SPACEMAKER SERIAL SURGICAL BALLOON DISSECTOR, WITH CANNULA, WITH VISUALIZATION AND WITH VISUALIZATION AND CANNULA

K962702 · General Surgical Innovations · GCJ · Feb 19, 1997 · Gastroenterology, Urology

Device Facts

Record IDK962702
Device NameSPACEMAKER SERIAL SURGICAL BALLOON DISSECTOR, WITH CANNULA, WITH VISUALIZATION AND WITH VISUALIZATION AND CANNULA
ApplicantGeneral Surgical Innovations
Product CodeGCJ · Gastroenterology, Urology
Decision DateFeb 19, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The intended use of the Spacemaker® Serial Surgical Balloon Dissector and the predicate devices is to dissect layers of connective tissue along natural planes of separation. The cannula portion of the device is intended to provide access for operative and diagnostic instruments for the duration of the operative procedure.

Device Story

Disposable, handheld, manually manipulated surgical balloon dissector; consists of balloon, center rod, handle, and suction/fill assembly. Used in OR by surgeons to dissect connective tissue along natural planes. Device allows multiple inflations/deflations per procedure using saline or air (10-1000cc). Hollow center rod version accommodates endoscopic visualization accessories. Optional cannula provides access for operative/diagnostic instruments. Benefits include creating surgical space for minimally invasive procedures.

Clinical Evidence

Bench testing only. Performance bench tests verified that modifications (multiple inflation capability, new silicone material, elliptical balloon shape) do not affect safety or efficacy compared to predicate devices. Biocompatibility testing performed per ANSI/AAMI 10993-1.

Technological Characteristics

Materials: Plastic, stainless steel, and silicone (compliant with ANSI/AAMI 10993-1). Principle: Manual balloon inflation/deflation for tissue dissection. Form factor: Handheld, disposable, with optional cannula and endoscopic visualization port. Balloon volume: 10-1000cc. Inflation medium: Saline or air. Configurations: Solid or hollow center rod; various balloon profiles (kidney, clover, horseshoe, semi-round, U-shaped, elliptical).

Indications for Use

Indicated for any area of the body where it is necessary to separate layers of connective tissue for surgical access, including general, laparoscopic, endoscopic, plastic, reconstructive, aesthetic surgery, and vascular use for exposure of superficial veins of upper/lower extremities; excludes arterial, deep, or non-extremity structures.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{1} K962702 Attachment 4 Section 10.0: 510(K) Summary FEB 19 1997 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92(c). ## 10.1 Contact Person Cathleen Mantor General Surgical Innovations, Inc. 3172A Porter Drive Palo Alto, CA 94304 (415) 812-9730 (415) 812-9731 fax ## 10.2 Date Summary Prepared July 9, 1996 ## 10.3 Trade Name Spacemaker® Serial Surgical Balloon Dissector, Class II Spacemaker® Serial Surgical Balloon Dissector with Cannula, Class II Spacemaker® Serial Surgical Balloon Dissector with Visualization, Class II Spacemaker® Serial Surgical Balloon Dissector with Visualization and Cannula, Class II ## 10.4 Common Name Surgical Balloon Dissector with or without Cannula ## 10.5 Predicate devices K944418 Spacemaker® and Spacemaker® II Surgical Balloon Dissector and Cannula, Class II K951878 Spacemaker® and Spacemaker® II Surgical Balloon Dissector with and without a Cannula, Class II 01/16/97 23 (page number from original submission) {2} Attachment 4 ## 10.6 Device Description The Spacemaker® Serial Surgical Balloon Dissector device is a disposable, handheld, manually manipulated surgical balloon dissection device. The device consists of a balloon, center rod, handle, and suction/fill assembly. The device can be supplied either with a solid center rod or with a hollow center rod. The hollow center rod allows for the insertion of an endoscopic visualization accessory, if desired. Both versions can be inflated and deflated multiple times during a procedure. The device can be supplied with or without a cannula. The cannula provides access for operative and diagnostic instrumentation for the duration of the surgical procedure. This submission seeks clearance for the Spacemaker® Serial Surgical Balloon Dissector to be supplied in four configurations as follows: - Spacemaker® Serial Surgical Balloon Dissector, Class II - Spacemaker® Serial Surgical Balloon Dissector with Cannula, Class II - Spacemaker® Serial Surgical Balloon Dissector with Visualization, Class II - Spacemaker® Serial Surgical Balloon Dissector with Visualization and Cannula, Class II ## 10.7 Intended Use The intended use of the Spacemaker® Serial Surgical Balloon Dissector is the same as the predicate devices (K944418, K951878). The intended use of the Spacemaker® Serial Surgical Balloon Dissector and the predicate devices is to dissect layers of connective tissue along natural planes of separation. The cannula portion of the device is intended to provide access for operative and diagnostic instruments for the duration of the operative procedure. 01/16/97 24 (page number from original submission) {3} Attachment 4 # 10.8 Comparison to Predicate Devices | Point of Comparison | Spacemaker® and Spacemaker® II Surgical Balloon Dissector and Cannula (K944418) | Spacemaker® and Spacemaker® II Dissector with and without Cannula (K951878) | Spacemaker® Serial Surgical Balloon Dissector | | --- | --- | --- | --- | | Intended Use | Dissect layers of connective tissue along natural planes of separation. | Dissect layers of connective tissue along natural planes of separation. | Dissect layers of connective tissue along natural planes of separation. | | Indications | Any area of the body where it is necessary to separate the layers of connective tissue for surgical access. | Any area of the body where it is necessary to separate the layers of connective tissue for surgical access. | Any area of the body where it is necessary to separate the layers of connective tissue for surgical access. | | Examples where surgical dissection of natural tissue planes can provide surgical access: | General, laparoscopic, and endoscopic surgery | General, laparoscopic, endoscopic, plastic and reconstructive, including aesthetic surgery | General, laparoscopic, endoscopic, plastic and reconstructive, including aesthetic surgery and vascular use for the exposure of superficial veins of the upper and/or lower extremities; no arterial nor deep nor non-extremity structure would be targeted. | | Supplied | Single patient use; Disposable | Single patient use; Disposable | Single patient use; Disposable | | Balloon Inflations Per Procedure | Single | Single | Multiple | | Principle of Operation | Handheld, manually manipulated surgical instrument | Handheld, manually manipulated surgical instrument | Handheld, manually manipulated surgical instrument | | Balloon Volume Range | 200-1500cc saline inflation | 10-2500cc saline or air inflation | 10-1000cc saline or air inflation | | Packaging Configuration | Dissector with cannula | Dissector with cannula or Dissector without cannula | Dissector with cannula or Dissector without cannula | | Balloon Shapes | Include kidney, clover, horseshoe, semi-round, and U-shaped profiles | Include kidney, clover, horseshoe, semi-round, and U-shaped profiles | Include kidney, clover, horseshoe, semi-round, U-shaped and elliptical profiles | | Materials | Plastic, Stainless steel | Plastic, Stainless steel | Plastic, Stainless steel | | Balloon Cover | Yes | Yes | No | | Center Rod | Rigid or flexible | Rigid or flexible | Rigid or flexible | | Endoscopic Visualization | No (Spacemaker®) Yes (Spacemaker® II) | No (Spacemaker®) Yes (Spacemaker® II) | Optional | | Cannula Sizes (mm) | 10/11 or 12/13 | 10/11, 12/13, or No Cannula | 10/11, 12/13, or No Cannula | 01/16/97 25 (page number from original submission) {4} Attachment 4 ## 10.9 Testing in Support of Substantial Equivalence Determination Results of performance bench tests verify that the modifications do not affect the safety or efficacy of the Spacemaker® Serial Surgical Balloon Dissector compared to the predicate devices. There is one material used in the Spacemaker® Serial Surgical Balloon Dissector (which was not used in the predicate devices). The remainder of the materials were used in the predicate devices. The material (a well known and commonly used silicone) meets the biocompatibility test requirements as defined in ANSI/AAMI 10993-1 (ISO 10993-1). 01/16/97 26 (page number from original submission)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...