NEXSTAR LIFTOFF PET APPLICATION SOFTWARE SUITE

K081426 · Philips Medical Systems (Cleveland), Inc. · KPS · Jun 4, 2008 · Radiology

Device Facts

Record IDK081426
Device NameNEXSTAR LIFTOFF PET APPLICATION SOFTWARE SUITE
ApplicantPhilips Medical Systems (Cleveland), Inc.
Product CodeKPS · Radiology
Decision DateJun 4, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1200
Device ClassClass 2
AttributesSoftware as a Medical Device, 3rd-Party Reviewed

Intended Use

Software contained in the PET Application Suite process, analyze, display, and quantify medical images/data. The PET and CT images may be registered and displayed in a "fused" (overlaid in the same spatial orientation) format to provide combined metabolic and anatomical data at different angles. Trained professionals use the images in: The evaluation, detection and diagnosis of lesions, disease and organ function such as cancer. cardiovascular disease, and neurological disorders. The detection, localization, and staging of tumors and diagnosing cancer patients. Treatment planning and interventional radiology procedures. The PET Application Suite includes software that provides a quantified analysis of regional cerebral activity from PET images. Cardiac imaging software provides functionality for the quantification of cardiology images and datasets including but not limited to myocardial perfusion for the display of wall motion and quantification of left-ventricular function parameters from gated myocardial perfusion studies and for the 3D alignment of coronary artery images from CT coronary angiography onto the myocardium.

Device Story

NexStar Liftoff PET Application Software Suite processes, analyzes, displays, and quantifies PET and CT medical images. Operates as a Windows-based software suite on standard hardware platforms; used by trained professionals (radiologists, cardiologists) in clinical settings. Inputs include PET and CT datasets; outputs include fused metabolic/anatomical images, quantified regional cerebral activity, myocardial perfusion metrics, wall motion displays, left-ventricular function parameters, and 3D coronary artery alignments. Software assists clinicians in lesion detection, tumor staging, and treatment planning. Benefits include improved diagnostic clarity through image fusion and automated quantification of cardiac and neurological function.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Windows-based software suite for image processing, display, and quantification. Features include image fusion, metabolic analysis, and cardiac realignment. Operates on hardware platforms meeting minimum system requirements. Software-based processing; no specific algorithm architecture (e.g., ML/DL) disclosed.

Indications for Use

Indicated for use by trained professionals to evaluate, detect, and diagnose lesions, disease, and organ function (cancer, cardiovascular, neurological) in patients. Used for tumor detection, localization, and staging; treatment planning; interventional radiology; quantification of regional cerebral activity; and cardiology image quantification (myocardial perfusion, wall motion, left-ventricular function, and 3D alignment of coronary artery images).

Regulatory Classification

Identification

An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K081426 JUN - 4 2008 # 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS ## General Information A. Submitter/ Contact Person: Philips Medical Systems (Cleveland), Inc. 595 Miner Rd. Cleveland, OH 44143 Melinda Novatny Tel: (440) 483-4255 Fax: (440) 483-2300 - Device Trade Name: NexStar Liftoff PET Application Software Suite B. Common Name: PET Application Software Classification Name: System, Emission Computed Tomography, (892.1200) 21CFR 892.1200, Class II Device Class: Product Code: 90 KPS Classification Panel: Radiology - C. Date prepared: April 11, 2008 - D. Predicate Device: GEMINI Raptor System (GEMINI TF) (K052640) #### E. Performance Standards: No performance standards have been developed for process and display applications. F. Intended Use: > Software contained in the PET Application Suite process, analyze, display, and quantify medical images/data. The PET and CT images may be registered and displayed in a "fused" (overlaid in the same spatial orientation) format to provide combined metabolic and anatomical data at different angles. Trained professionals use the images in: - o The evaluation, detection and diagnosis of lesions, disease and organ function such as cancer. cardiovascular disease, and neurological disorders. - The detection, localization, and staging of tumors and ു diagnosing cancer patients. - Treatment planning and interventional radiology o procedures. The PET Application Suite includes software that provides a quantified analysis of regional cerebral activity from PET images. Section X .- 1 {1}------------------------------------------------ Cardiac imaging software provides functionality for the quantification of cardiology images and datasets including but not limited to myocardial perfusion for the display of wall motion and quantification of left-ventricular function parameters from gated myocardial perfusion studies and for the 3D alignment of coronary artery images from CT coronary angiography onto the myocardium. ### (1. Device Description/ Comparison with Predicate Device: The NexStar Liftoff PET Application Software Suite (referred to as NexStar or Liftoff within the submission) is software used to process, analyze and display medical images and may be sold with Philips nuclear medicine PET/CT Systems or systems marketed by Philips. The PET Software Application Suite is a full suite of applications, including both review and processing. The NexStar Liftoff PET Application Software Suite is basically the same as the processing and reconstruction software cleared with the predicate device (GEMINI TF, K052640), with the extension of Image Fusion Software to include Metabolic Analysis and Cardiac Realignment. NexStar software is a Windows®-based suite of image display and processing applications and is deployable on hardware platforms, which meet the minimum requirements needed to run the software. ### H. System Performance Test/ Summary of Studies: No performance standards have been developed for process and display applications. ### 』、 Comparison to Predicate Devices: The basic differences in the system include the following: - . Deployment of software on hardware platforms meeting minimum level of requirements - Enhancements to processing applications . In conclusion, the device is substantially equivalent to the predicate device based upon similar intended use, and technological comparison. {2}------------------------------------------------ Third Party Review Quality Assessment Section 1 – Submission Information 510(k) No.: K081426 Third Party Organization: Underwriters Laboratories, Inc. Third Party's Primary Reviewer(s): Jeff D. Rongero ODE/OIVD Division: DRARD Branch/Team: Radiological Devices Branch Section 2 - 510(k) Decision Third party recommendation: SE √_____ NSE _____ ODE/OIVD final decision: SE √_____ NSE ______________________________________________________________________ Other (specify): __ Other (specify): _____________________________________________________________________________________________________________________________________________________________ Section 3 - Assessment of Third Party Review | Review Element | Rating (check one) | | | |---------------------------------------------------------------------------------------------------------------------------------------------|--------------------|----------------|----------------| | | Adequate | Minor Issue(s) | Major Issue(s) | | a. Determination of device eligibility for third party review | √ | | | | b. Extent of pre-submission consultation with ODE/OIVD division | | | | | c. Organization and format of review documentation | √ | | | | d. Determination of 510(k) administrative completeness (screening review) | √ | | | | e. Summary of device characteristics, intended use, and performance (including accessories, if applicable) and reason for 510(k) submission | √ | | | | f. Comparison to legally marketed devices—identification and analysis of key similarities and differences | √ | | | | g. Rationale for conclusions and recommendation | √ | | | | h. Use of guidance documents and standards | √ | | | | i. Resolution of 510(k) deficiencies and FDA requests for additional information | √ | | | | j. Scope of reviewer expertise and use of consulting reviewers | √ | | | | k. Other (specify): | | | | Comments (explanation of ratings/issues): ____________________________________________________________________________________________________________________________________ ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Section 4 – ODE/OIVD Assessor Information Assessed by: Sunder Rajan_ Date: 28-May-2008 Tel. No.: 240 276 3968 Routing: Division--Clip completed assessment (this page only) to inside front cover of 510(k). DMC-Forward this page only to Eric Rechen, POS/ODE, Rm. 120J, Corp. Blvd. (HFZ-402). {3}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The eagle is facing right. The text "U.S. DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circle around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # JUN - 4 2008 Philips Medical Systems (Cleveland), Inc. % Mr. Jeff D. Rongero Senior Project Engineer Underwriters Laboratories, Inc. 12 Laboratory Drive Research Triangle Park, NC 27709 Re: K081426 Trade/Device Name: NexStar Liftoff PET Application Software Suite Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: II Product Code: KPS Dated: May 20, 2008 Received: May 21, 2008 Dear Mr. Rongero: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter. | 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |-----------------|----------------------------------|--------------| | 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 892.xxxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrb/industry/support/index.html. Sincerely yours, Nancy C Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use ## 510(k) Number (if known): Not Known ### Device Name: NexStar Liftoff PET Application Software Suite Software contained in the PET Application Suite process, analyze, display, and quantify medical images/data. The PET and CT images may be registered and displayed in a "fused" (overlaid in the same spatial orientation) format to provide combined metabolical data at different angles. Trained professionals use the images in: - The evaluation, detection and diagnosis of lesions, disease and organ function such as cancer, . cardiovascular disease, and neurological disorders. - The detection, localization, and staging of tumors and diagnosing cancer patients. . - Treatment planning and interventional radiology procedures. . - The PET Application Suite includes software that provides a quantified analysis of regional cerebral ♥ activity from PET images. Cardiac imaging software provides functionality for the quantification of cardiology images and datasets including but not limited to myocardial perfusion for the display of wall motion and quantification of leftventricular function parameters from gated myocardial perfusion studies and for the 3D alignment of coronary artery images from CT coronary angiography onto the myocardium ### Prescription Use √ Over-The-Counter Use AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page Jongu Mthang (Division Sign-C Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...