K112770 · Ninepoint Medical, Inc. · NQQ · Dec 16, 2011 · Radiology
Device Facts
Record ID
K112770
Device Name
NVISION VLE IMAGING SYSTEM, NVISION VLE CATHETER
Applicant
Ninepoint Medical, Inc.
Product Code
NQQ · Radiology
Decision Date
Dec 16, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Indications for Use
The Nvision VLE Imaging System is indicated for use as an imaging tool in the evaluation of human tissue microstructure by providing two-dimensional, cross- sectional, real-time depth visualization.
Device Story
Nvision VLE Imaging System utilizes optical coherence tomography (OCT) to provide real-time, two-dimensional, cross-sectional visualization of human tissue microstructure. System components include a console, catheter, and inflation accessory kit. Operates via low coherence interferometry and swept-source technology (SS-OCT/OFDI); optical scattering from tissue depths is measured to determine internal morphology. High-speed, high-sensitivity imaging enables clinicians to evaluate tissue microstructure. Used in clinical settings; intended for prescription use.
Clinical Evidence
In-vitro and clinical testing were performed. All components, subassemblies, and systems met required specifications for completed tests. No specific sensitivity, specificity, or AUC metrics were provided in the summary.
Technological Characteristics
Optical Coherence Tomography (OCT) / Optical Frequency Domain Imaging (OFDI) system. Utilizes swept-source technology and low coherence interferometry. Components: Console, catheter, inflation accessory kit. Complies with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-1-4 (programmable electrical medical systems), IEC 60601-2-18 (endoscope), IEC 60601-2-22/60825-1 (laser safety), and ISO 10993-1/7 (biocompatibility). Sterilization via ethylene oxide (ISO 11135-1).
Indications for Use
Indicated for use as an imaging tool in the evaluation of human tissue microstructure by providing two-dimensional, cross-sectional, real-time depth visualization in humans.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
k112770
DEC 1 6 2011
Image /page/0/Picture/2 description: The image shows the logo for NinePoint Medical. The logo consists of the word "NinePoint" in a large, bold font, with the word "MEDICAL" in a smaller font underneath. Above the word "NinePoint" is a circle made up of small dots. The logo is simple and professional, and the use of the circle of dots adds a touch of visual interest.
One Kendall Square, Suite B7501 Cambridge, Massachusetts 02139 ninepointmedical.com
## 510(k) Summary
1. Basic Information-Submitter:
| 510(k) Owner: | NinePoint Medical Inc. |
|-------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | One Kendall Square, Suite B7501<br>Cambridge, MA 02139<br>(617) 250-7190 (main number)<br>(617) 250-7199 (fax) |
| Official Contact: | Cindy Domecus, R.A.C. (US & EU)<br>Principal, Domecus Consulting Services LLC<br>Consultant to NinePoint Medical, Inc.<br>(650) 343-4813<br>(650) 343-7822 (fax)<br>DomecusConsulting@comcast.net |
Date Summary Prepared: 21 September 2011
- 2. Device Name:
Trade Name: Nvision VLE Imaging System (OCT) Common Name: Optical Coherence Tomography Imaging System/Optical Frequency Domain Imaging Classification Name: Ultrasonic pulsed echo imaging system Regulation Number: 21 CFR 892.1560 Product Code: NQQ Classification: Class II
- 3. Predicate Devices:
Imalux OCT Imaging System - K033783 Imalux OCT Niris System - K042894 Imalux OCT Probe Sheath - K041077 Tomophase OCTIS - K102599
> Premarket Notification, Nvision VLE Imaging System Proprietary and Confidential Information of NinePoint Medical Inc.
{2}------------------------------------------------
ll277u
Image /page/1/Picture/1 description: The image shows the logo for NinePoint Medical. The logo features the word "NinePoint" in a bold, sans-serif font, with the word "MEDICAL" in a smaller font below it. Above the word "NinePoint" is a circular design made up of nine dots. The dots are evenly spaced around the circle.
One Kendall Square, Suite B7501 ambridge. Massachusetts 02139 nine point medical. com
- 4. Device Description:
The NinePoint Medical Nvision VLE Imaging System is a general imaging system comprised of the Nvision VLE Console, Nvision VLE Catheter and the Nvision VLE Inflation Accessory Kit.
- 5. Indications for Use Statement:
The Nvision VLE Imaging System is indicated for use as an imaging tool in the evaluation of human tissue microstructure by providing two-dimensional, cross- sectional, real-time depth visualization.
- 6. Technological Characteristics:
Optical coherence tomography (OCT) is an imaging technique analogous to ultrasound; however, instead of producing an image from the scattering of sound waves, it utilizes optical scattering based on differences in tissue composition to form two-dimensional images. OCT systems use the principles of low coherence interferometry to generate high-resolution images of tissue microstructures. Interferometry and OCT enable the measurement of the optical back-reflected signals, and therefore the morphology of the internal microstructures can be determined by the back-reflected signal from the different depths of the tissue.
All of the predicate OCT systems employ low coherence interferometry as the basis of the system. The systems use broadband optical sources with either superluminescent diodes or swept sources, high dynamic range detection and either fixed or scanning reference arms. The scanned reference arms were utilized in the Imalux, time domain OCT systems, while the advent of fixed reference arm and swept source OCT technology (Tomophase and NinePoint Medical) has permitted high speed and high sensitivity image capability. The Tomophase predicate device (K102599) and the Nvision VLE Imaging System use a derivative of OCT utilizing swept source technology, referred to as SS-OCT or optical frequency domain imaging, OFDI. OFDI is a derivative
> Premarket Notification, Nvision VLE Imaging System Proprietary and Confidential Information of NinePoint Medical Inc.
{3}------------------------------------------------
"dวววw
Image /page/2/Picture/1 description: The image shows the logo for NinePoint Medical. Above the word "NinePoint" is a circle made of nine dots. Below the word "NinePoint" is the word "MEDICAL" in smaller, capitalized letters.
One Kendall Square, Suite B7501 Cambridge, Massachusetts 02139 ninepointmedical.com
development of the time-domain OCT imaging modality, which enables high speed, two-dimensional, cross-sectional, real-time imaging.
## 7. Performance data:
The NinePoint Medical Nvision VLE Imaging System will be tested against and comply with the following voluntary standards:
> IEC 60601-1, General Safety IEC 60601-1-2, Electromagnetic Compatibility IEC 60601-1-4, Programmable Electrical Medical Systems IEC 60601-2-18, Endoscope IEC 60601-2-22, Laser Safety IEC 60825-1, Laser Safety ISO 10993-1, Biological Evaluation of Medical Devices ISO 10993-7, Biological Evaluation of Medical Devices, Ethylene Oxide Sterilization residual testing
ISO 11135-1, Sterilization of Health Care Products, Ethylene Oxide
In-vitro and clinical testing have been performed and all components, subassemblies and/or full devices and systems have met the required specifications for the completed tests.
8. 510(k) Summary:
----
NinePoint Medical Inc. has demonstrated that the Nvision VLE Imaging System is substantially equivalent to the predicate devices listed above.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the United States Department of Health and Human Services. The logo features a stylized eagle with three curved lines representing its wings. The eagle is positioned within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" is arranged around the circumference of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
DEC 1 6 2011
NinePoint Medical, Inc. % Domecus Consulting Services, LLC Ms. Cindy Domecus 1171 Barroihet Avenue Hillsborough, California 94010
Re: K112770
Trade/Device Name: Nvision VLE Imaging System Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: Class II Product Code: NOO Dated: December 09, 2011 Received: December 12, 2011
Dear Ms. Domecus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{5}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
> Sincerely yours, Mark N. Melkerso Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
112770
## Section 5 Indications for Use Statement
510(k) Number (if known):
K112770 This application
Device Name:
Indications for Use:
Nvision VLE Imaging System
The Nvision VLE Imaging System is indicated for use as an imaging tool in the evaluation of human tissue microstructure by providing twodimensional, cross sectional, real-time depth visualization.
Prescription Use _____________________________________________________________________________________________________________________________________________________________
AND/OR
(21CFR 807
</12770
Over-the -Counter Use
(Part 21 CFR 801 Subpart D) Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | <div style="text-align: right;"> <i>Nil RPOglen</i> <br/> <i>for mxm</i> </div> |
|--------------------------------------------------------|---------------------------------------------------------------------------------|
| (Division Sign-Off) | |
| Division of Surgical, Orthopedic, | |
| and Restorative Devices | |
510(k) Number_ Premarket Notification, Nvision VLE Imaging System Proprietary and Confidential Information of NinePoint Medical Inc.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.