The Bard® Mission® Disposable Core Biopsy Instrument is intended for use in obtaining biopsy samples from soft tissues such as lung, liver, spleen, kidney, prostate, lymph nodes, breast, thyroid, and various soft tissue tumors.
Device Story
Single-use, semi-automatic core biopsy instrument; used by clinicians to obtain soft tissue samples. Device features color-coded plungers for needle gauge identification (14G-20G), centimeter markings, and echogenic enhancement for ultrasound guidance. Operation involves manual insertion and spring-loaded firing mechanism to capture tissue in a specimen notch. Used in clinical settings for diagnostic biopsy procedures. Output is a tissue core sample for pathological analysis. Benefits include standardized, minimally invasive tissue acquisition for disease diagnosis.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included penetration depth, tensile strength, sterile barrier integrity, and performance after shipping/storage. Biocompatibility testing performed per ISO 10993-1:2009, including cytotoxicity, sensitization, intracutaneous reactivity, and material-mediated pyrogenicity.
Technological Characteristics
Semi-automatic spring-loaded core biopsy instrument. Materials evaluated per ISO 10993-1:2009. Features include centimeter markings, echogenic enhancement, and color-coded plungers. Sterile, single-use. No electronic components or software.
Indications for Use
Indicated for obtaining biopsy samples from soft tissues including lung, liver, spleen, kidney, prostate, lymph nodes, breast, thyroid, and soft tissue tumors in patients requiring tissue sampling.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
{1}------------------------------------------------
Image /page/0/Picture/2 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of three faces in profile, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 14, 2017
Bard Peripheral Vascular, Inc. % Mark Job Regulatory Technology Services. LLC 1394 25th Street, NW Buffalo, MN 55313
Re: K171953
Trade/Device Name: Bard® Mission® Disposable Core Biopsy Instrument Regulation Number: 21 CFR§ 876.1075 Regulation Name: Gastroenterology-Urology-Biopsy Instrument Regulatory Class: II Product Code: KNW Dated: September 5, 2017 Received: September 6, 2017
Dear Mark Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{2}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Joyce M. Whang -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K171953
Device Name
Bard® Mission® Disposable Core Biopsy Instrument
Indications for Use (Describe)
The Bard® Mission® Disposable Core Biopsy Instrument is intended for use in obtaining biopsy samples from soft tissues such as lung, liver, spleen, kidney, prostate, lymph nodes, breast, thyroid, and various soft tissue tumors.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### BARD® MISSION® Disposable Core Biopsy Instrument 510(k) Summary 21 CFR 807.92
As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (I)(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is based is as follows:
## Submitter Information:
| Applicant: | Bard Peripheral Vascular, Inc |
|------------|-------------------------------|
| | 1625 West 3rd Street |
| | Tempe, Arizona 85281 |
| Phone: | 480-350-6067 |
|----------|-----------------------------------------------|
| Fax: | 480-449-2570 |
| Contact: | Susan Sheffield, Regulatory Affairs Associate |
| Date: | 15 May 2017 |
### Subject Device Name:
| Device Trade Name: | Bard® Mission® Disposable Core Biopsy Instrument |
|-----------------------|--------------------------------------------------|
| Common or Usual Name: | Core Biopsy Instrument |
| Classification: | Instrument, Biopsy (Product Code KNW); Class II |
| Review Panel: | Gastroenterology / Urology |
| Regulation Number: | 21 CFR 876.1075 |
#### Predicate Devices:
- M-Biopsy™ Semi-Automatic Biopsy Instrument by Mermaid ● Medical (K161409; cleared 2 August 2016)
- . BARD® MONOPTY® and BARD® MAX-CORE® Disposable Core Biopsy Instruments (K133948/S001; cleared 21 February 2014)
#### Reference Devices:
- POWERGLIDE® Midline Catheter (K121073; cleared 1 June 2012) .
- ULTRACLIP® DUAL TRIGGER Breast Tissue Marker (K042341; 20 September . 2004)
- ENCOR® Breast Biopsy System (K051158; 16 May 2005)
- . SIDEKICK® and USHER® Support Catheters (K131493; cleared 2 August 2013)
The purpose of the secondary predicate device BARD® MAX-CORE®/MONOPTY® is to highlight similarities to the branding for BARD® disposable core biopsy devices. The reference
Image /page/3/Picture/17 description: The image shows the word "BARD" in a stylized, sans-serif font. The letters are a solid green color. The "A" in "BARD" is unique, as it is formed by two triangles that meet at the top, creating a sharp, angular shape. The letters are bold and evenly spaced, giving the word a strong and balanced appearance.
{5}------------------------------------------------
devices are included to provide sufficient evidence for the safe usage of the materials found in BARD® MISSION®.
## Device Description:
The subject device BARD® MISSION® Disposable Core Biopsy Instrument is a single use core biopsy device. It is available in several needle gauge sizes and lengths. The plunger is color coded according to the various gauge sizes, e.g. yellow = 20 gauge, pink = 18 gauge, purple = 16 gauge, and green = 14 gauge.
## Indications for Use of Device:
The BARD® MISSION® Disposable Core Biopsy Instrument is intended for use in obtaining biopsy samples from soft tissues such as from the lung, liver, spleen, kidney, prostate, lymph nodes, breast, thyroid, and various soft tissue tumors.
## Technological Comparison to Predicate Device:
The subject device BARD® MISSION® has the following similarities to the primary predicate device M-Biopsy™:
- Same class and regulation number
- Same Intended Use ●
- Similar Indications for Use ●
- 0 Same Target Population/Conditions of Use (anatomical location of use, user interface, how device interacts with other devices, interaction with patient, etc.)
- Same Fundamental Scientific Technology, including Design, Mechanism/Mode of Action, and Energy Used/Delivered)
- . Same catalogue offerings, including Gauge Size, Needle Length, and Penetration Depth
- . Similar design features, such as Centimeter Markings, Grip Design, Color Coding, and Echogenicity
- . Same imaging capabilities
The subject device BARD® MISSION® Disposable Core Biopsy Instrument incorporates the following differences:
- Slight alterations to the language used in the Indications for Use in order to maintain . consistency in branding for the BARD® Disposable Core Biopsy products
- Different dimensions of the Specimen Notch used to collect the biopsy sample o
- Differences in design, such as the Finger Grips and Fire Ready Indicator .
- Unknown differences, such as: ●
- Color Coding o
- Echogenic Enhancement o
- Sterility Assurance o
Image /page/4/Picture/24 description: The image shows the word "BARD" in a stylized, sans-serif font. The letters are a solid green color. The letters are bold and have a uniform thickness. The overall impression is clean and modern.
{6}------------------------------------------------
### Performance Data:
To demonstrate that the subject device BARD® MISSION® is as safe and as effective as the predicate device M-Biopsy™, its technological characteristics and performance criteria were evaluated. Using FDA Guidance Documents on non-clinical testing of medical devices and internal Risk Assessment procedures, the following in vitro tests were performed on the subject device:
- Number of Samples
- · Penetration Depths
- Stylet / Cannula to Handle Tensile Strength
- Corresponding Working Needle Length and Cutting Cannula OD, and Stylet/Cannula Working Needle Lengths
- Integrity of the Sterile Barrier
- · Performance After Ship Testing
- Needle Protection After Shipping and Storage
The results from bench testing demonstrate that the subject device BARD® MISSION® performed as expected according to the technological characteristics and performance requirements of C.R. Bard; and thus, can be concluded as comparable to other semiautomatic core biopsy devices on the market, demonstrating the subject device BARD® MISSION® to be as safe, as effective, and performs as well as the legally marketed device M-Biopsy™.
## Biocompatibility:
Per ISO 10993-1:2009, "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process." The following biocompatibility tests were successfully performed and/or adopted in accordance with GLP regulations:
- Cytotoxicity
- Sensitization
- Intracutaneous Reactivity ●
- o Material Mediated Pyrogenicity
The results of biocompatibility testing demonstrate that the subject device BARD® MISSION® is considered biocompatible for its intended use.
## Conclusions:
The subject device, BARD® MISSION® Disposable Core Biopsy Instrument, met all predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs. Likewise, the subject device BARD® MISSION® and the predicate device M-Biopsy™ share the same or similar characteristics: intended use, indications for use, target population / conditions for use, and fundamental scientific technology. Therefore, Bard concludes that the subject device BARD® MISSION® Disposable Core Biopsy Instrument is substantially equivalent to the legally marketed predicate device M-Biopsy™ Semi-Automatic Biopsy Instrument.
Image /page/5/Picture/20 description: The image shows the word "BARD" in a stylized, sans-serif font. The color of the text is a dark green. The letters are bold and have a slightly geometric appearance. The background is white.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.