LOGIQ S7 Expert

K160182 · GE Healthcare · IYN · Feb 26, 2016 · Radiology

Device Facts

Record IDK160182
Device NameLOGIQ S7 Expert
ApplicantGE Healthcare
Product CodeIYN · Radiology
Decision DateFeb 26, 2016
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1550
Device ClassClass 2
AttributesPediatric

Intended Use

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Obstetrics (Fetal); Abdominal (includes renal, GYN/Pelvic); Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transrectal; Transvaginal; Transesophageal and Intraoperative (abdominal, thoracic, vascular).

Device Story

The LOGIQ S7 Expert is a mobile, full-featured diagnostic ultrasound system. It acquires, processes, and displays ultrasound data using various transducers. The system features a console with a keyboard, specialized controls, a 10.1-inch LCD touch screen, and a 23-inch LCD widescreen monitor. It supports multiple imaging modes including B-mode, M-mode, PW/CW Doppler, Color/Power Doppler, and 3D/4D imaging. The device is operated by qualified physicians in clinical settings to perform diagnostic evaluations. It incorporates software features such as STIC, OmniView, Cardiac AFI, and Email2MMS. The system provides real-time visualization and measurement capabilities, aiding clinicians in diagnosis and clinical decision-making across a wide range of anatomical regions and exam types.

Clinical Evidence

No clinical studies were required to support substantial equivalence. The device was evaluated through bench testing, including acoustic output, biocompatibility, cleaning/disinfection effectiveness, and thermal, electrical, electromagnetic, and mechanical safety testing.

Technological Characteristics

Mobile console with 23-inch LCD monitor and 10.1-inch touch screen. Employs ultrasonic pulsed Doppler and echo imaging. Supports B, M, PW/CW Doppler, Color/Power Doppler, and 3D/4D modes. Complies with AAMI/ANSI ES60601-1, IEC60601-1-2, IEC60601-2-37, NEMA UD 2/UD 3, ISO 10993-1, and ISO 14971. Connectivity via DICOM. Biocompatible patient-contact materials.

Indications for Use

Indicated for diagnostic ultrasound imaging or fluid flow analysis in patients requiring evaluation of fetal, abdominal, pediatric, small organ, neonatal/adult cephalic, cardiac, peripheral vascular, musculoskeletal, urological, transrectal, transvaginal, transesophageal, or intraoperative structures.

Regulatory Classification

Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract symbol that appears to be three stylized human profiles facing to the right. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 26, 2016 GE HEALTHCARE BRYAN BEHN SR. REGULATORY AFFAIRS MANAGER 9900 INNOVATION DR. WAUWATOSA WI 53226 Re: K160182 Trade/Device Name: Logiq S7 Expert Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: January 20, 2016 Received: January 27, 2016 Dear Mr. Behn: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Michael D.'Hara For Robert Ochs Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) #### 1601 Device Name LOGIQ S7 Expert #### Indications for Use (Describe) The device is intended for use by a qualified physician for ultrasound evaluation of Obstetrics (Fetal); Abdominal (includes renal, GYN/Pelvic); Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Cardiac (adult and pediatric); Peripheral Vasculo-skeletal Conventional and Superficial; Urology (including prostate); Transrectal: Transvaginal; Transesophageal and Intraoperative (abdominal, thoracic, vascular). | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------------------------------------------------------------|---------------------------------------------| | <span style="text-decoration: underline;"><b>X</b></span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular border. The letters and the border are both in a blue color. The logo is simple and recognizable, representing the company's brand identity. # Indications for Use Forms The following forms represent indications with clinical applications and exam types along with the modes of operation for the LOGIQ S7 Expert system and for all of its probe/mode combinations. Combinations identified by "P" represent those previously cleared with the unmodified LOGIQ S7 Expert (K14261) or on another GE ultrasound system in the case of the probes. "N" is new in this submission. The subject modification does not alter the previously cleared system level indications and clinical applications. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular border. The letters and the border are both a light blue color, set against a white background. The logo is simple, yet recognizable, representing the multinational conglomerate corporation. GE Healthcare 510(k) Premarket Notification Submission #### Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert Ultrasound System Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | |----------------------------------------------------|---|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------| | Ophthalmic | | | | | | | | | | | | | Fetal / Obstetrics[7] | P | P | P | P | P | P | P | P | P | P | 5,6 | | Abdominal[1] | P | P | P | P | P | P | P | P | P | P | 3,5,6 | | Pediatric | P | P | P | P | P | P | P | P | P | P | 3,5,6 | | Small Organ[2] | P | P | P | P | P | P | P | P | P | P | 3,5,6 | | Neonatal Cephalic | P | P | P | P | P | P | P | P | P | P | 5,6 | | Adult Cephalic | P | P | P | P | P | P | P | P | P | P | 5,6 | | Cardiac Adult & Pediatric | P | P | P | P | P | P | P | P | P | P | 3,5,6 | | Peripheral Vascular | P | P | P | P | P | P | P | P | P | P | 3,5,6 | | Musculo-skeletal Conventional | P | P | P | P | P | P | P | P | P | P | 3,5,6 | | Musculo-skeletal Superficial | P | P | P | P | P | P | P | P | P | P | 3,5,6 | | Other[4] | P | P | P | P | P | P | P | P | P | P | 3,5,6 | | Exam Type, Means of Access | | | | | | | | | | | | | Transesophageal | P | P | P | P | P | P | P | P | P | P | | | Transrectal [8] | P | P | P | P | P | P | P | P | P | P | 3,5,6 | | Transvaginal | P | P | P | P | P | P | P | P | P | P | 3,5,6 | | Transuretheral | P | P | P | P | P | P | P | P | P | P | 3,5,6 | | Intraoperative [8] | P | P | P | P | P | P | P | P | P | P | 3,5,6 | | Intraoperative Neurological | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | N = new indication; P = previously cleared by FDA [1] Abdominal includes Renal, GYN/Pelvic. Notes: [2] Small organ includes breast, testes and thyroid [3] Elastography Imaging - Elasticity. [4] Other use includes Urology/Prostate [5] 3D/4D Imaging mode [6] Needle guidance imaging [7] Includes infertility monitoring of follicle development, GYN [8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV) [*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD. System provides real-time 3D and 4D acquisition when used with special 4D probes (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle has decorative flourishes around its perimeter, giving it a classic and recognizable appearance. The logo is simple, yet distinctive, and is widely recognized as the symbol of the General Electric company. GE Healthcare 510(k) Premarket Notification Submission # Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with C1-5-D Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | |----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>(Notes) | | Ophthalmic | | | | | | | | | | | | | Fetal / Obstetrics[7] | P | P | P | P | P | P | P | P | P | P | 5,6 | | Abdominal[1] | P | P | P | P | P | P | P | P | P | P | 3,5,6 | | Pediatric | P | P | P | P | P | P | P | P | P | P | 3,5,6 | | Small Organ[2] | | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | Cardiac Adult & Pediatric | | | | | | | | | | | | | Peripheral Vascular | P | P | P | P | P | P | P | P | P | P | 3,5,6 | | Musculo-skeletal Conventional | | | | | | | | | | | | | Musculo-skeletal Superficial | | | | | | | | | | | | | Other[4] | P | P | P | P | P | P | P | P | P | P | 3,5,6 | | Exam Type, Means of Access | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | Transrectal[8] | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Transuretheral | | | | | | | | | | | | | Intraoperative[8] | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | N = new indication; P = previously cleared by FDA Notes: [1] Abdominal includes Renal, GYN/Pelvic. [2] Small organ includes breast, testes and thyroid [3] Elastography Imaging - Elasticity. [4] Other use includes Urology/Prostate [5] 3D/4D Imaging mode [6] Needle guidance imaging [7] Includes infertility monitoring of follicle development, GYN [8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV) [*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD. System provides real-time 3D and 4D acquisition when used with special 4D probes (Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle has decorative flourishes around its perimeter, giving it a classic and recognizable appearance. The logo is simple, yet distinctive, and is widely recognized as the symbol of the General Electric company. # Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with 9L-D Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>(Notes) | |----------------------------------------------------|---|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------| | Ophthalmic | | | | | | | | | | | | | Fetal / Obstetrics[7] | P | P | P | | P | P | P | P | P | P | 5,6 | | Abdominal[1] | P | P | P | | P | P | P | P | P | P | 3,5,6 | | Pediatric | P | P | P | | P | P | P | P | P | P | 3,5,6 | | Small Organ[2] | P | P | P | | P | P | P | P | P | P | 3,5,6 | | Neonatal Cephalic | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | Cardiac Adult & Pediatric | | | | | | | | | | | | | Peripheral Vascular | P | P | P | | P | P | P | P | P | P | 3,5,6 | | Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | P | 3,5,6 | | Musculo-skeletal Superficial | P | P | P | | P | P | P | P | P | P | 3,5,6 | | Other[4] | | | | | | | | | | | | | Exam Type, Means of Access | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | Transrectal[8] | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Transuretheral | | | | | | | | | | | | | Intraoperative[8] | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | N = new indication; P = previously cleared by FDA [1] Abdominal includes Renal, GYN/Pelvic. Notes: [2] Small organ includes breast, testes and thyroid [3] Elastography Imaging - Elasticity. [4] Other use includes Urology/Prostate [5] 3D/4D Imaging mode [6] Needle guidance imaging [7] Includes infertility monitoring of follicle development, GYN (8) Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV) [*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD. System provides real-time 3D and 4D acquisition when used with special 4D probes (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle has decorative swirls around the perimeter, giving it a classic and recognizable appearance. The logo is simple, yet distinctive, representing the brand identity of General Electric. GE Healthcare 510(k) Premarket Notification Submission # Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with ML6-15 Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | |----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>(Notes) | | Ophthalmic | | | | | | | | | | | | | Fetal / Obstetrics[7] | | | | | | | | | | | | | Abdominal[1] | | | | | | | | | | | | | Pediatric | P | P | P | | P | P | P | P | P | P | 3,5,6 | | Small Organ[2] | P | P | P | | P | P | P | P | P | P | 3,5,6 | | Neonatal Cephalic | P | P | P | | P | P | P | P | P | P | 3,5,6 | | Adult Cephalic | | | | | | | | | | | | | Cardiac Adult & Pediatric | | | | | | | | | | | | | Peripheral Vascular | P | P | P | | P | P | P | P | P | P | 3,5,6 | | Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | P | 3,5,6 | | Musculo-skeletal Superficial | P | P | P | | P | P | P | P | P | P | 3,5,6 | | Other[4] | | | | | | | | | | | | | Exam Type, Means of Access | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | Transrectal[8] | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Transuretheral | | | | | | | | | | | | | Intraoperative[8] | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | N = new indication; P = previously cleared by FDA Notes: [1] Abdominal includes Renal, GYN/Pelvic. [2] Small organ includes breast, testes and thyroid [3] Elastography Imaging - Elasticity. [4] Other use includes Urology/Prostate [5] 3D/4D Imaging mode [6] Needle guidance imaging [7] Includes infertility monitoring of follicle development, GYN (8) Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV) [*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD. System provides real-time 3D and 4D acquisition when used with special 4D probes (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized script, enclosed within a circular border. The color of the logo is blue, and the background is white. GE Healthcare 510(k) Premarket Notification Submission # Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with IC5-9-D Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | |----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>(Notes) | | Ophthalmic | | | | | | | | | | | | | Fetal / Obstetrics[7] | P | P | P | | P | P | P | P | P | P | 5,6 | | Abdominal[1] | P | P | P | | P | P | P | P | P | P | | | Pediatric | | | | | | | | | | | | | Small Organ[2] | | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | Cardiac Adult & Pediatric | | | | | | | | | | | | | Peripheral Vascular | | | | | | | | | | | | | Musculo-skeletal Conventional | | | | | | | | | | | | | Musculo-skeletal Superficial | | | | | | | | | | | | | Other[4] | P | P | P | | P | P | P | P | P | P | 3,5,6 | | Exam Type, Means of Access | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | Transrectal[8] | P | P | P | | P | P | P | P | P | P | 3,5,6 | | Transvaginal | P | P | P | | P | P | P | P | P | P | 3.5,6 | | Transuretheral | | | | | | | | | | | | | Intraoperative[8] | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | N = new indication; P = previously cleared by FDA Notes: [1] Abdominal includes Renal, GYN/Pelvic. [2] Small organ includes breast, testes and thyroid [3] Elastography Imaging - Elasticity. [4] Other use includes Urology/Prostate [5] 3D/4D Imaging mode [6] Needle guidance imaging [7] Includes infertility monitoring of follicle development, GYN [8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV) [*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD. System provides real-time 3D and 4D acquisition when used with special 4D probes (Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle has decorative swirls around the perimeter, giving it a classic and recognizable appearance. The logo is simple, yet distinctive, representing the brand identity of General Electric. # Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with 3CRF-D Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | |----------------------------------------------------|-------------------|---|------------|------------|---------------|-----------------|---------------|-----------------|------------------|-------------|---------------| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes* | Harmonic Imaging | Coded Pulse | Other (Notes) | | Ophthalmic | | | | | | | | | | | | | Fetal / Obstetrics[7] | P | P | P | | P | P | P | P | P | P | 5,6 | | Abdominal[1] | P | P | P | | P | P | P | P | P | P | 5,6 | | Pediatric | P | P | P | | P | P | P | P | P | P | 5,6 | | Small Organ [2] | | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | Cardiac Adult & Pediatric | | | | | | | | | | | | | Peripheral Vascular | | | | | | | | | | | | | Musculo-skeletal Conventional | | | | | | | | | | | | | Musculo-skeletal Superficial | | | | | | | | | | | | | Other[4] | P | P | P | | P | P | P | P | P | P | 5,6 | | Exam Type, Means of Access | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | Transrectal[8] | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Transuretheral | | | | | | | | | | | | | Intraoperative[8] | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | N = new indication; P = previously cleared by FDA Notes: [1] Abdominal includes Renal, GYN/Pelvic. [2] Small organ includes breast, testes and thyroid [3] Elastography Imaging - Elasticity. [4] Other use includes Urology/Prostate [5] 3D/4D Imaging mode [6] Needle guidance imaging [7] Includes infertility monitoring of follicle development, GYN (8) Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV) [*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD. System provides real-time 3D and 4D acquisition when used with special 4D probes (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) {10}------------------------------------------------ Image /page/10/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle has decorative flourishes around its perimeter, giving it a classic and recognizable appearance. The logo is simple, yet distinctive, and is widely recognized as the symbol of the General Electric company. GE Healthcare 510(k) Premarket Notification Submission ### Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with L8-18i-D Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | |----------------------------------------------------|-------------------|---|------------|------------|---------------|-----------------|---------------|-----------------|------------------|-------------|---------------| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes* | Harmonic Imaging | Coded Pulse | Other (Notes) | | Ophthalmic | | | | | | | | | | | | | Fetal / Obstetrics[7] | | | | | | | | | | | | | Abdominal[1] | | | | | | | | | | | | | Pediatric | P | P | P | | P | P | P | P | P | P | 5,6 | | Small Organ[2] | P | P | P | | P | P | P | P | P | P | | | Neonatal Cephalic | P | P | P | | P | P | P | P | P | P | | | Adult Cephalic | | | | | | | | | | | | | Cardiac Adult & Pediatric | | | | | | | | | | | | | Peripheral Vascular | P | P | P | | P | P | P | P | P | P | 3,5,6 | | Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | P | 3,5,6 | | Musculo-skeletal Superficial | P | P | P | | P | P | P | P | P | P | 3,5,6 | | Other[4] | | | | | | | | | | | | | Exam Type, Means of Access | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | Transrectal[8] | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Transuretheral | | | | | | | | | | | | | Intraoperative[8] | P | P | P | | P | P | P | P | P | P | 3,5,6 | | Intraoperative Neurological | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | N = new indication; P = previously cleared by FDA Notes: [1] Abdominal includes Renal, GYN/Pelvic. [2] Small organ includes breast, testes and thyroid [3] Elastography Imaging - Elasticity. [4] Other use includes Urology/Prostate [5] 3D/4D Imaging mode [6] Needle guidance imaging [7] Includes infertility monitoring of follicle development, GYN [8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV) [*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD. System provides real-time 3D and 4D acquisition when used with special 4D probes (Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) {11}------------------------------------------------ Image /page/11/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized script, enclosed within a circular shape. The circle is outlined with a thin blue line, and there are small, white, leaf-like shapes evenly spaced around the inside of the circle. #### Diagnostic Ultrasound Indications for Use Form #### GE LOGIQ S7 Expert with S4-10-D Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | | |----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|--| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | | | Ophthalmic | | | | | | | | | | | | | | Fetal / Obstetrics[7] | P | P | P | P | P | P | P | P | P | P | 5 | | | Abdominal[1] | P | P | P | P | P | P | P | P | P | P | 5 | | | Pediatric | P | P | P | P | P | P | P | P | P | P | 5 | | | Small Organ[2] | P | P | P | P | P | P | P | P | P | P | 5 | | | Neonatal Cephalic | P | P | P | P | P | P | P | P | P | P | 5 | | | Adult Cephalic | | | | | | | | | | | | | | Cardiac Adult & Pediatric | P | P | P | P | P | P | P | P | P | P | | | | Peripheral Vascular | | | | | | | | | | | | | | Musculo-skeletal Conventional | | | | | | | | | | | | | | Musculo-skeletal Superficial | | | | | | | | | | | | | | Other[4] | | | | | | | | | | | | | | Exam Type, Means of Access | | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | | Transrectal[8] | | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | | Transuretheral | | | | | | | | | | | | | | Intraoperative[8] | | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | N = new indication; P = previously cleared by FDA Notes: [1] Abdominal includes Renal, GYN/Pelvic. [2] Small organ includes breast, testes and thyroid [3] Elastography Imaging - Elasticity. [4] Other use includes Urology/Prostate [5] 3D/4D Imaging mode [6] Needle guidance imaging [7] Includes infertility monitoring of follicle development; GYN [8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV) [*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD. System provides real-time 3D and 4D acquisition when used with special 4D probes (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) {12}------------------------------------------------ Image /page/12/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle has decorative swirls around the perimeter, giving it a classic and recognizable appearance. The logo is simple, yet distinctive, representing the brand identity of General Electric. ## Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with P2D Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | | |----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|--| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | | | Ophthalmic | | | | | | | | | | | | | | Fetal / Obstetrics[7] | | | | | | | | | | | | | | Abdominal[1] | | | | | | | | | | | | | | Pediatric | | | | | | | | | | | | | | Small Organ[2] | | | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | | | Adult Cephalic | | | P | P | | | | | | | | | | Cardiac Adult & Pediatric | | | P | P | | | | | | | | | | Peripheral Vascular | | | P | P | | | | | | | | | | Musculo-skeletal Conventional | | | | | | | | | | | | | | Musculo-skeletal Superficial | | | | | | | | | | | | | | Other[4] | | | | | | | | | | | | | | Exam Type, Means of Access | | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | | Transrectal[8] | | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | | Transuretheral | | | | | | | | | | | | | | Intraoperative[8] | | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | N = new indication; P = previously cleared by FDA Notes: [1] Abdominal includes Renal, GYN/Pelvic. [2] Small organ includes breast, testes and thyroid [3] Elastography Imaging - Elasticity. [4] Other use includes Urology/Prostate [5] 3D/4D Imaging mode [6] Needle guidance imaging [7] Includes infertility monitoring of follicle development, GYN [8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV) [*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD. System provides real-time 3D and 4D acquisition when used with special 4D probes (Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) {13}------------------------------------------------ Image /page/13/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle has decorative swirls around the perimeter, giving it a classic and recognizable appearance. The logo is simple, yet distinctive, representing the brand identity of General Electric. # Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with P6D Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | | |----------------------------------------------------|-------------------|---|------------|------------|---------------|-----------------|---------------|-----------------|------------------|-------------|---------------|--| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes* | Harmonic Imaging | Coded Pulse | Other [Notes] | | | Ophthalmic | | | | | | | | | | | | | | Fetal / Obstetrics[7] | | | | | | | | | | | | | | Abdominal[1] | | | | | | | | | | | | | | Pediatric | | | | | | | | | | | | | | Small Organ[2] | | | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | | | Adult Cephalic | | | P | P | | | | | | | | | | Cardiac Adult & Pediatric | | | P | P | | | | | | | | | | Peripheral Vascular | | | P | P | | | | | | | | | | Musculo-skeletal Conventional | | | | | | | | | | | | | | Musculo-skeletal Superficial | | | | | | | | | | | | | | Other[4] | | | | | | | | | | | | | | Exam Type, Means of Access | | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | | Transrectal[8] | | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | | Transuretheral | | | | | | | | | | | | | | Intraoperative[8] | | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | N = new indication; P = previously cleared by FDA Notes: [1] Abdominal includes Renal, GYN/Pelvic. [2] Small organ includes breast, testes and thyroid [3] Elastography Imaging - Elasticity. [4] Other use includes Urology/Prostate [5] 3D/4D Imaging mode [6] Needle guidance imaging [7] Includes infertility monitoring of follicle development; GYN [8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV) [*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD. System provides real-time 3D and 4D acquisition when used with special 4D probes (Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) {14}------------------------------------------------ Image /page/14/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The letters are white, creating a contrast against the blue background. GE Healthcare 510(k) Premarket Notification Submission ## Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert with RAB4-8-D Transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation | | | | | | | | | | | | |----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|--| | Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>(Notes) | | | Ophthalmic | | | | | | | | | | | | | | Fetal / Obstetrics[7] | P | P | P | | P | P | P | P | P | P | 5,6 | | | Abdominal[1] | P | P | P | | P | P |…
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...