are spaced closely together, and the word is centered in the image. Below the word "OLYMPUS" is a thin, horizontal, yellow line. The background of the image is white.
The ENDOEYE 3D is used with a 3D video system and a 3D monitor with dedicated polarization glasses.
For laparoscopic applications, the video telescope is inserted via a trocar into the patient.
The ENDOEYE 3D is used with a video system center, light source and monitor to achieve its intended function.
In addition, the ENDOEYE 3D can be placed in compatible instrument trays for reprocessing.
The ENDOEYE 3D can provide an image with either white light or narrow band imaging.
# 5.6 INDICATIONS FOR USE
# A) VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300
This video system center is intended to be used with OLYMPUS camera heads. endoscopes, monitors, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis, treatment, and video observation.
# B) VISERAELITE II VIDEO TELESCOPE ENDOEYE 3D WA50080A/ WA50082A
This instrument has been designed to be used with a video system center, light source, documentation equipment, monitor, hand instruments, electrosurgical unit, and other ancillary equipment for endoscopy and endoscopic surgery within the thoracic and abdominal cavities including the female reproductive organs.
The device is also indicated for visualization during transanal and transvaginal natural orifice surgery.
The ENDOEYE 3D has the same indication of use as the predicate device except for the addition of the last sentence regarding visualization of transanal and transvaginal applications. This difference does not represent a different intended use.
> 510(k) Summary Page 5 of 12
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. A thin, horizontal, yellow line is underneath the word. The registered trademark symbol is located to the right of the word.
K193026
# 5.7 COMPARISON OF TECHNOLOGY CHARACTERISTICS WITH THE PREDICATE DEVICE
The VISERA ELITE II has the same technological characteristics and design as the predicate device except for the following new features:
-Integration of the video processor, light source, and 3D mixer (OTV-S300)
-New integrated video connector into one for 3D observation
-New LED construction
- -No fog free function
The detailed comparison chart between subject device and predicate devices are as following.
> 510(k) Summary Page 6 of 12
{9}------------------------------------------------
K193026 Page 7 of 12 Traditional 510(k) Notification VISERA ELITE II VIDEO SYSTEM CENTER, TELESCOPE
Image /page/9/Picture/1 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are slightly shadowed, giving them a three-dimensional appearance. A thin, curved, yellow line is underneath the word. There is a registered trademark symbol next to the S.
# Device Comparison Table of OTV-S300
| | <Subject device><br>OTV-S300 | <Primary Predicate<br>device 1><br>OTV-S190 | <Additional Predicate<br>device 2><br>CLV-S190 | <Additional Predicate<br>device 3><br>MAJ-Y0154<br>3D PROCESSOR |
|----------------------------|------------------------------|---------------------------------------------|------------------------------------------------|-----------------------------------------------------------------|
| 510(k)<br>Number | K193026 | K111425 | K111425 | K123365 |
| Rated voltage | 100V AC<br>50/60Hz | 100V AC<br>50/60Hz | 100V AC<br>50/60Hz | 100V AC<br>50/60Hz |
| Rated input | 400VA | 150VA | 500A | 100VA |
| Dimension<br>(maximum) | W383×H199×D506<br>(mm) | W382×H91×D489<br>(mm) | W383×H162×D536<br>(mm) | W375×H91×D478<br>(mm) |
| Weight | 19.3kg | 8.8kg | 14.9kg | 7.7kg |
| AGC (Auto<br>gain control) | Provided | Provided | Provided | Not provided |
| 3D observation | Provided | Not provided | Not provided | Provided |
| Front panel<br>(Operation) | Touch panel | Push button | Push button | Push button |
| Examination<br>Lamp | LED | Not provided | Xenon lamp | Not provided |
| NBI | Provided | Provided | Provided | Not provided |
{10}------------------------------------------------
K193026 Traditional 510 € Notification VISERA ELITE II VIDEO SYSTEM CENTER, TELESCOPE
Image /page/10/Picture/1 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are closely spaced and have a slight 3D effect. There is a registered trademark symbol to the right of the word. A yellow bar is located underneath the word.
| | <Subject device><br>WA50080A, WA50082A | <Predicate device 1><br>WA50050A/52A |
|--------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)<br>Number | K193026 | K190744 / K111788 |
| Outer diameter<br>of insertion<br>tube (distal<br>end) | WA50080A: max.10.07 mm<br>WA50082A: max.10.20 mm | WA50050A: max. 5.45 mm<br>WA50052A: max. 5.45 mm |
| Maximum<br>working<br>length | WA50080A: 330 mm<br>WA50082A: 335 mm | WA50050A: 300.5 mm<br>WA50052A: 303 mm |
| Direction of<br>view (DOV) | WA50080A: 0°<br>WA50082A: 30° | WA50050A: 0°<br>WA50052A: 30° |
| NBI<br>observation | Compatible | compatible |
| Field of view<br>(FOV) | Both models: 67° ± 9° | Both models: 80.4° ± 8° |
| Depth of field | 25 to 120 mm | 21 to 200 mm |
| Type of CCD<br>chip | Color CCD | Color CCD |
| | <Subject device> | <Predicate device 1> |
| | WA50080A, WA50082A | WA50050A/52A |
| Number of CCD chip | 2 | 1 |
| 2D function | Available | Available |
| 3D function | The video telescope can transmit 2D images and 3D images to the video system center. The video telescope is equipped with a second image sensor inside the distal end to create 3D images that can be viewed on compatible video system centers and video monitors using compatible 3D glasses. | The video telescope can only transmit 2D images to the video system center. A 3D image sensor is not included and therefore; no 3D mode is offered. |
# Device Comparison Table of WA50080A, WA50082A
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are evenly spaced and the word is centered. There is a yellow gradient below the word, and a registered trademark symbol is located to the right of the word.
{12}------------------------------------------------
K193026
All other technological characteristics of both the subject and predicate devices are identical. The differences in technological characteristics noted in the tables do not raise different questions of safety and effectiveness.
# 5.8 PERFORMANCE DATA
# 1) Risk analvsis
Risk analysis for the OTV-S300 and the WA50080A/ WA50082A was conducted in accordance with established in-house acceptance criteria based on ISO 14971:2007. The design verification tests and their acceptance criteria were identified and performed as a result of the risk analysis assessment.
The following performance data is provided in support of the substantial equivalence determination.
# 2) Reprocessing validation testing
Reprocessing instruction and reprocessing method validation testing for the WA50080A/WA50082A were conducted and documentation was provided as recommended by Guidance for Industry and Food and Drug Administration Staff, "Reprocessing Medical Devices in Health Care Setting: Validation Methods and Labeling".
# 3) Biocompatibility testing
Biocompatibility testing for the WA50080A/WA50082A were conducted in accordance with the FDA's Guidance for Industry and Food and Drug Administration Staff, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The biocompatibility testing included the following tests:
- in vitro cytotoxicity
- irritation i
- acute systemic toxicity ।
- material mediated pyrogenicity i
- dermal sensitization test in guinea pigs Magnusson and Kligman (M&K) ।
{13}------------------------------------------------
# 4) Software verification and validation testing
Software verification and validation testing for the he OTV-S300 and the WA50080A/WA50082A was conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices".
# 5) Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing was conducted on the OTV-S300 and the WA50080A/WA50082A. The system complies with the ANSI/AAMI ES 60601-1:2005/(R) 2012 and A1:2012 and IEC 60601-2-18:2009 standards for safety and the IEC 60601-1-2:2014 standards for EMC.
# 6) Performance testing - Bench
Bench testing for the OTV-S300 and the WA50080A/WA50082A as listed below was conducted to ensure that the subject device performs as intended and meets design specifications.
# [OTV-S300]
- Difference for Emergency Lamp (to ensure non-failure LEDs will functionas emergency lights)
- NBI Observation
- Observation function
- Image Quality, including:
- · Brightness
- · Image Intensity
- Color Performance
- · Signal to Noise ratio
-Usability
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are slightly shadowed, giving them a three-dimensional appearance. Below the word "OLYMPUS" is a thin, horizontal band of yellow. The registered trademark symbol is located to the right of the word.
## [WA50080A/ WA50082A]
- Real-use contamination
- Field of View
- Design Validation/Usability
- Evaluation of optical properties incl. comparison of optical system properties and image quality (predicate vs. subject devices)
- Thermal Safety
- Surfaces and Edges
- Working Length
- Evaluation of ghost effect
- Evaluation of still image
- Distortion
- Resolution
- Evaluation of performance after reprocessing
- Evaluation of illumination performance
- Mechanical stress resistance (bending/impact)
[OTV-S300 and WA50080A/ WA50082A]
- Photobiological safety
# 7) Performance testing - Animal
No animal study was performed to demonstrate substantial equivalence.
# 8) Performance testing - Clinical
No clinical study was performed to demonstrate substantial equivalence.
# 59 CONCLUSIONS
Based on the performance testing the OTV-S300 and the WA50080A/WA50082A are as safe and effective and are substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.