K082055 · Shenzhen Creative Industry Co., Ltd. · KNG · Feb 6, 2009 · Obstetrics/Gynecology
Device Facts
Record ID
K082055
Device Name
MODEL PC-860 FETAL DOPPLER
Applicant
Shenzhen Creative Industry Co., Ltd.
Product Code
KNG · Obstetrics/Gynecology
Decision Date
Feb 6, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.2660
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The device is an ultrasonic fetal heart beat detector, which can detect the Fetal Heart Rate. The built in speaker of the device allows for listening of the fetal heartbeat.
Device Story
Handheld ultrasonic fetal Doppler; detects fetal heart rate via Doppler effect. User applies ultrasonic gel to abdomen; places probe to locate fetal heart. Device transmits ultrasonic waves; receives reflected waves with frequency shifts caused by fetal heart movement. Internal processing calculates heart rate; built-in speaker provides audible heartbeat. Used by clinicians or patients to monitor fetal status. Provides real-time heart rate data and audible feedback to assist in fetal health assessment.
Clinical Evidence
Clinical testing was used to validate the effectiveness and accuracy of the device. All test results were satisfactory.
Technological Characteristics
Ultrasonic Doppler fetal monitor; operates via ultrasonic wave reflection and frequency shift detection. Complies with EN/IEC 60601-1 for electrical/mechanical safety. Includes ultrasonic probe and built-in speaker.
Indications for Use
Indicated for detection of fetal heart rate in pregnant women using ultrasonic Doppler technology.
Regulatory Classification
Identification
A fetal ultrasonic monitor is a device designed to transmit and receive ultrasonic energy into and from the pregnant woman, usually by means of continuous wave (doppler) echoscopy. The device is used to represent some physiological condition or characteristic in a measured value over a period of time (e.g., perinatal monitoring during labor) or in an immediately perceptible form (e.g., use of the ultrasonic stethoscope). This generic type of device may include the following accessories: signal analysis and display equipment, electronic interfaces for other equipment, patient and equipment supports, and component parts. This generic type of device does not include devices used to image some relatively unchanging physiological structure or interpret a physiological condition, but does include devices which may be set to alarm automatically at a predetermined threshold value.
{0}------------------------------------------------
K082055 page 1/3
FEB - 6 2009
## 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: May 14, 2008
1. Company and Correspondent making the submission:
Name - Shenzhen Creative Industry Co., Ltd.
Address - 2/F, Block 3, Nanyou Tian'an Industry Town, Shenzhen, China
Telephone - +86-755-2643 4955
Fax - +86-755-2643 5433
Contact - Jianbin Liu
Email - liujianbin@creative-sz.com
2. Device :
Trade/proprietary name: Fetal Doppler, Model PC-860 Common Name : Fetal Doppler Classification Name : Monitor, Ultrasonic, Fetal
Predicate Device: Edan Instruments, Inc., Ultrasonic Pocket Doppler Model: Sonotrax 510(K) number: K040480
3. Classifications Names & Citations : 21CFR 884.2660, KNG, Fetal Doppler, Class II
{1}------------------------------------------------
#### 4. Description
### 4.1 General
This Doppler tests the fetal heart rate through non-invasive ultrasonic Doppler Effect. As is known, ultrasonic wave propagating at a given frequency will be reflected when encountering an obstacle. If it is still an obstacle, the back wave will share the same frequency with the transmitted wave. Once the obstacle moves, the frequency of the back wave will be changed. The higher rate the object moves at, the bigger frequency change will take place. This is the so-called Doppler Effect. With the apparatus, the ultrasonic probe is placed on the abdomen of the pregnant woman. The ultrasonic probe can perceive the fetal heartbeat. When the transmitted wave encounters the fetal heart, the back wave will develop offset frequency. With the offset frequency, the fetal heart rate and frequency can be worked out.
### 4.2 Directions
As discussed in the General description, the Shenzhen Creative Industry Co., Ltd Fetal Doppler, model PC-860 is relatively simple to use. The Doppler is placed on the abdomen of the patient by first feeling out the fetal contour with hand to find the approximate position of fetal heart. Generally, fetal heart is at a location 113 of the lower abdomen (below the navel) during short pregnant weeks, and along with the pregnant weeks increasing, it moves upwards and lean to right or lean to left. To scan the abdomen, the user must first daub the acoustics surface of Doppler probe uniformly with the appropriate ultrasonic gel, and then put the probe on pregnant woman's abdomen (a location near fetal heart). It is important to ensure that probe contacts surface completely.
#### 5. Indication for use :
The device is an ultrasonic fetal heart beat detector, which can detect the Fetal Heart Rate. The built in speaker of the device allows for listening of the fetal heartbeat.
{2}------------------------------------------------
# 6. Comparison with predicate device :
Shenzhen Creative Industry Co., Ltd., believes that the Fetal Doppler, Model PC-860 is substantially equivalent to the Edan Instruments, Inc., Ultrasonic Pocket Doppler, Model: Sonotrax, 510(K) number: K040480
- 7. Safety and Performance Data :
Electrical, mechanical, environmental safety and performance testing according to standard EN/IEC 60601-1. Clinical testing was used to validate the effectiveness and accuracy of the device. Specific testing relating to Ultrasound equipment was use to verify performance to recognized standards. All test results were satisfactory.
8. Conclusions:
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Shenzhen Creative Industry, Inc., Ltd concludes that Fetal Doppler, Model PC-860 is safe and effective and substantially equivalent to predicate devices as described herein. ·
- 9. Shenzhen Creative Industry Co., Ltd. will update and include in a summary any other information deemed seasonably necessary by the FDA.
END
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows a logo for the Department of Health & Human Services. The logo features a stylized eagle with three lines representing its wings. The text "DEPARTMENT OF HEALTH &" is vertically oriented on the left side of the logo, and the text "HUMAN SERVICES" is vertically oriented on the right side of the logo.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB - 6 2009
Shenzhen Creative Industry Co., Ltd. c/o Mr. Mark Job Reviewer Regulatory Technology Services LLC 1394 25th Street, N.W. BUFFALO MN 55313
Re: K082055
Trade/Device Name: PC-860 Fetal Doppler Regulation Number: 21 CFR §884.2660 Regulation Name: Fetal ultrasonic monitor and accessories Regulatory Class: II Product Code: KNG Dated: January 22, 2009 Received: January 23, 2009
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter,
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | (240) 276-0115 |
|----------------|----------------------------------|----------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | (240) 276-0115 |
| 21 CFR 892.xxx | (Radiology) | (240) 276-0120 |
| Other | | (240) 276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry.suppor/index.html.
Sincerely yours,
Janine M. Morris
Acting Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### Indications for Use
# Indications for Use
510(k) Number (if known):
K082055
Device Name: PC-860 Fetal Doppler
Indications For Use:
The device is an ultrasonic fetal heart beat detector, which can detect the Fetal Heart Rate. The built in speaker of the device allows for listening of the fetal heartbeat.
Prescription Use ... . AND/OR S (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Aogn M. Whing
---
(Division Sign-Off)
Division of Reproductive, Abdominal, and Radiological Device 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.