ImagingRing System

K161400 · Medphoton GmbH · IYE · Dec 8, 2016 · Radiology

Device Facts

Record IDK161400
Device NameImagingRing System
ApplicantMedphoton GmbH
Product CodeIYE · Radiology
Decision DateDec 8, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5050
Device ClassClass 2

Indications for Use

The ImagingRing System (IRS) is an X-ray device consisting of (1) imaging components, supports and software and (2) patient support structure (couch top), designed to support patient localization in the Couch Coordinate System (CCS) and intended to be used as image guidance equipment as part of the radiation therapy (RT) treatment process, in all areas of the body where such image guidance is determined by a licensed physician. IRS facilitates patient localization by means of imaging patient anatomy including the therapeutic target volume, critical structures, bone and soft tissue with or without the use of implanted markers or contrast agents. IRS provides 2D planar imaging (including streaming and recording sequences of 2D frames for motion analysis) and 3D volumetric imaging before, during and after irradiation, and is intended to support patient positioning, monitoring and management of internal target motion, and decision making as a function of target position, size, shape and displacement resulting from patient set-up deviation, organ deformation and anatomical movement. With the goal to accurately perform patient alignment with respect to an external treatment beam and to spare critical organs surrounding the target volume, IRS is intended to be: . embedded in the radiotherapy workflow, i.e. being able to communicate with third party systems used in particle therapy or LINAC based RT such as Oncology Information Systems (OIS) or Radiotherapy Control Systems (RTCS) in order to receive commands and data from and provide information to (e.g., receiving command from RTCS to start an image acquisition and providing image data to RTCS for the purpose to calculate and execute the patient (re)positioning vector externally in the third party system) interfaced with a Patient Positioning Systems (PPS), like - but not limited to - a robotic . multi-axis system, to allow the calculated (re)-positioning vector to be applied with respect to Coordinate Systems the treatment delivery unit relates to.

Device Story

ImagingRing System (IRS) is an X-ray imaging device for radiotherapy patient localization. Inputs: X-ray projections acquired via ring-mounted source/detector. Operation: System moves along couch top; source/detector move independently for non-isocentric trajectories and variable FOV. Software (ImRiSS) orchestrates acquisition/processing; PLC (IRCS) controls hardware. Outputs: 2D planar images (streaming/recording) and 3D CBCT volumes. Used in clinical radiotherapy environments; operated by trained personnel. Integrates with third-party Oncology Information Systems (OIS) or Radiotherapy Control Systems (RTCS) to receive commands and export data for external patient re-positioning calculations. Interfaces with robotic Patient Positioning Systems (PPS) to apply positioning vectors. Benefits: Enables accurate patient alignment relative to treatment beams, monitors internal target motion, and facilitates sparing of critical organs.

Clinical Evidence

No clinical data. Bench testing only. Verification and validation testing performed at module, integration, and system levels to demonstrate conformance to safety standards (IEC 60601-1, IEC 60601-1-3, IEC 60601-2-54, IEC 62304, IEC 62366, ISO 14971).

Technological Characteristics

X-ray imaging system with digital flat panel detector (1024x1024 matrix, 400 μm pixel size). Features motorized lateral/longitudinal movement, 360° radial rotation, and independent source/detector positioning. Generator range 40-120 kVp. Connectivity: Communicates with third-party OIS/RTCS. Materials: Advanced fiber composites (no metal in treatment area). Software: ImRiSS (orchestration) and IRCS (PLC control).

Indications for Use

Indicated for patients undergoing radiotherapy workflows for conditions including malignant and benign brain tumors, brain metastases, spine lesions, squamous cell carcinoma of the head and neck, lung, breast, pancreatic, hepatic malignancies, prostate, and bone metastases.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image contains a circular logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling an emblem. The emblem is a stylized representation of three human profiles facing right, stacked on top of each other, with flowing lines suggesting movement or connection. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 medPhoton GmbH % Mr. Daniel Schaffarzick Official Correspondent Strubergasse 16 Salzburg, 5020 AUSTRIA Re: K161400 Trade/Device Name: ImagingRing System Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: November 4, 2016 Received: November 8, 2016 Dear Mr. Schaffarzick: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. Michael D'Hara Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### Indications for Use 510(k) Number (if known) K161400 Device Name ImagingRing System Indications for Use (Describe) The ImagingRing System (IRS) is an X-rav device consisting of (1) imaging components, supports and (2) patient support structure (couch top), designed to support patient localization in the Couch Coordinate System (CCS) and intended to be used as image guidance equipment as part of the radiation therapy (RT) treatment process, in all areas of the body where such image guidance is determined by a licensed physician. IRS facilitates patient localization by means of imaging patient anatomy including the therapeutic target volume, critical structures, bone and soft tissue with or without the use of implanted markers or contrast agents. IRS provides 2D planar imaging (including streaming and recording sequences of 2D frames for motion and 3D volumetric imaging before, during and after irradiation, and is intended to support pationing, monitoring and management of internal target motion, and decision making as a function, size, shape and displacement resulting from patient set-up deviation. organ deformation and anatomical movement. With the goal to accurately perform patient alignment with respect to an external treatment beam and to spare critical organs surrounding the target volume, IRS is intended to be: · embedded in the radiotherapy workflow, i.e. being able to communicate with third party systems used in particle therapy or LINAC based RT such as Oncology Information Systems (OIS) or Radiotherapy Control Systems (RTCS) in order to receive commands and data from and provide information to (e.g., receiving command from RTCS to start an image acquisition and providing image data to RTCS for the purpose to calculate and execute the patient (re)-positioning vector externally in the third party system) · interfaced with a Patient Positioning Systems (PPS), like - but not limited to - a robotic multi-axis system, to allow the calculated (re)-positioning vector to be applied with respect to Coordinate Systems the treatment delivery unit relates to. Type of Use (*Select one or both, as applicable*) | <div> <span style="font-size: 10pt;"> <span style="font-family: Symbol;"> <span style="color: black;">☑</span> </span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span style="font-size: 10pt;"> <span style="font-family: Symbol;"> <span style="color: black;">☐</span> </span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995.#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and comblete and review the collection of information, Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/14) Page 1 of 1 PSC Publishing Services (301) 443-6740 FF medPhoton GmbH Traditional 510(k) Submission Image /page/2/Picture/22 description: The image shows the text "Chapter 04 - 2/2". The text is written in a simple, sans-serif font and is left-aligned. The numbers indicate that this is chapter 4, and it is page 2 of 2. {3}------------------------------------------------ Image /page/3/Picture/2 description: The image shows the logo for medPhoton. The logo consists of the letters 'm' and 'p' connected in a stylized, cursive font, with the 'p' resembling the Greek letter Phi (Φ). The letters are in blue. Below the symbol, the word "medPhoton" is written in a smaller, gray sans-serif font. # 510(k) Summary In accordance with the requirements of the Safe Medical Device Act, medPhoton GmbH herewith submits a 510(k) Summary. This 510(k) summary for the ImagingRing System meets the requirements of 21 CFR 807.92. | Date Prepared: | November 04th, 2016 | |---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter Information: | medPhoton GmbH<br>Strubergasse 16<br>5020 Salzburg<br>Austria<br>Phone: 0043 699 161 053 86<br>Registration Number: not currently registered<br>Owner/Operator Number: - | | Official Correspondent: | Mr. Daniel Schaffarzick<br>medPhoton GmbH<br>Strubergasse 16<br>5020 Salzburg<br>Austria<br>Phone: 0043 699 161 054 06 | | US Agent (Contact): | Dr. Christian Kesberg<br>Österr. AußenwirtschaftsCenter New York<br>120 West 45th Street, 9th Floor<br>New York, NY 10036<br>Phone: T +1 212 421 52 50<br>E-mail: newyork@wko.at | | Device(s) Identification:<br>Device Trade Name:<br>Device common name:<br>Device Classification Name:<br>Regulation Description:<br>Product Code:<br>Reference:<br>Review Panel:<br>Device Class: | ImagingRing System<br>ImagingRing System<br>Accelarator, Linear, Medical<br>Medical charged-particle radiation therapy system<br>IYE<br>per 21 CFR 892.5050<br>Radiology<br>II | | medPhoton GmbH | Traditional 510(k) Submission | Chapter 05 – 2/9 | |----------------|-------------------------------|------------------| {4}------------------------------------------------ Image /page/4/Picture/2 description: The image contains a logo for a company called "medPhoton". The logo features the company name in a simple, sans-serif font, with "med" in gray and "Photon" in a darker shade. Above the company name is a stylized graphic that appears to be a combination of a lowercase "m" and the Greek letter "Phi", both in a blue color. The "m" is drawn with a flowing, curved line, while the "Phi" symbol is more geometric. ## Device Description: The medPhoton IRS is a standalone imaging system designed to obtain TwoDimensional (2D) and Three-Dimensional (3D) X-ray images. The images visualize patient's anatomy whose information can be used to accurately position or re-position the patient undergoing radiotherapy treatment. The IRS consists of - · hardware called ImagingRing (IR) consisting of couch top, ring, detector, X-ray source, generator, etc .; - · software called ImagingRing Software Suite (ImRiSS), to orchestrate image acquisition and data processing: - . programmable logic controller (PLC) components and related software (e.g., ImagingRing Control System - IRCS), to execute and control image acquisition. The system is designed in a way such that it provides the following key features: - 1. During imaging acquisition, the IR can be moved along the couch top in order to acquire images of any anatomical region of interest. - 2. The X-ray source and detector can move independently in order to enable nonisocentric acquisition trajectories (i.e., to focus on a selectable center of interest within the patient in an axial plane) and variable size of the axial Field Of View (FOV). - 3. The combination of imaging arm and couch top is designed to be directly attached to a range of Patient Positioning Systems (PPS) including multi-axis robotic devices. The imaging device is integrated with a couch top. The independently moveable components allow for acquisition of X-ray projective images. These provide the foundation for 2D planar imaging as well as acquisition and reconstruction of 3D ConeBeam Computed Tomography (CBCT) images around customizable centers of interest. # Intended Use: The ImagingRing System (IRS) is an X-ray device consisting of (1) imaging components, supports and software and (2) patient support structure (couch top), designed to support patient localization in the Couch Coordinate System (CCS) and intended to be used as image guidance equipment as part of the radiation therapy (RT) treatment process, in all areas of the body where such image quidance is determined by a licensed physician. IRS facilitates patient localization by means of imaging patient anatomy including the therapeutic target volume, critical structures, bone and soft tissue with or without the use of implanted markers or contrast agents. IRS provides 2D planar imaging (including streaming and recording sequences of 2D frames for motion analysis) and 3D volumetric imaging before, during and after irradiation, and is intended to support patient positioning, monitoring and management of internal target motion, | medPhoton GmbH | Traditional 510(k) Submission | Chapter 05 – 3/9 | |----------------|-------------------------------|------------------| |----------------|-------------------------------|------------------| {5}------------------------------------------------ Image /page/5/Picture/2 description: The image shows the logo for medPhoton. The logo features a stylized blue symbol resembling a lowercase 'm' connected to a Greek letter 'Phi'. Below the symbol, in gray text, is the word 'medPhoton'. and decision making as a function of target position, size, shape and displacement resulting from patient set-up deviation, organ deformation and anatomical movement. With the goal to accurately perform patient alignment with respect to an external treatment beam and to spare critical organs surrounding the target volume, IRS is intended to be: . embedded in the radiotherapy workflow, i.e. being able to communicate with third party systems used in particle therapy or LINAC based RT such as Oncology Information Systems (OIS) or Radiotherapy Control Systems (RTCS) in order to receive commands and data from and provide information to (e.g., receiving command from RTCS to start an image acquisition and providing image data to RTCS for the purpose to calculate and execute the patient (re)positioning vector externally in the third party system) interfaced with a Patient Positioning Systems (PPS), like - but not limited to - a robotic . multi-axis system, to allow the calculated (re)-positioning vector to be applied with respect to Coordinate Systems the treatment delivery unit relates to. # Predicate devices: 1. Primary Predicate | Device Trade Name: | XVI R5.0 | |--------------------|----------------------| | Applicant: | Elekta Limited | | 510(k) No.: | K131965 | | 2. | | | Device Trade Name: | kVue™ IGRT Couch top | | Applicant: | Aquaplast/Qfix | | 510(k) No.: | K060671 | The ImagingRing System is considered substantial equivalent to the Elekta XVI R5.0 (K131965) and the kVue™ IGRT Couch top (K060671). - 1. Intended use, medical application and treatment method as well as the basic parameter settings are equivalent for the ImagingRing System and the predicate devices. - 2. There is no significant difference in intended use or technology. | medPhoton GmbH | Traditional 510(k) Submission | Chapter 05 – 4/9 | |----------------|-------------------------------|------------------| |----------------|-------------------------------|------------------| {6}------------------------------------------------ Image /page/6/Picture/6 description: The image shows the logo for medPhoton. The logo consists of the letters 'm' and 'p' in a stylized, cursive font, with the 'p' resembling the Greek letter phi. Below the symbol is the word 'medPhoton' in a smaller, sans-serif font. # Summary of Technological Characteristics: | Characteristics | ImagingRing System | XVI R5.0 | |----------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------| | 510(k) | - | K131965 | | Manufacturer | medPhoton GmbH | Elekta Limited. | | Product code | IYE | IYE | | <b>Indications for use</b> | | | | Used with a charged particle or photon<br>radiation therapy system | ✓ | ✓ | | Verification of patient setup position | ✓ | ✓ | | <b>Image acquisition</b> | | | | X-ray imaging with digital flat panel detector | ✓ | ✓ | | Stereoscopic acquisition | ✓ | ✓ | | CBCT | ✓ | ✓ | | Temperature | 15°C to 30°C | 10°C to 35°C | | Humidity | 30 to 65 % | 30% to 70% | | Atmospheric Pressure | 80 kPa to 110 kPa | 80 kPa to 106 kPa | | Motorized movements laterally and<br>longitudinally | ✓ | ✓ | | Touchguard prevents collision | ✓ | ✓ | | source just delivers radiation when the<br>source arm is situated in a locked position | - | ✓ | | Touchguard prevents collision | ✓ | ✓ | | Movements are continuously variable | ✓ | ✓ | | Beam rotates radially through 360° in the<br>direction of the isocenter | ✓ | ✓ | | Speed | 0 to 1 revolutions/min | 0 to 1 revolutions/min | | Position Accuracy | $\pm0.2°$ | $\pm0.3°$ | | Inherent filtration | 1.4 | 1.4 | | Added filtration | 5.2 | 5.6 | | Total filtration at<br>100 kV | 6.6 | 7.0 | | Nominal voltage | 230 V | 400 V, 415 V, 440 V,<br>or 480 V | | Voltage variation | ±10% | ±10% | | Phases | Single-phase and earth | Three-phase and earth | | Frequency | 50 - 60 Hz | 48 Hz or 62 Hz, ±2% | | Impedance | 0.2 Ohms | 0.348 ohms to 0.501 ohms | | Fuse rating | 32 A | 63 A circuit recommended | | confinement of extra focal radiation | extra focal radiation is not more<br>than 15 cm on each side of the<br>largest X-ray field | extra focal radiation is not<br>more than 15 cm on each<br>side of the largest X-ray field | | beam limits | Independently moveable X and<br>Y jaws to shape the x-ray field of<br>irradiation | collimator cassettes that you<br>can manually change to set<br>the size of the X-ray beam | | leakage radiation in the loading state | 0.1979 mGy/h | < 1.0 mGy/h | | leakage radiation when not in the loading<br>state | 0 mGy/h | < 20 mGy/h | | Relation between the X-ray field and image<br>receptor area | At the image receptor plane, the<br>X-ray beam must not be larger<br>than the boundaries by more | At the image receptor plane, the<br>X-ray beam must not be<br>larger than the boundaries by | | medPhoton GmbH | Traditional 510(k) Submission | Chapter 05 – 5/9 | |----------------|-------------------------------|------------------| |----------------|-------------------------------|------------------| {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for medPhoton. The logo consists of the letters 'm' and 'p' in lowercase, with the 'p' stylized to resemble the Greek letter phi (Φ). The letters are in a blue color. Below the letters, the word 'medPhoton' is written in a smaller, gray font. | | than 3% of the source-to-image<br>distance along the two major<br>axes | more than 3% of the source-<br>to-image distance along the<br>two major axes | |--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------| | Radiation warning lamp | ✓ | ✓ | | 80 kV | 4.4 mm Al | 6.1 mm Al | | 120 kV | 6.3 mm Al | 8.5 mm Al | | Generator operating range | 40 - 120 kVp | 70 - 150 kVp1 | | Flat panel detector pixel size | 400 μm | 400 μm1 | | Flat panel matrix | 1024 x 1024 | 1024 x 10241 | | 2D low contrast visibility | ≥12 disks visible | ≥12 disks visible2 | | 2D spatial resolution | ≥1.6 line/mm | ≥1.4 line/mm2 | | CBCT spatial resolution | ≥10 lp/cm (up to 18) | ≥7 lp/cm1 | | CBCT low contrast resolution | < 1.5% | < 1.5%2 | | CBCT uniformity | < 2% | < 2%2 | | Transverse geometric accuracy | < 1 mm | < 1mm1 | | FlexMap3 | Yes | yes1 | | Presets and automatic setup | The ImagingRing System<br>receives information such as<br>patient and preset selection,<br>from a third party system (e.g.<br>the Radiotherapy Control<br>System of a treatment delivery<br>device) for automatic set up of<br>image acquisition parameters | SYNERGISTIQ4 | | 2D planar imaging | ✓ | PlanarView5 | | Fluoroscopy mode | ✓ | MotionView6 | | | Recording of frame sequences<br>for motion analysis (pulsed<br>fluoroscopy-like acquisition<br>mode) | | | 3D reconstruction | ✓ | VolumeView7 | | 4D volume view | N/A | Symmetry8 | <sup>1</sup> Data from http://www.aapm.org/meetings/amos2/pdf/34-8077-2750-150.pdf | medPhoton GmbH | Traditional 510(k) Submission | Chapter 05 – 6/9 | |----------------|-------------------------------|------------------| |----------------|-------------------------------|------------------| <sup>2</sup> http://chapter.aapm.org/seaapm/symposia/2015/Weatherford.pdf . Phantom used: Catphan 503. <sup>3</sup> The term FlexMap indicates a technique to compensate for the mounted components due to the difference in gravitational force along the rotation trajectory, leading to a difference between nominal and actual values as a function of the position in the trajectory. <sup>4</sup> In SYNERGISTIQ mode, the information from the patient management information system is used to automatically set up the image acquisition parameters, such as patient selection, reference data, and preset selection. <sup>5</sup> In the PlanarView mode. XVI uses presets to acquire one 2D planar image. The digital accelerator qantry does not move, a sequence of frames is acquired, and XVI does an average of these frames to create the image. The user can acquire images at applicable gantry angles in this mode. <sup>6</sup> In the MotionView mode, XVI uses presets to acquire a sequence of 2D planar images. This can be while the qantry rotates or with no gantry movement. <sup>7</sup> In the VolumeView mode, XVI acquires a sequence of 2D projection images while the digital accelerator gantry rotates. XVI uses the acquired images to reconstruct a 3D anatomical volume, which the operator can use for registration with imported CT reference data. This makes sure that the patient position is correct, and shows the target movement. <sup>8</sup> Symmetry is the acquisition, reconstruction, and registration of a 4D VolumeView. During a Symmetry VolumeView, XVI acquires the images and reconstructs them in relation to the breathing cycle over time. This lets the operator see the different phases of the breathing cycle for a patient. {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for medPhoton. The logo features a stylized lowercase 'm' connected to the Greek letter Phi (Φ) in blue. Below the symbol, the word "medPhoton" is written in gray, with "med" on the first line and "Photon" on the second line. | Intrafraction Imaging | The ImagingRing System,<br>provided that no collision is<br>detected, is capable of acquiring<br>images while the treatment<br>beam is ON. | Intrafraction imaging is an<br>option that lets you acquire<br>kV images during an MV<br>treatment field delivery. | |-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------| | Segmental Technique | The ImagingRing System is<br>capable of interrupting image<br>acquisitions using the proper<br>command. | Lets you use the<br>INTERRUPT button on the<br>FKP to do a pause in a<br>MotionView or VolumeView<br>acquisition. | | Distributed Review | The ImagingRing System<br>provides a third party system<br>(e.g. the Radiotherapy Control<br>System of a treatment delivery<br>device) called requestor, with<br>the requested image dataset for<br>image review and further<br>processing. | Distributed Review lets you<br>send data to the patient<br>management information<br>system in which you can do<br>the CBCT image review. | | IEC 60601-1 | ✓ | ✓ | | IEC 60601-2-1 | - | ✓ | | IEC 60601-1-3 | ✓ | ✓ | | IEC 60601-2-44 | ✓ | ✓ | | IEC 60601-2-54 | ✓ | ✓ | | IEC 62304 | ✓ | ✓ | | IEC 62366 | ✓ | ✓ | | ISO 14971 | ✓ | ✓ | Table 1: Comparison of technological characteristics of IRS and the predicate XVI R5.0 | Characteristics | ImagingRing System | KVue | |-----------------------------------------------------------------------------------|----------------------------------------------------|----------------------------| | 510(k) | - | K060671 | | Manufacturer | medPhoton GmbH | WFR/AQUAPLAST<br>CORP/Qfix | | Product code | IYE | JAI | | <i>Indications for use</i> | | | | Used with a charged particle or photon<br>radiation therapy system | ✓<br>limited for use in particle therapy<br>only | ✓ | | Patient setup position | ✓ | ✓ | | <i>Material</i> | | | | Advanced fiber composites that contain no<br>metal parts in the treatment area. | ✓ | ✓ | | Biocompatibility | ✓ | ✓ | | <i>Interface functionalities</i> | | | | Mountable on treatment tables using plates<br>with mechanical fixation. | ✓ | ✓ | | <i>Mechanical properties</i> | | | | Meets the deflection requirements of IEC<br>60976 (<5 mm in the specified setup). | < 1 mm effect in imaging volume<br>accuracy<br>N/A | ✓ | | Virtual Indexing | ✓ | ✓ | | medPhoton GmbH | Traditional 510(k) Submission | Chapter 05 - 7/9 | |----------------|-------------------------------|------------------| |----------------|-------------------------------|------------------| {9}------------------------------------------------ Image /page/9/Picture/0 description: The image contains a logo for a company called medPhoton. The logo features the letters "m" and "Φ" (phi) in blue, with the word "medPhoton" written in gray below. The "m" and "Φ" are stylized and connected, creating a unique visual element for the brand. | Radiolucent: Aluminum Equivalence<br>Thickness (AET): ~0.6 mm @ 100kVp;<br>Water Equivalence Thickness (WET): 2-6<br>mm @ 6 MV) | AET = 1.9 mm<br>WET≤16.5 mm | ✓ | |---------------------------------------------------------------------------------------------------------------------------------|-----------------------------|---| |---------------------------------------------------------------------------------------------------------------------------------|-----------------------------|---| Table 2: Comparison of technological characteristics of IRS and the predicate KVue The ImagingRing System and the two predicate devices target the same population, i.e., patients undergoing radiotherapy workflows in a hospital environment. Therefore, they are subject to the same human factors and must provide the same compatibility with the environment and interaction with other devices such as - but not limited to positioning/immobilization devices, radiotherapy treatment delivery systems and patient positioning systems. The ImagingRing System, like two predicate devices, can be used for treatment that includes but is not limited to malignant and benign brain tumors, brain metastases, spine lesions, squamous cell carcinoma of the head and neck, lung, breast, pancreatic, hepatic malignancies, prostate, and bone metastases. The devices create the same functional outputs that support procedures during radiotherapy. This is reflected in the similarity of the indications for use of the devices. The differences in the environmental conditions for operation, gantry movement, electrical data, beam limiting devices, technical specifications, view mode options and attenuation properties do not have impact on safety and effectiveness. Therefore the IRS is considered substantial equivalent for the main functional characteristics and the indications for use. The ImagingRing System is considered substantial equivalent to the Elekta XVI R5.0 (K131965) and the Qfix kVue (K060671). There is no significant difference in intended use or technology. {10}------------------------------------------------ Image /page/10/Picture/2 description: The image shows the logo for medPhoton. The logo features the letters 'm' and 'Phi' in a stylized, connected design, with the 'm' appearing in lowercase and the 'Phi' symbol in uppercase. The color of the letters is a bright, vibrant blue. Below the symbol is the word 'medPhoton' in a smaller, sans-serif font. # Summary of performance testing: The system is subject to compliance testing to voluntary consensus safety standards. Details of the standards employed in the design are specified in chapter 9 (Declaration of Conformity and Summary Reports) which includes but not limited to IEC 60601-1, IEC 60601-1-3, IEC 60601-2-54, IEC 62304, IEC 62366 and ISO 14971. Testing was performed to evaluate the respective safety and performance requirements in the form of module and integration verification as well as system level validation. The results from verification and validation testing prove the conformance to applicable standards and demonstrate that safety & effectiveness have been achieved. # Animal Testing: medPhoton GmbH did not perform any animal testing for the ImagingRing System. Clinical Testing: medPhoton GmbH did not perform any clinical testing for the ImagingRing System. Conclusion: medPhoton GmbH believes that the ImagingRing System is substantially equivalent to the currently legally marketed devices. The ImagingRing System does not introduce new indications for use, has the same technological characteristics and does not introduce new potential hazards or safety risks.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...