LEKSELL GAMMA KNIFE TARGET SYSTEM, MODEL 24001

K984328 · Elekta Instrument AB · IWB · May 21, 1999 · Radiology

Device Facts

Record IDK984328
Device NameLEKSELL GAMMA KNIFE TARGET SYSTEM, MODEL 24001
ApplicantElekta Instrument AB
Product CodeIWB · Radiology
Decision DateMay 21, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5750
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Leksell® Gamma Knife Target System Model 24001 is a teletherapy device indicated for use in the stereotactic irradiation of intracranial structures

Device Story

Leksell® Gamma Knife Target System Model 24001 is an upgraded stereotactic radiosurgery device for intracranial structures. It utilizes Cobalt-60 sources through 201 collimators to deliver radiation. The system integrates a Computerized Control System, Automatic Positioning System, patient couch, radiation unit, and patient surveillance system. Patient head is fixated via Leksell Stereotactic Coordinate Frame. The device receives treatment planning data from Leksell® Gamma Plan software. Operated by clinical staff, the system automates positioning and radiation delivery based on pre-planned coordinates. Output is precise, targeted radiation to intracranial lesions. Benefits include accurate, repeatable stereotactic irradiation for therapeutic purposes.

Clinical Evidence

Bench testing only. No clinical data provided. Performance was validated through system and subsystem-level testing, including software and hardware integration testing, to demonstrate accuracy and repeatability.

Technological Characteristics

Teletherapy device using Cobalt-60 radiation source with 201 collimators. Components include Computerized Control System, Automatic Positioning System, couch, radiation unit, and patient surveillance system. Patient fixation via Leksell Stereotactic Coordinate Frame. Receives treatment planning data from Leksell® Gamma Plan software. System is an upgrade to the predicate Model 23004.

Indications for Use

Indicated for stereotactic irradiation of intracranial structures in patients requiring such therapy.

Regulatory Classification

Identification

A radionuclide radiation therapy system is a device intended to permit an operator to administer gamma radiation therapy, with the radiation source located at a distance from the patient's body. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts (including beam-limiting devices), and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ MAY 2 1 1999 K984328 # SUMMARY OF SAFETY AND EFFECTIVENESS SECTION 19: This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and CFR 807.92. ## SUBMITTER INFORMATION 19.1 | a. | Company Name: | Elekta Instruments, AB | |----|--------------------------------|---------------------------------------------------------------------------------------------| | b. | Company Address: | Birger Jarlsgatan 53, S-103, 97<br>Stockholm, Sweden | | c. | Company Phone:<br>Company Fax: | (011) 46 8 5872 54 00<br>(011) 46 8 5872 55 00 | | d. | Contact Person: | Sverker Glans<br>Vice President<br>Quality and Regulatory Affairs<br>Elekta Instruments, AB | | e. | Date Summary Prepared: | November 20, 1998 | # 19.2. DEVICE IDENTIFICATION | a. | Trade/Proprietary Name: | Leksell® Gamma Knife Target System<br>Model 24001 | |----|-------------------------|----------------------------------------------------------| | b. | Classification Name: | Radionuclide Radiation Therapy Device<br>21 CFR 892.5750 | # IDENTIFICATION OF PREDICATE DEVICE 19.3 | Company | Device | 510(k) No. | Date Cleared | |--------------------|-------------------------------------------|------------|-------------------| | Elekta Instruments | Leksell Gamma Unit<br>Model 23004, Type B | K924849 | November 20, 1995 | {1}------------------------------------------------ #### 19.4 DEVICE DESCRIPTION The Leksell® Gamma Knife Target System is an upgraded version of the currently available Leksell® Gamma Knife device. The device is intended for the stereotactic irradiation of intracranial structures and is composed of seven basic components. The Leksell® Gamma Knife Target System receives data from the Leksell Gamma Plan software. The Target System is composed of a Computerized Control System, Automatic Positioning System, Couch, Radiation Unit and Patient Surveillance System. The patient's head is fixated by the Leksell Stereotactic Coordinate Frame. #### 19.5 SUBSTANTIAL EQUIVALENCE The Leksell® Gamma Knife Target System Model 24001 is substantially equivalent to previous version of the Leksell® Gamma Knife Model 23004 currently in commercial distribution by Elekta Instruments. The Leksell® Gamma Knife Target System and the predicate device are both indicated for the stereotactic irradiation of intracranial structures. The fundamental technical characteristics are similar to those of the predicate devices and are listed on the comparison charts provided in this 510(k) submission. #### 19.6 INTENDED USE The Elekta Instruments Leksell® Gamma Knife Target System Model 24001 is intended for the stereotactic irradiation of intracranial structures. {2}------------------------------------------------ #### 19.7 TECHNOLOGICAL CHARACTERISTICS A complete comparison of the technological characteristics of the predicate is provided within this submission. Both the predicate and Target System are composed of the same radiation unit using Cobalt 60 through 201 collimator sources. The Target System and the predicate receive treatment planning data from the Leksell® Gamma Plan software program. In both devices, the patient's head is fixated by the Leksell® Stereotactic Coordinate Frame. #### 19.8 PERFORMANCE DATA The Leksell® Gamma Knife Target System has been demonstrated to perform as intended with accuracy and repeatability. The Leksell® Gamma Knife Target System has been tested on the system and subsystem level. Through testing of the software and hardware components of the device have also been completed. Integration testing and complete system testing have also been performed. Results of performance testing, software and hardware testing of the Leksell® Gamma Knife Target System have been included in Sections 13, 15, and 16 of this submission. ### 19.9 510(K) CHECKLIST This notification contains all information required by 21 CFR 807.87. A completed copy of the Premarket Notification 510(k) Reviewer's Checklist is provided in this submission. {3}------------------------------------------------ Rockville MD 20850 Food and Drug Administration 9200 Corporate Boulevard Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three stylized human figures or birds in flight. MAY 2 1 1999 Elektra Instruments AB C/O Carol Patterson Consultant 18140 Smokesignal Drive San Diego, California 92127 RE: K984328 Gamma Knife: Dated: May 17, 1999 Received: May 18, 1999 Regulatory Class: II 21 CFR 892.5750/Procode: 90 IWB Dear Ms. Patterson: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any oblication you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # INDICATIONS FOR USE | 510(k) Number: | K984328<br> | |-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name: | Elekta Leksell® Gamma Knife Target System Model 24001 | | Indications for Use: | The Leksell® Gamma Knife Target System Model 24001 is<br>a teletherapy device indicated for use in the stereotactic<br>irradiation of intracranial structures | | To Be Assigned By FDA | | PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) David h. Segerson (Division Sign-Off) (Division of Reproductive, Abdominal, ENT, and Radiological Deyig 510(k) Number Prescription Use OR Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801.109)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...