LEKSELL GAMMA KNIFE PERFEXION

K133565 · Elekta Instrument AB · IWB · May 9, 2014 · Radiology

Device Facts

Record IDK133565
Device NameLEKSELL GAMMA KNIFE PERFEXION
ApplicantElekta Instrument AB
Product CodeIWB · Radiology
Decision DateMay 9, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5750
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Leksell Gamma Knife® Perfexion™ is a teletherapy device intended for stereotactic irradiation of head structures ranging from very small target sizes of a few millimeters to several centimeters. Indications include but are not limited to; metastatic tumors, arteriovenous malformations, trigeminal neuralgia, essential tremor, meningiomas, vestibular schwannomas, pituitary adenomas and glioblastoma.

Device Story

Teletherapy system for stereotactic irradiation of intracranial structures; utilizes 192 Cobalt-60 sealed sources to deliver ionizing radiation shots to precise targets. System comprises medical unit (radiation source, patient positioning system, control/power units) and office side (operator console, computer). Operated by clinicians in treatment room/control room environment. Preoperative radiological exams inform treatment planning. Device enables highly accurate dose delivery to targets ranging from millimeters to centimeters. Benefits include non-invasive treatment of tumors and neurological conditions.

Clinical Evidence

Clinical literature review conducted to support safety and efficacy for expanded indications (essential tremor, pituitary adenoma, glioblastoma). Scientific advisors confirmed extensive publications supporting clinical use based on experience with previous system models.

Technological Characteristics

Teletherapy system; 192 Cobalt-60 sealed sources; stereotactic irradiation; patient positioning system; operator console with computer; radiation unit with integrated control/power units.

Indications for Use

Indicated for stereotactic irradiation of intracranial structures including metastatic tumors, arteriovenous malformations, trigeminal neuralgia, medically refractory essential tremor, meningiomas, post-surgical pituitary adenomas, and recurrent glioblastomas.

Regulatory Classification

Identification

A radionuclide radiation therapy system is a device intended to permit an operator to administer gamma radiation therapy, with the radiation source located at a distance from the patient's body. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts (including beam-limiting devices), and accessories.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ MY 0 9 2014 K133565 Page 1 of 3 Dokumentnamn/Name of document ELEKTA INSTRUMENT AB Traditional 510(k) | Utfärdare/Issuer | Matilda Forsberg | |-----------------------------|-------------------------------| | Ref nr/Dok nr/Ref no/Doc no | -- | | Utgava /Edition | -- | | Avser/Regarding | Leksell Gamma Knife Perfexion | | Directory | -- | # Section 4- 510(k) Summary As Required by 21 CFR 807.92(c) 510 (k) Summary ### Subscribers Name & Address 1. Elekta Instrument AB Kungstensgatan 18, P.O. Box 7593 SE-103 93 Stockholm, Sweden Tel: (011) 46 8 587 254 00 Fax: (011) 46 8 587 255 00 Official Correspondent: Ms Louise Wachtmeister Date summary prepared: 2013-11-14 ### 2. Trade Name Leksell Gamma Knife® Perfexion™ ### Device Classification 3. | Common Name | Product Code | Class | Regulation Number | |---------------------------------|--------------|-------|-------------------| | Leksell Gamma Knife® Perfexion™ | IWB | II | 21 CFR 892.5750 | ### Predicate Device Identification 4. | Legally marketed devices to which equivalence is being claimed | 510(k) # | |----------------------------------------------------------------|----------| | Leksell Gamma Knife® Perfexion ™™ | K120811 | ### 5. Other relevant submissions | Devices | 510(k) # | |-----------------------------|----------| | Leksell Stereotactic System | K080355 | | Leksell GammaPlan® PFX | K103093 | {1}------------------------------------------------ # ELEKTA INSTRUMENT AB Dokumentnamn/Name of document Traditional 510(k) | Utfärdare/Issuer<br>Matilda Forsberg | Ref nr/Dok nr/Ref no/Doc no<br>-- | Utgåva /Edition<br>-- | |---------------------------------------------------|-----------------------------------|-----------------------| | Avser/Regarding:<br>Leksell Gamma Knife Perfexion | -- | Directory<br>-- | ### Device Description 6. Leksell Gamma Knife Perfexion is a radiosurgery system for use in the stereotactic irradiation of intra-cranial structures. Surgery is achieved by delivering a prescribed dose as one or more shots of ionizing radiation to the exact site of the target. Based on preoperative radiological examinations, the Leksell Gamma Knife Perfexion unit provides highly accurate external irradiation of intra-cranial structures using collimated beams of ionizing radiation. Each unit contains 192 sealed sources of Cobalt 60. Leksell Gamma Knife Perfexion systems consists of several units, physically separated in an Office side in the control room and a Medical side in the treatment room. - The medical side consists basically of the radiation unit with a patient positioning . system. The system control and power units are placed inside the cover of the radiation unit. - The office side consists of the operator console with control panel and office computer. . #### 7. Intended Use Leksell Gamma Knife® Perfexion™ is a teletherapy device intended for stereotactic irradiation of head structures ranging from very small target sizes of a few millimeters to several centimeters. Indications include but are not limited to; metastatic tumors, arteriovenous malformations, trigeminal neuralgia, essential tremor, meningiomas, vestibular schwannomas, pituitary adenomas and glioblastoma. #### 8. Nonclinical testing Non-clincal testing was not required to support substantial equivalence with the predicate device. The fundamental technical characteristics are the same to those of the predicate device. A risk analysis review was performed to check the impact from the new intended use. The investigation resulted in no changes to any assessments regarding the safety of the Leksell Gamma Knife® Perfexion. The conclusion was that the Leksell Gamma Knife® Perfexion is as safe and as effective as its predicate for the new indications. {2}------------------------------------------------ # A INSTRUMENT AR ELEKTA INSTRUMENT AB Traditional 510(k) | Utfärdare/Issuer | Matilda Forsberg | |-----------------------------|-------------------------------| | Ref nr/Dok nr/Ref no/Doc no | -- | | Utgåva /Edition | -- | | Avser/Regarding | Leksell Gamma Knife Perfexion | | | Directory | | | -- | Dokumentnamn/Name of docume ### 9. Summary of clinical tests A clinical literature review was conducted to demonstrate that the Leksell Gamma Knife® Perfexion is safe and effective for the new indications added to the intended use. Reports have been approved by physicians regarding clinical relevance, based on experience of previous models of the system. The scientific advisors found extensive scientific publications supporting the safety and efficacy of using Leksell Gamma Knife Perfexion for the added indications (essential tremor, pituitary and glioblastoma). ### 10. Substantial Equivalence The functionality for the Leksell Gamma Knife Perfexion is equivalent to its predicate device K120811 in safety and effectiveness and the fundamental technical characteristics are the same to those of the predicate device. {3}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows a circular logo with text around the perimeter and a stylized symbol in the center. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA". The central symbol resembles a stylized bird or abstract shape with three curved lines forming its body and wing. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 9, 2014 Elekta Instrument AB % Ms. Matilda Forsberg Regulatory Affairs Engineer Kungstensgatan 18 SE-103 93 Stockholm SWEDEN . Re: K133565 Trade/Device Name: Leksell Gamma Knife® Perfexion™ Regulation Number: 21 CFR 892.5750 Regulation Name: Radionuclide radiation therapy system Regulatory Class: II Product Code: IWB Dated: April 16, 2014 Received: April 28, 2014 Dear Ms. Forsberg: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the cnacument date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ Page 2-Ms. Forsberg If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Smh.7) for Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page. 510(k) Number (if known) K133565 Device Name Leksell Gamma Knife Perfexion ### Indications for Use (Describe) Leksell Gamma Knife® Perfexion™ is a teletherapy device intended for stereotactic irradiation of head structures ranging from very small target sizes of a few millimeters to several centimeters e.g. metastatic tumors, arteriovenous malformations, trigeminal neuralgia, medically refractory essential tremor, meningiomas, post-surgical pituitary adenomas and recurrent glioblastomas. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C) ਂ ਦ ## PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED. | | FOR FDA USE ONLY | |--------------------------------------------|------------------| | | | | Call Concession Company Control Concession | | Concurrence of Center for Devices and Radiological Health (CDRH) (Signalure) h.7) FORM FDA 3881 (6/13) ( {6}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...