The Xpress BCD is a reusable device indicated for use with stereotactic biopsy platforms to apply temporary focal mechanical compression to a stereotactic breast biopsy site to achieve hemostasis. Rx Only. Federal (USA) law restricts this device to sale by or on the order of a physician.
Device Story
XPress BCD is a reusable accessory for stereotactic breast biopsy platforms; provides temporary focal mechanical compression to biopsy sites to achieve hemostasis. Device consists of three configurations (flat, pyramidal, spherical) made of Delrin (acetal homopolymer resin). Used in clinic/radiology settings by physicians/technicians. Device is enclosed in a sterile, single-use polyethylene cover and used with sterile gauze pads to prevent direct patient contact. Attached to existing mammographic x-ray system compression paddles; pressure applied mechanically via the paddle. Replaces manual compression; improves workflow by reducing manual labor for staff. Benefits include effective hemostasis and reduced hematoma risk post-biopsy.
Clinical Evidence
Prospective, single-arm historical control study (NCT02327598) of 120 biopsies in 118 female subjects. Primary endpoint: hematoma formation (defined as > 0.52 cm³ mass on post-biopsy mammography). Results: 7.5% (9/120) immediate post-compression hematoma rate; 3.3% (4/120) delayed hematoma rate. No fat necrosis observed at one-year follow-up in 65 evaluable sites. No unanticipated adverse events reported. Study confirms device is safe and effective for hemostasis.
Technological Characteristics
Materials: Delrin (acetal homopolymer resin). Configurations: Large flat, modified pyramidal, modified spherical. Principle: Mechanical focal compression via attachment to existing stereotactic biopsy paddle. Connectivity: None (mechanical accessory). Sterilization: Non-sterile (reusable; requires intermediate-level disinfection with tuberculocidal agent).
Indications for Use
Indicated for female patients 18+ years undergoing stereotactic-guided mammographic breast biopsy procedures to achieve hemostasis at the biopsy site. Contraindications include pregnancy, lactation, cognitive impairment, active blood thinners (excluding low-dose aspirin/NSAIDs), and known bleeding disorders or uncorrected coagulopathy (platelets ≤ 80,000 or INR ≥ 1.5).
Regulatory Classification
Identification
A mammographic x-ray system is a device intended to be used to produce radiographs of the breast. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features the department's name encircling a symbol of three stylized human profiles. The profiles are stacked and connected by flowing lines, creating a sense of unity and collaboration. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 19, 2017
XPress BCD, LLC Peter Kremers, M.D. 6024 White Flint Drive Frederick, MD 21702
Re: K163388
Trade/Device Name: XPress BCD Breast Compression Device Regulation Number: 21 CFR 892.1710 Regulation Name: Mammographic x-ray System Regulatory Class: Class II Product Code: POY Dated: June 9, 2017 Received: June 12, 2017
Dear Dr. Kremers:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours. Jennifer R. Stevenson -23
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K163388
Device Name XPress BCD®
Indications for Use (Describe)
The Xpress BCD is a reusable device indicated for use with stereotactic biopsy platforms to apply temporary focal mechanical compression to a stereotactic breast biopsy site to achieve hemostasis.
Rx Only. Federal (USA) law restricts this device to sale by or on the order of a physician.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------|--------------------------------------------------|
| \[x] Prescription Use (Part 21 CFR 801 Subpart D) | \[ ] Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K163388 510(K) SUMMARY
July 17, 2017
#### 1.0 SPONSOR
Peter Kremers, MD (Contact person) 6024 White Flint Drive Frederick, MD 21702 T: 301.401.5853 E: peterkremers@comcast.net
#### NAME OF DEVICE 2.0
Trade name: XPress BCD® Breast Compression Device Common name: Post Breast Biopsy Hemostatic Breast Compression Device
#### CLASSIFICATION OF DEVICE 3.0
Classification: Class II Product Code: POY
Regulation Number: 892.1710
Classification Panel: Radiology
#### 4.0 PREDICATE DEVICE
Device Trade Name: Biopsy Digit S and Biopsy Digit SL 510(k) Control Number: K113607 Product Code: IZH Manufacturer: Giotto USA, LLC, Wichita, KS
#### 5.0 INDICATIONS FOR USE
The Xpress BCD is a reusable device indicated for use with stereotactic biopsy platforms to apply temporary focal mechanical compression to a stereotactic breast biopsy site to achieve hemostasis. Rx Only. Federal (USA) law restricts this device to sale by or on the order of a physician.
#### DEVICE DESCRIPTION 6.0
The XPress BCD device is an accessory used in conjunction stereotactic breast biopsy devices.
The XPress BCD is comprised of Delrin® DuPont (acetal homopolymer resin, Delrin 150NC010), or similar material. There are three (3) independent XPress BCD device configurations that are provided as a set: configuration 1 (large flat), configuration 2 (modified pyramidal), and configuration 3 (modified spherical). These different configurations are used based on the biopsy site being treated. The Xpress design includes a mechanical slot for attaching to compression paddles of mammographic x-ray systems
{4}------------------------------------------------
The XPress BCD is used in conjunction with the XPress BCD Cover (K863853; classification JAA; manufactured by Preferred Medical Products, Ducktown, TN). It is used to enclose the XPress BCD for use as intended. The XPress BCD Cover, is a sterile, singleuse covering. The XPress BCD Cover is a 4 X 5 inches covering of a 0.002 mm thick polyethylene blend material, and its closure system is a round or oblong synthetic rubber band made of a polyisoprene polymer, measuring 2 X 1/8 X 1/16 inches. The XPress BCD Cover is packaged with the XPress BCD or may be purchased separately.
#### 7.0 BIOCOMPATIBILITY
The Xpress BCD device does not come in direct or indirect contact with the patient as intended for use. Biocompatibility testing has not been performed on the XPress BCD.
#### 8.0 STERILIZATION AND CLEANING
The XPress BCD is a non-sterile, reusable device. The Xpress BCD device does not come in direct or indirect contact with patients during breast biopsy procedures. The Xpress BCD compression device is to be covered by a sterile, single-use XPress BCD Cover prior to use as intended.
The XPress BCD device, without protective covering, must be cleaned, disinfected, and inspected prior to use on a patient, including between biopsies on the same patient, using an EPA-registered disinfectant that achieves intermediate level of disinfection with a tuberculocidal claim.
#### 9.0 CLINICAL TESTING
A prospective historical control single arm clinical study (NCT02327598) was conducted at one investigational site to evaluate the performance of the XPress BCD breast compression device to obtain hemostasis after stereotactic-guided breast biopsy following 120 biopsies in 118 female subjects with mammographic findings that warranted stereotactic biopsy. Five investigators and six assistants participated in this study; the 118 enrolled and treated subjects met the following inclusion criteria:
- Adult women: eighteen years of age and older. ●
- . Sites of calcification (size not restricted) and/or breast lesions (masses) ≤ 1.5 cm in diameter deemed suspicious by radiologist/ breast surgeon.
- Provided written informed consent.
None of the 118 enrolled and treated subjects met the study's criteria for exclusion:
- Pregnancy.
- Lactation.
- Cognitive impairment.
- . On active blood thinners including, but not limited to, warfarin, clopidogrel and heparin. Low dose aspirin or non-steroidal medication are not subject to this requirement.
- A known bleeding disorder or uncorrected coagulopathy with a platelet count of ≤ . 80,000 and/or an international normalized ratio of ≥ 1.5.
{5}------------------------------------------------
- Note : Coagulation parameters were not routinely performed unless suspicion of . coagulopathy including known deficiencies such as Factor VIII and XI, Diffuse Intravascular Coagulation, Uremia, Liver Failure, Myeloproliferative Disorder, Alpha 2 Antiplasmin deficiency, Monoclonal Gammopathy, and Lupus Anticoagulant).
Follow-up after discharge included phone contact after 24 hours and 7 days. Patients were asked to return if they developed significant bruising greater than 3 cm in diameter, a new lump at the biopsy site, increasing pain, or signs of infection including redness, increased warmth, or increased temperature. Returning patients out to one week were evaluated clinically as well as with ultrasound of the biopsy site.
Additionally, all 65 of the 120 original biopsy sites deemed evaluable for evidence of fat necrosis were reviewed at one year. Evaluable sites required the absence of interval surgical biopsy/lumpectomy and mammograms available for review at one year following biopsy and mechanical compression.
The primary clinical performance measure / adverse effect were hematoma formation. Hematoma was defined as a new three dimensional mass (> 0.52 cm') detected on standard post-biopsy full- field mammography. Incidence and size of delayed hematomas were also noted. Unanticipated adverse events were defined as any serious adverse effect on health, safety, or any life-threatening problem or death caused by or associated with the device. No unanticipated adverse events were noted during the course of the study and follow-up evaluation.
Table 9-1 summarizes demographics and performance measures for 120 biopsy procedures performed on 118 enrolled and treated clinical study patients. Immediate post-compression hematoma was detected by primary investigators in 9/120 biopsies (7.5%). Additional delayed hematomas detected out to one week after biopsy were found following 4/120 (3.2%) of the biopsies. No mammographic evidence of fat necrosis was noted at one-year post-biopsy in a subgroup of 65 subjects who had not been subject to open biopsy or lumpectomy and were evaluated with mammography at one year post-biopsy. All biopsy sites meeting the criteria for evaluation were assessed for mammographic for evidence of fat necrosis by both the primary study and independent mammographers.
Table 9-1. Demographics and Performance Measures across 118 Patients and 120 Biopsy Sites
| Measur<br>e | Age<br>(yrs) | Breast<br>Density | Breast<br>Compression<br>(mm) | Calcs/<br>Other* | Cores | Compression<br>Time (min) | Hematoma | Hematoma<br>Size (cm³) | Delayed<br>Hematoma | Fat<br>Necrosis |
|-------------|--------------|-------------------|-------------------------------|------------------|-------|---------------------------|-----------------|------------------------|--------------------------------|-----------------|
| Total | | | | 80/ 40 | | | 9/120<br>(7.5%) | | 4/120 (3.3%) | 0/65 |
| Averag<br>e | 57 | 2 | 65 | | 9.8 | 11 | | 4.4 | | |
| Range | 34-<br>79 | 1-4 | 30-103 | | 5-24 | 10-20 | | 1.0-12.1 | Occurred on<br>days 2, 5, 7, 7 | |
*80 sites were clusters of calcification. 40 sites were other targets, including masses, densities, and sites of architectural distortion.
Table 9-2 provides the results from two independent mammographers' examinations; one found hematomas on immediate post-biopsy imaging following 7/120 (5.8%) of biopsy procedures and the other in 9/120 (7.5%). Neither noted any mammographic evidence of fat necrosis in the 65 subjects who did not proceed to open biopsy or lumpectomy and for whom mammography at one year post-biopsy was available for review.
{6}------------------------------------------------
| Reviewer | Hematoma | Average<br>Hematoma<br>Size (cm3) | Fat Necrosis |
|----------|------------------|-----------------------------------|--------------|
| A | 7/120<br>(5.80%) | 5.6 | 0/65 |
| B | 9/120<br>(7.5%) | 5.7 | 0/65 |
| | | Table 9-2. Independent Imaging Review across 118 Patients and 120 Biopsy Sites | |
|--|--|--------------------------------------------------------------------------------|--|
| | | | |
Of the 120 original biopsy sites. 65 were evaluable for evidence of fat necrosis as surgical biopsy/lumpectomy had not been performed and mammograms were available for review at one year following biopsy and mechanical compression.
Table 9-3 presents assessments reported by study staff and the 118 subjects. Most assessments are reported per biopsy site (N=120), however, post-study satisfaction was reported per patient (N=118).
Table 9-3. Staff and Subject Reported Outcomes across 118 Patients and 120 Biopsy Sites
| | Staff-Reported Evaluation | Subject-Reported Pain | | | Subject-Reported Satisfaction | | |
|-------------------|---------------------------------------------------|--------------------------------|-------------|------|---------------------------------------------|---------------------|---------------------|
| | | Procedure | Compression | Post | 0-hour | 24-hour | 1-7 day |
| # Biopsy<br>Sites | 120 | 120 | 120 | 120 | 120 | 115/118<br>patients | 113/118<br>patients |
| Average | 5 | 2 | 0.9 | 0.3 | 1.3 | 1.3 | 1.2 |
| Criteria<br>Rang | -5= extreme degradation<br>5= extreme enhancement | 0 = no pain<br>10 = sever pain | | | 1 = great experience<br>5 = poor experience | | |
Note: Staff evaluation compared their impressions on the use of mechanical compression on workflow relative to the use of standard manual compression using a 10 point scale (-5 extreme degradation to +5 extreme enhancement of workflow). The effect on workflow closely mirrored staff estimates of net time saved or lost (in minutes) relative to standard manual compression (assuming that they would have been responsible for performing manual breast compression).
#### SUBSTANTIAL EQUIVALENCE FEATURES 10.0
During breast biopsy procedures, the XPress BCD is used to apply direct mechanical compression to the biopsy site, as a replacement for standard manual compression. The XPress BCD is inserted into the stereo guide compression paddle, and pressure is applied mechanically by the paddle and attached XPress BCD device. Before insertion into the compression paddle, the XPress BCD is placed inside, and enclosed by, a sterile XPress BCD Cover, so that the device does not come in direct contact with the patient. Further, sterile 4 X 4 gauze pads are placed between the covered XPress BCD device and the patient, which is an additional barrier to direct contact with the patient.
The XPress BCD Breast Compression Device, is substantially equivalent to the legally marketed, commercially available Biopsy Digit S and Biopsy Digit SL (K113607). manufactured by Giotto USA, LLC, Wichita, KS 67226. The XPress device does not introduce any technological characteristic differences that raise questions of safety or effectiveness; to establish this, a list of comparative features and their similarities and differences is provided in the table below:
{7}------------------------------------------------
| Table 10 -1. Comparison of Similarities and Differences between Biopsy Digit S and | |
|------------------------------------------------------------------------------------|--|
| Biopsy Digit SL (K113607) and XPress BCD | |
| COMPARISON | PREDICATE DEVICE:<br>BIOPSY DIGIT S AND<br>BIOPSY DIGIT SL | SUBJECT DEVICE:<br>XPRESS BCD | SUBSTANTIALLY<br>EQUIVALENT |
|---------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The Biopsy Digit S and<br>SL are intended to be<br>used for mammographic<br>procedures requiring<br>stereotactic guidance,<br>such as fine needle<br>aspiration, needle biopsy<br>and guide wire placement. | The XPress BCD is intended<br>to be used for mammographic<br>procedures requiring<br>stereotactic guidance. The<br>device is indicated for focal<br>mechanical compression of<br>biopsy sites to achieve and<br>maintain hemostasis. | Yes. Intended for use in<br>mammographic<br>procedures requiring<br>stereotactic guidance.<br>SE demonstrated by<br>well-controlled clinical<br>trial. |
| Intended<br>Population | The target population<br>consists of female patients<br>undergoing stereotactic<br>guided mammographic<br>biopsy procedures. | The target population consists<br>of female patients undergoing<br>stereotactic guided<br>mammographic biopsy<br>procedures. | Yes. |
| Prescription Use | Yes | Yes | Yes |
| Device Class | Class II | Class II | Yes |
| Device | System, X-Ray,<br>Mammographic | Post Breast Biopsy Hemostatic<br>Breast Compression Device | Yes. Regulation<br>Description is<br>Mammographic x-ray<br>system. SE<br>demonstrated by well-<br>controlled clinical trial. |
| Regulation Number | 21 CFR 892.1710<br>(a) Identification. A<br>mammographic x-ray<br>system is a device<br>intended to be used to<br>produce radiographs of<br>the breast. This generic<br>type of device may<br>include signal analysis<br>and display equipment,<br>patient and equipment<br>supports, component<br>parts, and accessories.<br>(b) Classification. Class<br>II. | 21 CFR 892.1710<br>(a) Identification. A<br>mammographic x-ray system<br>is a device intended to be used<br>to produce radiographs of the<br>breast. This generic type of<br>device may include signal<br>analysis and display<br>equipment, patient and<br>equipment supports,<br>component parts, and<br>accessories.<br>(b) Classification. Class II. | Yes |
| Product Code | IZH | POY | Yes. Product Codes<br>included in Regulation<br>Number 892.1710. SE<br>demonstrated by well-<br>controlled clinical trial. |
| Used for<br>mammographic<br>procedures<br>requiring<br>stereotactic<br>guidance | Yes | Yes | Yes |
| Regulation<br>Description | Mammographic x-ray<br>system. | Mammographic x-ray system. | Yes |
| Provided Sterile or<br>Non-Sterile | Non-sterile | Non-sterile | Yes |
| COMPARISON | PREDICATE DEVICE:<br>BIOPSY DIGIT S AND<br>BIOPSY DIGIT SL | SUBJECT DEVICE:<br>XPRESS BCD | SUBSTANTIALLY<br>EQUIVALENT |
| User interface<br>software | Yes | No | Yes. No differences in<br>safety or effectiveness<br>are introduced |
| Electro-mechanical<br>mechanisms | Yes | Yes* | Yes. |
| Computer interface<br>to store images | Yes | No | Yes. No differences in<br>safety or effectiveness<br>are introduced |
| Computer interface<br>to determine<br>coordinates for<br>needle positioning | Yes | No | Yes. No differences in<br>safety or effectiveness<br>are introduced |
{8}------------------------------------------------
"Though the subject device does not contain any electro-mechanisms, when attached to a stereotactic breast biopsy device it may be moved either manually or electro-mechanically via that device.
#### 11.0 SUBSTANTIAL EQUIVALENCE
The XPress BCD, and the claimed predicate device, the Biopsy Digit S and Biopsy Digit SL device, are accessories used during stereotactic-guided breast biopsy procedures including breast biopsies. The XPress BCD device and the predicate device have the same general intended use – they are accessories for mammographic procedures requiring stereotactic guidance including breast biopsies. The XPress device does not introduce any technological characteristic differences that raise questions of safety or effectiveness. The XPress BCD device has been demonstrated to be as safe and effective as the predicate device for intended use in a prospective, single site single arm historical control clinical study (NCT02327598). There are no additional significant questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.