da Vinci Xi Surgical System, da Vinci X Surgical System
K182140 · Intuitive Surgical, Inc. · NAY · Oct 24, 2018 · Gastroenterology, Urology
Device Facts
Record ID
K182140
Device Name
da Vinci Xi Surgical System, da Vinci X Surgical System
Applicant
Intuitive Surgical, Inc.
Product Code
NAY · Gastroenterology, Urology
Decision Date
Oct 24, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Xi Surgical System Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically- assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use. The Intuitive Surgical Endoscopic Instrument Control System (da Vinci X Surgical System Model IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically- assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
Device Story
Computer-controlled, electro-mechanical surgical system; consists of surgeon console, patient-side cart, and vision-side cart. Surgeon operates console to control endoscopic instruments (EndoWrist) and rigid endoscopes. System translates surgeon hand movements into precise instrument motion for minimally invasive surgery. Used in operating rooms by trained physicians. Provides high-definition visualization and instrument control to facilitate tissue manipulation, dissection, and suturing. Benefits include potential for reduced blood loss and shorter hospital stays compared to traditional methods. This submission adds 'Gastrectomy' to the labeled indications for use.
Clinical Evidence
Literature-based clinical evidence; 16 publications (2 randomized controlled trials, 10 systematic reviews/meta-analyses, 4 database studies). Compared da Vinci-assisted gastrectomy to laparoscopic and open cohorts. Results showed comparable mortality, complication rates, blood transfusion, and reoperation rates. da Vinci-assisted procedures showed comparable or lower estimated blood loss and comparable or shorter hospital stays compared to laparoscopic/open cohorts. Operative times were comparable or longer, without impacting mortality or complication rates.
Technological Characteristics
Software-controlled, electro-mechanical system. Components: Surgeon Console, Patient Side Cart, Vision Side Cart. Uses rigid endoscopes and EndoWrist instruments. No changes to technological characteristics from predicate devices.
Indications for Use
Indicated for adult and pediatric patients undergoing urologic, general laparoscopic, gynecologic laparoscopic, general thoracoscopic, and thoracoscopically-assisted cardiotomy procedures, including coronary anastomosis via adjunctive mediastinotomy. Used for tissue manipulation (grasping, cutting, dissection, approximation, ligation, suturing, electrocautery) and delivery of ablation probes.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 24, 2018
Intuitive Surgical, Inc % Cindy Domecus Principal, Domecus Consulting Services, LLC/ISI Chief Regulatory Advisor Domecus Consulting Services LLC 1171 Barroihet Drive Hillsborough, California 94010
Re: K182140
Trade/Device Name: da Vinci Xi Surgical System, da Vinci X Surgical System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: NAY Dated: October 8, 2018 Received: October 10, 2018
## Dear Cindy Domecus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Image /page/1/Picture/5 description: The image shows the closing of a letter with the word "Sincerely,". Below that is the name "Jennifer R. Stevenson-S3". The following line says "For Binita S. Ashar, M.D., M.B.A., F.A.C.S.". The last line says "Director".
Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known)
#### K182140
#### Device Name
Intuitive Surgical® da Vinci® Xi Endoscopic Instrument Control System
(da Vinci Xi System, Model IS4000) and Endoscopic Instruments and Accessories
#### Indications for Use (Describe)
The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Xi Surgical System Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically- assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
Type of Use (Select one or both, as applicable):
| <span style="font-family: DejaVu Sans, sans-serif">✘</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known)
#### K182140
#### Device Name
Intuitive Surgical® da Vinci® X Endoscopic Instrument Control System
(da Vinci X System, Model IS4200) and Endoscopic Instruments and Accessories
#### Indications for Use (Describe)
The Intuitive Surgical Endoscopic Instrument Control System (da Vinci X Surgical System Model IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically- assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
Type of Use (Select one or both, as applicable)
| <div> <span> <svg height="16" width="16"> <polygon points="0,0 0,16 16,16 16,0" style="fill:none;stroke:black;stroke-width:2"></polygon> <polyline points="2,8 6,12 14,4" style="fill:none;stroke:black;stroke-width:2"></polyline> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg height="16" width="16"> <polygon points="0,0 0,16 16,16 16,0" style="fill:none;stroke:black;stroke-width:2"></polygon> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary (21 CFR § 807.92(c))
## I. SUBMITTER INFORMATION
| Submitter: | Intuitive Surgical, Inc.<br>1266 Kifer Road<br>Sunnyvale, CA 94086 |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Cindy Domecus, R.A.C. (US & EU)<br>Principal, Domecus Consulting Services LLC<br>Chief Regulatory Advisor to Intuitive Surgical<br>Telephone: 650.343.4813<br>Fax: 650.343.7822<br>Email: domecusconsulting@comcast.net |
| Date Summary Prepared: | October 23, 2018 |
| II. SUBJECT DEVICE INFORMATION | |
| Device Trade Name: | da Vinci® Xi and X Surgical Systems, Model IS4000 and M |
| Device Trade Name: | da Vinci® Xi and X Surgical Systems, Model IS4000 and Model IS4200 |
|----------------------|--------------------------------------------------------------------|
| Common Name: | System, Surgical, Computer Controlled Instrument |
| Classification Name: | Endoscope and Accessories (21 CFR §876.1500) |
| Regulatory Class: | II |
| Product Code: | NAY |
| Submission Type: | Traditional 510(k) |
#### III. PREDICATE DEVICE INFORMATION:
Intuitive Surgical da Vinci Xi and X Surgical Systems, Models IS4000 and Predicate Device: IS4200 (K131861, K152578, K153276, K161178, K170713, K171632, K171294 K172643, K173842, K173585) Intuitive Surgical da Vinci Si Surgical System, Model IS3000 (K081137, K123463, K090993)
## IV. DEVICE DESCRIPTION:
This 510(k) is for a labeling modification only, to include the following additional representative, specific procedure under the cleared "general laparoscopic surgical procedures" Indication for Use of the da Vinci Xi Surgical System, Model IS4000 (K131861) and the da Vinci X Surgical System, Model IS4200 (K171294) and the associated labeling claims: Gastrectomy. There are no changes to the technological characteristics of the cleared da Vinci Xi or X Surgical Systems (Models IS4000 and IS4200) proposed in this submission. The da Vinci Xi and X Surgical Systems, Models IS4000 and IS4200, are softwarecontrolled, electro-mechanical systems designed for surgeons to perform minimally invasive surgery. The Model IS4000 and Model IS4200 Surgical Systems consist of a Surgeon Console, a Patient Side Cart (PSC), and a Vision Side Cart (VSC) and are used with an Endoscope, EndoWrist Instruments, and Accessories.
{5}------------------------------------------------
## V. INDICATIONS FOR USE
The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Surgical System, Models: IS4000 and IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
## Precaution for Representative Uses
The demonstration of safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of cancer (overall survival, disease-free survival, local recurrence) or treatment of the patient's underlying disease/condition. Device usage in all surgical procedures should be guided by the clinical judgment of an adequately trained surgeon.
## VI. COMPARISON OF INTENDED USE, INDICATIONS FOR USE AND TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
There are no changes to the technological characteristics for the subject devices compared to the cleared predicate devices, da Vinci Xi Surgical System, Model IS4000 (K131861) and the da Vinci X Surgical System, Model IS4200 (K171294). This 510(k) is for a labeling modification to include Gastrectomy as a labeled use under the cleared "general laparoscopic surgical procedures" Indication for Use of the cleared predicate devices, do Vinci Xi Surgical System, Model IS4000 and the da Vinci X Surgical System, Model IS4200. The subject devices differ from the predicate devices by this modification to the labeling. Results of clinical data from literature demonstrated that the subject devices have the same intended use as the predicate devices.
## VII. PERFORMANCE DATA
There were no technological changes to the subject devices, thus no bench testing, electromagnetic compatibility testing, sterilization testing or biocompatibility testing was required.
## Clinical Study Data
Published clinical data were provided to support use of the da Vinci Xi and X Surgical Systems (Models IS4000 and IS4200) in "Gastrectomy" procedures to demonstrate that the intended use of the devices is the same as the predicate devices. Sixteen (16) publications were identified for this procedure based on specific search criteria and filters. These publications included two (2) randomized controlled trials (1b), ten (10) systematic reviews/meta-analysis (2a), and four (4) database studies (2c) comparing da Vinci
{6}------------------------------------------------
assisted procedures with laparoscopic and/or open cohorts. A detailed summary of the published clinical data on this procedure is provided in Tables 1A and 1B below.
The findings from the Gastrectomy publications show that da Vinci-assisted procedures as compared to laparoscopic procedures are associated with:
- Mortality: comparable mortality rates;
- Estimated Blood Loss (EBL): comparable or lower2 EBL volumes;
- Blood Transfusion: comparable blood transfusion rates;
- Lengths of Hospital Stay (LOS): comparable or shorter² lengths of hospital stay;
- Intraoperative Complication Rates: comparable intraoperative complication rates; ●
- Postoperative Complication Rates: comparable postoperative complication rates;
- Conversion Rates: comparable conversion rates;
- Reoperation Rates: comparable reoperation rates;
- Readmission Rates: comparable readmission rates; 3
- Anastomotic Leak Rates: comparable anastomotic leak rates;
- Anastomotic Stenosis Rates: comparable anastomotic stenosis rates; and,
- Operative Time: comparable or longer operative times. However, this increase was not associated with an increase in the mortality or complication rates.
The findings from the Gastrectomy publications show that da Vinci-assisted procedures as compared to open procedures are associated with:
- Mortality: comparable mortality rates; ●
- Estimated Blood Loss (EBL): lower EBL4 volumes;
- Blood Transfusion: comparable blood transfusion rates;
- Lengths of Hospital Stay (LOS): comparable or shorter 1engths of hospital stay;
- Intraoperative Complication Rates: comparable intraoperative complication rates;
- . Postoperative Complication Rates: comparable postoperative complication rates;
- Reoperation Rates: comparable reoperation rates;
- Readmission Rates: comparable readmission rates; 6 ●
- Anastomotic Leak Rates: comparable anastomotic leak rates;
<sup>1</sup>Nine (9) publications reported EBL data. One (1) publication (Reference #9) reported comparable values between the da Vinci and laparoscopic cohorts; seven (7) publications reported lower EBL values for the da Vinci cohort as compared to the laparoscopic cohort (References #1, 3, 8, 10, 11, 13, 15); and, one (1) publication reported higher EBL values for the da Vinci cohort as compared to the laparoscopic cohort (Reference #6). The higher EBL value in the da Vinci cohort was not associated with increased blood transfusion, complication or mortality rates. EBL values are provided in Table 1A.
²Eleven (11) publications reported LOS data. Three (3) publications (References #1, 9, 11) reported shorter LOS values for the da Vinci cohort; and, seven (7) publications reported comparable LOS values (References #5, 6, 10, 12, 13, 15) for the da Vinci and laparoscopic cohorts. One publication (Reference #4) reported comparable LOS values for the da Vinci, laparoscopic and open cohorts. LOS values are provided in Table 1A.
<sup>3</sup> One (1) publication (Reference #4) reported comparable readmission rates for the da Vinci, laparoscopic and open cohorts. "Six (6) publications reported EBL data. All six (6) publications reported lower EBL values (References #2, 3 ,6, 7, 14, 15) for the da Vinci cohorts as compared to the open cohorts. EBL values are provided in Table 1A.
<sup>්</sup> Eight (8) publications reported LOS data. Five (5) publications (References #2, 6, 7, 14, 15) reported shorter LOS values for the da Vinci cohort as compared to the open cohort; and, two (2) publications reported comparable LOS values (References #5, 12) for the da Vinci and open cohorts. One publication (Reference #4) reported comparable LOS values across the da Vinci, laparoscopic and open cohorts. LOS values are provided in Table 1A.
رَ One (1) publication (Reference #4) reported comparable readmission rates for the da Vinci, laparoscopic and open cohorts.
{7}------------------------------------------------
- Anastomotic Stenosis Rates: comparable anastomotic stenosis rates; and,
- . Operative Time: longer operative times. However, this increase was not associated with an increase in the mortality or complication rates.
## VIII. CONCLUSION
The da Vinci Xi and X Surgical Systems (models IS4000 and IS4200) have the same intended use as the predicate devices, as demonstrated by the clinical data from literature to support the safety and effectiveness for the new labeled use of Gastrectomy procedures under the "general laparoscopic surgical procedure" indications compared to the predicate devices. In addition, the subject devices have the same technological characteristics as the predicate devices. The da Vinci Xi and X Surgical Systems (Models IS4000 and IS4200) are substantially equivalent to the cleared predicate devices.
{8}------------------------------------------------
TABLE 1A: da Vinci vs. Open and da Vinci vs. Laparoscopic Gastrectomy Procedures
| Author | Study Size (N) | Operation Time (minutes) | EBL (ml) | Length of Stay (days) | Transfusion Rate (%) | Intraop Comp Rate (%) | Postop Comp Rate (%) | Mortality (in-hospital - 30 days %) | Reoperation Rate (%) | Conversion Rate (%) | Author | Study Size (N) | Positive Surgical<br>Margin Rate (%) | Proximal resection<br>margin (cm) | Distal resection<br>margin (cm) | Lymph Node Yield (n) | Anastomotic<br>Leak Rate (%) | Anastomotic Stenosis<br>Rate (%) |
|--------------------|------------------------------------------|---------------------------------------|-----------------------------------------------|----------------------------------------|-------------------------------------|-----------------------|-------------------------------|------------------------------------------------------|----------------------|-----------------------------|--------------------|---------------------------------------------------|--------------------------------------|-----------------------------------------|------------------------------------------|----------------------------------------------------------------|------------------------------|----------------------------------|
| 1. Pan 2017 | Robotic 102<br>Lap 61 | 153.11<br>151.97 | 41.27<br>83.69 | 3.75<br>5.36 | Not Reported | | 4.9<br>19.7 | 0<br>0 | Not Reported | 0<br>0 | 1. Pan 2017 | Robotic<br>102<br>Lap<br>61 | Not Reported | | | 36.06<br>30.03 | 0<br>3.3 | Not Reported |
| 2. Wang 2016 | Robotic 151<br>Open 145 | 242.7<br>192.4 | 94.2<br>152.8 | 5.7<br>6.4 | 0.66<br>0.69 | Not Reported | 9.3<br>10.3 | 0<br>0 | Not Reported | | 2. Wang 2016 | Robotic<br>151<br>Open<br>145 | Not Reported | $5.3 \pm 1.5$ | $5.5 \pm 1.7$ | 30.1<br>29.1 | 2.6<br>2.1 | Not Reported |
| 3. Eom 2016* | Robotic 53<br>Lap 541<br>Open 2654 | 268.4<br>227.3<br>200.4 | 74.2<br>137.7<br>272.4 | | Not Reported | | | | | | 3. Eom 2016* | Robotic<br>53<br>Lap<br>541<br>Open<br>2654 | Not Reported | Not Reported | | Not Reported | Not Reported | Not Reported |
| 4. Glenn 20156 | Robotic 223<br>Lap 789<br>Open 8585 | Not Reported | | 12.2<br>12.1<br>13.0 | Not Reported | | 33.6***<br>27.8***<br>31.7*** | 4.9<br>1.9<br>2.6 | Not Reported | | 4. Glenn 20156 | Robotic<br>223<br>Lap<br>789<br>Open<br>8585 | Not Reported | Not Reported | | Not Reported | Not Reported | Not Reported |
| 5. Greenleaf 20176 | Robotic 223<br>Lap 1487<br>Open 4717 | Not Reported | | 8<br>8<br>9 | Not Reported | | | Not Available1,2<br>Not Reported<br>Not Available1,2 | Not Reported | 7.6<br>19<br>N/A | 5. Greenleaf 20176 | Robotic<br>223<br>Lap<br>1487<br>Open<br>4717 | 5.8<br>10.9<br>14.6 | Not Reported | Not Reported | 129/223 (57.8%)**<br>791/1487 (53.2%)**<br>2352/4717 (49.9%)** | Not Reported | Not Reported |
| 6. Parisi 2017 | Robotic 151<br>Lap 151<br>Open 302 | 365.44<br>220.37<br>198.67 | 117.91<br>95.93<br>127.26 | 8.85<br>9.07<br>12.68 | 2.6/8.6**<br>4.6/8.6**<br>3.0/6.3** | 1.3<br>2.0<br>1.7 | 17.9<br>11.9<br>19.5 | 0<br>0<br>0 | 1.3<br>3.3<br>3.6 | 4.6<br>5.3<br>N/A | 6. Parisi 2017 | Robotic<br>151<br>Lap<br>151<br>Open<br>302 | 2.0<br>4.0<br>5.0 | Not Reported | Not Reported | 27.78<br>24.58<br>25.82 | 2.6<br>2.6<br>3.6 | 0.7<br>0.7<br>0.3 |
| 7. Caruso 2017 | Robotic 689<br>Open 5434 | 226 - 415.9<br>126.7 - 331.8 | 30.3 - 208.2<br>78.8 - 564.6 | 5.1 - 11.3<br>6.4 - 16.5 | Not Reported | | 12.77<br>11.94 | 0.44<br>0.64 | Not Reported | 1.4<br>N/A | 7. Caruso 2017 | Robotic<br>689<br>Open<br>5434 | Not Reported | Not Reported | Not Reported | 22 - 41.1<br>25.2 - 43.3 | 3.0<br>1.6 | Not Reported |
| 8. Chen 2017 | Robotic 1830<br>Lap 4123 | 202.1 - 439<br>166.7 - 361 | 30.3 - 176.6<br>34 - 212.5 | 5.1 - 14<br>5.9 - 15 | Not Reported | | 10.60<br>10.11 | 0.84<br>0.52 | 1.9<br>1.11 | 0.57<br>0.73 | 8. Chen 2017 | Robotic<br>1830<br>Lap<br>4123 | Not Reported | 2.6 - 5.9<br>2.9 - 5.7 | 4 - 7.4<br>3.8 - 7.7 | 23.1 - 47.2<br>20 - 46.5 | Not Reported | Not Reported |
| 9. Chuan 2015 | Robotic 551<br>Lap 1245 | 226 - 361<br>176 - 345 | 30.3 - 130.48<br>44.7 - 131.3 | 5.1 - 12.1<br>6.5 - 17.3 | Not Reported | | 12.70<br>13.01 | | Not Reported | | 9. Chuan 2015 | Robotic<br>551<br>Lap<br>1245 | Not Reported | 3.8 - 5.6<br>3.68 - 5.7 | 7.3 - 12.4<br>6.21 - 11.6 | 32.8 - 44.3<br>32.6 - 43.2 | 2.55<br>2.68 | Not Reported |
| 10. Duan 2017 | Robotic 993<br>Lap 2510 | 202 - 439<br>170.7 - 345 | 30.3 - 131.3<br>44.7 - 173.45 | 5.1 - 12.1<br>6.5 - 17.3 | Not Reported | | 12.22<br>12.68 | | Not Reported | | 10. Duan 2017 | Robotic<br>993<br>Lap<br>2510 | Not Reported | Not Reported | Not Reported | 25 - 44.3<br>26 - 43.2 | Not Reported | Not Reported |
| 11. Hu 2016 | Robotic 1096<br>Lap 2484 | 178 - 439<br>137.6 - 319.8 | 47 - 171.3<br>87.1 - 152.8 | 4.5 - 11<br>5.7 - 13.2 | Not Reported | | UK<br>UK | 0.26<br>0.60 | Not Reported | 1.91<br>1.30 | 11. Hu 2016 | Robotic<br>1096<br>Lap<br>2484 | Not Reported | UK<br>UK | UK<br>UK | 25.3 - 44<br>20.7 - 40 | Not Reported | Not Reported |
| 12. Kostakis 2017 | Robotic 85<br>Lap 1281<br>Open 11194 | Not Reported | | 9.6<br>8 - 16.9<br>11.5 - 24 | Not Reported | | 10.71<br>7.99<br>5.83 | 3.45<br>3.06<br>5.35 | 3.45<br>5.98<br>4.04 | 0 - 18.2<br>0 - 27.7<br>N/A | 12. Kostakis 2017 | Robotic<br>85<br>Lap<br>1281<br>Open<br>11194 | 0<br>4.63<br>8.07 | Not Reported | Not Reported | Not Reported | 0-3.4<br>0-5.6<br>0-5.8 | 0<br>2.4 - 5.6<br>0 - 1.4 |
| 13. Wang 2017 | Robotic 1134<br>Lap 2610 | 202.1 - 415.9<br>173.5 - 362.2 | 30.3 - 214.2<br>44.7 - 212.5 | 4.5 - 12.1<br>4.3 - 17.3 | Not Reported | | 11.7<br>10.7 | | Not Reported | | 13. Wang 2017 | Robotic<br>1134<br>Lap<br>2610 | Not Reported | range (3.8 - 5.8)<br>range (3.68 - 5.7) | range (5.1 - 12.4)<br>range (5.6 - 11.6) | 25 - 44.3<br>26 - 43.2 | 2.4<br>2.3 | 0.50<br>1.20 |
| 14. Yang 2017 | Robotic 606<br>Open 5364 | Not Available1,3 | Not Available1,4 | Not Available1,5 | Not Reported | | 12.71<br>12.08 | | Not Reported | | 14. Yang 2017 | Robotic<br>606<br>Open<br>5364 | Not Reported | Not Available1,2<br>Not Available1,2 | Not Available1,3<br>Not Available1,3 | Not Available1,4<br>Not Available1,4 | 3.0<br>1.6 | Not Reported |
| 15. Zong 2014 | Robotic 520/997<br>Lap 2207<br>Open 5260 | 202 - 344<br>134 - 235<br>126.7 - 222 | 30.3 - 197.6<br>39.5 - 173.45<br>78.8 - 386.1 | 5.1 - 10.5<br>6.2 - 11.9<br>6.7 - 13.4 | Not Reported | | 11.54<br>12.01<br>11.50 | 0.50 - 0.58<br>0.32<br>0.67 | Not Reported | | 15. Zong 2014 | Robotic<br>520/997<br>Lap<br>2207<br>Open<br>5260 | Not Reported | Not Reported | Not Reported | 25 - 42.8<br>26 - 42.7<br>31.7 - 43.3 | 2.7<br>1.6<br>2.3 | Not Reported |
| 16. Guerra 2018 | Robotic 570<br>Lap 1456 | 206.4 - 439.0<br>167.1 - 361.0 | Not Reported | 6.75 - 10.5<br>6.7 - 13.0 | Not Reported | | | Not Reported | Not Reported | | 16. Guerra 2018 | Robotic<br>570<br>Lap<br>1456 | Not Reported | Not Reported | Not Reported | Not Available1,5<br>Not Available1,5 | Not Reported | Not Reported |
*Data reported for "total" gastrectorny proced sinilar results for "listal" gastrectory procedures; ** Intra-op and post-op reported; *** M-hopital overal complication rate. ¹Absolute rates/means not available; ²OR 1.07 [0.42, 2.75]; ³WMD 129.74 [-178.31, -81.16]; ³WMD -2.39 [-2.92] -1.87]
"The years 2010, 2011 and 202 overlap the two (2) databases used in the Glenn and Greenleaf publications. As such, the eported results.
{9}------------------------------------------------
TABLE 1B: da Vinci vs. Open and da Vinci vs. Laparoscopic Gastrectomy Procedures
*Data reported for "total" gastrectory procedures her results for "distal" gastrectomy procedures, ** V reported for cases with > 15 UV
¹Absolute rates/means not available; ׳WMD 0.10 [-0.43, 0.64]; ³WMD 0.52 [-0.76, 1.79); 'WMD -1.36 [-3.69, 0.97]; ³MD 2.92 [0.32, 5.53]
"The years 2010, 2011 and 2012 verləp between the Glenn and Greenleaf publications. As such, there is a possibility of data overlap in the eported results.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.