K170713 · Intuitive Surgical, Inc. · NAY · Jun 13, 2017 · Gastroenterology, Urology
Device Facts
Record ID
K170713
Device Name
da Vinci Xi Surgical System
Applicant
Intuitive Surgical, Inc.
Product Code
NAY · Gastroenterology, Urology
Decision Date
Jun 13, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K170713 · Jun 13, 2017
da Vinci Xi Surgical System
Intuitive Surgical, Inc.
Retrospective multi-center clinical study (medical records/chart review); Published clinical literature (Bittner, et al. 2015)
The retrospective study was used to support the safety and effectiveness of the da Vinci Surgical System for inguinal hernia repair by comparing outcomes (length of stay, complications, readmission, reoperation, mortality) against open surgery and historical laparoscopic data.
Patients undergoing inguinal hernia repair (652 robotic-assisted, 602 open); Sample Size: 1,254 total (652 robotic-assisted; 602 open); Number of Sites: 6
Open surgical procedures and historical laparoscopic data
Length of stay, intraoperative/postoperative complications, transfusions, readmission rates, reoperation rates, mortality, operative time
Indications for Use
The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Surgical System. Model 184000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endloscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparossopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatic use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
Device Story
da Vinci Xi (Model IS4000) is a software-controlled, electro-mechanical surgical system; consists of Surgeon Console, Patient Side Cart, and Vision Side Cart. Surgeon operates console to control endoscopic instruments (dissectors, scalpels, forceps, needle holders, retractors) and endoscopes. System translates surgeon hand movements into precise instrument motion at surgical site. Used in operating room by trained physicians for minimally invasive procedures. Provides visualization and instrument control for tissue manipulation, suturing, and ablation probe placement. Benefits include potential for minimally invasive access in complex procedures. This submission adds Inguinal Hernia Repair to existing general laparoscopic indications.
Clinical Evidence
Retrospective, multi-center, non-randomized study (N=652 robotic, N=602 open). Primary endpoints: length of stay, complications, readmission, reoperation, mortality. Results: Robotic cohort showed comparable intraoperative complications, transfusions, and readmission rates to open surgery. Robotic cohort had lower postoperative complications (discharge to 30 days) and lower reoperation rates (discharge to 30 days) compared to open. Operative time was longer for robotic-assisted procedures. No long-term outcomes (recurrence, chronic pain) assessed.
Technological Characteristics
Software-controlled, electro-mechanical system. Components: Surgeon Console, Patient Side Cart, Vision Side Cart, Endoscope, EndoWrist instruments. Connectivity: System-integrated. Sterilization: Compatible with standard endoscopic instrument sterilization protocols. No changes to technological characteristics from predicate.
Indications for Use
Indicated for adult and pediatric patients undergoing urologic, general laparoscopic, gynecologic, general thoracoscopic, and thoracoscopically-assisted cardiotomy procedures, including coronary anastomosis via adjunctive mediastinotomy. Used for tissue manipulation (grasping, cutting, dissection, suturing, ligation) and delivery of ablation probes.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{1}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an abstract human figure or a caduceus, composed of three interconnected profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 13, 2017
Intuitive Surgical, Inc. % Ms. Cindy Domecus Chief Regulatory Advisor, Intuitive Surgical Domecus Consulting Services LLC 1171 Barroilhet Drive Hillsborough, California 94010
Re: K170713
Trade/Device Name: da Vinci Xi Surgical System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: NAY Dated: March 7, 2017 Received: March 9, 2017
Dear Ms. Domecus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{2}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Image /page/1/Picture/9 description: The image shows the name "Jennifer R. Stevenson -S3" in a large, sans-serif font. The name is likely a signature or identification of a person. The text is clear and easy to read, with a simple and professional appearance.
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
#### 510(k) Number (if known)
K170713
#### Device Name
da Vinci Surgical System, Model IS4000, and EndoWrist Instruments and Accessories
#### Indications for Use (Describe)
The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Surgical System. Model 184000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endloscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparossopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatic use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary (21 CFR § 807.92(c))
#### I. SUBMITTER INFORMATION
| Submitter: | Intuitive Surgical, Inc. | | |
|--------------------------------|--------------------------------------------------|--|--|
| | 1266 Kifer Road | | |
| | Sunnyvale, CA 94086 | | |
| Contact: | Cindy Domecus, R.A.C. (US & EU) | | |
| | Principal, Domecus Consulting Services LLC | | |
| | Chief Regulatory Advisor to Intuitive Surgical | | |
| | Telephone: 650.343.4813 | | |
| | Fax: 650.343.7822 | | |
| | Email: domecusconsulting@comcast.net | | |
| Date Summary Prepared: | June 1, 2017 | | |
| II. SUBJECT DEVICE INFORMATION | | | |
| Device Trade Name: | da Vinci® Surgical System, Model IS4000 | | |
| Common Name: | System, Surgical, Computer Controlled Instrument | | |
| Classification Name: | Endoscope and Accessories (21 CFR §876.1500) | | |
| Regulatory Class: | II | | |
| Product Code: | NAY | | |
Traditional 510(k)
#### III. PREDICATE DEVICE INFORMATION:
Following are the two predicate devices for this 510(k):
Intuitive Surgical da Vinci Surgical System, Model IS4000 (K131861, K152578, K153276, K161178), and Intuitive Surgical da Vinci Surgical System, Model IS3000 (K081137, K123463, K090993)
#### IV. DEVICE DESCRIPTION:
Submission Type:
This 510(k) is for a labeling modification only, to include "Inguinal Hernia Repair" procedures under the cleared "general laparoscopic surgical procedures" indication for the da Vinci Xi Surgical System (K131861). There are no changes to the technological characteristics of the cleared da Vinci Xi Surgical System proposed in this submission. The da Vinci Xi Surgical System, Model IS4000, is a softwarecontrolled, electro-mechanical system designed for surgeons to perform minimally invasive surgery. The Model IS4000 Surgical System consists of a Surgeon Console, a Patient Side Cart (PSC), and a Vision Side Cart (VSC) and is used with an Endoscope, EndoWrist Instruments, and Accessories.
#### V. INDICATIONS FOR USE
The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Surgical System, Model IS4000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy
{5}------------------------------------------------
procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.
#### Precaution for Representative Uses
The demonstration of safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of cancer (overall survival, disease-free survival, local recurrence) or treatment of the patient's underlying disease/condition . Device usage in all surgical procedures should be guided by the clinical judgment of an adequately trained surgeon.
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
There are no changes to the technological characteristics of the cleared da Vinci Xi Surgical System (IS4000) proposed in this submission.
## VII. PERFORMANCE DATA
## Pre-Clinical Animal Study Data
This premarket notification is supported by animal study data including the results from six (6) evaluations in a total of 24 animals demonstrating use of da Vinci Xi Surgical System in the following procedures: Hysterectomy and Salpingectomy/Oophorectomy (Adnexectomy), Pyeloplasty, Nephrectomy, Nissen Fundoplication, Colectomy and Mitral Valve Repair. These data were submitted in support of clearance of the da Vinci Xi Surgical System (K131861).
## Clinical Study Data
A retrospective, multi-center, non-randomized controlled clinical study evaluating the use of the da Vinci Surgical System in Inguinal Hernia Repair procedures compared with open surgical procedures was conducted. This study included 652 and 602 subjects in the robotic-assisted and open cohorts, respectively, and was conducted at six (6) investigational sites in United States. Post operative data was collected throughout the thirty (30) day follow up period. Longer term endpoints such as inguinal hernia repair recurrence and chronic pain were not assessed in this study, and thus no claims regarding these clinical outcomes data should be inferred. The results demonstrated the safety and effectiveness of robotic-assisted inguinal hernia repair procedures. Additionally, such data provides information comparing the use of the da Vinci System in Inguinal Hernia Repair procedures with open surgery in the following key measures :
- Length of Stay: comparable outpatient and inpatient lengths of hospital stay were reported for both the robotic-assisted and open cohorts. A statistically lower inpatient length of stay was found when propensity matched analysis was conducted.
- Intraoperative Complications: comparable intraoperative complication rates were reported for both the robotic-assisted and open cohorts.
- Transfusions: comparable blood transfusion rates were reported in both cohorts.
- Postoperative Complications through Discharge: comparable postoperative through discharge complication rates were reported for both the robotic-assisted and open cohorts.
<sup>4</sup> Propensity matched results are the same as the reported outcomes unless stated otherwise.
{6}------------------------------------------------
- Postoperative Complications (Discharge through 30 days) -: a lower rate of postoperative complications from discharge through 30 day follow up was reported for the robotic-assisted cohort as compared to the open cohort.
- Readmission Rates: comparable readmission rates were reported for both the robotic-assisted and open cohorts.
- Reoperation Rates (Postoperative to Discharge): comparable reoperation rates were reported for the robotic-assisted cohort as compared to the open cohort.
- Reoperation Rates (Discharge through 30 days): lower reoperation rates were reported for the robotic-assisted cohort as compared to the open cohort. This statistically lower rate of reoperations was not observed in the propensity matched analysis. The results of the propensity matched analysis showed comparable reoperation rates.
- Mortality: comparable mortality rates were reported in both cohorts.
- Operative Time: longer operative times were reported for the robotic-assisted cohort as compared to open surgery .
The data from both cohorts of the retrospective study are compared to data on laparoscopic procedures from eleven (11) publications referenced in Bittner, et al., "Update of guidelines on laparoscopic (TAPP) and endoscopic (TEP) treatment of inguinal hernia (International Endohernia Society", Chapter 4: "TEP versus TAPP: which is better?" Surg Endosc 29:289-321 (2015).
² Based on the results of the multivariate analysis, age greater than 65 years, presence of comorbidities and open inguinal hernia repair are associated with a higher likelihood of a postoperative complication within 30 days of surgery. Furthermore, propensity matched analyses showed that age greater than 65 years and open inguinal hernia are associated with a higher likelihood of a postoperative complication within 30 days of surgery.
The longer operative time for the robotic-assisted cohort was not associated with increases in the complication, readmission, reoperation or mortality rates.
{7}------------------------------------------------
| Outcome | Robotic-Assisted<br>Cohort N=6521 | Open Cohort<br>N=6021 | Laparoscopic<br>Cohort N = 20 – N<br>= 3,4572 |
|-------------------------------------|-----------------------------------|------------------------|-----------------------------------------------|
| Patient Demographics | | | |
| Mean age, y (± SD or range) | 55.8 ± 15.6 | 57.2 ± 16.6 | 34.91 – 62.3 |
| Female Gender, n (%) | 64 (9.8) | 60 (10.0) | 0 – 6.1% |
| Male Gender, n (%) | 588 (90.2) | 542 (90.0) | 88 – 100% |
| Mean BMI, kg/ m2 (± SD or range) | 27.3 ± 5.1 | 26.7 ± 5.0 | 22.4 – 26.8 |
| Length of Stay | | | |
| Inpatient (days) | 3.01 ± 4.65 (n=52) | 4.866 ± 5.35<br>(n=51) | 0.8 – 5 days |
| Outpatient (hours) | 7.16 ± 3.01 | 6.89 ± 2.77 | |
| Intraoperative Complications, n (%) | 2 (0.3) | 0 | 0 – 8% |
| Transfusions, n (%) | | | |
| Intraoperative | 0 | 0 | Not Reported |
| Perioperative | 2 (0.3) | 2 (0.3) | |
| Postoperative Complications, n (%) | | | |
| Postop to discharge | 15 (2.3) | 10 (1.7) | 0 – 36%2a |
| Post discharge to 30 days | 28 (4.3) | 44 (7.3)7 | |
| Readmission Rates, n (%)* | 23 (3.5) | 26 (4.3)9 | Not Reported |
| Reoperation Rates | | | |
| Postop to discharge | 3 (0.5)* | 2 (0.3)* | 0 – 2.5%2a |
| Post discharge to 30 days | 0* | 8 (1.3)*8 | |
| Mortality, n | 0 | 1 | 0 |
| Operative Time, min (± SD or range) | 79.7 ± 31.7 | 45.0 ± 21.7 | 32.6 – 110 |
| Adverse Events, n3 | 44 | 4910 | Not Applicable |
| All Complications, n (%)4 | 45 (6.9) | 51 (8.5)11 | 7.9 – 8.7% |
| Conversion Rate, n (%) | 6 (0.9) | Not Applicable | 0 – 4% |
| TABLE 1: da Vinci, Open and Laparoscopic Inguinal Hernia Repair | | | | |
|-----------------------------------------------------------------|--|--|--|--|
|-----------------------------------------------------------------|--|--|--|--|
*Related to Inguinal Hernia Repair Procedure
් Data from a retrospective, multi-center, non-randomized clinical study evaluating the use of the da Vinci Surgical System in Inguinal Hernia Repair procedures compared with open surgical procedures. This table includes all subjects with complete data (missing data is omitted), despite propensity scores showing that one third of investigational subjects did not have an appropriate match among control subjects and large site to site differences.
් Data from eleven (11) publications referenced in Bittner, et al., "Update of guidelines on laparoscopic (TAPP) and endoscopic (TEP) treatment of inguinal hernia (International Endohernia Society", Chapter 4: "TEP versus TAPP: which is better?" Surq Endosc 29:289-321 (2015).
4 Post operative complications and reoperation rates were not segregated by specific timeframes. [Note: One publication (Bracale, et al (2012) reported only relative risk and mean differences for post operative complications and operative time / length of stay.]
് Adverse Events defined as death or serious injury (i.e., Clavien Grade Illa or higher). Publications cited in Bittner, et al. did not categorize complications/events by severity (i.e., serious).
All complications defined as all intraoperative complications for the robotic-assisted and open cohorts. In the laparoscopic cohort, one publication cited in Bittner, et al reported "overall morbidity".
3Inguinal hernia recurrence and chronic pain were not assessed in this study.
6 The results are heavily influenced by one site. See also other footnotes below. Without this site, the mean number of inpatient days for open procedures is 3.00.
728/44 complications came from one of the 6 sites.
8 Of the 10 reoperations, 80% (8/10)were done at one site.
ి14/26 readmissions were at one site
10 30/49 AEs occurred at one site
14 30/51 were reported at one site.
{8}------------------------------------------------
## VIII. CONCLUSION
Based on the information provided in this premarket notification, the inclusion "Inguinal Hernia Repair" procedures under the da Vinci Xi Surgical System "general laparoscopic surgical procedure" indication is substantially equivalent to the predicate devices. The basis of this substantial equivalence determination was an assessment of the immediate post-operative outcomes comparing the robotic-assisted surgery inguinal hernia repair versus laparoscopic surgery inguinal hernia repair.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.