Segmentation performance test: images from healthcare institutions in China and US covering three major vendors (GE, Siemens, Philips)
>1 (licensed physicians)
Multimodality image registration
—
—
Non-inferior to predicate device (NMI)
—
—
Registration performance test: images from healthcare institutions in China and US covering three major vendors (GE, Siemens, Philips)
—
Indications for Use
It is used by radiation oncology department to register multimodality images and segment (non-contrast) CT images, to generate needed information for treatment planning, treatment evaluation and treatment adaptation.
Device Story
AccuContour is standalone software for radiation oncology departments. Inputs: medical images and DICOM data (specifically non-contrast CT for segmentation; CT, MRI, PET for registration). Processing: utilizes deep learning for automatic contouring of organs-at-risk (head, neck, thorax, abdomen, pelvis) and intensity-based algorithms for image registration. Outputs: segmented images and registered multimodality image sets. Workflow: clinicians use software to receive, review, and edit images/contours to support treatment planning, evaluation, and adaptation. Benefits: automates time-consuming contouring and registration tasks, facilitating efficient radiation therapy workflows.
Clinical Evidence
No clinical data. Bench testing only. Segmentation performance evaluated using DICE similarity coefficients (DSC) on CT images from US and China across three major scanner vendors. Registration performance evaluated using Normalized Mutual Information (NMI) on multi-modality image sets (CT/CT, CT/MR, CT/PET). Results demonstrated non-inferiority to the predicate device.
Technological Characteristics
Standalone software running on Windows OS. Compatible with DICOM 3.0 compliant medical images. Segmentation: Deep Learning-based. Registration: Intensity-based. No specific hardware requirements beyond DICOM 3.0 compliance for scanners and treatment planning systems.
Indications for Use
Indicated for use by radiation oncology departments to register multimodality images and segment non-contrast CT images for treatment planning, evaluation, and adaptation.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Xiamen Manteia Technology LTD. % Ms. Diana Hong General Manager Mid-Link Consulting Co. Ltd. P.O Box 120-119 Shanghai, 200120 CHINA
February 28, 2020
## Re: K191928
Trade/Device Name: AccuContour™ Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: QKB Dated: January 10, 2020 Received: January 17, 2020
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191928
Device Name AccuContour
Indications for Use (Describe)
It is used by radiation oncology department to register multimodality images and segment (non-contrast) CT images, to generate needed information for treatment planning, treatment evaluation and treatment adaptation.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number: K191928
- 1. Date of Preparation: 02/26/2020
- 2. Sponsor Identification
## Xiamen Manteia Technology LTD.
1903, B Tower, Zijin Plaza, No.1811 Huandao East Road, Xiamen, China
Establishment Registration Number: Not registered yet.
Contact Person: Lu Xie Position: Research Management Tel: +86 592-6100813 Email: xielu@manteiatech.com
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Mr. Lee Fu (Alternative Contact Person)
### Mid-Link Consulting Co., Ltd
P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: +1-360-925-3199 Email: info@mid-link.net
{4}------------------------------------------------
#### 4. Identification of Proposed Device
Trade Name: AccuContour™ Common Name: Medical Imaging Software
### Regulatory Information
Classification Name: System, Imaging processing, Radiological Classification: II; Product Code: QKB Regulation Number: 21CFR 892.2050 Review Panel: Radiology;
Indication for Use Statement:
It is used by radiation oncology department to register multimodality images and segment (non-contrast) CT images, to generate needed information for treatment evaluation and treatment adaptation.
### Device Description:
The proposed device, AccuContour™, is a standalone software which is used by radiation oncology department to register multimodality images and segment (non-contrast) CT images, to generate needed information for treatment planning, treatment evaluation and treatment adaptation.
The product has two image process functions:
- (1) Deep learning contouring: it can automatically contour the organ-at-risk, including head and neck, thorax, abdomen and pelvis (for both male and female),
- (2) Automatic Registration, and
- (3) Manual Contour.
It also has the following general functions:
- > Receive, add/edit/delete, transmit, input/export, medical images and DICOM data;
- A Patient management;
- > Review of processed images;
- > Open and Save of files.
{5}------------------------------------------------
- ર. Identification of Predicate Device
510(k) Number: K182624 Product Name: MIM-MRT Dosimetry
- 6. Non-Clinical Test Conclusion
### Segmentation performance test
The segmentation performance test was performed on proposed device and predicate device to evaluate the automated segmentation accuracy. Two separate tests were performed. One test involved images generated in healthcare institutions in China using scanner models available in China covering three major vendors. The other involved images generated in healthcare institutions in US using scanner models available in US covering three major vendors. The three major vendors were GE, Siemens and Philips. For each body parts, all intended organs were included in images of the US and China. Ground truthing of each image was generated from the consensus of at least three licensed physicians. DICE similarity coefficients (DSC) was used for evaluation. DSC values were calculated on two sets of images for test group and control group, respectively. According to the results, it could be concluded that the DSC of proposed device was non-inferiority compared with that of the predicate device.
### Registration performance test
The registration performance test was performed on proposed device and predicate device to evaluate the automated registration accuracy. Two separate tests were performed. One test involved images generated in healthcare institutions in China using scanner models available in China covering three major vendors. And the image registration feature is tested on multi-modality image sets from same patients. The other involved most images generated in healthcare institutions in U.S. All fixed image and most moving images came from U.S and a small amount of moving images adopted from online database were originally from non-US sources. All the scanner models covered three major vendors. And the image registration feature is only tested on multi-modality image sets from different patients. Both tests covered various modalities, including CT/CT, CT/MR and CT/PET. The Normalized Mutual Information (NMI) was used for evaluation. NMI values were calculated on two sets of images for both the proposed device and predicate device, respectively. The NMI value of proposed device was compared with that of the predicate device. According to the results, it could be concluded that the NMI of proposed device was non-inferiority compared with that of the predicate device.
#### 7. Clinical Test Conclusion
No clinical study is included in this submission.
{6}------------------------------------------------
#### 8. Substantially Equivalent (SE) Comparison
| ITEM | Proposed Device | Predicate Device K182624 |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulatory Information | | |
| Regulation No. | 21CFR 892.2050 | 21CFR 892.2050 |
| Product Code | QKB | LLZ |
| Class | II | II |
| Indication for Use | It is used by radiation oncology department to register multimodality images and segment (non-contrast) CT images, to generate needed information for treatment planning, treatment evaluation and treatment adaptation. | MIM software is used by trained medical professionals as a tool to aid in evaluation and information management of digital medical images. The medical image modalities include, but are not limited to, CT, MRI, CR, DX, MG, US, SPECT, PET and XA as supported by ACR/NEMA DICOM 3.0. MIM assists in the following indications:<br>• Receive, transmit, store, retrieve, display, print, and process medical images and DICOM objects.<br>• Create, display and print reports from medical images.<br>• Registration, fusion display, and review of medical images for diagnosis, treatment evaluation, and treatment planning.<br>• Evaluation of cardiac left ventricular function and perfusion, including left ventricular enddiastolic volume, end-systolic volume, and ejection fraction.<br>• Localization and definition of objects such as tumors and normal tissues in medical images.<br>• Creation, transformation, and modification of contours for applications including, but not limited to, quantitative analysis, aiding adaptive therapy, transferring contours to radiation therapy treatment planning systems and |
| | | | Table 1 Comparison of Technology Characteristics<br>J | |
|--|--|--|-------------------------------------------------------|--|
|--|--|--|-------------------------------------------------------|--|
{7}------------------------------------------------
| ITEM | Proposed Device | Predicate Device K182624 |
|--------------------------------------------|-------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | archiving contours for patient follow-up<br>and management. |
| | | • Quantitative and statistical analysis of<br>PET/SPECT brain scans by comparing to<br>other registered PET/SPECT brain scans. |
| | | • Planning and evaluation of permanent<br>implant brachytherapy procedures (not<br>including radioactive microspheres). |
| | | • Calculating absorbed radiation dose as<br>a result of administering a radionuclide. |
| | | When using device clinically, the user<br>should only use FDA approved<br>radiopharmaceuticals. If using with<br>unapproved ones, this device should only<br>be used for research purposes. |
| | | Lossy compressed mammographic<br>images and digitized film screen images<br>must not be reviewed for primary image<br>interpretations. Images that are printed to<br>film must be printed using an<br>FDA-approved printer for the diagnosis<br>of digital mammography images.<br>Mammographic images must be viewed<br>on a display system that has been cleared<br>by the FDA for the diagnosis of digital<br>mammography images. The software is<br>not to be used for mammography CAD. |
| Label/labeling | Conform with 21CFR Part 801 | Conform with 21CFR Part 801 |
| Operating System | Windows | Windows and MAC system |
| Segmentation Features | | |
| Algorithm | Deep Learning | Atlas-based |
| Compatible<br>Modality | Non-Contrast CT | Non-Contrast CT |
| Compatible<br>Scanner Models | No Limitation on scanner model,<br>DICOM 3.0 compliance required. | No Limitation on scanner model,<br>DICOM3.0 compliance required. |
| Compatible<br>Treatment<br>Planning System | No Limitation on TPS model, DICOM<br>3.0 compliance required. | No Limitation on TPS model, DICOM<br>3.0 compliance required. |
| Contraindications | None | None |
| ITEM | Proposed Device | Predicate Device K182624 |
| Registration Features | | |
| Algorithm | Intensity Based | Intensity Based |
| Compatible<br>Modality | CT, MRI, PET | CT, MRI, CR, DX, MG, US, SPECT,<br>PET and XA |
| Compatible<br>Scanner Models | No Limitation on scanner model,<br>DICOM 3.0 compliance required. | No Limitation on scanner model,<br>DICOM3.0 compliance required. |
| Compatible<br>Treatment<br>Planning System | No Limitation on TPS model, DICOM<br>3.0 compliance required. | No Limitation on TPS model, DICOM<br>3.0 compliance required. |
{8}------------------------------------------------
- 9. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.