CEDARS-SINAI CARDIAC SUITE NUCLEAR MEDICINE SOFTWARE
K141652 · Cedars-Sinai Medical Center · KPS · Jul 22, 2014 · Radiology
Device Facts
Record ID
K141652
Device Name
CEDARS-SINAI CARDIAC SUITE NUCLEAR MEDICINE SOFTWARE
Applicant
Cedars-Sinai Medical Center
Product Code
KPS · Radiology
Decision Date
Jul 22, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1200
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Cedars-Sinai Cardiac Suite of applications is intended to enable an automated display, review, and quantification of Nuclear Medicine Cardiology medical images and datasets. Cedars-Sinai Cardiac Suite may be used in multiple settings including the hospital, clinic, doctors office, or remotely. The results provided should be reviewed by qualified healthcare professionals (e.g., radiologists, cardiologists, or general nuclear medicine physicians) trained in the use of medical imaging devices.
Device Story
Cedars-Sinai Cardiac Suite is a standalone software suite for processing/reviewing cardiac SPECT and PET images. Inputs: camera-independent reconstructed SPECT/PET image files. Processing: automated quantitative/qualitative analysis via modules including QPS (Quantitative Perfusion SPECT), QGS (Quantitative Gated SPECT), QBS (Quantitative Blood Pool SPECT), QARG (reporting), Fusion (SPECT/CT/CTA or PET/CT/CTA), Motion Correction (MOCO), AutoRECON, and OPET (including viability quantification/databases). Output: quantitative/qualitative cardiac imaging data displayed for clinician review. Used in hospitals, clinics, or remote settings by radiologists/cardiologists to assist in clinical decision-making regarding cardiac health. Benefits: provides standardized, automated quantification of nuclear cardiology studies to aid diagnostic assessment.
Clinical Evidence
No clinical data. Substantial equivalence supported by software validation confirming all device features were compliant with software requirements.
Technological Characteristics
Standalone software for SPECT/PET image processing. Compatible with Windows (XP, Vista, 7, 8, Server) and Mac OS X. Viewer component compatible with iPad (2, 3, 4, Mini). Features include QPS, QGS, QBS, QARG, Fusion, MOCO, AutoRECON, and OPET. Operates on camera-independent reconstructed image files.
Indications for Use
Indicated for patients undergoing Nuclear Medicine Cardiology SPECT or PET procedures. Intended for use by radiologists, cardiologists, or general nuclear medicine physicians in hospitals, clinics, or remote settings.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
Predicate Devices
CSMC Cardiac Suite- Nuclear Medicine Software (K112474)
Submission Summary (Full Text)
{0}------------------------------------------------
### Section 6
510(k) Summary
K141652
Page 1 of 3
## 6. 510(k) Summary
This 510(k) summary information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| APPLICANT: | Cedars-Sinai Medical Center, Department of Medicine,<br>Artificial Intelligence in Medicine Program |
|------------------------|-----------------------------------------------------------------------------------------------------|
| DATE PREPARED: | 31 October 2012 |
| CONTACT PERSON: | Geoff Pollard, Program Manager<br>8700 Beverly Blvd<br>Los Angeles, CA 90048<br>Phone: 310.423.4663 |
| TRADE NAME: | Cedars-Sinai Cardiac Suite Nuclear Medicine Software |
| CLASSIFICATION NAME: | Emission Computed Tomography System |
| DEVICE CLASSIFICATION: | Class II |
| REGULATION NUMBER | 892.1200 |
| PRODUCT CODE | KPS |
| PREDICATE DEVICES: | CSMC Cardiac Suite- Nuclear Medicine Software (K112474) |
#### Substantially Equivalent To:
The CSMC Cardiac Suite v2013- Nuclear Medicine Software is substantially equivalent in intended use, principal of operation and technological characteristics to the previously cleared CSMC Cardiac Suite v2012- Nuclear Medicine Software (K112474).
#### Description of the Device Subject to Premarket Notification:
The Cedars-Sinai Cardiac Suite is a stand-alone software solution for Cardiac SPECT and PET imaging processing and review. Cedars-Sinai Cardiac Suite (non-viewer) minimum system requirements include a computer with at least IGB RAM, 50MB hard disk space for software installation. a display resolution of at least 1024x768 with 16-bit color, a network adapter, a mouse (or other pointer device; trackpad, trackball, etc ... ) and one of the following operating systems: Windows XP, Windows Vista, Windows 7, Windows 8, Windows Server 2003. Windows Server 2008. Windows Server 2012. Mac OS X (10.6. 10.7 & 10.8). The viewer component of the Cedars-Sinai Cardiac Suite minimum system requirements is iPad 2, iPad (3rd generation), iPad (4th generation) and iPad Mini. The Cedars-Sinai Cardiac Suite operates on camera independent reconstructed SPECT and/or PET image files. CSMC Cardiac Suite will be marketed as a comprehensive application suite that includes QPS+QGS (Quantitative Perfusion SPECT + Quantitative Gated
Page 11 of 46 Premarket Notification
{1}------------------------------------------------
SPECT) and CSImport applications. This allows automatic processing and review of quantitative and qualitative information generated by nuclear medicine studies. Purchasable Options consist of Quanitative Blood Pool SPECT (QBS), QARG (for reporting purposes), Fusion (SPECT/CT/CTA and/or PET/CT/CTA), Motion Correction (MOCO), Automated Reconstruction (AutoRECON), Remote Viewer (Viewer) and OPET. OPET also includes viability quantification and two additional databases (rubidium and ammonia) for processing PET studies.
## Indication for Use:
The Cedars-Sinai Cardiac Suite of applications is intended to enable an automated display, review, and quantification of Nuclear Medicine Cardiology medical images and datasets. Cedars-Sinai Cardiac Suite may be used in multiple settings including the hospital. clinic, doctors office, or remotely. The results provided should be reviewed by qualified healthcare professionals (e.g., radiologists, or general nuclear medicine physicians) trained in the use of medical imaging devices.
#### Technological Characteristics:
The modified CSMC Cardiac Suite- Nuclear Medicine Software has the same technological characteristics and is similar in data display and analysis compared to the predicates. The table below illustrates the similarities and differences in Technological Characteristics between the devices.
| | Modified CSMC Cardiac Suite | CSMC Cardiac Suite<br>(K117474) |
|-------------------------------|--------------------------------------------------------------------------------------|---------------------------------|
| Use | | |
| Patient population | Patients who are undergoing<br>Nuclear Medicine Cardiology<br>SPECT or PET procedure | SAME |
| User | Radiologists, cardiologists, or<br>general nuclear medicine<br>physicians | SAME |
| Use environment | Multiple settings including the<br>hospital, clinic, doctors office, or<br>remotely | SAME |
| Features | | |
| Quantitative Gated SPECT | Yes | SAME |
| Quantitative Perfusion SPECT | Yes | SAME |
| Quantitative Blood Pool SPECT | Yes | SAME |
| Quantitative PET | Yes | SAME |
| Correction of motion artifact | Yes | SAME |
| Reconstruction Package | Yes | SAME |
#### Performance
Support of the substantial equivalence of the CSMC Cardiac Suite was provided as a
Cedars-Sinai Medical Center CSMC Cardiac Suite- Nuclear Medicine Software
Page 12 of 46 Premarket Notification
{2}------------------------------------------------
result of software validation, which confirms all features of the device were compliant with the software requirements.
## Basis for Determination of Substantial Equivalence:
Upon reviewing and comparing intended use, design, principle of operation and overall technological characteristics, the CSMC Cardiac Suite is determined by Cedars-Sinai Medical Center to be substantially equivalent to existing legally marketed device.
Cedars-Sinai Medical Center CSMC Cardiac Suite- Nuclear Medicine Software
Page 13 of 46 Premarket Notification
.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drue Administration 10903 New Hamnshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 22, 2014
Cedars-Sinai Medical Center Department of Medicine Artificial Intelligence in Medicine Program % Mr. Geoff Pollard Program Manager 8700 Beverly Blvd. LOS ANGELES CA 90048
Re: K141652
CSMC Cardiac Suite Nuclear Medicine Software Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: II Product Code: KPS Dated: June 17, 2014 Received: June 20, 2014
Dear Mr. Pollard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (2 l CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2-Mr. Pollard
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael D. O'Hara
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K141652
Device Name: Cedars-Sinai Medical Center (CSMC) Cardiac Suite v2015- Nuclear Medicine Software
Indications for Use:
The Cedars-Sinai Cardiac Suite of applications is intended to enable an automated display, review, and quantification of Nuclear Medicine Cardiology medical images and datasets. Cedars-Sinai Cardiac Suite may be used in multiple settings including the hospital, clinic, doctors office, or remotely. The results provided should be reviewed by qualified healthcare professionals (e.g., radiologists, cardiologists, or general nuclear medicine physicians) trained in the use of medical imaging devices.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Richard D. O'Hara
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health
510(k) k141652
Page 1 of __ 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.