K131493 · C.R. Bard, Inc. · DQY · Aug 2, 2013 · Cardiovascular
Device Facts
Record ID
K131493
Device Name
SIDEKICK AND USHER SUPPORT CATHETERS
Applicant
C.R. Bard, Inc.
Product Code
DQY · Cardiovascular
Decision Date
Aug 2, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The Sidekick® and Usher® Support Catheters are single lumen catheters intended to create a pathway for other devices in the peripheral vasculature.
Device Story
Single lumen support catheters; used to create pathway for other devices in peripheral vasculature; supports Crosser® CTO Recanalization Catheters. Features standard luer fitting hub; optional attachable hemostatic valve; guidewire introducer included. Available in straight, angled, tapered, and non-tapered configurations; radiopaque markers at distal tip for fluoroscopic visualization. Used by physicians in clinical settings. Provides stable access for interventional devices; facilitates navigation through peripheral vasculature.
Clinical Evidence
Bench testing only. No clinical data. Testing included ablation efficiency, catheter burst, leak, inner/outer diameter, effective length, hub testing, kink resistance, tracking force, tip visualization, guidewire loading, particulate evaluation, coating integrity, and tensile strength. Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, pyrogen, hemolysis, complement activation, in-vivo thrombogenicity).
Technological Characteristics
Single lumen catheter; radiopaque markers; luer fitting hub; hemostatic valve; guidewire introducer. Materials evaluated per ISO 10993-1. Shelf life validated per ASTM F1980-07. Sterilization method consistent with predicates.
Indications for Use
Indicated for creating a pathway for other devices in the peripheral vasculature. Contraindicated for use with cutting/scoring balloons, pediatrics, neonatal, and neurovascular patients.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
# Sidekick® and Usher® Support Catheters
#### 510(k) Summary 21 CFR 807.92
As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (I)(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is based is as follows:
#### Submitter Information:
| Applicant: | Bard Peripheral Vascular, Inc<br>1625 West 3rd Street<br>Tempe, Arizona 85281 | |
|------------|-------------------------------------------------------------------------------|-------------|
| Phone: | 480-638-2939 | AUG 02 2013 |
| Fax: | 480-449-2546 | |
| Contact: | Timothy Wade, Regulatory Affairs Associate | |
| Date | March 19, 2013 | |
# Subject Device Name:
| Device Trade Name: | Sidekick® and Usher® Support Catheters |
|-----------------------|--------------------------------------------------------------|
| Common or Usual Name: | Percutaneous Catheter (21 CFR 870.1250,<br>Product Code DQY) |
| Classification: | Class II |
| Classification Panel: | Cardiovascular |
# Predicate Devices:
- MicroSheath XL Support Catheter (K073289; cleared January 11, 2008) .
- MicroSheath LP Support Catheter (K080849; cleared July 25, 2008) .
Image /page/0/Picture/13 description: The image shows the word "BARD" in a stylized, bold font. The letters are outlined in black, giving them a distinct and prominent appearance. The overall design is simple yet impactful, with the unique font choice adding a touch of character to the word.
{1}------------------------------------------------
#### Device Description:
The Sidekick® and Usher® Support Catheters are single lumen catheters with a standard luer fitting hub and separate attachable hemostatic valve. The catheters are recommended to support the Crosser® CTO Recanalization Catheters 14S/14P and S6.
The Sidekick® Support Catheter is available in straight, angled, tapered and non-tapered configurations in 70cm and 110cm effective lengths. The Usher® Catheter is tapered and is available in straight and angled configurations in 83cm effective lengths. The Sidekick® and straight Usher® Catheters have a single radiopaque marker 1mm from the distal tip. The angled Usher® Support Catheter configurations have three radiopaque markers at the distal tip for enhanced visualization of the catheter tip and angle under fluoroscopy. The most proximal radiopaque marker is located 15mm from the distal tip.
The product hub identifies SD for Sidekick® Catheter, USH for Usher® Catheter, A for Angled and T for Tapered; in addition to the sheath profile and effective length in centimeters. A guidewire introducer is provided to facilitate the guidewire passage through the optional hemostatic valve. The guidewire introducer shaft color matches the shaft color of the recommended support catheter.
#### Indications for Use of Device:
The Sidekick® and Usher® Support Catheters are single lumen catheters intended to create a pathway for other devices in the peripheral vasculature.
#### Contraindications:
The Sidekick® and Usher® Catheters are contraindicated for use with cutting/scoring balloons, pediatrics, neonatal and neurovascular patients.
# Technological Comparison to Predicate Devices:
The Sidekick® and Usher® Support Catheters have the following similarities to the predicate devices:
- . Same intended use (MicroSheath® XL Predicate)
- . Same indications for use (MicroSheath® XL Predicate)
- . Same target population (MicroSheath® XL Predicate)
- Same fundamental scientific technology (both predicates) .
- . Same operating principle (both predicates)
- . Same sterility assurance level and method of sterilization (both predicates)
Image /page/1/Picture/19 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are outlined in black, giving them a distinct and prominent appearance. The overall design is simple and clean, with a focus on readability.
{2}------------------------------------------------
The Sidekick® and Usher® Support Catheters differ from the predicate devices in dimensional characteristics, tapered catheter configurations, modifications in the shaft design, and new materials including the new hemostasis valve and guidewire introducer components. Additionally, as compared to the MicroSheath® LP Support Catheter predicate, the subject Usher® Support Catheter is not indicated for use in the coronary vasculature. The use of the Usher® Support Catheter in the peripheral vasculature is fully within the scope of both predicate devices and therefore does not affect the safety or effectiveness when used as indicated.
#### Performance Data:
To demonstrate substantial equivalence of the subject devices, the Sidekick® and Usher® Support Catheters, to the predicate devices, the technological characteristics and performance criterion were evaluated. Using FDA Guidance Documents on non-clinical testing of medical devices and internal Risk Assessment procedures, the following in vitro tests were performed:
- . Ablation Efficiency
- . Catheter Burst
- . Catheter Leak
- . Catheter Inner Diameter
- . Catheter Outside Diameter
- Catheter Effective Length .
- Catheter Hub Testing .
- Hub Leak Testing .
- Hub Luer Taper .
- . Kink Resistance
- Track Without Damage .
- . Tensile Strength
- Delivery of Recommended Devices .
- . Tracking Force
- Catheter Tip Visualization .
- . Ease of Guidewire Loading
- . Particulate Evaluation
- . Coating Integrity
- . Pouch Visual Inspection
- . Dye Penetration
Bard Peripheral Vascular, Inc.
Image /page/2/Picture/25 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are outlined in black, giving them a distinct and modern appearance. The overall design is simple yet impactful, making the word easily readable.
{3}------------------------------------------------
Pouch Tensile .
Product will have a viable shelf life based upon successful completion of testing performed in accordance with ASTM F1980-07, "Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices", FDA Guidance, "Shelf Life of Medical Devices", issued April 1991, and the Bard internal stability program. Additionally, biocompatibility tests were performed in accordance with ISO 10993-1:2009, "Biological Evaluation of Medical Devices Part 1: Evaluation and testing within a risk management process." Testing conducted includes Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, Pyrogen Testing, Hemolysis, Complement Activation, and in-vivo Thrombogenicity.
The results from these tests demonstrate that the technological characteristics and performance criteria of the Sidekick® and Usher® Support Catheters are comparable to the predicate devices and performs at least as safely and effectively as the legally marketed device.
#### Conclusions:
The Sidekick® and Usher® Support Catheters, met all the predetermined acceptance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs. The Sidekick® and Usher® Support Catheters are substantially equivalent to the legally marketed predicate devices, the MicroSheath® XL and MicroSheath® LP Support Catheters.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a human figure embracing a globe or sphere.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 2, 2013
Bard Peripheral Vascular, Inc. c/o Mr. Mark Job Regulatory Technology Services, LLC 1394 25th Street NW Buffalo, MN 55313
Re: K131493
Trade/Device Name: Sidekick® and Usher® Support Catheters Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: II Product Code: DOY Dated: July 18, 2013 Received: July 19, 2013
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 - Mr. Job
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kenneth J. Cavanaugh -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: Sidekick® and Usher® Support Catheter
Indications for Use: The Sidekick® and Usher® Support Catheters are single lumen catheters intended to create a pathway for other devices in the peripheral vasculature.
Prescription Use X (Part21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Image /page/6/Picture/12 description: The image shows the name "Kenneth J. Cavanaugh-S" in a bold, sans-serif font. The text is arranged in two lines, with "Kenneth J." on the first line and "Cavanaugh-S" on the second line. The letters are black, and the background is white. The image appears to be a close-up of the name, possibly extracted from a document or sign.
Bard Peripheral Vascular, Inc.
Image /page/6/Picture/14 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are outlined in black, and the interior of the letters is white. The letters are evenly spaced and appear to be of the same size. The image is simple and clear, with a focus on the word itself.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.