K152592 · Hitachi, Ltd., Health Care Company · LHN · Nov 2, 2015 · Radiology
Device Facts
Record ID
K152592
Device Name
PROBEAT-V
Applicant
Hitachi, Ltd., Health Care Company
Product Code
LHN · Radiology
Decision Date
Nov 2, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PROBEAT-V is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.
Device Story
Proton beam therapy system; generates/delivers proton beams for radiation therapy. System components: accelerator (LINAC, synchrotron), beam transport system, rotating gantry, scanning nozzle, patient positioning system, and cone beam CT/X-ray imaging. Used in clinical settings; operated by trained medical professionals. Device delivers prescribed dose/distribution to patient site. Healthcare providers use imaging and beam delivery controls to target tumors; output facilitates precise radiation delivery to treat localized tumors. Benefits include targeted radiation therapy for susceptible conditions.
Clinical Evidence
Bench testing only. Performance data included mechanical testing of rotating gantry and patient couch, beam performance testing (dose shape/distribution), and safety interface testing (beam stop control, dose monitor, interlocks). Electrical safety and electromagnetic compatibility testing performed per IEC 60601-1 and IEC 60601-1-2.
Indicated for patients with localized tumors and other conditions susceptible to treatment by radiation.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Predicate Devices
Hitachi Ltd. PROBEAT with Discrete Spot Scanning System (K073059)
K151132 — PROBEAT-V · Hitachi, Ltd., Health Care Company · Jun 1, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The caduceus is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 2, 2015
Hitachi, Ltd., Health Care Company % Mr. Jonathan Kahan Regulatory Counsel Hogan Lovells US LLP 555 13th Street NW WASHINGTON DC 20016
Re: K152592
Trade/Device Name: PROBEAT-V Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: LHN Dated: September 10, 2015 Received: September 10, 2015
Dear Mr. Kahan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D.'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known)
K152592 Device Name
PROBEAT-V Indications for Use (Describe)
The PROBEAT-V is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{3}------------------------------------------------
### 510(k) SUMMARY
### Hitachi PROBEAT-V
## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Hitachi, Ltd., Health Care Company 4-14-1 Sotokanda, Chiyodaku, Tokyo, 101-8010, Japan Telephone: +81 (3) 45643563 Facsimile: +81 (3) 45642882
Contact Person: Naoya Nishimura
Date Prepared: September 10, 2015
#### Name of Device and Name/Address of Sponsor
PROBEAT-V Proton Beam Therapy System
Hitachi, Ltd. Power Systems Company 3-1-1 Saiwai-cho, Hitachi-shi, Ibaraki-ken, 317-8511 Japan
#### Common or Usual Name
Proton beam therapy system
#### Classification Name
Medical Charged-Particle Radiation Therapy System, 21 CFR 892.5050, Product Code LHN
#### Predicate Devices
Hitachi Ltd. PROBEAT with Discrete Spot Scanning System (K073059)
Hitachi Ltd. PROBEAT-V (K151132)
#### Intended Use / Indications for Use
The PROBEAT-V is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.
{4}------------------------------------------------
# Technological Characteristics
The PROBEAT-V is a proton beam irradiation system, which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam with the prescribed dose, dose distribution and directed to the prescribed patient site. The equipment to perform the above work is composed of two main components: (1) a beam delivery system whose primary responsibility is to ensure that the desired prescription parameters are properly delivered and (2) equipment necessary to generate the proton beam and direct it to the beam delivery system for patient treatment.
The beam delivery system is composed of the following components"
- o Gantry Room
- . Rotating Gantry
- . Scanning Nozzle
- . Patient Positioning System
- . Cone Beam CT / X-ray Imaging System
- o Fixed Beam Room
- Scanning Nozzle
- . Patient Positioning System
- . Cone Beam CT
The beam production system is composed of the following components"
- Accelerator system (LINAC, Synchrotron) o
- Beam transport system (Low/High Energy Beam Transport systems) о
## Performance Data
The company performed testing, as follows:
- The mechanical performance of the rotating gantry and patient couch O
- Beam performance testing to evaluate beam dose shape and beam dose O
- Safety interface testing to evaluate beam stop control, dose monitor, area safety, O and mechanical interlocks.
Further, electrical safety and electromagnetic compatibility testing was also performed in accordance with IEC 60601-1 and IEC 60601-1-2.
In all instances, the PROBEAT-V functioned as intended and met its specifications.
## Substantial Equivalence
The PROBEAT-V has the same intended use and indications for use, as well as substantially similar principles of operation and technological characteristics, as compared to Hitachi's cleared PROBEAT with DSSS (K073059) and a previously cleared version of the PROBEAT-V (K151132). Each of the systems comprises the same components of the beam production and beam delivery subsystems. The minor differences between the PROBEAT-V that is the
{5}------------------------------------------------
subject of this submission and the cleared PROBEAT and PROBEAT-V do not raise different questions of safety or effectiveness. Thus, the PROBEAT-V is substantially equivalent to its predicate devices.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.