DOPPLER FETAL HEARTBEAT RATE DETECTOR

K090510 · Shenzhen Bestman Instrument Co., Ltd. · KNG · Apr 29, 2010 · Obstetrics/Gynecology

Device Facts

Record IDK090510
Device NameDOPPLER FETAL HEARTBEAT RATE DETECTOR
ApplicantShenzhen Bestman Instrument Co., Ltd.
Product CodeKNG · Obstetrics/Gynecology
Decision DateApr 29, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.2660
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

This Doppler can be used for the detection of average fetal heartbeat rate.

Device Story

Doppler fetal heartbeat rate detector uses continuous wave (CW) ultrasound Doppler principle to detect fetal heart rate. Device utilizes split D piezoelectric transducer; high-frequency oscillator transmits ultrasound acoustic waves into body via coupling medium; reflected waves from fetal heart received by second transducer; converted to high-frequency electronic signal; amplified and filtered to produce baseband audio Doppler signal; output via loudspeaker; fetal heart rate displayed on liquid crystal arithmometer. Used in clinical settings to assist healthcare providers in monitoring fetal status; provides real-time auditory and visual feedback of fetal heart rate; aids in clinical assessment of fetal well-being.

Clinical Evidence

Bench testing only. Device subjected to safety and performance testing to ensure compliance with functional requirements and industry safety standards. No clinical data presented.

Technological Characteristics

Ultrasonic fetal monitor; split D piezoelectric transducer; CW ultrasound (2.0 MHz for BF-500B, 2.5 MHz for BF-500+/BF-500++); liquid crystal arithmometer display; loudspeaker output; water-based coupling medium; Class II device.

Indications for Use

Indicated for detection of average fetal heartbeat rate in pregnant patients.

Regulatory Classification

Identification

A fetal ultrasonic monitor is a device designed to transmit and receive ultrasonic energy into and from the pregnant woman, usually by means of continuous wave (doppler) echoscopy. The device is used to represent some physiological condition or characteristic in a measured value over a period of time (e.g., perinatal monitoring during labor) or in an immediately perceptible form (e.g., use of the ultrasonic stethoscope). This generic type of device may include the following accessories: signal analysis and display equipment, electronic interfaces for other equipment, patient and equipment supports, and component parts. This generic type of device does not include devices used to image some relatively unchanging physiological structure or interpret a physiological condition, but does include devices which may be set to alarm automatically at a predetermined threshold value.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(K) SUMMARY This summary of 5 10(k) safety a nd ef fectiveness information is be ing submitted i n accordance with 21 CFD part 807.92. The assigned 510(k) number is: K090570 APR 2 9 2010 1. Submitter's name, address, phone number, contact person and ### preparation data: Name: Shenzhen Bestman Instrument Co., Ltd. Address: 4/F, Block 210, 2nd Industrial Area of Nanyou, Xiangnan Rd., Nanshan Dist., Shenzhen, P.R.China Phone: 86-755-26713783/26713784 Fax: 86-755-26495167/26713783 Contact Person: Bai Yong Official correspondent: Bai Yong General Manager Shenzhen Bestman Instrument Co., Ltd. Address: 4/F, Block 210, 2nd Industrial Area of Nanyou, Xiangnan Rd., Nanshan Dist., Shenzhen, P.R.China Tel: 86-755-26713783/26713784 Fax: 86-755-26495167/26713783 Email: bestm@public.szptt.net.cn . Date of Preparation: April 27, 2010 ### 2. Device: - . Proprietary Name: Doppler fetal heartbeat rate detector - Ultrasonic Fetal Monitor � Common Name: - Classification Name: Fetal Ultrasonic Monitor and Accessories ● l - Product Code: KNG {1}------------------------------------------------ - . Manufactured By: Shenzhen Bestman Instrument Co., Ltd., China - 3. Predicate Device: K040480 SONOTRAX #### 4. Classification Names: Class II as per 21CFR 884-2660, Ultrasonic Fetal Monitor and accessories. #### 5. Description: Doppler fetal heartbeat rate detector uses Doppler principle of ultrasound signal to detect the fetal heart rate. Doppler fetal heartbeat rate detector uses a split D piezoelectric transducer. A high frequency oscillator supplies a continuous high frequency voltage to one half of the split D transmitter transducer. The high frequency voltage is converted to an ultrasound acoustic wave by the transducer and is transmitted to biophysical objects through an applied coupling water based medium and moves through biophysical objects. The acoustic ultrasound is reflected by body and moving objects such as the fetal heart. The reflected ultrasound is received by the second split D receiver transducer and is converted via the piezoelectric effect into a high frequency electronic signal. The received electronic signal is amplified and detected. The result is a base band audio Doppler shifted signal which is filtered and converted to audio via a loudspeaker. At the same time the fetal heart rate is displayed on a liquid crystal arithmometer display. ### 6. Indications for use: This Doppler can be used for the detection of average fetal heartbeat rate. Models List Form: (Please see the attached "Diagnostic Ultrasound Indications For Use Format" forms for detail.) | Series | Models | Attached Transducers | |--------|---------------------|----------------------| | 1 | BF-500B | CW20<br>(fc=2.0MHz) | | 2 | BF-500+<br>BF-500++ | CW25<br>(fc=2.5MHz) | ### 7. Contra-indications: Not be discovered till now. {2}------------------------------------------------ #### 8. Comparison to Predicate Devices: Doppler fetal heartbeat rate detector has the same device characteristics as the above predicate approved device. #### 9. Test Data: Doppler fetal heartbeat rate detector has been subjected to extensive safety, performance test and validations before release. Final test of the Doppler fetal heartbeat rate detector includes various performance tests designed to ensure that the device meets all of its functional requirements and performance specifications. Safety tests have further been performed to ensure the devices comply with applicable industry and safety standards. Doppler fetal heartbeat rate detector includes instructions for safe and effective use, warnings, cautions and guidance for use. ### 10. Literature Review: A review of the literature pertaining to the safety of fetal Doppler has been conducted and appropriate safeguards h ave been incorporated in the design of the Doppler fetal heartbeat rate detector. #### 11. Conclusion: The c onclusions drawn fr om the te st of the Doppler fetal hea rtbeat ra te det ector demonstrates that the device is substantially equivalent to the predicate device. 3 {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is a stylized image of an eagle with its wings spread. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Shenzhen Bestman Instrument Co., Ltd % Mr. Marc M. Mouser Manager/FDA Office Coordinator Underwriters Laboratories, Inc. 2600 N.W. Lake Road CAMAS WA 98607-8542 APR 2 9 2010 Re: K090510 Trade/Device Name: Doppler fetal heartbeat detector Models: BF-500B, BF-500+, and BF500++ Regulation Number: 21 CFR 884.2660 Regulation Name: Fetal ultrasonic monitor and accessories Regulatory Class: II Product Code: KNG Dated: April 2, 2010 Received: April 8, 2010 Dear Mr. Mouser: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. This determination of substantial equivalence applies to the following transducers intended for use with the Doppler fetal heartbeat detector - Models: BF-500+, and BF500++as described in your premarket notification: Transducer Model Number Model BF-500+ & BF-500++ CW25 > Model BF-500B CW20 {4}------------------------------------------------ If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. If you have any questions regarding the content of this letter, please contact Shahram Vaezy at (301) 796-6242. Sincerely yours. Robe RBeater J (fa) Donald St. Pierre Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure(s) {5}------------------------------------------------ ## Indications for Use 510(k) Number (if known): KO905 /0 Device Name: Doppler fetal heartbeat rate detector Indications for Use: This Doppler can be used for the detection of average fetal heartbeat rate. Models List Form: | Series | Models | Attached<br>Transducers | |--------|----------|-------------------------| | 1 | BF-500B | CW20<br>(fc=2.0MHz) | | 2 | BF-500+ | CW25<br>(fc=2.5MHz) | | | BF-500++ | | Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Robert A. Zecker (Division Sign, Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety **510(k) Number** K090510 {6}------------------------------------------------ X System: Transducer: __________________________________________________________________________________________________________________________________________________________________ BF-500B Model: Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | | Mode of Operation | | | | | | | |-----------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------| | General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) | | Ophthalmic | Ophthalmic | | | | | | | | | | Fetal | | | | N | | | | | | Abdominal | | | | | | | | | | Intra-operative (Specify) | | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | | Laparoscopic | | | | | | | | | Fetal<br>Imaging<br>& Other | Pediatric | | | | | | | | | | Small Organ (Specify) | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | Adult Cephalic | | | | | | | | | | Trans-rectal | | | | | | | | | | Trans-vaginal | | | | | | | | | | Trans-urethral | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | Musculo-skeletal<br>(Conventional) | | | | | | | | | | Musculo-skeletal<br>(Superficial) | | | | | | | | | | Intravascular | | | | | | | | | | Other (Specify) | | | | | | | | | Cardiac | Cardiac Adult | | | | | | | | | | Cardiac Pediatric | | | | | | | | | | Intravascular (Cardiac) | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | Intra-cardiac | | | | | | | | | | Other (Specify) | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | | | | | | | | | | Other (Specify) | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under this appendix * Examples of other modes of operation may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, and Color Velocity Imaging Koba K090510 {7}------------------------------------------------ System: ______________________________________________________________________________________________________________________________________________________________________ Transducer: __________________________________________________________________________________________________________________________________________________________________ Model: BF-500+ Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | | Mode of Operation | | | | | | | |-----------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------| | General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) | | | | | | | | | | | | Ophthalmic | Ophthalmic | | | | | | | | | | Fetal | | | | N | | | | | | Abdominal | | | | | | | | | | Intra-operative (Specify) | | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | | Laparoscopic | | | | | | | | | Fetal<br>Imaging<br>& Other | Pediatric | | | | | | | | | | Small Organ (Specify) | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | Adult Cephalic | | | | | | | | | | Trans-rectal | | | | | | | | | | Trans-vaginal | | | | | | | | | | Trans-urethral | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | Musculo-skeletal<br>(Conventional) | | | | | | | | | | Musculo-skeletal<br>(Superficial) | | | | | | | | | | Intravascular | | | | | | | | | | Other (Specify) | | | | | | | | | | Cardiac Adult | | | | | | | | | Cardiac | Cardiac Pediatric | | | | | | | | | | Intravascular (Cardiac) | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | Intra-cardiac | | | | | | | | | | Other (Specify) | | | | | | | | | Peripheral | Peripheral vessel | | | | | | | | | Vessel | Other (Specify) | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under this appendix * Examples of other modes of operation may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, and Color Velocity Imaging Robert M. Becker K090510 {8}------------------------------------------------ X System: ___________________ Transducer: Model: BF-500++ Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | | Mode of Operation | | | | | | | |-----------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------| | General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) | | Ophthalmic | Ophthalmic | | | | | | | | | | Fetal | | | | N | | | | | Fetal<br>Imaging<br>& Other | Abdominal | | | | | | | | | | Intra-operative (Specify) | | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | | Laparoscopic | | | | | | | | | | Pediatric | | | | | | | | | | Small Organ (Specify) | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | Adult Cephalic | | | | | | | | | | Trans-rectal | | | | | | | | | | Trans-vaginal | | | | | | | | | | Trans-urethral | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | Musculo-skeletal<br>(Conventional) | | | | | | | | | | Musculo-skeletal<br>(Superficial) | | | | | | | | | | Intravascular | | | | | | | | | | Other (Specify) | | | | | | | | | Cardiac | Cardiac Adult | | | | | | | | | | Cardiac Pediatric | | | | | | | | | | Intravascular (Cardiac) | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | Intra-cardiac | | | | | | | | | | Other (Specify) | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | | | | | | | | | | Other (Specify) | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under this appendix * Examples of other modes of operation may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, and Color Velocity Imaging > 8 . Robert M. Bestery K090510 {9}------------------------------------------------ System: Transducer: CW25(This probe applies to BF-500+ & BF-500++ Model) Model: X Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | | Mode of Operation | | | | | | | |-----------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------| | General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) | | Ophthalmic | Ophthalmic | | | | | | | | | | Fetal | | | | N | | | | | | Abdominal | | | | | | | | | | Intra-operative (Specify) | | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | Fetal<br>Imaging<br>& Other | Laparoscopic | | | | | | | | | | Pediatric | | | | | | | | | | Small Organ (Specify) | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | Adult Cephalic | | | | | | | | | | Trans-rectal | | | | | | | | | | Trans-vaginal | | | | | | | | | | Trans-urethral | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | Musculo-skeletal<br>(Conventional) | | | | | | | | | | Musculo-skeletal<br>(Superficial) | | | | | | | | | | Intravascular | | | | | | | | | | Other (Specify) | | | | | | | | | Cardiac | Cardiac Adult | | | | | | | | | | Cardiac Pediatric | | | | | | | | | | Intravascular (Cardiac) | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | Intra-cardiac | | | | | | | | | | Other (Specify) | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | | | | | | | | | | Other (Specify) | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under this appendix * Examples of other modes of operation may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, and Color Velocity Imaging Robert M. Beckerf K090510 {10}------------------------------------------------ System: Transducer: CW20(This probe applies to BF-500B Model) Model: X Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | | Mode of Operation | | | | | | Other*<br>(Specify) | |-----------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------| | General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | | | Ophthalmic | Ophthalmic | | | | | | | | | Fetal<br>Imaging<br>& Other | Fetal | | | | N | | | | | | Abdominal | | | | | | | | | | Intra-operative (Specify) | | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | | Laparoscopic | | | | | | | | | | Pediatric | | | | | | | | | | Small Organ (Specify) | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | Adult Cephalic | | | | | | | | | | Trans-rectal | | | | | | | | | | Trans-vaginal | | | | | | | | | | Trans-urethral | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | Musculo-skeletal<br>(Conventional) | | | | | | | | | | Musculo-skeletal<br>(Superficial) | | | | | | | | | | Intravascular | | | | | | | | | | Other (Specify) | | | | | | | | | Cardiac | Cardiac Adult | | | | | | | | | | Cardiac Pediatric | | | | | | | | | | Intravascular (Cardiac) | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | Intra-cardiac | | | | | | | | | | Other (Specify) | | | | | | | | | Peripheral<br>Vessel | Peripheral vessel | | | | | | | | | | Other (Specify) | | | | | | | | N = new indication; P = previously cleared by FDA; E = added under this appendix * Examples of other modes of operation may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, and Color Velocity Imaging Tobe M. Beckey K090510
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...