GREATBATCH SCIENTIFIC MR COMPATIBLE LAPAROSCOPES AND OPERATING LAPAROSCOPES

K972280 · Wilson Greatbatch Technologies, Inc. · GCJ · Dec 11, 1997 · Gastroenterology, Urology

Device Facts

Record IDK972280
Device NameGREATBATCH SCIENTIFIC MR COMPATIBLE LAPAROSCOPES AND OPERATING LAPAROSCOPES
ApplicantWilson Greatbatch Technologies, Inc.
Product CodeGCJ · Gastroenterology, Urology
Decision DateDec 11, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

For use to permit direct viewing of the organs within the peritoneal cavity for the purpose of performing diagnostic and surgical procedures in a MR or an interventional MR environment, not to exceed a 1.5 Tesla static magnetic field.

Device Story

Laparoscope and operating laparoscope designed for use in MR or interventional MR environments; static magnetic field limit 1.5 Tesla. Device provides direct visualization of peritoneal cavity organs during diagnostic and surgical procedures. Operated by physicians in clinical/surgical settings. MR compatibility verified via testing for magnetic attraction, torque, and RF heating. Benefits include enabling surgical visualization during MR-guided interventions.

Clinical Evidence

Bench testing only. MR safety testing performed using a 1.5 Tesla GE Signa 64 MHz system. Results: very slight string deflection, 5 degrees torque, and 0.5°C maximum temperature rise during 60 minutes of RF exposure at 2.1 W/kg SAR. Artifact distortion levels were assessed.

Technological Characteristics

Rigid laparoscope available in 8.0mm and 10.0mm diameters with 0°, 30°, or 45° angles. Autoclavable. Designed for MR compatibility in static fields up to 1.5 Tesla.

Indications for Use

Indicated for patients requiring diagnostic or surgical procedures within the peritoneal cavity where direct visualization is necessary in an MR or interventional MR environment (up to 1.5 Tesla).

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ . # DEC I I laar ## - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -510(k) SUMMARY GREATBATCH SCIENTIFIC MR COMPATIBLE LAPAROSCOPE AND OPERATING LAPAROSCOPE K972280 | Submitter Name: | Greatbatch Scientific<br>a division of Wilson Greatbatch Ltd. | |---------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter Address: | 4100 Barton Road<br>Clarence, New York 14031 | | Contact Person: | Gary J. Sfeir, RAC | | Phone Number: | 716.759.5277 | | Facsimile Number: | 716.759.5280 | | Date Prepared: | 8 December, 1997 | | Device Trade Name: | Greatbatch Scientific MR Compatible Laparoscope and Operating<br>Laparoscope | | Device Common Name: | Laparoscope | | Classification Name: | Laparoscope | | Predicate Devices: | Optus Laparoscope and Operating Laparoscope | | Device Description: | The Greatbatch Scientific MR Compatible Laparoscope and Operating<br>Laparoscopes are available in standard and autoclavable 8.0mm,<br>10.0mm sizes, with 0°, 30° or 45° angles. | | Intended Use: | For use to permit direct viewing of the organs within the peritoneal<br>cavity for the purpose of performing diagnostic and surgical procedures<br>in a MR or an interventional MR environment, not to exceed a 1.5<br>Tesla static magnetic field. | | Device Technological<br>Characteristics and<br>Comparison to<br>Predicate Devices(s): | The device technological characteristics are similar in design to the<br>predicate device. | | Performance Data: | The device was tested for MR Compatibility and was found to be<br>acceptable for use in a 1.5 Tesla static magnet field. See attached MR<br>Safety Testing summary. | | Conclusion: | The Greatbatch Scientific MR Compatible Laparoscope and Operating<br>Laparoscope as designed, can be used in a MR or an interventional MR<br>environment, not to exceed a 1.5 Tesla static magnetic field. | {2}------------------------------------------------ ## GREATBATCH SCIENTIFIC MR COMPATIBLE LAPAROSCOPE MR SAFETY TESTING SUMMARY A Greatbatch Scientific MR Compatible 8.0mm endoscope was tested in conjunction with a 1.5 Tesla GE Signa Magnet for the presence of magnetic attractive and torque, artifact, and RF Heating. Based on intended use of the device and the results of extensive MR safety testing listed below, this device is MR Compatible. #### Magnetic Attraction and Torque 1. | Static field strength- | 1.5 Tesla / GE Signa 64 MHz MR System | |------------------------|---------------------------------------| | Type of test- | String deflection | | Observed deflection - | Very slight | | Observed torque - | 5 degrees | ### 2. Artifact 1.5 Tesla / GE Signa 64 MHz MR System Static field strength -Sequences l Amount of distortion: | Sequence | axial plane | sagittal plane | |----------|-------------|----------------| | FSPGR | ++ | ++ | artifact same as the device 44 #### 3. Heating RF Phantom: A . | Type of phantom - | A 50 lb., fluid filled rectangular shaped<br>plastic Plexiglas phantom | |----------------------------------------|------------------------------------------------------------------------| | Type of RF Coil - | Body coil | | SAR applied - | 2.1 W/kg | | Length of time - | 60 minutes | | Maximum temperature<br>rise observed - | 0.5°C | {3}------------------------------------------------ Image /page/2/Picture/2 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with three heads, representing the department's focus on health, human services, and science. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The seal is black and white. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC 1 1997 Mr. Gary J. Sfeir, RAC Director, Regulatory Affairs Greatbatch Scientific 4100 Barton Road Clarence. New York 14031 Re: K972280 > Trade Name: Greatbatch Scientific MR Compatible Laparoscopes and Operating Laparoscopes Regulatory Class: II Product Code: GCJ Dated: September 10, 1997 Received: September 12, 1997 Dear Mr. Sfeir: We have reviewed your Section 510(k) notification of intent to market the devices referenced above and we have determined the devices are substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the devices, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your devices are classified (see above) into either class II (Special Controls) or class III (Premarket Approval), they may be subject to such additional controls. Existing major regulations affecting your devices can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements , as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. In addition, FDA may publish further announcements concerning your devices in the Federal Register. Please note: this response to your premarket notification submissions does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4}------------------------------------------------ ### Page 2 - Mr. Gary J. Sfeir, RAC This letter will allow you to begin marketing your devices as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your devices to a legally marketed predicate device results in a classification for your devices and thus, permits your devices to proceed to the market. If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your devices, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Styph Rhodes Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ . # DEC I I laar ## - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -510(k) SUMMARY GREATBATCH SCIENTIFIC MR COMPATIBLE LAPAROSCOPE AND OPERATING LAPAROSCOPE K972280 | Submitter Name: | Greatbatch Scientific<br>a division of Wilson Greatbatch Ltd. | |---------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter Address: | 4100 Barton Road<br>Clarence, New York 14031 | | Contact Person: | Gary J. Sfeir, RAC | | Phone Number: | 716.759.5277 | | Facsimile Number: | 716.759.5280 | | Date Prepared: | 8 December, 1997 | | Device Trade Name: | Greatbatch Scientific MR Compatible Laparoscope and Operating<br>Laparoscope | | Device Common Name: | Laparoscope | | Classification Name: | Laparoscope | | Predicate Devices: | Optus Laparoscope and Operating Laparoscope | | Device Description: | The Greatbatch Scientific MR Compatible Laparoscope and Operating<br>Laparoscopes are available in standard and autoclavable 8.0mm,<br>10.0mm sizes, with 0°, 30° or 45° angles. | | Intended Use: | For use to permit direct viewing of the organs within the peritoneal<br>cavity for the purpose of performing diagnostic and surgical procedures<br>in a MR or an interventional MR environment, not to exceed a 1.5<br>Tesla static magnetic field. | | Device Technological<br>Characteristics and<br>Comparison to<br>Predicate Devices(s): | The device technological characteristics are similar in design to the<br>predicate device. | | Performance Data: | The device was tested for MR Compatibility and was found to be<br>acceptable for use in a 1.5 Tesla static magnet field. See attached MR<br>Safety Testing summary. | | Conclusion: | The Greatbatch Scientific MR Compatible Laparoscope and Operating<br>Laparoscope as designed, can be used in a MR or an interventional MR<br>environment, not to exceed a 1.5 Tesla static magnetic field. | {6}------------------------------------------------ ## GREATBATCH SCIENTIFIC MR COMPATIBLE LAPAROSCOPE MR SAFETY TESTING SUMMARY A Greatbatch Scientific MR Compatible 8.0mm endoscope was tested in conjunction with a A Greatoath Sciencific NA Companie of magnetic attraction and torque, artifact, and RF Heating. Based on intended use of the device and the results of extensive MR safety testing listed below, this device is MR Compatible. #### Magnetic Attraction and Torque 1. 1.5 Tesla / GE Signa 64 MHz MR System Static field strength-.. String deflection Type of test-Very slight Observed deflection -Observed torque -5 degrees ### 2. Artifact . . 1.5 Tesla / GE Signa 64 MHz MR System Static field strength -Sequences l Amount of distortion: | Sequence | axial plane | sagittal plane | |----------|-------------|----------------| | FSPGR | ++ | ++ | artifact same as the device 44 #### 3. RF Heating Phantom: A . | Type of phantom - | A 50 lb., fluid filled rectangular shaped<br>plastic Plexiglas phantom | |----------------------------------------|------------------------------------------------------------------------| | Type of RF Coil - | Body coil | | SAR applied - | 2.1 W/kg | | Length of time - | 60 minutes | | Maximum temperature<br>rise observed - | 0.5°C | {7}------------------------------------------------ Image /page/6/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC 1 1997 Mr. Gary J. Sfeir, RAC Director, Regulatory Affairs Greatbatch Scientific 4100 Barton Road Clarence. New York 14031 Re: K972280 > Trade Name: Greatbatch Scientific MR Compatible Laparoscopes and Operating Laparoscopes Regulatory Class: II Product Code: GCJ Dated: September 10, 1997 Received: September 12, 1997 Dear Mr. Sfeir: We have reviewed your Section 510(k) notification of intent to market the devices referenced above and we have determined the devices are substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the devices, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your devices are classified (see above) into either class II (Special Controls) or class III (Premarket Approval), they may be subject to such additional controls. Existing major regulations affecting your devices can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements , as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. In addition, FDA may publish further announcements concerning your devices in the Federal Register. Please note: this response to your premarket notification submissions does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {8}------------------------------------------------ ### Page 2 - Mr. Gary J. Sfeir, RAC This letter will allow you to begin marketing your devices as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your devices to a legally marketed predicate device results in a classification for your devices and thus, permits your devices to proceed to the market. If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your devices, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Stupt Rhodes Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {9}------------------------------------------------ ## SECTION 4 ### INDICATIONS FOR USE 510 (k) Number (if known): Device Name: Greatbatch Scientific MR Compatible Laparoscopes and Operating Laparoscopes Indications For Use: The Greatbatch Scientific MR Compatible Laparoscopes and Operating Laparoscopes are used to permit viewing of the organs within the peritoneal cavity for the purpose of performing diagnostic and surgical procedures in a MR or an interventional MR environment, not to exceed a shielded 1.5 Tesla magnet. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED CONCURRENCE OF CDRH; OFFICE OF DEVICE EVALUATION (ODE) | PRESCRIPTION USE | X | OR... | OVER-THE-COUNTER USE | | |------------------|---|-------|----------------------|--| |------------------|---|-------|----------------------|--| | (Division Sign-Off) | | (OPTIONAL FORMAT 1-2-96) | |-----------------------------------------|---------|--------------------------| | Division of General Restorative Devices | | | | 510(k) Number | K972280 | |
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...