Disposable Biopsy Needle

K243296 · Carbon (Shenzhen) Medical Device Co., Ltd. · KNW · Jul 11, 2025 · Gastroenterology, Urology

Device Facts

Record IDK243296
Device NameDisposable Biopsy Needle
ApplicantCarbon (Shenzhen) Medical Device Co., Ltd.
Product CodeKNW · Gastroenterology, Urology
Decision DateJul 11, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1075
Device ClassClass 2

Indications for Use

The Disposable Biopsy Needle is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate. It is not intended for use in bone.

Device Story

Disposable, sterile, spring-loaded, percutaneous soft tissue biopsy instrument; consists of cannula, inner stylet, mechanical power unit, and protective cover. Used by physicians under ultrasound guidance to obtain tissue samples. Device features side-notch design for automatic extraction of small tissue samples. Mechanical power unit includes safety lock to prevent activation when locked. Color-coded per ISO 6009 (18G = Pink). Used in clinical settings for pathological testing of soft tissue organs. Benefits include minimally invasive tissue sampling for diagnostic purposes.

Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including biocompatibility (ISO 10993 series), packaging verification (ASTM F1886, F1929, F88), sterilization validation (ISO 11135, ISO 11737), and performance testing (ISO 9626, ISO 7864, ISO 6009). Sampling verification performed in vitro using porcine liver/kidney and canine prostate tissues.

Technological Characteristics

Materials: Stainless steel (cannula/stylet), Polydimethylsiloxane (fluid), Polyoxymethylene copolymer (POM) and Acrylonitrile Butadiene Styrene (ABS) (mechanical unit), Polypropylene (protective cover). Principle: Spring-operated side-notch biopsy gun. Dimensions: 18G; 100mm, 150mm, 200mm lengths. Sterilization: Ethylene oxide (SAL 10^-6). Connectivity: None (mechanical).

Indications for Use

Indicated for individuals requiring biopsy for sampling of soft tissue abnormalities, specifically liver, kidney, and prostate. Not intended for use in bone.

Regulatory Classification

Identification

A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION July 11, 2025 Carbon (Shenzhen) Medical Device Co., Ltd. % Jishan Jin Regulatory Affairs Linnwell International Certification Consulting Co., Ltd. First Floor, Building 1, No. 333 Wanfang Road, Minhang District Shanghai, 201112 China Re: K243296 Trade/Device Name: Disposable Biopsy Needle Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology Biopsy Instrument Regulatory Class: Class II Product Code: KNW Dated: October 13, 2024 Received: October 18, 2024 Dear Jishan Jin: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food &amp; Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K243296 - Jishan Jin Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K243296 - Jishan Jin Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Jessica Carr -S Jessica Carr, PhD Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K243296 | | | Device Name Disposable Biopsy Needle | | | Indications for Use (Describe) The Disposable Biopsy Needle is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate. It is not intended for use in bone. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} 510(k) Summary [As required by 21 CFR 807.92(c)] 1. Submission Information 510(k) Number: K243296 Date: July 9, 2025 Type of 510(k) Submission: Traditional submission Applicant: Carbon (Shenzhen) Medical Device Co., Ltd. Room 203, Building B, 5# Skyworth Innovation Vally, No. 1 Tangtou Road, Tangtou Community, Shiyan Street, Bao'an District, 518102 Shenzhen, PEOPLE'S REPUBLIC OF CHINA Contact: Ye Min Title: Quality Manager Tel: +86 13622306079 Email: ye.min@carbonmed.cn 2. Device Description Trade/Device Name: Disposable Biopsy Needle Common Name: Disposable Biopsy Needle Device Classification Name: Gastroenterology-urology biopsy instrument Regulation Number: 21 CFR 876.1075 Product Code: KNW Device Class: Class II 3. Predicate Device Trade/Device Name: Bard® Max-Core® Disposable Core Biopsy Instrument 510(k) Number: K133948 Regulation number: 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulation Class: Class II Product Code: KNW Manufacture: Carbon (Shenzhen) Medical Device Co., Ltd. 4. Indication for use The Disposable Biopsy Needle is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate. It is not intended for use in bone. 5. Device Description The disposable biopsy needle is a sterile, spring loaded, disposable percutaneous soft tissue biopsy instrument. The disposable biopsy needle consists of a cannula, an inner stylet, a mechanical power device and a protective cover. It is used to obtain tissue samples for tissue biopsy, suitable for tissue biopsy of various organs such as the kidney, liver, prostate etc. The needle need to be inserted by a qualified physician under ultrasound guidance. The button of the mechanical power unit comply with the color coding requirements of ISO 6009, i.e. 18 gauge = Pink. The disposable biopsy needle includes three models: CMBNA/1810, CMBNA/1815, CMBNA/1820. {5} 6. Comparison of Technological Characteristics With Predicate Device | Parameters | Subject Device | Predicate Device | Comparison | | --- | --- | --- | --- | | 510(k) Number | K243296 | K133948 | -- | | Trade/Device Name | Disposable Biopsy Needle | Bard® Max-Core® Disposable Core Biopsy Instrument | -- | | Model | CMBNA/1810, CMBNA/1815, CMBNA/1820 | MC1810, MC1816, MC1820 (Other models are not included in this substantial equivalence discussion) | -- | | Owner | Carbon (Shenzhen) Medical Device Co., Ltd. | C. R. Bard, Inc. | -- | | Regulation number | 21 CFR 876.1075 | 21 CFR 876.1075 | Same | | Product Code | KNW | KNW | Same | | Classification | Class II | Class II | Same | | Indications for use | The Disposable Biopsy Needle is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate. It is not intended for use in bone. | The core needle biopsy device is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lymph nodes and various soft tissue tumors. It is not intended for use in bone. | Subset Note 1 | | Soft Tissue Organ Examples | Soft tissue organs such as liver, kidney, prostate. | Soft tissue organs such as liver, kidney, prostate, spleen, lymph nodes, abdomen, thyroid, testes, bladder, lung, breast. | | | Type of use | Prescription use | Prescription use | Same | | Patient population | Individuals requiring biopsy for sampling of soft tissue abnormalities | Individuals requiring biopsy for sampling of soft tissue abnormalities | Same | | Gauge x Length | 18G×100mm, 18 G×150mm, 18 G×200mm | 18G×10 cm, 18G×16 cm, 18G×20 cm | Similar Note 2 | | Penetration depth | 21mm | 22 mm | Similar Note 2 | | Single use | Yes | Yes | Same | | Sterile supply | Yes | Yes | Same | | Sterilization method | Ethylene oxide | Ethylene oxide | Same | | SAL | 10⁻⁶ | 10⁻⁶ | Same | | Mechanics of Action | Spring operated | Spring operated | Same | | Sample notch | 18mm | 18mm | Same | {6} | size | | | | | --- | --- | --- | --- | | Mode of action | Single puncture and samples | Single puncture and samples | Same | | Number of Samples | One or more | One or more | Same | | Method of placement | Percutaneous puncture | Percutaneous puncture | Same | | Visualization applications | Ultrasound | X-ray, ultrasound, CT, etc. | Subset Note 3 | | Safety features | Locking function: The needle cannot be activated after the slide is locked | Undisclosed | Same Note 5 | | Principle of operation | It is a side-notch biopsy gun. A hand-operated, non-electronic, surgical instrument designed for the automatic extraction of a small tissue sample from an anatomical structure, while causing minimal surrounding tissue damage, for tissue pathological testing. It cannot be use for bone tissue biopsy. | It is a side-notch biopsy gun. A hand-operated, non-electronic, surgical instrument designed for the automatic extraction of a small tissue sample from an anatomical structure, while causing minimal surrounding tissue damage, for tissue pathological testing. It cannot be use for bone tissue biopsy. | Same | | Package | 5 blister packs with Tyvek lids in a cardboard shelf box with the IFU | 5 blister packs with Tyvek lids in a cardboard shelf box with the IFU | Same | | Material (Contact with human body) | Outer cannula: Stainless steel | Cannula: Stainless steel | Same Note 4 | | | Inner stylet: Stainless steel | Undisclosed | | | | Medical Fluid: Polydimethylsiloxane | Undisclosed | | | | Left and right half tubes: Polyoxymethylene copolymer (POM) | | | | | Mechanical power unit: Acrylonitrile Butadiene Styrene (ABS)+ Polyoxymethylene copolymer (POM) | Undisclosed | | | | Protective cover: Polypropylene (PP) | Undisclosed | | | Nature of body contact | Outer cannula, Inner stylet, Medical Fluid, Left and right half tubes: Tissue/bone/dentin, Limited | Cannula & Stylet: Tissue/bone/dentin, Limited | Same | | | Mechanical power unit, protective cove: Intact skin, Limited | Handle: Intact skin, Limited | | | EO residue | Meet ISO 10993-7 | Meet ISO 10993-7 | Same | | ECH residue | Meet ISO 10993-7 | Meet ISO 10993-7 | | | Shelf life | 2 years | 3years | Similar Note 4 | {7} | Bacterial endotoxin | <20EU/set | <20EU/set | Same | | --- | --- | --- | --- | ## Differences between New device and Predicate Devices: ### Note 1: The indications of the subject device are fewer than those of Bard® Max-Core® because the current declared indications for the predicate device cover all specifications of 14G-20G biopsy needles, while the subject device only has one specification of 18G. Therefore, the indicated soft tissues are only liver, kidney and prostate. ### Note 2: The subject device is available in lengths of 100mm, 150mm and 200mm, while the predicate device is available in 100mm, 160mm and 200mm. There is no difference between the two in the maximum and minimum length, only a slight difference in the middle specifications. This difference does not affect clinical use, as the physician will choose the appropriate length of the biopsy needle according to the patient's situation. The width and length of the biopsy needle can be found on the label and labeling to ensure that the physician is aware of it. There are also subtle differences in the penetration depth between the subject device and the predicate device. The penetration depth can also be found in the user manual, and it is up to the doctor to decide whether the advance distance is appropriate, so this difference does not raise any safety or effectiveness issues. ### Note 3: The subject device is only visible under ultrasound and has fewer visualization technology than the predicate device. We have clarified the use conditions of the subject device in the user manual and provided a B-ultrasound image display test report to prove its visibility, so this difference does not have any safety or effectiveness issues. ### Note 4: The shelf life and materials of the subject device may differ from those of the predicate device. We have demonstrated the biosafety, stability, and packaging integrity of the subject device through non-clinical bench testing. Therefore, these differences do not raise any safety or effectiveness issues. ## 7. Non-clinical Testing All nonclinical testing performed on new devices is to demonstrate the substantial equivalence to the predicate devices. Tests setup and execution are performed in accordance with applicable standards. Results of the testing demonstrate the compliance to the standards and matching the performance of new devices to the predicate devices. The following performance data were provided in support of the substantial equivalence determination. ### Biocompatibility test - ISO 10993-5 Third edition 2009-06-01 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity; - ISO 10993-10 Fourth edition 2021-11 Biological evaluation of medical devices - Part 10: Tests for skin sensitization; - ISO 10993-23 First edition 2021-01 Biological evaluation of medical devices - Part 23: Tests for irritation; - ISO 10993-11 Third edition 2017-09 Biological evaluation of medical devices - Part 11: Tests for {8} systemic toxicity; - ISO 10993-4 Third edition 2017-04 Biological evaluation of medical devices--Part 4: Selection of tests for interactions with blood; - ASTM F756-17 Standard Practice for Assessment of Hemolytic Properties of Materials; - USP-NF M98900_01_01 &lt;151&gt; Pyrogen Test (USP Rabbit Test); ## Packaging verification - ASTM F1886/F1886M-16 Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection; - ASTM F1929-23 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration; - ASTM F88/F88M-2021 Standard Test Method for Seal Strength of Flexible Barrier Materials; - DIN 58953-6:2023 Sterilization- Sterile supply -Part 6: Microbial barrier testing of packaging materials for medical devices which are to be sterilized; - USP-NF M98810_01_01 &lt;71&gt; Sterility Tests; ## Sterilization and shelf life - ISO 11135 Second edition 2014-07-15 Sterilization of health-carISOe products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices [Including: Amendment 1 (2018)]; - ISO 10993-7 Second edition 2008-10-15 Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals [Including: Technical Corrigendum 1 (2009), AMENDMENT 1: Applicability of allowable limits for neonates and infants (2019)]; - USP-NF M98830_02_01 &lt;85&gt; Bacterial Endotoxins Test; - ANSI AAMI ST72:2019 Bacterial endotoxins - Test methods, routine monitoring, and alternatives to batch testing; - ISO 11737-2 Third edition 2019-12 Sterilization of medical devices - Microbiological methods - Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process; - ASTM F1980-21 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices - ASTM D4169-23 Standard Practice for Performance Testing of Shipping Containers and Systems; ## Performance Test - ISO 9626 Second edition 2016-08-01 Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods; - ISO 7864 Fourth edition 2016-08-01 Sterile hypodermic needles for single use - Requirements and test methods; - ISO 6009 Fourth edition 2016-08-01 Hypodermic needles for single use - Colour coding for identification; ## Sampling Verification Testing In order to verify the sampling ability of the disposable biopsy needle, the biopsy needle was inserted into three different in vitro tissues (porcine liver, porcine kidney and canine prostate), and the mass and quality of the tissues were examined. The samples collected using the subject device were comparable to those collected using the predicate device. The mechanical power unit of the biopsy needle was tested on in vitro tissues and pork, and 24 repeated sampling steps were completed as instructed in the instructions to evaluate whether the device could fire {9} correctly and whether the safety lock function was normal. All operations are smooth and meet the requirements of the acceptance criteria, which indicates that the disposable biopsy needle can work as intended. ## 8. Animal Study Substantial equivalence does not depend on animal test data. ## 9. Clinical Testing Substantial equivalence does not depend on clinical test data. ## 10. Substantial Equivalent (SE) Conclusion Based on the device comparison information and non-clinical testing, the subject device is Substantially Equivalent (SE) to the predicate device which is US legally market device. Therefore, the subject device is determined as safe and effective. 6
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...