K131891 · Varian Medical Systems, Inc. · MUJ · Sep 25, 2013 · Radiology
Device Facts
Record ID
K131891
Device Name
ECLIPSE TREATMENT PLANNING SYSTEM
Applicant
Varian Medical Systems, Inc.
Product Code
MUJ · Radiology
Decision Date
Sep 25, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Eclipse Treatment Planning System (Eclipse TPS) Is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (brachytherapy) treatments. In addition, the Eclipse Proton Eye algorithm is specifically indicated for planning proton treatment of neoplasms of the eye.
Device Story
Eclipse 12 is a software-based treatment planning system (TPS) for radiation therapy. It processes medical images to design and simulate external beam (photon, electron, proton) and internal (brachytherapy) radiation treatments. Operated by trained medical professionals in clinical settings, the system utilizes various dose calculation algorithms (e.g., AcurosXB, AcurosBV, PRO, BAO, DVO) to generate treatment plans. It integrates with hospital infrastructure via SQL Server, supporting RT prescription, plan validation, and status tracking. The software provides tools for image segmentation, dose-volume histogram (DVH) estimation, and modulated scanning for proton therapy. Clinicians use the output to finalize radiation delivery parameters, aiming to optimize dose distribution to target tissues while sparing healthy anatomy. The system supports remote deployment and utilizes the Eclipse Scripting API for workflow integration.
Clinical Evidence
No clinical data. Evidence consists of non-clinical verification and validation testing performed on a production-equivalent device under clinically representative conditions to confirm design input requirements and functional integrity.
Technological Characteristics
Software-based TPS; 64-bit architecture. Algorithms: AcurosXB, AcurosBV, PRO, BAO, DVO, and fluence calculation via LMC. Connectivity: SQL Server database, Eclipse Scripting API (ESAPI), DICOM-compatible. Features: Deformable registration (SmartAdapt), DVH estimation, modulated proton scanning (spot/line), and support for various linac platforms (TrueBeam, Elekta VMAT).
Indications for Use
Indicated for patients with malignant or benign diseases requiring radiotherapy. Used for planning external beam irradiation (photon, electron, proton) and internal irradiation (brachytherapy). Proton Eye algorithm indicated for proton treatment planning of ocular neoplasms.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{1}------------------------------------------------
## PREMARKET NOTIFICATION
### 510(k) Summary
Eclipse Treatment Planning System
As required by 21 CFR 807.92
Submitter's Name:
Varian Medical Systems 3100 Hansen Way, m/s E110 Palo Alto CA94304
Contact Name: Peter J. Coronado Phone: 650/424.6230 Fax:650/842.5040 Date: 21 June 2013
Eclipse Treatment Planning System
Proprietary Name:
Classification Name:
Common/Usual Name:
Predicate Devices:
Device Description:
Indications for Use:
system,planning,radiation therapy treatment 21CFR892.5050, MUJ, Class II
Eclipse, TPS, Eclipse TPS, Eclipse 12, Eclipse 12 TPS, Eclipse TPS 12.
Eclipse Treatment Planning System (K102011)
The Varian Eclipse™ Treatment Planning System (Eclipse TPS) provides software tools for planning the treatment of malignant or benign diseases with radiation. Eclipse TPS is a computer-based software device used by trained medical professionals to design and simulate radiation therapy treatments. Eclipse TPS is capable of planning treatments for external beam irradiation with photon, electron, and proton beams, as well as for internal irradiation, (brachytherapy) treatments.
The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (brachytherapy) treatments. In addition, the Eclipse Proton Eye algorithm is specifically indicated for planning proton treatment of neoplasms of the eye.
{2}------------------------------------------------
## Changes in Technological Characteristics:
| FEATURE AND/OR<br>SPECIFICATION OF<br>NEW/MODIFIED<br>DEVICE | CLEARED DEVICE<br>FEATURE/SPECIFICATION<br>(ECLIPSE 10- K102011) | DEVICE WITH CHANGE<br>ECLIPSE 12 |
|--------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Supported<br>External Beams<br>& Accessories | • TrueBeam support | • TrueBeam support |
| Graphical User<br>Interface | • Nexus Phase 0 - Home<br>screen integration and<br>navigation not present | • Nexus Phase 0- Home screen integration<br>and navigation |
| Image<br>Segmentation | • Automatic on-demand<br>and pre-processing tools<br>for multiple<br>organs/structures | • No Automatic on-demand and pre-<br>processing tools for multiple<br>organs/structures<br>(SmartAdapt) toolset utilizing changed CT-<br>MR and MR-MR deformable registration |
| Dose<br>Calculation | • Photon calculation<br>○ RapidArc:<br>intermediate<br>dose calculation<br>○ RapidArc: Varian<br>linac and support<br>○ FFF support for<br>TrueBeam | • Photon calculation<br>○ RapidArc: enhancements in<br>intermediate dose calculation<br>○ IMRT: intermediate dose<br>calculation<br>○ RapidArc Varian linac, and Elekta<br>VMAT support<br>○ FFF: Support for C3 and<br>TrueBeam<br>• Dose-Volume Histogram (DVH) Estimation |
| | • Proton calculation<br>○ Modulated<br>scanning (spot<br>scanning)<br>technique<br>○ Spot editor | • Proton calculation<br>○ Modulated scanning technique<br>(spot and line scanning), support<br>block and MLC<br>○ Range uncertainty feature<br>○ Spot editor user interface<br>improvements<br>○ Dosimetrically equivalent<br>treatment units (for different<br>gantries)<br>○ Block drill bit corrections for<br>milling machines |
| | • Brachytherapy calculation<br>○ AcurosBV dose<br>calculation<br>algorithm<br>○ AcurosBV<br>calculates dose to<br>transport media<br>but reports it in<br>water | • Brachytherapy calculation<br>○ AcurosBV dose calculation<br>algorithm in 64 bit environment<br>○ AcurosBV calculates does to<br>transport media<br>○ Nexus phase 0 support |
| FEATURE AND/OR<br>SPECIFICATION OF<br>NEW/MODIFIED<br>DEVICE | CLEARED DEVICE<br>FEATURE/SPECIFICATION<br>(ECLIPSE 10- K102011) | DEVICE WITH CHANGE<br>ECLIPSE 12 |
| | 64 bit External Beam<br>Planning, BrachyVision<br>and Proton, PRO and<br>AcurosXB algorithms Fluence calculation by<br>LMC algorithm | 64 bit External Beam Planning,<br>BrachyVision and Proton, PRO, AcurosXB<br>AcurosBV and BAO & DVO algorithms and<br>DVH Estimation. Unified fluence calculation in Eclipse &<br>DCF by the final 3D dose calculation<br>algorithm |
| Import/Export<br>Interfaces | Film Scanner import | No Film Scanner import Eclipse Scripting API (ESAPI) read only<br>access |
| Infrastructure | Sybase Server | SQL Server migration Zero Clinical Downtime: Faster DB<br>upgrades Zero Clinical Downtime: Remote<br>deployment of the clients Nexus Phase 0 RT Prescription integration Plan validation and status change<br>service Dosimetrically Equivalent<br>machine change service Approval modifications |
.
:
{3}------------------------------------------------
and the commend of the comments of the country
:
. . . . . .
:
{4}------------------------------------------------
K131891
Page 4 of 4
#### Summary of Non-clinical Testing
Verification testing was performed to demonstrate that the performance and functionality of the new and existing features met the design input requirements.
Regression testing was performed to verify the integrity of any changes. Validation testing was performed on a production equivalent device, under clinically representative conditions by qualified personnel.
#### Conclusion of Non-Clinical testing
Results from Verification and Validation testing demonstrate that the product met defined user needs and defined design input requirements. Varian therefore considers Eclipse 12 to be safe and to perform at least as well as the predicate device.
{5}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Contral Center - WO66-Gold Silver Spring, MI) 20993-0002
September 25, 2013
DEPARTMENT OF HEALTH & HUMAN SERVICES
Varian Medical Systems. Inc. % Mr. Peter J. Coronado Director, Regulatory Affairs 911 Hansen Wav, M/S E110 PALO ALTO CA 94303
Re: K131891
Trade/Device Name: Eclipse 12 Treatment Planning System Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system. Regulatory Class: 11 Product Code: MUJ, L.HN Dated: June 21, 2013 Received: June 28, 2013
Dear Mr. Coronado:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendnents, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
ff your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean I hat FDA has made a determination that your device complies with other requirements of the Act mar + Breaks intact and regulations administered by other Federal agencies. You must or any with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
Image /page/4/Picture/11 description: The image shows a logo with a stylized bird-like figure composed of three curved lines, suggesting movement or flight. The figure is positioned to the right of a circular arrangement of text, which appears to be a company name or slogan. The logo is simple, using only black lines on a white background, giving it a clean and modern look.
{6}------------------------------------------------
Page 2-Mr. Coronado
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industrv/default.htm.
Sincerely yours,
Mechal D. Offara
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known): K131891
Eclipse 12 Treatment Planning System Device Name:
Indications for Use:
The Eclipse Treatment Planning System (Eclipse TPS) Is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (brachytherapy) treatments. In addition, the Eclipse Proton Eye algorithm is specifically indicated for planning proton treatment of neoplasms of the eye.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Michael D. O'Hara
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health
510(k) K 131891
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.