Non-inferiority demonstrated via NMI and HD values
—
—
Multi-modality image sets from different patients across five major scanner vendors
—
Radiotherapy dose calculation
Monte Carlo method
—
Validated for accuracy in clinically relevant settings
—
—
18 patients' treatment planning including lung, liver, brain, and Head & Neck cancers
—
Indications for Use
The MOZI Treatment Planning System (MOZI TPS) is used to plan radiotherapy treatments with malignant or benign diseases. MOZI TPS is used to plan external beam irradiation with photon beams.
Device Story
MOZI TPS is a standalone software platform for radiotherapy treatment planning. It processes medical images (CT) to perform structure delineation, plan design, optimization, and evaluation. Inputs include patient imaging data; the system utilizes deep learning for automatic contouring and Monte Carlo methods for dose calculation. It supports rigid and deformable image registration. Operated by trained medical professionals in clinical settings, the software provides tools for IMRT and VMAT planning. Output includes treatment plans and dose distributions, which assist clinicians in designing and simulating radiation therapy. The device benefits patients by enabling precise, optimized radiation delivery for malignant or benign conditions.
Clinical Evidence
No clinical trials performed. Evidence consists of bench testing, including software verification/validation, commissioning per AAPM TG-119, and end-to-end testing on 18 patient cases. Automatic segmentation performance was validated against reference device K191928 using 560 training and 187 testing image sets, reporting Mean Dice Similarity Coefficient (DSC) values across various anatomical structures (e.g., Lung 0.99, Heart 0.98, Brainstem 0.88). Deformable registration was validated against reference device K182624 using NMI and HD metrics, showing non-inferiority.
Technological Characteristics
Standalone software; Windows-based. Features: Monte Carlo dose calculation (1-20 MeV), deep learning-based automatic segmentation, intensity-based rigid/deformable registration. Supports DICOM RT standards (Image, Structure, Plan, Dose). Complies with IEC 62083-2009, IEC 62304, and ISO 14971. No patient contact; no sterilization or biocompatibility requirements.
Indications for Use
Indicated for patients with malignant or benign diseases requiring external beam photon radiation therapy planning.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Predicate Devices
Eclipse Treatment Planning System (Eclipse TPS) (K172163)
{0}------------------------------------------------
Manteia Technologies Co., Ltd. Chao Fang RA 1903-1904, B Tower, Zijin Plaza, No.1811 Huandao East Road Xiamen. Fujian 361001 CHINA
Re: K223724
Trade/Device Name: MOZI TPS Regulation Number: 21 CFR 892.5050 Regulation Name: Medical Charged-Particle Radiation Therapy System Regulatory Class: Class II Product Code: MUJ Dated: January 1, 2023 Received: January 3, 2023
Dear Mr. Chao Fang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
July 10, 2023
Image /page/0/Picture/9 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/5 description: The image shows a digital signature. The signature is for Lora D. Weidner. The signature was created on July 10, 2023 at 17:31:17 -04'00'.
Lora D. Weidner, Ph.D. Assistant Director Radiation Therapy Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223724
Device Name MOZI TPS
Indications for Use (Describe)
The MOZI Treatment Planning System (MOZI TPS) is used to plan radiotherapy treatments with malignant or benign diseases. MOZI TPS is used to plan external beam irradiation with photon beams.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
The following information is provided as required by 21 CFR 807.92. The assign 510(k) Number: K223724
## 1. SUBMITTER
| Name and Address: | Manteia Technologies Co., Ltd. |
|---------------------------------------|----------------------------------|
| | 1903-1904, B Tower, Zijin Plaza, |
| | No.1811 Huandao East Road, |
| | Xiamen, China |
| Establishment Registration<br>Number: | 3016686005 |
| Contact Person: | Chao Fang |
| | RA&QA Manager |
| | Email: fangchao@manteiatech.com |
| Date of Preparation: | July 9, 2023 |
## 2. DEVICE
| Device/Trade Name: | MOZI TPS |
|-------------------------|---------------------------------------------------|
| Common Name: | MOZI Treatment Planning System (MOZI TPS) |
| Product Classification: | Class II |
| Classification Name: | System, Planning, Radiation Therapy Treatment |
| Product Code: | MUJ |
| Regulation Number: | 21CFR 892.5050 |
| Regulation Description: | Medical charged-particle radiation therapy system |
### 3. PREDICATE DEVICE
| Predicate Device: | K172163 (Eclipse Treatment Planning System (Eclipse TPS)) |
|---------------------|-----------------------------------------------------------|
| Reference Device 1: | K191928 (AccuContour™) |
| Reference Device 2: | K182624 (MIM-MRT Dosimetry) |
# 4. Device Description:
The proposed device, MOZI Treatment Planning System (MOZI TPS), is a standalone software which is used to plan radiotherapy treatments (RT) for patients with malignant or benign diseases.
Its core functions include image processing, structure delineation, plan design, optimization and evaluation. Other functions include user login, graphical interface, system and patient management. It can provide a platform for completing the related work of the whole RT plan.
The product uses generally accepted methods for below automatic function:
{4}------------------------------------------------
> automatic rigid and deformable registration of image processing
- > dose calculation with Monte Carlo method of plan design and optimization
And it also uses deep learning method for automatic contouring of structure delineation.
### 5. INDICATIONS FOR USE
The MOZI Treatment Planning System (MOZI TPS) is used to plan radiotherapy treatments for patients with malignant or benign diseases. MOZI TPS is used to plan external beam irradiation with photon beams.
## 6. SUBSTANTIALLY EQUIVALENT(SE) COMPARISON
The Subject Device, MOZI TPS, makes use of a Predicate Device, Eclipse TPS(K172163), as the Predicate Device for substantial equivalence comparison, and Reference Device 1, AccuContour™(K191928), and Reference Device 2, MIM-MRT Dosimetry(K182624), as Reference Device for performance comparison on automatic function of rigid registration, deformable registration and contouring. In addition, the subject device adopts the same algorithm as AccuContour™(K191928) on automatic function of rigid registration and contouring.
At a high level, both the predicate device and the subject device are based on the same characteristics:
· Both the subject device and the predicate provide software tools for planning the treatment of malignant or benign diseases with radiation.
· They are computer-based software devices used by trained medical professionals to design and simulate radiation therapy treatments.
- · They are both capable of planning treatments for external beam irradiation with photon.
The significant differences compared with the predicate device are as follows:
- 1. The intended use range of subject device is less than predicate device, which has been clearly stated in the intended use of subject device.
- 2. For dose calculation, the subject device incorporates a Monte Carlo based algorithm while the predicate device an AAA/AXB based algorithm. Both have been verified to be substantially equivalent by a performance test.
- 3. Performance test report on automatic rigid registration and contouring has been performed in subject device and reference device 1, AccuContour™(K191928). The test result is acceptable.
- 4. Performance test on automatic deformable has been performed in subject device and reference device 2, MIM-MRT Dosimetry(K182624). The test result is acceptable.
The following comparison table "Device Comparison Table" provides a detailed comparison.
{5}------------------------------------------------
| 510(k) Summary |
|----------------|
| |
## Device Comparison Table
| ITEM | Subject Device | Predicate Device<br>K172163 | Reference Device<br>K191928 | Reference Device<br>K182624 |
|----------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulatory Information | | | | |
| Regulation No. | 21CFR 892.2050 | 21CFR 892.2050 | 21CFR 892.2050 | 21CFR 892.2050 |
| Product Code | MUJ | MUJ | QKB | LLZ |
| Class | II | II | II | II |
| Indication for Use | The MOZI Treatment<br>Planning System (MOZI<br>TPS) is used to plan<br>radiotherapy treatments<br>for patients with<br>malignant or benign<br>diseases. MOZI TPS is<br>used to plan external<br>beam irradiation with<br>photon beams. | The Eclipse Treatment<br>Planning System (Eclipse<br>TPS) is used to plan<br>radiotherapy treatments<br>for patients with<br>malignant or benign<br>diseases. Eclipse TPS is<br>used to plan external<br>beam irradiation with<br>photon, electron and<br>proton beams, as well as<br>for internal Irradiation<br>(brachytherapy)<br>treatments.<br>In addition, the Eclipse<br>Proton Eye algorithm is<br>specifically indicated for<br>planning proton treatment<br>of neoplasms of the eye. | It is used by radiation<br>oncology department to<br>register multi-modality<br>images and segment<br>(non-contrast) CT<br>images, to generate<br>needed information for<br>Treatment planning,<br>treatment evaluation and<br>treatment adaptation. | MIM software is used by trained<br>medical professionals as a tool to aid<br>in evaluation and Information<br>management of digital medical<br>images. The medical image<br>modalities include, but are not limited<br>to, CT, MRI, CR, DX, MG, US,<br>SPECT, PET and XA as supported by<br>ACR/NEMA DICOM 3.0. MIM<br>assists in the following indications:<br>- Receive, transmit, store,<br>retrieve, display, print, and<br>process medical images and<br>DICOM objects.<br>- Create, display and print reports<br>from medical images.<br>- Registration, fusion display, and<br>review of medical images for<br>diagnosis, treatment evaluation, |
| | | | | and treatment planning.<br>• Evaluation of cardiac left<br>ventricular function and<br>perfusion.<br>• Localization and definition of<br>objects such as tumors and<br>normal tissues in medical<br>images. |
| Label/labeling | Conform with 21CFR<br>Part 801 | Conform with 21CFR<br>Part 801 | Conform with 21CFR<br>Part 801 | Conform with 21CFR<br>Part 801 |
| Operating System | Windows | Windows | Windows | Windows and MAC |
| Technological Characteristics | | | | |
| Dose calculation algorithms | Monte Carlo (photon) | AAA、AXB | N/A | N/A |
| Auto rigid registration algorithms | Intensity based | Intensity based | Intensity based | Intensity based |
| Auto deformable registration algorithms | Intensity based | Intensity based | Intensity based | Intensity based |
| Auto segmentation algorithms | Deep learning | Atlas based | Deep learning | Atlas based |
| Graphical User Interface Features | | | | |
| Multiple-instance application | Yes | Yes | N/A | N/A |
| Multiple-workspace layout | Yes | Yes | N/A | N/A |
| Graphical display/editing of field<br>parameters | Yes | Yes | N/A | N/A |
| Beam's-Eye-View display | Yes | Yes | N/A | N/A |
| 3D patient image display | Yes | Yes | N/A | N/A |
| Dose-Volume Histogram display | Yes | Yes | N/A | N/A |
| Patient Management Features | | | | |
| DICOM RT | Yes | Yes | N/A | N/A |
| (the device supports RT<br>Image/ RT Structure/ RT<br>Plan/ RT Dose) | | (the device supports RT<br>Image/ RT Structure/ RT<br>Plan/ RT Dose) | | |
| Import/Delete/Export/Edit/Retrieve<br>patient data | Yes | Yes | N/A | N/A |
| Image Processing Features | | | | |
| CT/MR manual image registration | Yes | Yes | Yes | Yes |
| Auto rigid registration | Yes (Intensity Based) | Yes (Intensity Based) | Yes (Intensity Based) | Yes (Intensity Based) |
| Auto deformable registration | Yes (Intensity Based) | Yes (Intensity Based) | Yes (Intensity Based) | Yes (Intensity Based) |
| Structure Delineation Features | | | | |
| Automatic CT segmentation tool | Yes (deep learning) | Yes (Atlas based) | Yes (deep learning) | Yes (Atlas based) |
| Manual CT segmentation tool | Yes | Yes | Yes | Yes |
| Plan Design & Plan Optimization Features | | | | |
| Photon Calculation | Yes | Yes | N/A | N/A |
| - Energy Range | 1 MeV – 20 MeV | 1 MeV – 50 MeV | N/A | N/A |
| - 3-dimensional conformal radiation<br>therapy Planning | Yes | Yes | N/A | N/A |
| - Intensity modulated radiotherapy<br>Planning | Yes | Yes | N/A | N/A |
| - Volumetric modulated arc therapy<br>Planning | Yes | Yes | N/A | N/A |
| - Directional heterogeneity<br>correction | Yes | Yes | N/A | N/A |
| - Treatment Head modelling | Yes | Yes | N/A | N/A…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.