VERISUITE 1.8, VERISUITE-PARTICLE 1.8, VERISUITE, VERISUITE-PARTICLE

K133914 · Medcom GmbH · LHN · Apr 3, 2014 · Radiology

Device Facts

Record IDK133914
Device NameVERISUITE 1.8, VERISUITE-PARTICLE 1.8, VERISUITE, VERISUITE-PARTICLE
ApplicantMedcom GmbH
Product CodeLHN · Radiology
Decision DateApr 3, 2014
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 892.5050
Device ClassClass 2
AttributesTherapeutic

Indications for Use

VeriSuite is an active therapeutic medical device for verification of the patient position and calculation of a correction vector for the treatment of tumors during a radiation therapy with photons, electrons (from a linear accelerator) or particles (protons, heavy ions). The VeriSuite system calculates digitally reconstructed radiographs (DRRs) based on a high-resolution CT data set for a treatment position. With these DRRs and Xray images acquired during the performance of the position verification procedure a correction vector for the patient position can be calculated. The VeriSuite® system with fluoroscopy extension provides the possibility to a medical user to observe the position of the patient and the treatment area during the treatment, mainly the movements of the patient caused by breathing. This information can be used to interrupt the treatment. An authorized person must evaluate the correctness of the calculation and approve the result for further usage. The system shall only be used after correct installation in appropriate treatment rooms by trained personnel. Legal regulations especially requlation for the operation of X-ray devices must be regarded. VeriSuite must not be used for diagnostic purposes.

Device Story

VeriSuite is an image processing system for radiotherapy patient positioning. Inputs include real-time stereoscopic X-ray images and DRRs derived from patient CT scans and treatment planning data. The system calculates a correction vector to align patient position with the treatment plan. The fluoroscopy extension allows clinicians to monitor patient/treatment area movement, specifically respiratory motion, to manually interrupt treatment if necessary. Used in treatment rooms by trained personnel; requires authorized user evaluation and approval of calculated corrections. Benefits include improved accuracy of radiation delivery to tumors and real-time monitoring of patient movement.

Clinical Evidence

Bench testing only. Non-clinical verification and validation software tests were conducted to confirm the system meets intended use and safety requirements.

Technological Characteristics

System includes VeriSuite software, Ralco 302 collimators, Sedecal SHF 835 X-ray generator, Varian A277/A272 X-ray tubes, and Varian 4030E flat panel digital imagers. Operates via stereoscopic X-ray imaging and DRR calculation. Complies with IEC 61217 (coordinates/movements), ISO 14971 (risk management), and IEC 62304 (software lifecycle).

Indications for Use

Indicated for patient position verification and correction during radiotherapy treatment with external beams or charged particles. Used for patients undergoing radiation therapy for tumors. Not for diagnostic use.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Page 6 of 44 K133914 Page 1 of 4 # V. 510(k) Summary of Safety and Effectiveness #### A. Submitter | Name: | MedCom GmbH | | |--------------------------------------|------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------| | Address: | Rundeturmstr. 12<br>Darmstadt, HE 64283<br>Germany | | | Establishment Reg.: | 3006579682 | | | Telephone: | +49 (6151) 95147-18 | | | Fax: | +49 (6151) 95147-20 | | | Contact: | Mr. Luca Salvatore<br>lsalvatore@medcom-online.de | | | Date: | December 20, 2013 | | | <b>B. Device</b> | | | | Trade Name: | VeriSuite also marketed as VeriSuite 1.8 and VeriSuite-<br>Particle and VeriSuite-Particle 1.8 | | | Common name: | Patient position verification system | | | Classification: | Regulatory Class: | II | | | Product Code: | LHN / Classification Name<br>system, radiation therapy,<br>charged-particle, medical | | | Subsequent Product Code: | IYE / Classification Name:<br>accelerator, linear, medical | | | CFR Section: | 892.5050 | | | Panel: | Radiology | | <b>C. Predicate Devices</b> | | | | Device trade name:<br>510(k) number: | VeriSuite 1.8<br>K092653 | | 892.5050 Class II LHN MedCom GmbH Company name: Classification: Product code: Classification Number: {1}------------------------------------------------ IYE Subsequent code: Device trade name: EXACTRAC 5.5 / ExacTrac X-RAY 6D 510(k) number: K072506 Company name: BRAINLAB AG 892.5050 Classification Number: Class II Classification: Product code: IYE ### D. Reason for Submission Changed device: Fluoroscopy extension #### E. Standards 1. IEC 61217 (2008), Radiotherapy equipment - Coordinates, movements, and scales Consolidated Edition 1.2. (Radiology) 2. ISO 14971:2007, Medical devices - Application of risk management to medical devices. (General I (QS/RM)) 3. IEC 62304 First edition 2006-05, Medical device software - Software life cycle processes. (Software/Informatics) #### F. Description VeriSuite is an image processing system for verification and correction of the patient position during a radiation therapy treatment. The verification or correction is performed by a comparison of X-ray images that are acquired during the treatment with DRRs (digital reconstructed radiographs) calculated from a CT image series of the patient and information from the radiation therapy planning. The correction can also be based on fiducial, radio-opaque markers that are implanted in the patient. With the fluoroscopy extension it is possible to observe the position of the patient and the treatment area during the treatment. Mainly the movements of the patient caused by breathing are in the focus. This information can be used to interrupt the treatment by the user. VeriSuite is a system of devices consisting of the VeriSuite software and a number of hardware devices: | Device | Type | 510(k) / registration number | |------------------------------------|----------------------|------------------------------| | Beam limiting<br>collimator device | Ralco 302 | K946320 | | X-ray generator | Sedecal SHF 835 (RF) | 9617251 | | X-ray tubes | Varian A277 or A272 | 1717855 | | Flat panel digital<br>imager | Varian 4030E, | K024147 | {2}------------------------------------------------ Image /page/2/Figure/2 description: The image shows a system overview diagram. The diagram includes components such as Collimators Ralco R302 1, 2, and 3, Xray Tubes 1, 2, and 3 Varian A277, and Digital Flat Panels Varian 4030E 1 and 2. The diagram also includes a VeriSuite Workstation running Windows XP Professional, with components such as VeriSuite Application, Varian DLLs PaxScan API L04, and Network Cards Intel PRO 1000 GT. All these hardware devices are legally marketed in the US as listed in the table above. #### G. Intended Use VeriSuite is an active therapeutic medical device for verification of the patient position and calculation of a correction vector for the treatment of tumors during a radiation therapy with photons, electrons (from a linear accelerator) or particles (protons, heavy ions). The VeriSuite system calculates digitally reconstructed radiographs (DRRs) based on a high-resolution CT data set for a treatment position. With these DRRs and Xray images acquired during the performance of the position verification procedure a correction vector for the patient position can be calculated. The VeriSuite® system with fluoroscopy extension provides the possibility to a medical user to observe the position of the patient and the treatment area during the treatment, mainly the movements of the patient caused by breathing. This information can be used to interrupt the treatment. An authorized person must evaluate the correctness of the calculation and approve the result for further usage. The system shall only be used after correct installation in appropriate treatment rooms by trained personnel. Legal regulations especially requlation for the operation of X-ray devices must be regarded. VeriSuite must not be used for diagnostic purposes. #### H. Technological Comparison to Predicate Devices The changed device VeriSuite 1.8 (build number B641.4, subject of this submission) is substantially equivalent to the predicate device VeriSuite 1.8 (K092653) except the fluoroscopy extension. {3}------------------------------------------------ Image /page/3/Picture/1 description: This image shows handwritten text on a white background. The text includes "K133914" at the top, followed by "Page 4 of 4" in a cursive style. Additionally, there is a printed line of text at the bottom left that reads "Page 9 of 44". The handwriting appears to be done with a dark ink pen. The fluoroscopy extension of VeriSuite is substantially equivalent to the fluoroscopy functionality of the predicate device EXACTRAC 5.5 / ExacTrac X-RAY 6D (K072506) in terms of intended use and technology. In contrast to EXACTRAC 5.5, VeriSuite does not support optical prepositioning and internal respiratory gating signal display. Because of the reproducibility of the patient preposition due to the fixation of the patient, optical prepositioning is not required in the VeriSuite environment. Respiratory gating can be displayed using external devices. Minor differences exist in the hardware used in terms of resolution of the flat panels, x-ray image and pixel size, which are higher for VeriSuite. Refer to section XII for a detailed predicate device comparison. #### l. Non-clinical Performance Data Non-clinical verification and validation software tests were conducted to confirm that VeriSuite 1.8 with fluoroscopy meets its intended use and is safe and effective. See section XVI.I for details. #### J. Conclusion Based on the information provided in this Premarket Notification MedCom concludes that VeriSuite 1.8 with fluoroscopy extension is as safe and effective and substantially equivalent to the predicate devices described herein. {4}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the emblem. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 3, 2014 MedCom GmbH % Mr. Luca Salvatore Quality Manager Rundeturmstrabe 12 64283 Darmstadt GERMANY Re: K133914 Trade/Device Name: VeriSuite 1.8, VeriSuite-Particle 1.8, VeriSuite, VeriSuite-Particle Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: LHN Dated: January 6, 2014 Received: January 8, 2014 Dear Mr. Salvatore: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {5}------------------------------------------------ ### Page 2-Mr. Salvatore If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.lda.gov/MedicalDevices/Resources/orYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Michael D. O'Hara for Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page. #### 510(k) Number (if known) #### K133914 #### Device Name VeriSuite #### Indications for Use (Describe) The VeriSuite patient position verification and correction of the patient's position during a radiotherapy treatment with external beams or charged particles. It is based on stereoscopic X-ray images and DRRs calculated from a CT image series of the treatment region of the patient and information from the treatment planning. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Mechal D. O'Hara FORM FDA 3881 (9/13) Ser nees ( 80) : 441-6746 Pat Iraka {7}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond fo, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...