syngo VSim

K151887 · Siemens Medical Solutions USA, Inc. · MUJ · Sep 16, 2015 · Radiology

Device Facts

Record IDK151887
Device Namesyngo VSim
ApplicantSiemens Medical Solutions USA, Inc.
Product CodeMUJ · Radiology
Decision DateSep 16, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5050
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

V Sim is a software solution to prepare anatomical data (CT) for the planning of external beam radiation therapy treatment. V Sim provides features to define anatomical regions, reference points and geometric treatment plans prior to dosimetric planning with a treatment planning system.

Device Story

syngo VSim is a 3D post-processing software application for radiation therapy planning. It takes CT images as input; allows clinicians to perform segmentation and contouring of target volumes and organs at risk; manages reference points; and enables placement of geometric treatment beams. The device outputs DICOM-RT Structures and DICOM-RT Plans for use in downstream dosimetric treatment planning systems. It operates on a syngo-based workspace. The software assists radiation oncologists or dosimetrists in visualizing soft-tissue structures and defining treatment borders, facilitating more accurate radiation beam placement compared to conventional X-ray imaging. It is intended for use in clinical settings where radiation therapy planning is performed.

Clinical Evidence

No clinical data. The device underwent non-clinical bench testing, including integration and functional testing, to verify software performance against specifications. Risk management and hazard analysis were performed. Testing confirmed the device performs as intended and is comparable to the predicate.

Technological Characteristics

Software-based 3D post-processing application. Operates on syngo-based workspace. Connectivity via DICOM (PS 3.1-3.18). Adheres to IEC 62304 (software life cycle), ISO 14971 (risk management), IEC 62366 (usability), and IEC 61217 (radiotherapy coordinates).

Indications for Use

Indicated for preparation of anatomical CT data for external beam radiation therapy treatment planning, including definition of anatomical regions, reference points, and geometric treatment plans.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The caduceus is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the logo. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 16, 2015 Siemens Medical Solutions USA, Inc. % Ms. Kimberly Mangum Regulatory Affairs Specialist 51 Valley Stream Parkway MALVERN PA 19355 Re: K151887 Trade/Device Name: syngo VSim Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: MUJ Dated: June 15, 2015 Received: July 10, 2015 Dear Ms. Mangum: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Michael D'Hara For Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K151887 Device Name syngo VSim Indications for Use (Describe) V Sim is a software solution to prepare anatomical data (CT) for the planning of external beam radiation therapy treatment. V Sim provides features to define anatomical regions, reference points and geometric treatment plans prior to dosimetric planning with a treatment planning system. | Type of Use (Select one or both, as applicable) | | |-----------------------------------------------------------------|----------------------------------------------------------------| | <div> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## SIEMENS #### 510(k) Summary FOR syngo VSim Submitted by: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355 Date Prepared: September 2, 2015 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92. #### 1. General Information Importer/Distributor Establishment: Registration No: 2240869 Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355 ### Manufacturing Facility: Siemens AG Medical Solutions Doris-Ruppenstein Straße 4 D-91052 Erlangen, Germany Establishment Registration Number: 3004977335 ### 2. Contact Person: Kimberly Mangum Requlatory Affairs Specialist Siemens Medical Solutions. Inc. USA 40 Liberty Boulevard Malvern, PA 19355 Phone: (610) 448-6477 Fax: (610) 640-4481 Email: kimberly.mangum@siemens.com ### 3. Device Name and Classification Product Name: svnao VSim Propriety Trade Name: syngo VSim System, Planning, Radiation Therapy Treatment Classification Name: Classification Panel: Radiology 21 CFR §892.5050 CFR Section: Class II Device Class: Product Code: MUJ {4}------------------------------------------------ #### 4. Legally Marketed Primary Predicate Device: | Product Name: | syngo Dosimetrist Workspace v2.7 | |-----------------------|-----------------------------------------------| | Propriety Trade Name: | syngo Dosimetrist Workspace v2.7 | | Classification Name: | System, Planning, Radiation Therapy Treatment | | Classification Panel: | Radiology | | CFR Section: | 21 CFR § 892.5050 | | Device Class: | Class II | | Product Code: | MUJ | | 510(k) Number: | K101119 | #### 5. Indications for Use VSim is a software solution to prepare anatomical data (CT) for the planning of external beam radiation therapy treatment. VSim provides features to define anatomical regions, reference points and qeometric treatment plans prior to dosimetric planning with a treatment planning svstem. #### 6. Substantial Equivalence: The subject device syngo VSim is substantially equivalent to following medical devices in commercial distribution as listed in Table 1: #### Table 1: Predicate Devices | Manufacturer | Predicate Device | 510(k) | Clearance Date | |--------------|----------------------------------|---------|----------------| | Siemens | syngo Dosimetrist Workspace v2.7 | K101119 | June 16, 2010 | #### 7. Device Description: Virtual simulation, the most basic form of planning, allows more accurate placement of radiation beams than is possible using conventional X-ray imaging, where soft-tissue structures are often difficult to assess and normal tissues difficult to protect. The virtual simulation application syngo VSim allows the use of contouring and segmentation tools to assist in the drawing of the borders of the treatment volume and organ at risk. syngo VSim is a software application that runs on a syngo® based workspace. It is a 3D-post-processing software application that uses CT planning images as input and creates as output DICOM-RT Structures and DICOM RT Plans for treatment planning. syngo VSim is used to create 3D-models of target volumes and organs at risk. The main functionalities of syngo VSim are segmentation of organs, management of reference points and placement of geometrical treatment beams. {5}------------------------------------------------ # SIEMENS - 8. Summary of Technical Characteristics of the Subject Device as Compared with the Predicate Device: The subject device syngo VSim application does not differ in intended use from the 510(k) cleared predicate device syngo® Dosimetrist Workspace 2.7 (K101119, clearance date 06/16/2010). A comparison of the technical characteristics for the subject device and the predicate device is provided in Table 2 below. Predicate and Subject Device Comparable Technological Table 2: Characteristics | Property | Subject Device | Primary Predicate Device<br>K101119 | |---------------------|--------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------| | Indications for Use | Modified to reflect syngo VSim as a<br>standalone software device. There is no<br>change in the intended use from the<br>predicate device. | Indications for use reflects syngo<br>VSim as a component of syngo®<br>Dosimetrist Workspace 2.7 | | VSim software | Stand alone software device to be<br>marketed as syngo VSim. | Component of syngo® Dosimetrist<br>Workspace 2.7 | The basic operating principle remains unchanged from the predicate device. Furthermore, there are no changes in regard to the features of VSim (Segmentation, Reference Point Management, Simulation) compared to the predicate device. Testing and validation is completed. Test results show that the subject device, syngo VSim, is comparable to the VSim task card of the predicate device and therefore Siemens believes it substantially equivalent. #### 9. Nonclinical Testing: syngo VSim is designed to fulfill the requirements of the following safety and performance standards listed in Table 3 below: | Recognition<br>Number | Product<br>Area | Title of Standard | Reference<br>Number and Date | Publication<br>Date | Standards<br>Development<br>Organization | |-----------------------|-----------------|---------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|---------------------|------------------------------------------| | N/A | General | General Requirements<br>for Collateral Standard:<br>Programmable<br>Electrical Medical<br>Systems | IEC 60601-1-<br>4:2000, Consol.<br>Ed. 1.1, Medical<br>Electrical<br>Equipment - Part<br>1-4: | 09/08/2009 | IEC | | 12-217 | Radiology | Medical Electrical<br>Equipment -<br>Requirements for the | 62083 Edition 2.0<br>2009-09 | 03/18/2011 | IEC | | Table 3: Performance Standards | | |---------------------------------|--| | | | {6}------------------------------------------------ | Recognition<br>Number | Product<br>Area | Title of Standard | Reference<br>Number and Date | Publication<br>Date | Standards<br>Development<br>Organization | |-----------------------|-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|---------------------|------------------------------------------| | N/A | | Safety of Radiotherapy<br>Treatment Planning<br>Systems | | | | | N/A | Software | Medical device software<br>- Software life cycle<br>processes | IEC 62304 First<br>Edition 2006-05 | 09/09/2008 | IEC | | N/A | General | Medical devices -<br>Application of usability<br>engineering to medical<br>devices (General) | EN/IEC 62366 | 09/08/2009 | IEC | | N/A | Radiology | Digital Imaging and<br>Communications in<br>Medicine (DICOM) Set | PS 3.1 - 3.18 | 03/16/2012 | NEMA | | 5-40 | General | Medical devices -<br>Application of risk<br>management to medical<br>devices | 14971 Second<br>Edition 2007-03-<br>01 | 08/20/2012 | ISO | | N/A | General | Medical electrical<br>equipment -- Part 1-6:<br>General requirements<br>for basic safety and<br>essential performance --<br>Collateral Standard:<br>Usability | 60601-1-6:2004 | 1/30/2014 | IEC | | 12-267 | Radiology | Radiotherapy<br>equipment -<br>Coordinates,<br>movements, and scales | 61217 Edition 2.0<br>2011-12 | 01/30/2014 | IEC | This submission contains performance data to demonstrate continued conformance with special controls for medical devices containing software. Non clinical tests (integration and functional) were conducted for syngo VSim during product development. The modifications described in this Premarket Notification were supported with verification/validation testing. The risk analysis was completed and risk control implemented to mitigate identified hazards. The testing results support that all the software specifications have met the acceptance criteria. Testing for verification and validation of the device was found acceptable to support the claims of substantial equivalence. #### Software Verification and Validation Software Documentation for a Moderate Level of Concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for {7}------------------------------------------------ # SIEMENS Software Contained in Medical Devices" issued on May 11, 2005 is also included as part of this submission. #### Non-Clinical Testing Summary Performance tests were conducted to test the functionality of the syngo VSim. These tests have been performed to test the ability of the included features of the subiect device. The results of these tests demonstrate that the subject device performs as intended. The result of all conducted testing was found acceptable to support the claim of substantial equivalence. #### 10. General Safety and Effectiveness Concerns: The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device. Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled during development. verification and validation testing. To minimize electrical, mechanical, and radiation hazards, Siemens adheres to recognized and established industry practice and standards. #### 11. Conclusion as to Substantial Equivalence syngo VSim has the same intended use and comparable indications for use as the predicate device. The technical characteristics of the VSim component remain unchanged from the predicate device. Any technological differences between the subject device and predicate device do not raise different questions of safety or effectiveness. The predicate device was cleared based on non-clinical supportive information. The subject device non-clinical test data similarly supports the safety of the software with verification and validation testing. Verification and validation testing demonstrates that the subject device syngo VSim performs as intended. The non-clinical test data demonstrates that syngo VSim device performance is comparable to the predicate device that is currently marketed for the same intended use. In summary, Siemens is of the opinion that the syngo VSim does not introduce any new potential safety risk and is substantially equivalent to and performs as well as the predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...