ENDOEYE HD II (WA50040A, WA50042A, WA50050A, WA50052A)
K190744 · Olympus Winter & Ibe GmbH · HET · Aug 29, 2019 · Obstetrics/Gynecology
Device Facts
Record ID
K190744
Device Name
ENDOEYE HD II (WA50040A, WA50042A, WA50050A, WA50052A)
Applicant
Olympus Winter & Ibe GmbH
Product Code
HET · Obstetrics/Gynecology
Decision Date
Aug 29, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 884.1720
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This instrument has been designed to be used with a video system center, documentation equipment, monitor, hand instruments, electrosurgical unit, and other ancillary equipment for endoscopic surgery within the thoracic and abdominal cavities including the female reproductive organs.
Device Story
ENDOEYE HD II is a rigid video telescope for endoscopic diagnosis, treatment, and surgery. Device inputs include optical images from thoracic/abdominal cavities; outputs are high-definition video signals displayed on a monitor. Used in OR settings by surgeons; requires connection to video system center, light source, and monitor. Features include white light or narrow band imaging (NBI) and distal tip heater (select models) to prevent lens fogging. Provides visualization to assist surgeons in performing minimally invasive procedures. Benefits include high-definition visualization of internal anatomy to guide surgical interventions.
Clinical Evidence
Bench testing only. Testing included electrical safety (IEC 60601-1, IEC 60601-2-18), EMC (IEC 60601-1-2), thermal safety, mechanical performance (ISO 8600-3/4/5), transport/shipping (ASTM D4169-16), illumination system testing, software verification (IEC 62304), and usability validation of reprocessing instructions. All samples met predefined acceptance criteria.
Technological Characteristics
Rigid video telescope; color CCD sensor (2 chips); NBI observation capability; distal tip heater (select models). Connectivity: connects to video system center/light source. Sterilization: autoclavable. Standards: IEC 60601-1, IEC 60601-2-18, IEC 60601-1-2, ISO 8600-3/4/5, ISO 14971. Software: includes PCB with supervisor circuitry.
Indications for Use
Indicated for patients undergoing endoscopic surgery within the thoracic and abdominal cavities, including the female reproductive organs.
Regulatory Classification
Identification
A gynecologic laparoscope is a device used to permit direct viewing of the organs within the peritoneum by a telescopic system introduced through the abdominal wall. It is used to perform diagnostic and surgical procedures on the female genital organs. This generic type of device may include: Trocar and cannula, instruments used through an operating channel, scope preheater, light source and cables, and component parts.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
August 29, 2019
Olympus Winter & Ibe GmbH Sheri Musgnung Manager, Regulatory Affairs Olympus Corporation of the Americas 3500 Corporate Parkway Center Valley, PA 18034-061
Re: K190744
Trade/Device Name: ENDOEYE HD II (WA50040A, WA50042A, WA50050A, WA50052A) Regulation Number: 21 CFR 884.1720 Regulation Name: Gynecologic laparoscope and accessories Regulatory Class: II Product Code: HET, GCJ, NWB Dated: July 29, 2019 Received: July 30, 2019
Dear Sheri Musgnung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Sharon Andrews Assistant Division Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190744
Device Name ENDOEYE HD II (WA50040A, WA50042A, WA50050A, WA50052A)
Indications for Use (Describe)
This instrument has been designed to be used with a video system center, documentation equipment, monitor, hand instruments, electrosurgical unit, and other ancillary equipment for endoscopic surgery within the thoracic and abdominal cavities including the female reproductive organs.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary of Safety and Effectiveness
## August 26, 2019
#### General Information 1
| Manufacturer: | Olympus Winter & Ibe GmbH |
|---------------|--------------------------------------------|
| | Kuehnstr. 61 |
| | 22045 Hamburg |
| | Germany |
| | Establishment Registration Number: 9610773 |
| Official Correspondent: | Sheri L Musgnung |
|-------------------------|--------------------------------------------|
| | Manager, Regulatory Affairs |
| | Olympus Corporation of the Americas |
| | 3500 Corporate Parkway |
| | Center Valley PA 18034-061 |
| | Phone: 484 896-3147 |
| | Email: sheri.musgnung@olympus.com |
| | Establishment Registration Number: 2429304 |
#### Device Identification 2
| Proprietary /Trade Name: | ENDOEYE HD II (WA50040A, WA50042A,WA50050A, WA50052A) |
|--------------------------|-------------------------------------------------------|
| Classification name: | Laparoscope, Gynecologic and accessories |
| Regulations Number: | 21 CFR 884.1720 |
| Regulatory class: | II |
| Product code: | HET<br>GCJ / NWB |
| Review Panel: | Obstetrics and Gynecology |
#### Predicate Device 3
Olympus K11788, ENDOEYE HD II
The predicate has been subject to a design related recall.
{4}------------------------------------------------
#### 4 Product Description
The ENDOEYE HD II - High Definition Digital Video Laparoscope is a rigid video telescope used for endoscopic diagnosis, treatment, video observation and surgery within the thoracic and abdominal cavities including female reproductive organs. For laparoscopic applications, the video telescope is inserted via a trocar into the patient.
The ENDOEYE HD II is used with a video system center, light source and monitor to achieve its intended function.
In addition, the ENDOEYE HD II can be inserted in compatible instrument trays for reprocessing.
The ENDOEYE HD II can provide an image with either white light or narrow band imaging and
WA50040A and WA50042A include a heater function at the distal tip to reduce fogging of the lens.
#### 5 Indications for Use
This instrument has been designed to be used with a video system center, light source, documentation equipment, monitor, hand instruments, electrosurgical unit, and other ancillary equipment for endoscopy and endoscopic surgery within the thoracic and abdominal cavities including the female reproductive organs.
The ENDOEYE HD II has the same intended use as the predicate device.
#### Technological Characteristics 6
The ENDOEYE HD II has the same technological characteristics as the predicate device.
| Device | Subject device | Predicate device |
|---------------|--------------------|--------------------|
| Device Name | EndoEYE HD II | EndoEYE HD II |
| Model Number | WA50040A, WA50042A | WA50040A, WA50042A |
| | WA50050A, WA50052A | WA50050A, WA50052A |
| 510(k) number | Not yet known | K111788 |
{5}------------------------------------------------
| Device | Subject device | Predicate device |
|------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Field of view | WA50040A, WA50042A:<br>90° $\pm$ 9° (at 16:9 aspect<br>ratio)<br>WA50050A, WA50052A:<br>80.4° $\pm$ 8° (at 5:4 aspect<br>ratio) | WA50040A, WA50042A:<br>90° $\pm$ 9° (at 16:9 aspect<br>ratio)<br>WA50050A, WA50052A:<br>80.4° $\pm$ 8° (at 5:4 aspect<br>ratio) |
| Depth of field | 21 to 200 mm | 21 to 200 mm |
| Direction of<br>viewing (forward) | WA50040A, WA50050A:<br>0° $\pm$ 10°<br>WA50042A, WA50052A:<br>30° $\pm$ 10° | WA50040A, WA50050A:<br>0° $\pm$ 10°<br>WA50042A, WA50052A:<br>30° $\pm$ 10° |
| Type of CCD chip | Color CCD | Color CCD |
| Number of CCD<br>chip | 2 | 2 |
| Outer diameter of<br>distal end | Ø 10 mm (WA50040A,<br>WA50042A)<br>Ø 5.4 mm (WA50050A,<br>WA50052A) | Ø 10 mm (WA50040A,<br>WA50042A)<br>Ø 5.4 mm (WA50050A,<br>WA50052A) |
| Outer diameter of<br>insertion tube<br>maximum | WA50040A: 10.07 mm<br>WA50042A: 10.2 mm<br>WA50050A: 5.45 mm<br>WA50052A: 5.45 mm | WA50040A: 10.07 mm<br>WA50042A: 10.2 mm<br>WA50050A: 5.45 mm<br>WA50052A: 5.45 mm |
| Working length | WA50040A: 325 $\pm$ 5 mm<br>WA50042A: 330 $\pm$ 1 mm<br>WA50050A: 300 $\pm$ 0.5 mm<br>WA50052A: 302 $\pm$ 1 mm | WA50040A: 325 $\pm$ 5 mm<br>WA50042A: 330 $\pm$ 1 mm<br>WA50050A: 300 $\pm$ 0.5 mm<br>WA50052A: 302 $\pm$ 1 mm |
| NBI observation | available | available |
| Heater<br>functionality | WA50040A,<br>WA50042A: Yes<br>WA50050A,<br>WA50052A: No | WA50040A,<br>WA50042A: Yes<br>WA50050A,<br>WA50052A: No |
| Autoclavability | yes | yes |
The subject devices have equivalent technology, performance, dimensions and materials. The differences to the predicate device ENDOEYE HD II are:
- . Addition of a printed circuit board (PCB) containing a supervisor circuitry and a second temperature sensor
- Updated Instructions for Use. .
The above described differences in technological characteristics do not raise different questions of safety and effectiveness.
{6}------------------------------------------------
#### 7 Performance Data
The following Design verification and validation testing was performed on the subject device:
- Electrical safety as per IEC 60601-1 Edition 3.1 and IEC 60601-2-18 Edition 3.0
- Electromagnetic compatibility (EMC) as per IEC 60601-1-2 Edition 4.0
- Thermal Safety as per IEC 60601-1 Edition 3.1 and IEC 60601-2-18 Edition ● 3.0
- Mechanical performance testing as per ISO 8600-3 Edition 1, ISO 8600-4 Edition 2 and ISO 8600-5 Edition 1
- Transport and shipping testing as per ASTM D4169-16 ●
- Tests related to the expected service life as per IEC 60601-1 Edition 3.1 and . IEC 60601-2-18 Edition 3.0
- Testing of the illumination system as per IEC 60601-1 Edition 3.1 and IEC 60601-2-18 Edition 3.0
- Software Verification as per IEC 62304 Edition 1.1 ●
- . Usability Validation of the Instructions for Use as per FDA Guidance Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling
Risk analysis was carried out in accordance with established internal acceptance criteria based on ISO 14971:2007.
All samples tested met their predefined acceptance criteria.
#### 8 Conclusion
The performance data support the safety of the device and demonstrate that the subject devices comply with the recognized standards as specified and support a substantial equivalence determination.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.