USA SERIES LAPAROSCOPES

K013165 · Circon Acmi · GCJ · Jan 18, 2002 · Gastroenterology, Urology

Device Facts

Record IDK013165
Device NameUSA SERIES LAPAROSCOPES
ApplicantCircon Acmi
Product CodeGCJ · Gastroenterology, Urology
Decision DateJan 18, 2002
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

CIRCON's USA Series™ Laparoscopes are intended for use by qualified physicians to provide access, illumination and visualization of body cavities, hollow organs, and canals during endoscopic and laparoscopic surgical procedures. These include, but are not limited to, laparoscopic procedures used in general surgical procedures, cholecystectomy, colon resection or therapeutic thoracoscopy.

Device Story

Rigid endoscope for visualization of body cavities and organs; utilizes fiber optics and optical lenses within a rigid shaft; provides illumination and access for surgical instrumentation via an optional working channel; operated by qualified physicians in surgical settings; enables minimally invasive procedures such as cholecystectomy and colon resection; benefits patients through reduced invasiveness compared to open surgery.

Clinical Evidence

Bench testing only; no clinical data provided. Substantial equivalence is based on design, material, and intended use comparisons to legally marketed predicate devices.

Technological Characteristics

Rigid shaft endoscope; fiber optic illumination; optical lens imaging; medical-grade biocompatible materials; optional working channel for accessory instrumentation; non-powered/passive optical device.

Indications for Use

Indicated for use by qualified physicians for access, illumination, and visualization of body cavities, hollow organs, and canals during endoscopic and laparoscopic surgical procedures, including general surgery, cholecystectomy, colon resection, and therapeutic thoracoscopy.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # JAN 1 8 2002 #### 510(k) SUMMARY ## CIRCON's USA Series™ Laparoscopes ## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared CIRCON Corporation 6500 Hollister Avenue Santa Barbara, CA 93117 Phone: (805) 961-3290 Facsimile: (805) 968-7385 Mr. Wayne B. Sterner Contact Person: Corporate Director Regulatory Affairs 9-20-01 Date Prepared: ## Name of Device and Name/Address of Sponsor CIRCON's USA Series™ Laparoscopes with various Tradenames CIRCON Corporation 6500 Hollister Avenue Santa Barbara, CA 93117 #### Common or Usual Name Laparoscopes #### Classification Name Laparoscope, General and Plastic Surgery #### Predicate Devices Richard Wolf's Operating Laparoscopes (K991718); Smith and Nephew Dyonics 2.7 mm Microlaparoscope (K982149); Karl Storz's MVM 3.3 mm Microendoscope (K972504); Surgical Image Laboratories, Inc.'s Model LL100, 103, and 104 Laparoscopes (K955845); and Surgical Image Laboratories, Inc.'s 3.5 mm Laparoscope (K001594); {1}------------------------------------------------ ### Intended Use CIRCON's USA Series™ Laparoscopes have the same general intended use as the previously s Oblicate devices. The USA Series™ Laparoscopes are use as the proviously crowned physicians to provide access, illumination and mended for use by qualified physical pryans, and canals during endoscopic and visualization or body carrites, nonon organs, dass, dass not limited to, laparoscopic laparoscopic surgical procedures, cholecystectomy, colon resection, and therapeutic thoracoscopy. # Technological Characteristics and Substantial Equivalence CIRCON's USA Series™ Laparoscopes are substantially equivalent to the other currently marketed laparoscopes which are referenced above. CIRCON's the other carrences and their predicate devices are all telescopes that incorporate fiber optics and optical lenses within a rigid shaft and are connected to morporate fiber optios and veral models of CIRCON's USA Series™ Laparoscopes and predicate devices also incorporate a working channel within the diameter of the and production as 1200s throughput of accessory instrumentation for use during Itgia share that anow b vironging in Thus, CIRCON's USA Series™ Laparoscopes raise no new issues of safety or effectiveness. ### Biocompatibility All CIRCON USA Series™ Laparoscopes are made of medical grade biocompatible materials and are of the same type of material as used in other blocompatible fracticate laparoscopes that have the same body contact, contact duration, and intended use and indications. {2}------------------------------------------------ Public Health Service Image /page/2/Picture/2 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JAN 1 8 2002 ACMI Circon Corporation Mr. Wayne B. Sterner Corporate Director Regulatory Affairs 6500 Hollister Avenue Santa Barbara, California 93117-3019 Re: K013165 Trade Name: USA Series Laparoscopes Regulation Number: 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: GCJ Dated: January 4, 2002 Received: January 7, 2002 Dear Mr. Sterner: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Mr. Wayne Sterner This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, «Misbranding by reference to premarket notification» (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely vours. Sincerely yours, Mark N Mulkeren Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 510(k) Number (if known): _ Ho43 U43 U66 CIRCON's USA Series™ Laparoscopes Device Name: Indications for Use: CIRCON's USA Series™ Laparoscopes are intended for use by qualified physicians to provide access, illumination and visualization of body cavities, hollow organs, and canals during endoscopic and laparoscopic surgical procedures. These include, but are not limited to, laparoscopic procedures used in general surgical procedures, cholecystectomy, colon resection or therapeutic thoracoscopy. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) X OR Over-The-Counter Use _ (Optional Format 1-2-96) A. Muli N. Millerson Division Sign-Off) (Division Sign of General, Restorative and Neurological Devices 510(k) Number - \\\DC - 66763/1 - #1380464 v2
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...